The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.) and Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: • The PAWP monitoring is intended for adult and pediatric patients only; • C.O. monitoring is intended for adult patients only; The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The uMEC 60/uMEC 70/ uMEC 80/uMEC 100/uMEC 120/uMEC 150 monitors are not intended for helicopter transport, hospital ambulance, or home use.
Device Story
Multi-parameter patient monitors; input physiological signals via sensors (ECG leads, SpO2 probes, NIBP cuffs, IBP/CO/CO2 modules); process signals to display waveforms, numeric values, and alarm states; used in healthcare facilities by clinicians; provide real-time monitoring and data storage; support clinical decision-making through arrhythmia detection, ST-segment analysis, and physiological parameter trending; benefit patients by enabling continuous surveillance and timely intervention for physiological abnormalities.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via functional and system-level testing against consensus standards (IEC 60601-1, 60601-2-27, 80601-2-30, 60601-2-34, 80601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61).
Technological Characteristics
Multi-parameter patient monitors; display sizes 10.1"-15.6"; rechargeable Li-ion battery or AC power; eMMC data storage; internal thermal recorder; ECG (3/5/6/12-lead), NIBP, IBP, SpO2, Temp, CO2, C.O. modules; alarm system (visual/audible); software-based signal processing; compliant with IEC 60601-1, 60601-1-2, and relevant collateral standards.
Indications for Use
Indicated for monitoring, displaying, reviewing, storing, alarming, and transferring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, C.O., CO2) in adult, pediatric, and neonatal patients. PAWP limited to adult/pediatric; C.O. limited to adult. Used in healthcare facilities by trained clinical professionals.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2026
Shenzhen Mindray Bio-medical Electronics Co., LTD.
Lei Li
Manager Regulatory Affairs
Mindray Building, Keji 12th Road South
Hi-tech Industrial Park, Nanshan
Shenzhen, Guangdong 518057
China
Re: K253170
Trade/Device Name: uMEC 60/ uMEC 70/ uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)
Regulatory Class: Class II
Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, DXG, DSJ
Dated: September 20, 2025
Received: September 26, 2025
Dear Lei Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253170 - Lei Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253170 - Lei Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
KIMBERLY N. Digitally signed
CYROWLEY -S by KIMBERLY N. CROWLEY -S
For: Jennifer Kozen
Assistant Director
DHT2A: Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253170 | ? |
| Please provide the device trade name(s). | | ? |
| uMEC 60/ uMEC 70/ uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors | | |
| Please provide your Indications for Use below. | | ? |
| The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors are intended for
monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters
including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis,
QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen
Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP),
Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.) and Carbon Dioxide (CO2).
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the
following:
• The PAWP monitoring is intended for adult and pediatric patients only;
• C.O. monitoring is intended for adult patients only;
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They
should only be used by persons who have received adequate training in their use. The uMEC 60/uMEC 70/
uMEC 80/uMEC 100/uMEC 120/uMEC 150 monitors are not intended for helicopter transport, hospital
ambulance, or home use. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
K253170
Page 1 of 12
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Mindray uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors is provided below.
# 1. SUBMITTER
Applicant: SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Mindray Building, Keji 12th Road South High-tech Industrial Park, Nanshan Shenzhen 518057, P.R. China
Tel: +86 755 81888998
Fax: +86 755 26582680
Contact: Li Lei
Title: Manager Regulatory Affairs
Phone: +86 755 81885953
Fax: +86 755 26582680
E-mail: lilei.js@mindray.com
Date Prepared: September 25, 2025
# 2. DEVICE
Device Trade Name: uMEC 60/ uMEC 70/ uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors
Device Common Name: Patient Monitor
Classification Name: 21 CFR 870.1025, Class II, Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: II
Primary Product Code: MHX
Subsequent Product Codes Refer to Tabel 1
Panel: Cardiovascular
Table 1: Secondary Product Codes
| Regulation Number/Class | Product Codes | Regulation description | Device Common Name |
| --- | --- | --- | --- |
| 870.1025, II | DSI | Arrhythmia detector and alarm (including ST-segment measurement and alarm) | Detector and alarm, arrhythmia |
{5}
510(k) Summary
K253170
Page 2 of 12
| 870.1025, II | MLD | Arrhythmia detector and alarm (including ST-segment measurement and alarm) | Monitor, st segment with alarm |
| --- | --- | --- | --- |
| 870.2300, II | DRT | Cardiac Monitor (including cardiotachometer and rate alarm) | Monitor, cardiac (incl. cardiotachometer & rate alarm) |
| 870.1130, II | DXN | Noninvasive blood pressure measurement system | System, measurement, blood-pressure, non-invasive |
| 870.1110, II | DSK | Blood pressure computer | Computer, blood-pressure |
| 880.2910, II | FLL | Clinical electronic thermometer | Thermometer, electronic, clinical |
| 870.2700, II | DQA | Oximeter | Oximeter |
| 868.1400, II | CCK | Carbon dioxide gas analyzer | Analyzer, gas, carbon-dioxide, gaseous-phase |
| 870.1435, II | DXG | Single-function, preprogrammed diagnostic computer | Computer, diagnostic, pre-programmed, single-function |
| 870.1100, II | DSJ | Blood pressure alarm | Alarm, blood-pressure |
## 3. PREDICATE DEVICES
Primary predicate device:
- K200015- ePM Series Patient monitors, ShenZhen Mindray Bio-Medical Electronics CO., LTD.
## 4. DEVICE DESCRIPTION
The subject uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors includes six monitors:
- uMEC 60 Patient Monitor
- uMEC 70 Patient Monitor
- uMEC 80 Patient Monitor
- uMEC 100 Patient Monitor
- uMEC 120 Patient Monitor
- uMEC 150 Patient Monitor
{6}
510(k) Summary
K253170
Page 3 of 12
The above mentioned Patient Monitors are Mindray’s new generation monitoring product family with ergonomic and flexible design in platform of both software and hardware to meet the clinical needs of monitoring.
## 5. INTENDED USE/INDICATIONS FOR USE
The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO₂), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.) and Carbon Dioxide (CO₂).
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- The PAWP monitoring is intended for adult and pediatric patients only;
- C.O. monitoring is intended for adult patients only;
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 monitors are not intended for helicopter transport, hospital ambulance, or home use.
## 6. SUBSTANTIAL EQUIVALENCE
### Comparison of Indications
Both the predicate devices and the subject devices are multi parameter patient monitors intended to be used under the direction of clinical professionals. Comparing with the primary predicate ePM Series Patient Monitors (K200015), the indications for use for the subject device uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors are equivalent.
The subject and Primary predicate devices have the same use environment (healthcare facilities), and patient population (adult, pediatric, and neonatal patients).
In conclusion, the intended use of the subject device uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors is substantially equivalent to that of the primary predicate ePM Series Patient Monitors (K200015).
### Comparison of Technological Characteristics
The table below compare the key technological features of the subject devices to the primary predicate devices (K200015, ePM Series Patient Monitors). The features in grey are features which are different between the predicate devices and the subject devices. Where appropriate, the specific differences of the features are bolded.
{7}
510(k) Summary
K253170
Page 4 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| Primary display and touch screen | 10.1”
1024
×800
pixels | 12.1”
1280*
800
pixels | 15.6
1366
×768
pixels | 10.1”
1024
×800
pixels | 12.1”
1280*
800pixels | 15.6
1366
×768
pixels | 10.1”
1024×600
pixels | | 12.1”
1280×800
pixels | | 15.6
1366×768
pixels | |
| Secondary display | Mirrored display | | | | | | Mirrored display | | | | | |
| Power supply | Battery or AC power | | | | | | Battery or AC power | | | | | |
| Battery | Rechargeable Lithium-Ion, 10.8VDC, 5600 mAh.
Rechargeable Lithium-Ion, 10.95 VDC, 4500 mAh.
Rechargeable Lithium-Ion, 10.95 VDC, 2600 mA | | | | | | Rechargeable Lithium-Ion,10.95V, 5200mAh | | | | | |
| Data storage | Embedded Multi Media Card(eMMC) | | | | | | Embedded Multi Media Card(eMMC) | | | | | |
| Data recorder | Supports internal thermal recorder | | | | | | Supports internal thermal recorder | | | | | |
| Speaker | Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulation | | | | | | Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulation | | | | | |
| Alarm system | The alarm lamp is cyan, yellow, or red depending on alarm type | | | | | | The alarm lamp is yellow, or red depending on alarm type | | | | | |
| ECG | 3-lead, 5-lead, 6-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, an interpretation of resting 12-lead ECG, J-point Auto detection, Dual Channel Pace detection, adjustable QRS threshold and heart rate (HR).
Arrhythmia detection is intended for adult, pediatric and neonate.
Supports intelligent arrhythmia alarm.
ST segment analysis is intended for adult, pediatric and neonate. | | | | | | 3-lead, 5-lead, 6-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, J-point Auto detection, Dual Channel Pace detection, adjustable QRS threshold and heart rate (HR).
Arrhythmia detection is intended for adult, pediatric and neonate.
Supports intelligent arrhythmia alarm.
ST segment analysis is intended for adult, pediatric and neonate.
The measurement module has circuit board layout differences. | | | | | |
| Arrhythmia Analysis | Asystole, VFib/Vtac, Vtac, Vent.Brady, Extreme Tachy, Extreme Brady, PVCs, Couplet, Bigeminy, Trigeminy, R on T, Run PVCs, PVCs/min, Tachy, Brady, Missed Beats, Vent Rhythm, Pacer Not Pacing, Pacer Not Capture, Multif.PVC, | | | | | | Asystole, VFib/Vtac, Vtac, Vent.Brady, Extreme Tachy, Extreme Brady, PVCs, Couplet, Bigeminy, Trigeminy, R on T, Run PVCs, PVCs/min, Tachy, Brady, Missed Beats, Vent Rhythm, Pacer Not Pacing, Pacer Not Capture, Multif.PVC, | | | | | |
{8}
510(k) Summary
K253170
Page 5 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| | Nonsus.Vtac, Pause, Vent.Rhythm, Afib, Pauses/min, Pauses/min | | | | | | Nonsus.Vtac, Pause, Vent.Rhythm, Afib, Pauses/min, Pauses/min | | | | | |
| Respiration rate (Resp) | Measurement range:
Adult: 0 to 120 rpm;
Pediatric, neonate: 0 to 150 rpm.
Accuracy:
7 to 150 rpm: ±2 rpm or ±2%, whichever is greater;
0 to 6 rpm: Not specified | | | | | | Measurement range:
Adult: 0 to 120 rpm;
Pediatric, neonate: 0 to 150 rpm.
Accuracy:
7 to 150 rpm: ±2 rpm or ±2%, whichever is greater;
0 to 6 rpm: Not specified
The measurement module has circuit board layout differences. | | | | | |
| Temperature (Temp) | Measurement range: 0 to 50°C (32 to 122°F)
Accuracy: ±0.1°C or ±0.2°F (without probe). | | | | | | Measurement range: 0 to 50°C (32 to 122°F)
Accuracy: ±0.1°C or ±0.2°F (without probe).
The measurement module has circuit board layout differences. | | | | | |
| Pulse oxygen saturation (SpO_{2}) | Supports Mindray SpO2 function, Masimo SpO2 function and Nellcor SpO2 function from multi parameter module.
Mindray SpO2 function
Measurement range: 0 to 100%;
Accuracy:
70 to 100%: ±2% (adult/pediatric mode);
70 to 100%: ±3% (neonate mode);
0% to 69%: Not specified.
Masimo SpO2 function
Measurement range: 1 to 100%;
Accuracy:
70 to 100%: ±2% (measured without motion in adult/pediatric mode);
70 to 100%: ±3% (measured without motion in neonate mode);
70 to 100%: ±3% (measured with motion);
1% to 69%: Not specified.
Nellcor SpO2 function | | | | | | Only support Mindray SpO2 function from multi parameter module.
Measurement range: 0 to 100%;
Accuracy:
70 to 100%: ±2% (adult/pediatric mode);
70 to 100%: ±3% (neonate mode);
0% to 69%: Not specified.
The measurement module has circuit board layout differences. | | | | | |
{9}
510(k) Summary
K253170
Page 6 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Cleared in K200015 | | | | | | Subject Device | | | | | |
| | Measurement range: 0 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric); 70 to 100%: ±3% (neonate); 0% to 69%: Not specified. | | | | | | | | | | | |
| Pulse rate (PR) | PR from built-in Mindray SpO2 Module Measurement range: 20 to 254 bpm; Accuracy: ±3 bpm. PR from built-in Masimo SpO2 Module Measurement range: 25 to 240 bpm; Accuracy: ±3 bpm (measured without motion); ±5 bpm (measured with motion). PR from built-in Nellcor SpO2 Module Measurement range: 20 to 300 bpm; Accuracy: 20 to 250 bpm: ±3 bpm; 251 to 300 bpm, not specified | | | PR from built-in Mindray SpO2 Module Measurement range: 20 to 254 bpm; Accuracy: ±3 bpm. PR from built-in Masimo SpO2 Module Measurement range: 25 to 240 bpm; Accuracy: ±3 bpm (measured without motion); ±5 bpm (measured with motion). PR from built-in Nellcor SpO2 Module Measurement range: 20 to 300 bpm; Accuracy: 20 to 250 bpm: ±3 bpm; 251 to 300 bpm, not specified | | | PR from built-in Mindray SpO2 Module Measurement range:20 to 254 bpm; Accuracy: ±3 bpm. PR from IBP Module Measurement range: 25 to 350 bpm; Accuracy: ±1 bpm or ±1%, whichever is greater. | | | | | |
| Non-invasive blood pressure (NIBP) | Measurement range: | | | | | | Measurement range: | | | | | |
| | | | Adult | Pediatric | Neonate | | | | Adult | Pediatri c | Neonate | |
| | Systolic | | 25-290 | 25-240 | 25-140 | | | | | | | |
| | Diastolic | | 10-250 | 10-200 | 10-115 | | Systolic | | 25-290 | 25-240 | 25-140 | |
{10}
510(k) Summary
K253170
Page 7 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| | Accuracy:
Max mean error: ±5 mmHg;
Max standard deviation: 8 mmHg. | Diastolic | 10-250 | 10-200 | 10-115 | | | | | | | |
| | | | | | | | | Mean | 15-260 | 15-215 | 15-125 | |
| | | | | | | | | Accuracy:
Max mean error: ±5 mmHg;
Max standard deviation: 8 mmHg.
The measurement module has circuit board layout differences. | | | | |
| Invasive blood pressure (IBP) | Uses an internal IBP module to measure invasive blood pressure. The monitor can monitor invasive blood pressures and displays systolic, diastolic and mean pressures and a waveform for each pressure. Support PPV function. Measurement range: -50 to 300 mmHg;
Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | | | Uses an internal IBP module or stand-alone IBP Module to measure invasive blood pressure. The monitor can monitor invasive blood pressures and displays systolic, diastolic and mean pressures and a waveform for each pressure. Support PPV function.
Measurement range: -50 to 300 mmHg;
Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | | | Not support | | | Uses an internal IBP module to measure invasive blood pressure. The monitor can monitor invasive blood pressures and displays systolic, diastolic and mean pressures and a waveform for each pressure. Support PPV function.
Measurement range: -50 to 300 mmHg; Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | | |
| Cardiac output (C.O.) | Use internal C.O. module.
The cardiac output (C.O.)
measurement invasively measures
cardiac output and other hemodynamic parameters using the right heart (atria) thermodilution method. The temperature change | | | Use internal or external C.O. module.
The cardiac output (C.O.)
measurement invasively measures
cardiac output and other hemodynamic parameters
using the right heart (atria) thermodilution | | | Not support | | | Use internal C.O. module.
The cardiac output (C.O.)
measurement invasively measures
cardiac output and other
hemodynamic parameters using
the right heart | | |
{11}
510(k) Summary
K253170
Page 8 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| | is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The monitor is capable of storing 6 measurements. Measurement range: C.O: 0.1 to 20 L/min; TB: 23 to 43°C; TI: 0 to 27°C; Accuracy: C.O: ±5% or ±0.1 L /min, whichever is greater; TB, TI: ±0.1°C (without sensor). | | | method. The temperature change is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The monitor is capable of storing 6 measurements. Measurement range: C.O: 0.1 to 20 L/min; TB: 23 to 43°C; TI: 0 to 27°C; Accuracy: C.O: ±5% or ±0.1 L /min, whichever is greater; TB, TI: ±0.1°C (without sensor). | | | | | | (atria) thermodilution method. The temperature change is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The monitor is capable of storing 6 measurements. Measurement range: C.O: 0.1 to 20 L/min; TB: 23 to 43°C; TI: 0 to 27°C; Accuracy: C.O: ±5% or ±0.1 L /min, whichever is greater; TB, TI: ±0.1°C (without sensor). | | |
| Carbon dioxide (CO_{2}) | Compatible with 3 internal CO2 modules: Sidestream CO2 2.0 module Mainstream CO2 Module MicroStream CO2 module | | | Compatible with 4 internal or external CO2 modules: Sidestream CO2 1.0 Module Sidestream CO2 2.0 module Mainstream CO2 module MicroStream CO2 module *10M does not support external CO2 modules | | | Not support | | | Compatible with 1 internal CO2 modules: Sidestream CO2 2.0 module | | |
{12}
510(k) Summary
K253170
Page 9 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| | Type: Sidestream CO2 Module (CO2 2.0):
Measurement range: 0~150mmHg
Accuracy:
0~40 mmHg: ±2mmHg,
41~76 mmHg: ±5% of reading,
77~99 mmHg: ±10% of reading,
100~150mmHg: ±(3mmHg + 8% of reading),
ISO accuracy mode:
Add ±2mmHg to the full accuracy mode
AwRR measurement:
awRR measurement range: 0 to 150rpm;
awRR:
<60rpm, ±1rpm,
60~150rpm, ±2rpm | | | | | | | Not support | | Type: Sidestream CO2 Module (CO2 2.0):
Measurement range:
0~150mmHg
Accuracy:
0~40 mmHg: ±2mmHg,
41~76 mmHg: ±5% of reading,
77~99 mmHg: ±10% of reading,
100~150mmHg: ±(3mmHg + 8% of reading),
ISO accuracy mode:
Add ±2mmHg to the full accuracy mode
AwRR measurement:
awRR measurement range: 0 to 150rpm;
awRR:
<60rpm, ±1rpm,
60~150rpm, ±2rpm | | |
| | Type: Microstream CO2 Module
Measurement range:
CO2: 0~99mmHg;
awRR: 0~150rpm;
Accuracy:
CO2:
0~38mmHg: ±2mmHg;
39~99mmHg: ±5% of the reading+0.08% of (the | | | | | | | Not support | | | | |
{13}
510(k) Summary
K253170
Page 10 of 12
| Feature | ePM 10 | ePM 12 | ePM 15 | ePM 10M | ePM 12M | ePM 15M | uM E C 10 0 | uM E C 60 | uM EC 120 | uM EC 70 | uM EC 150 | uM EC 80 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cleared in K200015 | | | | | | | Subject Device | | | | | |
| | reading-38).
awRR:
0~70rpm: ±1rpm,
71~120rpm: ±2rpm,121~150rpm: ±3rpm | | | | | | | | | | | |
| | Type: Mainstream CO2
Module Measurement range:
CO2: 0~150mmHg;
awRR: 0~150rpm;
Accuracy:
CO2:
0~40mmHg: ±2mmHg,
41~70mmHg: ±5% of the reading,
71~100mmHg: ±8% of the reading,
101~150mmHg: ±10% of the reading;
awRR: ±1rpm. | | | | | | Not support | | | | | |
In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness.
## 7. PERFORMANCE DATA
To establish the substantial equivalence of the uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its specifications and is substantially equivalent to the predicate device.
In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
## Biocompatibility Testing
The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors are not patient contacting. There are no new patient contacting accessories or components. There have been no material changes to the previously cleared patient contacting devices, therefore biocompatibility testing is not required.
## Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of
{14}
510(k) Summary
K253170
Page 11 of 12
Premarket Submissions for Software Contained in Medical Devices.” Verification of the uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
## Electromagnetic Compatibility and Electrical Safety
The uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply:
- IEC 60601-1:2005/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005 MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC/TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
## Bench Testing
To establish the substantial equivalence of the uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device.
In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
- ANSI AAMI IEC 60601-2-27:2011(R)2016 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
- ANSI AAMI IEC 80601-2-30:2018 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-2-34 Edition 3.0 2011-05 Medical electrical equipment - Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
{15}
510(k) Summary
K253170
Page 12 of 12
- IEC 80601-2-49:2018 Medical electrical equipment –Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
- ISO 80601-2-55 Second edition 2018-02 [Including AMD1:2023] Medical electrical equipment--Part2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
- ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]
- ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) Medical electrical equipment--Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ANSI AAMI IEC 60601-1-8:2006 and A1:2012 [Including AMD 2:2021] Medical Electrical Equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems [Including Amendment 2 (2021)]
## 8. CONCLUSION
Based on the detailed comparison of specifications for each of the characteristics to the predicate devices, the performance testing and conformance with applicable standards, the uMEC 60/uMEC 70/uMEC 80/uMEC 100/uMEC 120/uMEC 150 Patient Monitors can be found substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.