The sponsor provided peer-reviewed literature and an RCT to support the safety and effectiveness of the elongated flexible cryoprobe variant for use in the GI tract.
GI tract application; Safety and effectiveness; Elongated probe
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Peer-reviewed literature (including randomized controlled trial); Randomized controlled trial and peer-reviewed clinical literature
Patients undergoing interventional procedures in the GI tract
Not applicable for this study
Safety and effectiveness of the elongated cryoprobe
Indications for Use
The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization (destruction) of tissue by the application of extreme cold. The Erbe Flexible Cryoprobes are intended for palliative devitalization (destruction) of tissue during interventional procedures by the application of extreme cold and cryoadhesion for applications such as the removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization) and tissue biopsies.
Device Story
ERBECRYO 2 is a cryosurgical unit consisting of a steel/aluminum housing, user interface, and gas control module; it connects to a CO2 source, footswitch, and flexible cryoprobes. The system operates via the Joule-Thomson effect, delivering regulated CO2 flow to the probe tip to achieve extreme cold. Used in clinical settings by physicians for interventional procedures (e.g., GI tract). The device enables cryoadhesion for tissue removal or devitalization. The subject device introduces an elongated flexible cryoprobe (2550mm) compared to the predicate (1150mm). The provider uses the probe under endoscopic guidance to target tissue; the resulting cryo-effect allows for biopsy or recanalization. Benefits include improved reach for GI applications while maintaining the same safety profile as the predicate.
Clinical Evidence
Bench testing included side-by-side tissue freezing performance comparisons, sterilization validation (ISO 11135, SAL 10^-6), EO residual testing (ISO 10993-7), biocompatibility (ISO 10993-1), packaging/shelf-life (ISO 11607-1, ASTM F 1980), EMC (IEC 60601-1-2), and electrical safety (IEC 60601-1). Real-world evidence, including a randomized controlled trial, was provided to support the safety and effectiveness of the elongated probe for GI tract applications.
Indicated for palliative tissue devitalization and cryoadhesion during interventional procedures, including removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tumors (palliative recanalization), and tissue biopsies.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
ERBECRYO 2 Cryosurgical Unit and Accessories (K190651)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 22, 2026
Erbe Elektromedizin GmbH
Matthias Kollek
Regulatory Affairs Specialist
Waldhoernlestrasse 17
Tuebingen, 72072
Germany
Re: K253230
Trade/Device Name: ERBECRYO 2 Cryosurgical Unit and Accessories: ERBECRYO 2 cryosurgical unit; Erbe Flexible Cryoprobe
Regulation Number: 21 CFR 878.4350
Regulation Name: Cryosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: GEH
Dated: September 29, 2025
Received: March 19, 2026
Dear Matthias Kollek:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253230 - Matthias Kollek
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253230 - Matthias Kollek
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K. Chen -S
Digitally signed by Colin
K. Chen -S
Date: 2026.04.22
21:59:05 -04'00"
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253230 | ? |
| Please provide the device trade name(s). | | ? |
| ERBECRYO 2 Cryosurgical Unit and Accessories: ERBECRYO 2 cryosurgical unit; Erbe Flexible
Cryoprobe | | |
| Please provide your Indications for Use below. | | ? |
| Indications for Use statement for the ERBECRYO 2 cryosurgical unit:
The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion and devitalization
(destruction) of tissue by the application of extreme cold.
Indications for Use statement for the Erbe Flexible Cryoprobes:
The Erbe Flexible Cryoprobes are intended for palliative devitalization (destruction) of tissue during
interventional procedures by the application of extreme cold and cryoadhesion for applications such as the
removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization)
and tissue biopsies. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Erbe Elektromedizin GmbH
K253230 - Traditional 510(k) for Flexible Cryoprobe
# 510(k) SUMMARY
## Applicant
Erbe Elektromedizin GmbH
Waldhoernlestrasse 17
72072 Tuebingen
Germany
Tel: 0049-7071-755-0
Fax: 0049-7071-755-179
## Contact Person
Dr. Matthias Kollek
Regulatory Affairs Specialist
E-Mail: Matthias.Kollek@erbe-med.com
## Date Prepared
April 21, 2026
## Device Information
Trade/Proprietary Name: ERBECRYO 2 Cryosurgical Unit and Accessories; Erbe Flexible Cryoprobes
Common Name: Cryosurgical Unit; Endoscopic cryosurgical probe
Classification Name: Cryosurgical Unit & Accessories
Regulation Number: 21 CFR 878.4350
Class: II
Product Code: GEH
## Legally Marketed Predicate Devices
ERBECRYO® 2 Cryosurgical Unit and Accessories - K190651
Page 1 of 5
{5}
Erbe Elektromedizin GmbH
K253230 - Traditional 510(k) for Flexible Cryoprobe
## Device Description
The ERBECRYO 2 unit consists of a steel casing with an aluminum casing button. The plastic face frame comprises the power switch, a plastic receptacle, and a LED display with a membrane keypad. Inside the housing the hardware components are installed. These are the electrical mains supply, the main board (central control unit), the user interface, the gas flow and control module, the flow sensor module and connection sockets for the Flexible Cryoprobes, the footswitch and ECB (Erbe communication bus).
The Unit is connected to an electrical power source, a Carbon Dioxide (CO₂) source, the footswitch, and an instrument. Upon activation via the pedal of the footswitch, the Unit delivers the regulated CO₂ flow to the end of the connected instrument.
The Flexible Cryoprobes are sterile, single-use flexible probes that are connected to the parent device ERBECRYO 2 for palliative devitalization (destruction) of tissue during interventional procedures by the application of extreme cold and cryoadhesion for applications such as the removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization) and tissue biopsies in combination with a cryosurgical unit.
## Device Modification
Compared to the initial clearance (K190651), an elongated variant of the Flexible Cryoprobes is added. The new variant has the same materials, operating principle, and fundamental design as the already cleared variants. The only difference is a longer distal end.
## Indications for Use
The Erbe Flexible Cryoprobes are intended for palliative devitalization (destruction) of tissue during interventional procedures by the application of extreme cold and cryoadhesion for applications such as the removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization) and tissue biopsies.
## Comparison of Technological Characteristics
### ERBECRYO 2
Besides labeling changes due to updated risk management (not related to the elongated cryoprobe), there were no changes to the design, operating principle, software or performance specifications of the parent device ERBECRYO 2 compared to the last clearance K190651.
Page 2 of 5
{6}
Erbe Elektromedizin GmbH
K253230 - Traditional 510(k) for Flexible Cryoprobe
Flexible Cryoprobes
| Characteristics | Subject Device | | | Predicate Device | | | Comparison |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Flexible Cryoprobe | | | Flexible Cryoprobes K190651 | | | |
| Manufacturer | Erbe Elektromedizin GmbH (Germany) | | | Erbe Elektromedizin GmbH (Germany) | | | Same |
| Regulation number | 878.4350 | | | 878.4350 | | | Same |
| Regulatory class | II | | | II | | | Same |
| Product code | GEH | | | GEH | | | Same |
| Indications for use | The Erbe Flexible Cryoprobes are intended for palliative devitalization (destruction) of tissue during interventional procedures by the application of extreme cold and cryoadhesion for applications such as the removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization) and tissue biopsies. | | | The Erbe Flexible Cryoprobes are intended for palliative devitalization (destruction) of tissue during interventional procedures by the application of extreme cold and cryoadhesion for applications such as the removal of foreign bodies, mucus plugs, blood clots, necrotic tissue, tissue tumors (palliative recanalization) and tissue biopsies. | | | Same |
| Prescription or OTC | Prescription | | | Prescription | | | Same |
| Compatibility | ERBECRYO 2 | | | ERBECRYO 2 | | | Same |
| Materials | Stainless steel, silicone, plastics, printing color, adhesive | | | Stainless steel, silicone, plastics, printing color, adhesive | | | Same |
| Probe Dimensions | OD [mm] | Length [mm] | Over-sheath | OD [mm] | Length [mm] | Over-sheath | Increased length based on cleared probe with OD of 1.7mm. |
| | 1.7 | 2550 | N/A | 1.1 | 1150 | OD 2.6mm L 817mm | |
| | | | | 1.1 | 1150 | OD 2.6mm L 757mm | |
| | | | | 1.7 | 1150 | N/A | |
| | | | | 2.4 | 1150 | N/A | |
| | | | | 2.4 | 1150 | N/A | |
| Input pressure | 769 psi to 943 psi (53 bar to 65 bar) | | | 769 psi to 943 psi (53 bar to 65 bar) | | | Same |
| Energy delivery | Cooling by means of CO₂ (Joule-Thomson Effect) | | | Cooling by means of CO₂ (Joule-Thomson Effect) | | | Same |
| Condition Provided/Use Condition | Sterile, single-use | | | Sterile, single-use | | | Same |
| Shelf-life | 3 years | | | 3 years | | | |
| Sterilization Method | Ethylene Oxide | | | Ethylene Oxide | | | Same |
Page 3 of 5
{7}
Erbe Elektromedizin GmbH
K253230 - Traditional 510(k) for Flexible Cryoprobe
The modified Flexible Cryoprobe is based on the cleared variant with an OD of 1.7mm. Compared to the cleared variant, the only difference is an increased distal length (Modified device: 2550mm compared to cleared device: 1150mm). All other technological characteristics such as materials, operating principle, energy type and fundamental design are identical. The modified and cleared probe variants have the same intended use. Non-clinical bench performance testing and safety considerations as described below demonstrate that the modified and cleared probe variants are substantially equivalent.
## Safety Considerations
Due to the length disparity between the subject and the predicate device, a benefit-risk discussion was provided in compliance with FDA Guidance "Benefit-Risk Factors to Consider When Determining Substantial Equivalence in Premarket Notifications (510(k)) with Different Technological Characteristics", demonstrating an improved benefit-risk profile (partially improved benefits while sharing the same risks) compared to the predicate device for the anticipated use in the GI tract.
To substantiate the anticipated application in the GI tract, real-world evidence in form of peer reviewed literature including a randomized controlled trial was provided to demonstrate safety and effectiveness of the elongated probe.
## Summary of non-clinical bench performance testing
Functional testing and design controls to verify both safety and performance of the subject device based on the risk analysis according to ISO 14971 was performed in compliance with ISO 13485, clause 7.3 to ensure that the subject device performs as intended and meets design specifications.
Side-by-side tissue testing was conducted to demonstrate equivalent freezing performance of the subject device compared to the predicate device by measuring the mean biopsy size in a standardized ex vivo setup including common activation times for cryoprobes and the simulation of the application conditions in the human body. Testing was performed on different tissue types to cover a range of tissue densities.
Sterilization validation was performed in compliance with ISO 11135 showing an SAL of 10-6. EO residual testing and limits are in compliance with ISO 10993-7.
Biocompatibility testing was performed in compliance with ISO 10993-1 and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" to demonstrate biocompatibility of the used materials.
Packaging and shelf-life validation was performed in compliance with ISO 11607-1 and accelerated aged (ASTM F 1980) devices.
Electromagnetic compatibility (EMC) was tested in compliance with IEC 60601-1-2 and FDA
Page 4 of 5
{8}
Erbe Elektromedizin GmbH
K253230 - Traditional 510(k) for Flexible Cryoprobe
Guidance "Electromagnetic Compatibility (EMC) of Medical Devices".
Electrical safety was tested in compliance with IEC 60601-1.
Software verification was provided for an enhanced documentation level in compliance with IEC 62304 and FDA Guidance "Content of Premarket Submissions for Device Software Functions" and the device complies with FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".
# Conclusion
The subject device has the same intended use, the same fundamental design, substantially equivalent performance characteristics, and the same energy source as the predicate device. The subject device was tested as described above to demonstrate reasonable assurance of safety and effectiveness of the elongated cryoprobe. Taken together, the subject device does not raise new or different questions of safety and effectiveness, and the subject device is substantially equivalent to the predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.