The AirRay-mini is a handheld, portable X-ray generator. This device is intended for use by qualified and trained healthcare professionals on adult patients to produce diagnostic X-ray images of the extremities. This device is not intended for mammography.
Device Story
Handheld, battery-powered X-ray generator; produces controlled X-ray radiation for use with separately cleared solid-state detectors or digital radiography systems. Internal battery powers inverter; generates high-voltage output across X-ray tube; electrons accelerate from cathode to anode target to produce X-rays. User selects tube voltage (kV), tube current (mA), and exposure time (mAs) via control unit. Used by qualified healthcare professionals in professional healthcare environments. Device does not acquire or process images; image formation depends on paired imaging system. Benefits include portable, controlled radiation delivery for extremity diagnostics.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-28, and IEC 60601-2-54 standards demonstrated.
Technological Characteristics
Portable X-ray generator; inverter-based; battery/AC powered. Peak voltage 80-90 kV; tube current 5-10 mA. Total filtration 2.5-2.7 mm AL. Weight 3.5 kg. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54.
Indications for Use
Indicated for adult patients requiring diagnostic X-ray imaging of the extremities. Not for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 23, 2026
Poskom Co., Ltd.
% Seong Jeong Hyun
Manager
Poskom Tower, 227, Sowon-Ro Deogyang-Gu, Goyang-Si
Gyeonggi-Do, 10534
REPUBLIC OF KOREA
Re: K253244
Trade/Device Name: AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F (AirRay-mini)
Regulation Number: 21 CFR 892.1720
Regulation Name: Mobile X-Ray System
Regulatory Class: Class II
Product Code: IZL
Dated: January 22, 2026
Received: January 22, 2026
Dear Seong Jeong Hyun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K253244 - Seong Jeong Hyun
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253244 - Seong Jeong Hyun
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253244 | ? |
| Please provide the device trade name(s). | | ? |
| AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F (AirRay-mini) | | |
| Please provide your Indications for Use below. | | ? |
| The AirRay-mini is a handheld, portable X-ray generator.
This device is intended for use by qualified and trained healthcare professionals on adult patients to produce diagnostic X-ray images of the extremities.
This device is not intended for mammography. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253244
# 510(k) Summary
[As Required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR 807.92(a)(1)]
02.07.2025
## 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: POSKOM Co., Ltd.
- Address: POSKOM Tower, 227 Sowon-ro, Deogyang-gu, Goyang-si, Gyeonggi-do, 10534, Republic of Korea
- Contact Name: Jeong Hyun, Sung
- Telephone No.: +82 31-906-9007
- Email Address: jerryhana@poskom.com
- Registration No.: K253244
## 3. Identification of Proposed Device(s) [21 CFR 807.92(a)(2)]
| 510(k) Number | K253244 |
| --- | --- |
| Trade/Device/Model Name | AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F(AirRay-mini) |
| Product Name | AirRay-mini |
| Common Name | Mobile x-ray unit |
| Regulation Name | Mobile x-ray system. |
| Regulation Number | 21 CFR 892.1720 |
| Classification Product Code | IZL |
| Device Class | II |
| 510(k) Review Panel | Radiology |
510(k) Summary
AirRay-mini
{5}
510(k) Summary
AirRay-mini
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate device within this submission is shown as follow;
| 510(k) Number | K143494 |
| --- | --- |
| Trade/Device/Model Name | ADX6000 |
| Product Name | ADX6000 |
| Common Name | Mobile x-ray system |
| Regulation Name | Mobile x-ray system |
| Regulation Number | 21 CFR 892.1720 |
| Classification Product Code | IZL |
| Device Class | II |
| 510(k) Review Panel | Radiology |
These predicate devices have not been subject to a design-related recall.
# 5. Description of the Device [21 CFR 807.92(a)(4)]
The AirRay-mini is a portable, battery-powered X-ray generator intended to provide controlled X-ray radiation for use with separately cleared solid-state detectors or digital radiography systems. The device consists of an X-ray control unit, a high-voltage generator, an X-ray tube assembly, a hand switch, and associated electronics.
DC power from the internal battery is supplied to the inverter, which generates a high-voltage output applied across the X-ray tube. Electrons emitted from the cathode are accelerated toward the anode target, where their deceleration produces X-ray radiation. The device delivers X-rays according to user-selected parameters such as tube voltage (kV), tube current (mA), and exposure time (mAs).
The AirRay-mini does not include an image receptor and does not acquire or process diagnostic images. Diagnostic image formation and quality depend entirely on the separately paired imaging system.
{6}
510(k) Summary
AirRay-mini
# 6. Indications for Use [21 CFR 807.92(a)(5)]
The AirRay-mini is a handheld, portable X-ray generator.
This device is intended for use by qualified and trained healthcare professionals on adult patients to produce diagnostic X-ray images of the extremities.
This device is not intended for mammography.
{7}
# 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of these devices compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the AirRay-mini and the predicate device:
[Table 3. Comparison of Proposed Device to Predicate Devices]
| | Proposed Device | | Predicate Device | Note |
| --- | --- | --- | --- | --- |
| K Number | K253244 | | K143494 | - |
| Manufacturer | POSKOM | | DEXCOWIN | - |
| Trade Name | AirRay-mini 90A/90F | AirRay-mini 80A/80F | ADX6000 | - |
| Product Name | AirRay-mini | | ADX6000 | - |
| Product Code | IZL | | IZL | Identical |
| Regulation Number | 21 CFR 892.1720 | | 21 CFR 892.1720 | Identical |
| 510(k) Review Panel | RADIOLOGY | | RADIOLOGY | Identical |
| Indications for Use | The AirRay-mini is a handheld, portable X-ray generator.
This device is intended for use by qualified and trained healthcare professionals on adult patients to produce diagnostic X-ray images of the extremities.
This device is not intended for mammography. | | The ADX6000 is a handheld and portable general purpose X-ray system. The device uses a variable tube current with voltage from 50-80 kVp and, therefore, is limited to taking diagnostic x-rays of extremities.
It is intended to be used by a qualified and trained clinician on both adult and pediatric patients. This device is not intended for mammography. | Identical |
| Configuration | Portable | | Portable | Identical |
| Generator Type | Inverter | | Inverter | Identical |
| Max Power Output | 500W | 420W | 400 kW | Similar |
| Peak Voltage | 90kV | 80kV | 80 kV | Similar |
| Tube Current | 5-10mA | 5-8mA | 1-5mA | Similar |
| Tube Current | 5-10mA | 5-8mA | | Similar |
510(k) Summary
AirRay-mini
{8}
| | Proposed Device | | Predicate Device | Note |
| --- | --- | --- | --- | --- |
| Total Filtration | 2.5 mm AL | 2.7 mm AL | 2.5 mm AL | Identical |
| Power Source | AC Line, DC Battery
100-240V AC, DC 22.2-25.2V | | AC Line, DC Battery
100-220V AC, DC 18.5-21V | Identical |
| Electrical Safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-1-6
IEC 60601-2-54 | | IEC-60601-1
IEC-60601-1-2,
IEC 60601-1-3,
IEC 60601-2-7,
IEC 60601-2-28 | Similar |
| Photos | | | | Similar |
| Weight | 3.5 kg | | 3.6 kg | Similar |
The AirRay-mini is substantially equivalent to its predicate device, the ADX6000, in design concept, technological characteristics, and intended use. Both devices are portable X-ray generators intended for use by qualified healthcare professionals in professional healthcare environments and are limited to taking diagnostic X-ray exposures of extremities when used in conjunction with separately cleared solid-state detectors or digital radiography systems. Neither device includes an image receptor, and diagnostic image quality is determined solely by the paired imaging system.
The electrical safety, EMC, and performance characteristics of the AirRay-mini have been evaluated in accordance with applicable standards, and the software has been validated. The overall output characteristics—including peak voltage, tube current ranges, exposure parameters, and dose output—are comparable to those of the predicate device. These similarities demonstrate that the technological characteristics do not introduce new performance concerns.
The primary difference between the AirRay-mini and the predicate device lies in product design, including form factor and mechanical configuration. These differences do not affect the device's essential performance or risk profile, as confirmed through testing and risk management activities. No new or different questions of safety or effectiveness are introduced as a result of these design differences.
Therefore, the AirRay-mini is substantially equivalent to the ADX6000.
510(k) Summary
AirRay-mini
{9}
# 8. Non-Clinical Test Summary
The 'AirRay-mini' complies with voluntary standards for electrical safety, electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
1) Electrical Safety, Electromagnetic Compatibility and Performance:
The 'AirRay-mini' complies with the electrical safety and electromagnetic compatibility requirements established by the standards.
| Standards No. | Standards Organization | Standard Title | Version | Publication Year |
| --- | --- | --- | --- | --- |
| 60601-1 | IEC | Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance | 3.2 | 2020 |
| 60601-1-2 | IEC | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 4.1 | 2020 |
| 60601-2-28 | IEC | Medical electrical equipment – Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis | 3.0 | 2017 |
| 60601-2-54 | IEC | Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy | 2.0 | 2022 |
510(k) Summary
6 / 7
AirRay-mini
{10}
2) Software Validation
The 'AirRay-mini' contains software with a Basic Documentation Level. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance:
- Content of Premarket Submissions for Device Software Functions (June 14, 2023)
9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between the proposed device and the predicate device, K143494 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
10. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the 'AirRay-mini' is substantially equivalent in safety and effectiveness to the predicate device as described herein.
510(k) Summary
AirRay-mini
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.