K253255 · Foshan Pingchuang Medical Technology Co., Ltd. · MUI · Jun 17, 2026 · Radiology
Device Facts
Record ID
K253255
Device Name
Non-Sterile Ultrasound Gel Sterile Ultrasound Gel
Applicant
Foshan Pingchuang Medical Technology Co., Ltd.
Product Code
MUI · Radiology
Decision Date
Jun 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Pediatric
Indications for Use
For Non-Sterile Ultrasound Gel: Non-Sterile Ultrasound Gel is intended to be used on intact skin as a non-sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures. For Sterile Ultrasound Gel: Sterile Ultrasound Gel is intended to be used on intact skin as a sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures.
Device Story
Water-soluble, clear polymer gel; functions as acoustic coupling medium between patient intact skin and ultrasound transducer; eliminates air at contact surface; facilitates transmission of sound waves into body and reception of echoes from tissues. Used in healthcare facilities by clinicians during diagnostic ultrasound imaging. Non-sterile version provided in multiple-use bottles; sterile version provided in single-use packages. Output is acoustic pathway; enables visualization of internal structures. Benefits patient by ensuring effective signal transmission for diagnostic imaging.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including physical/chemical properties (pH, viscosity, density), acoustic performance (velocity, impedance, attenuation), microbiological testing, biocompatibility (ISO 10993-5, -10, -23), and shelf-life validation.
Technological Characteristics
Water-based polymer gel; non-sterile or sterile (Gamma radiation, SAL 10^-6). Physical specs: pH 5.5-7.5; viscosity ≥15 Pa·s; acoustic velocity 1520-1620 m/s; acoustic impedance 1.5-1.7x10^6 Pa·s/m. Biocompatibility per ISO 10993. Transportation testing per ASTM D4169-23.
Indications for Use
Indicated for use on intact skin in adults and pediatrics as a transmission medium for coupling sound waves between patient and ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Sterile and Non-Sterile Ultrasonic Coupling Agent (K242167)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 17, 2026
Foshan Pingchuang Medical Technology Co., Ltd.
% Grace Liu
Consultant
Shenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square
Nanshan District
Shenzhen, Guangdong 518000
CHINA
Re: K253255
Trade/Device Name: Non-Sterile Ultrasound Gel
Sterile Ultrasound Gel
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: MUI
Dated: September 29, 2025
Received: September 29, 2025
Dear Grace Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253255 - Grace Liu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253255 - Grace Liu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MARJAN NABILI -S** for
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253255 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Non-Sterile Ultrasound Gel Sterile Ultrasound Gel | | |
| Please provide your Indications for Use below. | | ? |
| For Non-Sterile Ultrasound Gel: Non-Sterile Ultrasound Gel is intended to be used on intact skin as a non-sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures. For Sterile Ultrasound Gel: Sterile Ultrasound Gel is intended to be used on intact skin as a sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# 510(k) Summary - K253255
## 1. Contact Details
### 1.1 Applicant information
| **Applicant Name** | Foshan Pingchuang Medical Technology Co., Ltd. |
| --- | --- |
| **Address** | Room 301, One of Huaye Road No.3, Lingang Venture Park, Shunjiang, BeijiaoTown, Shunde District Foshan City, Guangdong 528312 China |
| **Contact person** | Anthony Huang |
| **Phone No.** | +86-18927534836 |
| **E-mail** | sales@pingchuangyl.com |
| **Date Prepared** | 2026-06-18 |
### 1.2 Submission Correspondent
| 中国大陆 | Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China |
| --- | --- |
| **Phone No.** | +86-755-86069197 |
| **Contact person** | Grace Liu |
| **Contact person's e-mail** | grace@cefda.com |
| **Website** | http://www.cefda.com |
## 2. Device Information
| **Trade name** | Non-Sterile Ultrasound Gel Sterile Ultrasound Gel |
| --- | --- |
| **Classification** | II |
| **Classification name** | Diagnostic ultrasonic transducer |
| **Product code** | MUI |
| **Regulation No.** | 21 CFR 892.1570 |
## 3. Legally Marketed Predicate Device
| **Trade Name** | Ultra/Phonic® Scanning Gel |
| --- | --- |
| **Manufacturer** | Pharmaceutical Innovations, Inc. |
| **510(k) Number** | K163026 |
| **Product Code** | MUI |
| **Device Class** | II |
| **Regulation No.** | 21 CFR 892.1570 |
The predicate device has not been subject to any design-related recall.
Page 1 of 7
{5}
#### 4. Legally marketed reference device
| **Trade Name** | Sterile and Non-Sterile Ultrasonic Coupling Agent |
| --- | --- |
| **Manufacturer** | Anhui Deepblue Medical Technology Co., Ltd. |
| **510(k) Number** | K242167 |
| **Product Code** | MUI |
| **Device Class** | II |
| **Regulation No.** | 21 CFR 892.1570 |
The reference device has not been subject to any design-related recall.
#### 5. Device Description
Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Gel are water-soluble, clear gel, and they are a type of conductive medium used in ultrasound diagnostic imaging procedures.
When the ultrasound gel is placed between the intact skin of patients and the medical electronic transducers, it eliminates air at the contact surface while providing coupling, so that it provides an acoustic pathway between the intact skin of patients and the medical electronic transducers. This pathway allows the transducer to send sound waves which enter the body, and to receive echoes from tissues and structures.
Non-Sterile Ultrasonic Gel is provided non-sterile, and it is available in several specifications: 60g/bottle, 80g/bottle, 100g/bottle, 150g/bottle and 250g/bottle. Sterile Ultrasonic Gel is a single-use device, provided sterile, and it is available in several specifications: 10g/bag, 12g/bag, 15g/bag, 20g/bag and 30g/bag.
The shelf life of both Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Ge is 3 years.
#### 6. Intended Use/Indication for Use
For Non-Sterile Ultrasound Gel:
Non-Sterile Ultrasound Gel is intended to be used on intact skin as a non-sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures.
For Sterile Ultrasound Gel:
Sterile Ultrasound Gel is intended to be used on intact skin as a sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures.
Page 2 of 7
{6}
## 7. Substantial Equivalence Comparison
| Comparison Item | Proposed Device (K253255) | Predicate Device (K163026) | Reference Device (K242167) | Comment |
| --- | --- | --- | --- | --- |
| Manufacturer | Foshan Pingchuang Medical Technology Co., Ltd. | Pharmaceutical Innovations, Inc. | Anhui Deepblue Medical Technology Co., Ltd. | / |
| Product Name | Non-Sterile Ultrasound Gel Sterile Ultrasound Gel | Ultra/Phonic® Scanning Gel | Sterile and Non-Sterile Ultrasonic Coupling Agent | / |
| Regulation Number | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 | Same |
| Product code | MUI | MUI | MUI | Same |
| Classification | Class II | Class II | Class II | Same |
| Prescription Use | Yes | Yes | Yes | Same |
| Intended Use / Indications for Use | For Non-Sterile Ultrasound Gel: Non-Sterile Ultrasound Gel is intended to be used on intact skin as a non-sterile transmission media for coupling sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures. For Sterile Ultrasound Gel: Sterile Ultrasound Gel is intended to be used on intact skin as a sterile transmission media for coupling | Ultra/Phonic® Scanning Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. It is placed on the patient's skin or on the transducer prior to initiating an ultrasound examination. The product is indicated for prescription use only. | Sterile and Non-Sterile Ultrasonic Coupling Agent are ultrasound couplings intended to be used on intact skin during non-invasive medical ultrasound procedures to couple sound waves between a patient and the medical imaging electronics. The gels are intended for use in all diagnostic ultrasound procedures which require coupling gel or fluid. The gels can be used during the procedures that involved adults and | Equivalent to the predicate device. |
Page 3 of 7
{7}
| | | sound waves between a patient and the ultrasound transducer during non-invasive diagnostic ultrasound imaging procedures. | | pediatrics in professional healthcare facility. | |
| --- | --- | --- | --- | --- | --- |
| Target patients | | Adults and Pediatrics | Adults and Pediatrics | Adults and Pediatrics | Same |
| Environment of Use | | Healthcare facility | Healthcare facility | Healthcare facility | Same |
| Usage | | Non-Sterile: multiple use bottles Sterile: single use packages | Multiple use bottles | Non-Sterile: multiple use bottles Sterile: single use packages | Non-Sterile is the same with the predicate device; Non-sterile and Sterile are the same with the reference device. |
| Patient-contacting material | | Water-based polymer gel | Water-based polymer gel | Water-based polymer gel | Equivalent |
| Performances | Appearance | The product is colorless or light-colored transparent gel without insoluble foreign matter. | Green coloring, high clarity | The product shall be colorless or light-colored transparent gel without insoluble foreign matters. | Equivalent |
| | Density | 0.85-1.15g/ml | 1.035g/ml | 987-1049 kg/m^{3} | Different |
| | pH | 5.5-7.5 | 6.25±0.25 | 5.0-8.0 | |
| | Viscosity | ≥15 Pa·s | 200,000±100,000 cps | ≥15 Pa·s | Same with the reference device |
| | Acoustic Velocity | 1520-1620 m/s at 35°C | 1590 m/s at 22.5°C | 1520-1620 m/s at 35°C | |
| | Acoustic Impedance | 1.5×10^{6}-1.7×10^{6} Pa·s/m at 35°C | 1.65×10^{6} Pa·s/m at 22.5°C | 1.5×10^{6}-1.7×10^{6} Pa·s/m at 35°C | |
| | Acoustic | ≤0.1 dB/(cm·MHz) at 35°C | 0.06+0.01116f^{1.3706} dB/(cm·MHz) | ≤0.1 dB/(cm·MHz) at 35°C | |
Page 4 of 7
{8}
| | Attenuation | | | | |
| --- | --- | --- | --- | --- | --- |
| Biocompatibility | | Conform with ISO 10993 series standards (cytotoxicity, skin sensitization, and skin irritation) | Conform with ISO 10993 series standards (cytotoxicity, skin sensitization, and skin irritation) | Conform with ISO 10993 series standards (cytotoxicity, skin sensitization, and skin irritation) | Same |
| Sterility | | Non-Sterile: non-sterile Sterile: Gamma radiation, SAL 10^{-6} | Non-sterile: non-sterile | Non-sterile: non-sterile Sterile: Gamma radiation, SAL 10^{-6} | Non-Sterile is the same with the predicate device; Non-sterile and Sterile are the same with the reference device. |
In conclusion, the proposed device has the similar indication for use to the predicate device (K163026), and also has similar technical characteristics to the predicate device (K163026) and the reference device (K242167). The differences don't raise any additional questions for safety and effectiveness. Therefore, the proposed device is substantially equivalent to the predicate device (K163026).
Page 5 of 7
{9}
## 8. Summary of Non-clinical Testing
Non-clinical tests were conducted to demonstrate the safety and effectiveness of the proposed device and the substantial equivalence to the predicate device. The following tests were conducted.
# Performance testing:
The following performance tests have been performed on Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Gel, according to the technical specifications, and the results demonstrate the subject device can meet the technical specifications.
- Physical and chemical performances (including Appearance, pH, Viscosity and Relative density)
- Acoustic performances (including Acoustic velocity, Acoustic impedance and Acoustic attenuation)
- Microbiological performances
- Microbial limits and Antimicrobial effectiveness (only for Non-Sterile Ultrasonic Gel)
- Sterility (only for Sterile Ultrasonic Gel)
# Biocompatibility:
The following biocompatibility tests have been performed on Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Gel, according to FDA guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on Sep. 8, 2023, and the test results showed that the proposed device has no biocompatibility issues.
- Cytotoxicity as per ISO 10993-5:2009
- Skin Irritation as per ISO 10993-23:2021
- Skin Sensitization as per ISO 10993-10:2021
# Sterilization:
For Sterile Ultrasonic Gel, a Sterility Assurance Level (SAL) of 10⁻⁶ has been validated in accordance with the requirements of ISO 11137-1:2006, ISO 11137-2:2013 and 11137-3:2017.
# Shelf Life:
Real-time aging has been performed on Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Gel, to assure the shelf-life of 3 years.
# Simulated Transportation Testing:
The simulated transportation testing has been conducted on Non-Sterile Ultrasonic Gel and Sterile Ultrasonic Gel according to ASTM D4169-23.
## 9. Clinical Testing
This 510(k) submission does not utilize clinical study for establishing substantial equivalence, therefore this section
Page 6 of 7
{10}
does not apply.
## 10. Conclusions
A comparison of the intended use, design specifications, and results of non-clinical testing between the proposed device and the legally marketed predicate device (K163026) raised no new questions of safety and effectiveness and show that they are Substantially Equivalent (SE).
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.