Microlife Digital Peak Flow Meter, Model PF200B (PF200B)
Applicant
Microlife Corporation
Product Code
BZH · Anesthesiology
Decision Date
Jun 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1860
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Microlife Digital Peak Flow Meter (Model PF200B) is intended for monitoring Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1) for patient home use. The device is designed for pediatric (≥5 years of age) to adult patients.
Device Story
Digital peak flow meter; measures PEF and FEV1 via flow turbine with optical interruption detection; digital signal processing computes pulmonary parameters. Used in home environment by patients (pediatric ≥5 years to adult). Device features LCD for local display; USB 2.0 (USB-C) for data transfer to computer; Bluetooth 5.0 for data transfer to mobile app. Healthcare providers use stored/transferred data to monitor respiratory function and inform clinical decisions. Benefits include longitudinal tracking of pulmonary health.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 62366-1), biocompatibility (ISO 10993), and performance testing against ISO 23747 and ISO 26782 standards.
Technological Characteristics
Surface-contacting (oral mucosa) device; limited duration (<24h). Sensing: flow turbine with optical interruption. Energy: battery-powered. Connectivity: USB-C (wired) and Bluetooth 5.0 (wireless). Software: Basic documentation level. Conforms to ISO 23747, ISO 26782, and ATS/ERS 2019 standards.
Indications for Use
Indicated for monitoring Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1) in pediatric (≥5 years) and adult patients for home use.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
Predicate Devices
Microlife Electronic Peak Flow Meter with PEF and FEV1, Models PF-100 and PF-100-1 (K040723)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 23, 2026
Microlife Corporation
% Vaibhav Rajal
Official Correspondent for Microlife Corporation
Mdi Consultants Inc.
55 Northern Blvd, Suite 200
Great Neck, New York 11021
Re: K253180
Trade/Device Name: Microlife Digital Peak Flow Meter, Model PF200B (PF200B)
Regulation Number: 21 CFR 868.1860
Regulation Name: Peak-Flow Meter For Spirometry
Regulatory Class: Class II
Product Code: BZH,
Dated: May 15, 2026
Received: May 15, 2026
Dear Vaibhav Rajal:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253180 - Vaibhav Rajal
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253180 - Vaibhav Rajal
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**PRAKHYA** Digitally signed by
**T SINGH -S** PRAKHYAT SINGH -S
Date: 2026.06.23
11:38:19 -04'00'
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253180
Device Name
Microlife Digital Peak Flow Meter, Model PF200B
Indications for Use (Describe)
The Microlife Digital Peak Flow Meter (Model PF200B) is intended for monitoring Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1) for patient home use. The device is designed for pediatric (≥5 years of age) to adult patients.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu,
Taipei 11492, Taiwan, China
Tel: 886-2-87971288 Fax: 886-2-87971282
## 510(k) SUMMARY
The assigned 510(k) number is: K253180.
### 1. Submitter's Identification:
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu,
Taipei 11492, Taiwan, China
Date Summary Prepared: June 22, 2026
Contact: Mr. Jimmy Deng
Global Regulatory Affairs & Quality Management Director
Microlife Corporation
Tel: 886-2-87971288 # 376
Email: Jimmy.Deng@microlife.com.tw
### 2. Name of the Device:
Microlife Digital Peak Flow Meter, Model PF200B
Regulation Number: 21 CFR 868.1860
Regulation Name: Peak-flow meter for spirometry.
Regulatory Class: II
Product Code: BZH
### 3. Information for the 510(k) Cleared Device (Predicate Device):
K040723: Microlife Electronic Peak Flow Meter with PEF and FEV1, Models PF-100 and PF-100-1 (with software), K040723, Microlife Intellectual Property GmbH
### 4. Device Description:
Microlife digital peak flow meter PF200B is a medical device that utilizes the principles of flow turbine with optical interruption detection and digital signal processing to compute and provide a measurement of peak expiratory flow and other pulmonary functional parameters.
Page 1 of 9
{5}
This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric (≥5 years of age) to adult patients.
# **5. Indications for Use:**
This Microlife Digital Peak Flow Meter (Model PF200B) is intended for monitoring Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1) for patient home use. The device is designed for pediatric (≥5 years of age) to adult patients.
# **6. Comparison to the 510(k) Cleared Device (Predicate Device):**
| Item | Subject Device: Microlife Digital Peak Flow Meter PF200B Microlife Corporation | Predicate Device: Microlife Peak Flow Meter PF100 and PF100-1 (with software) (K040723) Microlife Intellectual Property GmbH | Substantial Equivalence |
| --- | --- | --- | --- |
| **Measuring principle** | Respiratory flow is determined using a combination of flow turbine, optical interruption detection, and digital signal processing. | Respiratory flow is determined using a combination of flow turbine, optical interruption detection, and digital signal processing. | Substantially equivalent |
| **Measurement Algorithm** | Turbine rotation frequency computation with empirical method algorithm | Turbine rotation frequency computation with empirical method algorithm | Substantially equivalent |
| **Measurement Site** | Mouth | Mouth | Substantially Equivalent |
| **Parameters measured** | Peak expiratory flow (PEF) Forced expiratory volume 1-second (FEV1) | Peak expiratory flow (PEF) Forced expiratory volume 1-second (FEV1) | Substantially Equivalent |
Page 2 of 9
{6}
| **Indications for Use** | The PF200B is intended for monitoring Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1) for patient home use. The device is designed for pediatric (≥5 years of age) to adult patients. | The PF100 is intended for monitoring PEF (Peak Expired Flow Rate) and FEV1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. | Different The device is designed for pediatric (≥5 years of age) to adult patients. |
| --- | --- | --- | --- |
| **Maximum record flow rate** | 900 L/min | Not Publicly Available | Different The maximum flow rate recorded of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 23747) and does not present additional risks. |
| **Flow rate measurement accuracy** | ±10 L/min or ±10% of the reading, whichever is greater | Not Publicly Available | Different The flow rate measurement accuracy of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 23747) and does not present additional risks. |
| **Flow measurement linearity** | ≤10 L/min or ±5% of the reading, whichever is greater. | Not Publicly Available | Different The flow measurement linearity of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 23747) and does not present additional risks. |
{7}
| **Flow measurement resistance to flow** | ≤0.36 kPa/l/s | Not Publicly Available | Different The flow measurement resistance to flow of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 23747) and does not present additional risks. |
| --- | --- | --- | --- |
| **Flow measurement frequency response** | ≤0.25 l/s, or 12 %, whichever is greater | Not Publicly Available | Different The flow measurement frequency response of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 23747) and does not present additional risks. |
| **Maximum recorded volume** | 9.99 liters | Not Publicly Available | Different The maximum recorded volume of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 26782) and does not present additional risks. |
| **Volume measurement accuracy** | ±0.05L or ±2.5% of the reading, whichever is greater. | ±0.1 liter or ±5% of the reading, whichever is greater | Different The volume measurement accuracy of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 26782) and does not present additional risks. |
| **Volume measurement repeatability** | ≤0.05L or 2.5% of the reading, whichever is | Not Publicly Available | Different The volume |
{8}
| | greater. | | measurement repeatability of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 26782) and does not present additional risks. |
| --- | --- | --- | --- |
| **Volume measurement linearity** | <2.5% | Not Publicly Available | Different The volume measurement linearity of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 26782) and does not present additional risks. |
| **Volume measurement impedance** | ≤0.15 kPa/(l/s) | Not Publicly Available | Different The volume measurement impedance of the subject device conforms to the latest relevant standards recognized by US FDA and internationally (ATS/ERS 2019 update and ISO 26782) and does not present additional risks. |
| **Visual output** | Liquid crystal display (LCD) | Not Publicly Available | Different This difference does not affect safety or effectiveness because the display serves only to present measured values to the user and does not alter the method of measurement, calculation algorithms, or performance of the device |
| **Biocompatibility Contact Type** | Surface/oral mucosa contact device | Surface/oral mucosa contact | Substantially Equivalent |
{9}
| | | device | |
| --- | --- | --- | --- |
| Biocompatibility Contact Duration | Limited (≤24 hours) | Limited (≤24 hours) | Substantially Equivalent |
| Device service life | 10,000 measurements | Not Publicly Available | Different The specified service life reflects device durability and longevity rather than measurement methodology. Performance testing demonstrates that the subject device maintains accuracy over its intended service life; therefore, this difference does not affect safety or effectiveness. |
| Wired data connection interface | USB 2.0 interface with USB-C connector for data transfer to computer-based software as “USB datalink function” | USB 2.0 interface with micro-USB connector for data transfer to computer-based software as “USB datalink function” | Different Both the predicate device and subject device have wired data connection to realize a “USB datalink” function, enabling transfer of measurement record data from device memory to computer-based software via USB 2.0 interface for the purpose of data storage and display on the software. The difference between the predicate device and subject device are the different components implemented in the design to realize the USB function (e.g., USB connector). |
| Wireless data connection interface | Bluetooth 5.0 interface for data transfer to mobile device-based software as “Bluetooth datalink function” | None | Different The main difference between the predicate device and subject device is the addition of Bluetooth datalink function, for transfer of measurement record data from device memory to a mobile |
{10}
| | | | device- (e.g., smartphone) based software application (i.e., mobile app) for the purpose of data storage and display. |
| --- | --- | --- | --- |
Based on information from the comparison chart:
The predicate device PF100-1 and subject device PF200B are equivalent in most specifications and characteristics; the differences between the devices have been justified.
7. **Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:**
Testing information demonstrating the safety and effectiveness of the Microlife Digital Peak Flow Meter, Model PF200B in the intended environment of use is supported by testing performed in accordance with the principles outlined in The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] as well as applicable FDA-recognized consensus standards.
The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate device:
The following Guidance documents and National/International Standards were utilized for testing the subject device:
**Electrical Safety, Electromagnetic Compatibility:**
1) ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
2) IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
3) IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
4) IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-
{11}
2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
5) Electromagnetic Compatibility (EMC) of Medical Devices Guidance Document (June 6, 2022)
Performance testing:
6) ISO 23747 Second edition 2015-08 Anesthetic and respiratory equipment - Peak expiratory flow meters for the assessment of pulmonary function in spontaneously breathing humans
7) ISO 26782 First edition 2009-07 Anesthetic and respiratory equipment - Spirometers intended for the measurement of time forced expired volumes in humans
8) ATS/ERS Task Force: Standardization of lung function testing - Standardization of spirometry 2019
Usability
9) IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
10) IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION Medical devices - Part 1: Application of usability engineering to medical devices
11) ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
Biocompatibility: The subject Microlife Digital Peak Flow Meter (Model PF200B) is a surface-contacting medical device with limited-duration patient contact (≤ 24 hours).
12) ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
13) ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
14) ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
15) ISO 10993-12 Fifth edition 2021-01 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
16) ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
{12}
17) Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
# Cleaning
18) ISO 17664-1:2021 Processing of Health Care products- Information to be provided by the medical device manufacturer for the processing of medical devices
19) Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance Document (March 2015)
# Software/Cybersecurity
20) The Documentation Level for the Microlife Digital Peak Flow Meter, Model PF200B is classified as Basic Documentation, in accordance with the FDA guidance document "Content of Premarket Submissions for Device Software Functions" (June 2023).
21) The cybersecurity documentation for the Microlife Digital Peak Flow Meter, Model PF200B was prepared in accordance with the FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 2023).
None of the testing demonstrated any design characteristics that violated the requirements of the reviewer guidance and standards or resulted in any safety hazards. It was our conclusion that Microlife Digital Peak Flow Meter, Model PF200B tested met all relevant requirements of the aforementioned tests.
# 8. Discussion of Clinical Tests Performed:
Not Applicable
# 9. Conclusions:
The subject device has substantially equivalent intended use and characteristics as the predicate device. Moreover, non-clinical testing provided in the submission demonstrates that there are no differences in their technological characteristics that raise new questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.