K253132 · Zhejiang Chuangxiang Medical Technology Co., Ltd. · FGE · Mar 6, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253132
Device Name
Single use stone retrieval balloons
Applicant
Zhejiang Chuangxiang Medical Technology Co., Ltd.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Mar 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Retrieval Balloon is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon. The device is for adult use only.
Device Story
Single-use biliary catheter with natural latex balloon; Pebax catheter body; 2 or 3 internal lumens for balloon inflation, guidewire, and contrast medium. Used during ERCP procedures; operated by physicians in hospitals/clinics. Device inserted endoscopically; balloon inflated via external pump to specific diameters (8.5mm–18mm) to occlude bile duct or retrieve stones. Radiopaque rings at balloon ends provide fluoroscopic visualization. Contrast medium injected through dedicated lumen (in 3-lumen models) to assist imaging. Enables stone removal and ductal visualization; benefits patient by facilitating minimally invasive biliary clearance.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included appearance, dimensions, flow rate, guidewire/endoscope compatibility, inflation volume, radiodetectability, tensile strength, sealing/leakage, corrosion resistance, and pump performance. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity) performed per ISO 10993-1.
Technological Characteristics
Natural latex balloon; Pebax catheter; Tantalum radiopaque rings. 2 or 3 internal lumens. EO sterilized. 200cm working length; compatible with ≥3.2mm endoscope channels. Inflation via external pump; two-way valve maintains pressure. Shelf life 2 years. Complies with ISO 11135 (sterilization), ISO 10993-7 (EO residuals), ISO 11607 (packaging), and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for adult patients requiring endoscopic removal of biliary stones or contrast medium injection during biliary duct occlusion.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 6, 2026
Lucius Long
Zhejiang Chuangxiang Medical Technology Co., LTD.
Room 101-1, 201-1, 301, 401, Building 50
No.650 Hongfeng Road, Donghu Street, Linping District
Hangzhou, 311100
China
Re: K253132
Trade/Device Name: Single use stone retrieval balloons
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter And Accessories
Regulatory Class: Class II
Product Code: FGE
Dated: February 3, 2026
Received: February 3, 2026
Dear Lucius Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253132 - Lucius Long
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253132 - Lucius Long
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony C. Lee Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253132 | ? |
| Please provide the device trade name(s). | | ? |
| Single use stone retrieval balloons | | |
| Please provide your Indications for Use below. | | ? |
| The Retrieval Balloon is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon. The device is for adult use only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 1 of 9)
# 510(k) Summary( 21CFR 807.92)
## 1. Submitter's information
Name: Zhejiang Chuangxiang Medical Technology Co., LTD.
Address: Room 101-1, 201-1, 301, 401, Building 50, No.650 Hongfeng Road, Donghu Street, Linping District, Hangzhou City,311100 Zhejiang Province, P.R. China.
Contact person: Lucius.Long
Email: lucius.long@med-nova.com
Telephone: 86-571-89167088
Fax: 86-571-89167086
## 2. Date of Submission
25-September- 2025
## 3. Device Identification
510(k) Number: K253132
Trade/Device Name: Single use stone retrieval balloons
Regulation name: Biliary catheter and accessories
Regulation class: II
Classification name: Biliary catheter and accessories
Regulation number:876.5010
Product code: FGE
## 4. Predicative device identification
510(k) Number: K211021
Device Name: Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible
Product Code: FGE
## 5. Device description
The proposed device Single use stone retrieval balloons includes Stone Retrieval Balloon, inflation pump and two-way valve. Stone Retrieval Balloon is commonly used in traditional ERCP surgery with a long guidewire.
The proposed Single use stone retrieval balloons is comprised of a natural latex balloon mounted at the distal end of a Pebax catheter with three internal lumens for ballooning, guidewire and contrast medium. The internal lumen for ballooning is used to inflate/deflate the balloon. Multiple syringes are included with the packaging, allowing balloon inflation to a specific diameter.
{5}
科技医学
MEDNOVA
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 2 of 9)
Following insufflation, the balloon surface lies flat against the bile duct wall, enabling efficient and complete cleaning of the bile duct. A separate lumen is designed for a 0.035 inch guidewire. Another separate lumen is designed for contrast agent.
The balloon can be inflated to $8.5 \mathrm{~mm}$ , $12 \mathrm{~mm}$ , $15 \mathrm{~mm}$ and $18 \mathrm{~mm}$ diameters using the inflation pump and be kept in an inflated state by two-way valve. Balloon can be inflated to two or three distinct sizes if a different diameter is needed without having to exchange devices. During stone removal process, the balloon can be filled to removal multiple stones in the biliary tract.
There are two developing rings placed at the distal and proximal ends of the balloon providing high fluoroscopic visualization of the balloon location.
| Model | Diagram of structure | Differences in Design | Clinical Procedures |
| --- | --- | --- | --- |
| D Double cavity | 1. Guide wire chamber catheter seat:2. Balloon catheter seat: | The double cavity model included the Balloon catheter seat and the Guide wire chamber catheter seat without the injection cavity. | The Balloon catheter seat was connected to the inflator pump for inflating the balloon, and the Guide wire chamber catheter seat was used for inserting the guide wire. The guide wire is inserted from distal end of sheathing to guide wire chamber catheter seat. |
{6}
科技医学
MEDNOVA
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 3 of 9)
| T Three- cavity | 1 2 3 1 2 | The three cavities model included Balloon catheter seat, Guide wire chamber catheter seat and Injection chamber catheter holder. | The Balloon catheter seat was connected with the inflator pump to inflate the balloon, the Guide wire chamber catheter seat was used for guide wire insertion, and the Injection chamber catheter holder was used for injection of contrast agent. The guide wire is inserted from distal end of sheathing to guide wire chamber catheter seat. |
| --- | --- | --- | --- |
{7}
科技医学
MEDNOVA
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 4 of 9)
| E Fast exchange | Canyiling Quick Exchange Pouch Balloon | The fast exchange model included Balloon catheter seat, Guide wire chamber catheter seat and Injection chamber catheter holder. A quick exchange notch was arranged at one end of the guide wire cavity near the balloon, which could shorten the time of the guide wire passing through the guide wire cavity. | The Balloon catheter seat was connected with the inflator pump to inflate the balloon, the Guide wire chamber catheter seat was provided with a quick exchange port for rapid insertion of the guide wire, and the Injection chamber catheter holder was used for injection of contrast agent. The guide wire is inserted from distal end of sheathing to quick exchange port or guide wire chamber |
| --- | --- | --- | --- |
{8}
创想医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 5 of 9)
| | | | catheter seat. |
| --- | --- | --- | --- |
# 6. Indications for use
The Retrieval Balloon is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon. The device is for adult use only.
# 7. Comparison of Technological Characteristics with the predicate device:
| Item | Proposed device | Predicate device | Comparison to Predicate Devices |
| --- | --- | --- | --- |
| Device name | Single use stone retrieval balloons | Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible | N/A |
| Product code | FGE | FGE | N/A |
| Regulation No. | 876.5010 | 876.5010 | N/A |
| Manufacturer | Zhejiang Chuangxiang Medical Technology Co., LTD. | Micro-Tech(Nanjing)Co.,Ltd. | N/A |
| Classification | II | II | Same |
| Sterilization | EO | EO | Same |
| Material | Natural Latex/Tantalum/Polyether Block Amide | Natural Latex, Platinum Iridium Alloy, Polyether Block Amide | Different Biocompatibility tests have been done for the difference. Biological risks are acceptable. |
| Balloon Size | 8.5mm/12m/15mm/18mm | 9mm/12mm/15mm/18mm | Different Performance tests have been done for the difference. The size different is acceptable |
{9}
创想医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 6 of 9)
| Indications for Use | The Retrieval Balloon is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon. The device is for adult use only. | The Retrieval Balloon Catheter is indicated for use endoscopically to remove stones from biliary system, or to facilitate injection of contrast medium while occluding the duct with the balloon. | Same |
| --- | --- | --- | --- |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch | Same |
| Shelf Life | 2 years | 2 years | Same |
| Device components | Balloon, Sheathing, Coupling, Injection chamber catheter holder, Guide wire chamber, Balloon catheter seat, Developing ring, Inflator pump Two-way valve | Natural latex balloon, Pebax catheter with three internal lumens, adiopaque bands Multiple syringes | Different It is only different in description. |
| Lumen number | 2 and 3 | 3 | Different The proposed device includes non-injection lumen, which does not increase the risk. |
| Injection function | The three-lumen device has a catheter with an internal lumens for contrast medium, which is injected into the lumen via the injection chamber catheter holder. | The device has a catheter with an internal lumens for contrast medium, which is injected into the lumen via the injection chamber catheter holder. | Different The proposed device offers non-injection models, which do |
{10}
创想医学
MEDNOVA
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 7 of 9)
| | The two-lumen device does not have an injection function. | | not increase the risk. |
| --- | --- | --- | --- |
| Use environment | This product is intended to be used in medical institutions, such as hospitals and clinics. | This product is intended to be used in medical institutions, such as hospitals and clinics. | Same |
| Work length | 200cm | 200cm | Same |
| Compatible endoscopy work channel | ≥3.2mm | ≥3.2mm | Same |
| device illustration | Figure 1 Product photo

Figure 2 Balloon diagram

Figure 3 sheathing anatomy | Figure 1 Product photo

Figure 2 Balloon diagram

Figure 3 sheathing anatomy | Same |
{11}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 8 of 9)
# 9. Performance Data
Bench testing was performed to support a determination of substantial equivalence. The Single use stone retrieval balloons performs as well as the predicate and is substantially equivalent to the predicate devices.
The following list of bench testing was conducted on the Single use stone retrieval balloons to determine the substantial equivalence with the predicates:
- Appearance Test
- Dimension Test
- Luer connector
- Flow rate
- Guidewire Compatibility
- Endoscope Compatibility
- inflatable volume
- Radio-detectability: The development ring shall be radio-detectable.
- Sheath Strength tension
- Balloon Strength tension
- Tip shaping
- Connection Strength Test
- sealing performance of connection part
- leakage
- Non-hydrated: The catheter should not manifest clinically significant hydration when subjected to an aqueous medium
- Corrosion resistance
- Performance of the Inflator pump
- Fast-exchange port performance
# 10. Sterilization
The proposed device is sold in a sterile package. The proposed device have been sterilized in a validated EO sterilization cycle. The EO residual was measured after sterilization of the device to meet the criteria defined in ISO 11135 Second edition 2014-07-15 "Sterilization of Health Care products Ethylene Oxide - Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices [Including: Amendment 1 (2018)]", and ISO 10993-7 Second Edition 2008-10-15 "Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)]".
# 11. Shelf Life
The shelf-life of 2 years has been validated through real-time aging testing and the
{12}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K253132(Page 9 of 9)
requirements on packaging for terminally sterilized medical device per ISO 11607-1 Second Edition 2019-02 and ISO 11607-2 Second Edition 2019-02 are also met. The test result can imply that the subject devices can provide and maintain a sterile barrier and its intended performance for at least the claimed shelf life.
## 12. Biocompatibility
The Biocompatibility testing was performed to show that all patient contacting materials meet applicable biocompatibility standards per ISO 10993-1:2018 and the FDA guidance: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process”.
The cytotoxicity, sensitization and intracutaneous reactivity were performed to show that the proposed devices are biocompatible.
## 13. Conclusions
There are no significant differences between the proposed device and the predicate device, the proposed device doesn't raise any new issues of safety and effectiveness. From a clinical perspective and comparing design specifications, the proposed device Single use stone retrieval balloons is substantially equivalent to
Micro-Tech(Nanjing)Co.,Ltd. currently marketed Sterile Biliary Stone Retrieval Balloon Catheter, Retrieval Balloon / short-wire compatible.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.