K253199 · CG Bio Co., Ltd. · OMP · Jun 23, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253199
Device Name
CURASYS 2 (EMDN0003)
Applicant
CG Bio Co., Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Jun 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CURASYS 2 Negative Pressure Wound Therapy (NPWT) System is an integrated wound management system for use in acute and extended care settings. When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
Device Story
Portable, software-controlled NPWT pump; delivers controlled negative pressure to wound site via dressing kit; collects exudate in disposable canister. Operates in continuous or cyclic modes; pressure range -10 to -200 mmHg. Powered by AC or internal rechargeable lithium-ion battery for mobility. Used in acute/extended care settings; operated by clinicians. System monitors pressure status; provides visual/audible alarms for canister full, air leak, low battery, and tube blockage. Output informs clinicians of therapy status; aids wound healing by promoting perfusion and removing infectious material.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including negative pressure accuracy, flow rate, noise, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software verification/validation, usability engineering, and 5-day full-duration simulated-use testing.
Technological Characteristics
Powered suction pump; -10 to -200 mmHg range; 5 mmHg adjustment increments. 350 mL disposable canister. Rechargeable lithium-ion battery. IP32 rating; Type BF applied part. Software-controlled with continuous/cyclic modes. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for patients with open wounds (chronic, acute, traumatic, subacute, dehisced, burns, ulcers, flaps, grafts) requiring wound bed preparation, edema reduction, or exudate removal; also indicated for closed surgical incisions requiring drainage management.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 23, 2026
CG Bio Co., Ltd.
Youngwook Moon
1, Sangwon 12-Gil, Seongdong-Gu
Seoul, 04791
Republic Of Korea
Re: K253199
Trade/Device Name: CURASYS 2 Negative Pressure Wound Therapy System
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered suction pump
Regulatory Class: Class II
Product Code: OMP
Dated: May 20, 2026
Received: May 20, 2026
Dear Youngwook Moon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253199 - Youngwook Moon
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253199 - Youngwook Moon
Page 3
Sincerely,
MUSTAFA A.
MAZHER -S
For Yu-Chieh Chiu PhD,
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253199
Device Name
CURASYS 2 Negative Pressure Wound Therapy System
Indications for Use (Describe)
The CURASYS 2 Negative Pressure Wound Therapy (NPWT) System is an integrated wound management system for use in acute and extended care settings.
When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# ---## K253199
### 1. Submitter Information
**Submitter:** CG Bio Co., Ltd.
**Address:** 1, Sangwon 12-gil, Seongdong-gu, Seoul, Republic of Korea, 04791
**Phone:** +82-2-550-8300
**Contact Person:** Youngwook Moon, Ph.D.
**Mobile:** +1-919-348-5567
**Email:** moon@cgbio.co.kr
**Date Prepared:** May 20, 2026
### 2. Device Information
**Trade/Device Name:** CURASYS 2 Negative Pressure Wound Therapy System
**Common Name:** Powered suction pump for negative pressure wound therapy
**Regulation Name:** Powered suction pump
**Regulation Number:** 21 CFR 878.4780
**Product Code:** OMP
**Device Class:** Class II
### 3. Predicate Device Information
ACTIV.A.C.™ NPWT System, K201571
KCI / 3M
### 4. Device Description
The CURASYS 2 Negative Pressure Wound Therapy (NPWT) System is a portable, software-controlled negative pressure suction device intended to deliver controlled negative pressure to a wound site through an NPWT dressing kit.
The system consists of the CURASYS 2 pump unit, disposable suction canister, connecting tubing, power cord, hanger, and user instructions. The CURASYS 2 pump unit provides negative pressure, monitors pressure status, and provides user alarms for conditions such as full canister, air leak, low battery, and tube blockage.
1
{5}
For performance testing, the CURASYS 2 system was evaluated with the CUDP0510 dressing configuration from the KulaVAC Foam Dressing Kit family, which was cleared under 510(k) K122490.
The CURASYS 2 pump unit may be operated using AC power or its rechargeable lithium-ion battery. AC power is the primary power configuration during normal use, while the internal battery supports temporary operation during patient movement, relocation, or temporary interruption of external power. The battery is not intended to be the sole normal-use power source for continuous therapy.
The device provides continuous and cyclic negative pressure therapy modes. In continuous mode, the device delivers and maintains the selected negative pressure continuously. In cyclic mode, the device alternates between a high-pressure setting and a low-pressure setting according to the programmed cycle.
## 5. Indications for Use
The CURASYS 2 Negative Pressure Wound Therapy (NPWT) System is an integrated wound management system for use in acute and extended care settings.
When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
2
{6}
## 6. Technological Characteristics
The CURASYS 2 NPWT System has the same general technological characteristics as the predicate device. The device uses an electrically powered suction pump to generate controlled negative pressure and deliver that negative pressure to the wound site through tubing and an NPWT dressing kit. Wound exudate is transported through the tubing and collected in a disposable suction canister.
The CURASYS 2 system includes pressure control, continuous and cyclic therapy modes, rechargeable battery support, visual and audible alarm functions, and a user interface for therapy setting and monitoring. The alarm functions include full canister, air leak, low battery, and tube blockage conditions.
A comparison of the technological characteristics of the subject device and predicate device is provided below.
| Feature | Subject Device — CURASYS 2 NPWT System | Primary Predicate — ACTIV.A.C.^{TM} NPWT System, K201571 | |
| --- | --- | --- | --- |
| Regulation / Class / Product Code | 21 CFR 878.4780 / Class II / OMP | 21 CFR 878.4780 / Class II / OMP | |
| General device type | Powered NPWT pump system | Powered NPWT pump system | |
| Use environment | Acute, extended, and home care settings | Acute, extended, and home care settings | |
| Therapy modes | Continuous and cyclic therapy modes | Continuous and intermittent therapy modes | |
| Negative pressure range | -10 to -200 mmHg | -25 to -200 mmHg | |
| Pressure adjustment | 5 mmHg increments | Predicate-specific settings | |
| Canister | Disposable suction canister, 350 | Disposable canister, 300 | |
3
{7}
| | mL | mL | |
| --- | --- | --- | --- |
| Exudate management | Exudate collected in disposable canister; backflow prevention function | Exudate collected in disposable canister | |
| Alarm functions | Canister full, air leak, tube blockage, low battery | Comparable alarm categories, including blockage, canister, low battery, and leak-related alarms | |
| Power source | AC power with rechargeable lithium-ion battery support | AC power / battery-powered portable operation | |
| Battery use | Rechargeable internal battery supports temporary operation during patient movement or temporary loss of AC power | Battery-supported portable operation | |
| Electrical safety / protection | IEC 60601-1 tested; Type BF applied part; IP32 | IEC 60601-1 tested | |
| Software-controlled operation | Yes | Yes | |
| Portability | Portable reusable pump unit with hanger accessory | Portable reusable pump unit | |
| Dressing compatibility | Compatible NPWT dressing kit; performance testing conducted using the CUDP0510 dressing configuration from the KulaVAC Foam Dressing Kit family, K122490 | Compatible NPWT dressing kit | |
4
{8}
# 7. Summary of Nonclinical Testing
Nonclinical testing was performed to support substantial equivalence of the CURASYS 2 NPWT System. The nonclinical testing included electrical safety, electromagnetic compatibility, software verification and validation, usability engineering, bench performance testing, performance qualification testing, full-duration simulated-use testing.
# Bench Performance Testing
Bench performance testing was performed to support the safety and performance of the CURASYS 2 system. The testing included the following:
- Negative pressure accuracy;
- Suction flow rate;
- Operating noise;
- Electrical safety testing in accordance with applicable IEC 60601-1 requirements;
- Electromagnetic compatibility testing in accordance with applicable IEC 60601-1-2 requirements, including emissions and immunity;
- Software verification and validation for the software-controlled pump unit;
- Performance qualification testing of final finished devices from three lots; and
- Full-duration simulated-use testing under cyclic negative pressure operation for a continuous 5-day simulated-use period.
# 8. Summary of Clinical Testing
Clinical testing was not performed. Clinical data were not required to demonstrate substantial equivalence.
# 9. Conclusion
The CURASYS 2 Negative Pressure Wound Therapy System has the same intended use and similar technological characteristics as the predicate device. The differences between CURASYS 2 and the predicate device do not raise different questions of safety or effectiveness.
Based on the nonclinical performance testing, electrical safety testing, electromagnetic compatibility testing, software verification and validation, usability engineering evaluation, full-duration simulated-use testing, and shelf-life rationale, the CURASYS 2 Negative Pressure Wound Therapy System is as safe and effective as, and performs as well as, the predicate device.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.