K253119 · Capnia, Inc. · CCJ · May 29, 2026 · Anesthesiology
Device Facts
Record ID
K253119
Device Name
CoSense ETCO Monitor
Applicant
Capnia, Inc.
Product Code
CCJ · Anesthesiology
Decision Date
May 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1430
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end-tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It can be used with the pediatric population. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation.
Device Story
CoSense ETCO Monitor is a portable, battery-operated device for measuring end-tidal carbon monoxide (ETCO) concentration in exhaled breath. Input consists of exhaled breath samples collected via a single-use nasal cannula. The device utilizes an infrared capnometer to identify the end-tidal portion of the breath and an electrochemical sensor to quantify CO concentration. The system features a touch screen interface for menu navigation, patient identification, and display of measurement results. It is intended for use by clinicians to monitor CO levels in conditions involving hemolysis, CO poisoning, or smoke inhalation, and for smoking cessation programs. The output provides quantitative CO data, assisting clinicians in assessing hemolysis rates or evaluating potential CO toxicity. The device is designed for clinical environments.
Clinical Evidence
Clinical evidence is supported by peer-reviewed studies written by clinical experts. No new prospective clinical trials were conducted; substantial equivalence is supported by non-clinical benchtop testing, usability testing, and existing clinical literature.
Technological Characteristics
Portable unit (246mm x 197mm x 68mm, 3.3lbs). Sensing: Infrared capnometer for end-tidal detection; electrochemical cell for CO measurement. Power: Li-Ion rechargeable battery. Patient interface: Non-DEHP PVC nasal cannula. Standards: ISO 10993 (biocompatibility), ASTM F1980-16/D4169-09 (shelf-life), IEC 62304 (software), IEC 60601-1/1-2 (electrical/EMC).
Indications for Use
Indicated for monitoring exhaled carbon monoxide (CO) from endogenous (e.g., hemolysis) and exogenous (e.g., CO poisoning, smoke inhalation) sources in pediatric and adult populations; also for smoking cessation programs and screening for CO poisoning/smoke inhalation.
Regulatory Classification
Identification
A carbon monoxide gas analyzer is a device intended to measure the concentration of carbon monoxide in a gas mixture to aid in determining the patient's ventilatory status. The device may use techniques such as infrared absorption or gas chromatography.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
Capnia, Inc.
% Rob Packard
Senior Regulatory Consultant
Medical Device Academy, Inc.
345 Lincoln Hill Rd
Shrewsbury, Vermont 05738
Re: K253119
Trade/Device Name: CoSense ETCO Monitor (C20112)
Regulation Number: 21 CFR 868.1430
Regulation Name: Carbon Monoxide Gas Analyzer
Regulatory Class: Class II
Product Code: CCJ
Dated: May 1, 2026
Received: May 1, 2026
Dear Rob Packard:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253119 – Rob Packard
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K253119 – Rob Packard
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253119 | |
| Device Name CoSense ETCO Monitor | |
| Indications for Use (Describe) The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end-tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It can be used with the pediatric population. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
# 510(k) SUMMARY – K253119
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
## I. SUBMITTER INFORMATION
21 CFR 807.92(a)(1)
### APPLICANT
Company Name: Capnia, Inc.
Address: 1101 Chess Drive
City, State, Zip: Foster City, CA 94404
Tel: +1.866.432.3788
Contact Name: Ed Baumann
Contact Email: ed.baumann@capnia.com
### CORRESPONDENT
Company Name: Medical Device Academy, Inc.
Address: 345 Lincoln Hill Road
City, State, Zip: Shrewsbury, VT 05738
Tel: +1.802.258.1881
Contact Name: Rob Packard, President
Contact Email: rob@fdaestar.com
Date Prepared: May 26, 2026
## II. NAME OF THE DEVICE
21 CFR 807.92(a)(2)
Device Trade Name: CoSense ETCO Monitor
Classification Name: Analyzer, gas, carbon-monoxide, gaseous-phase
Regulation: 21 CFR 868.1430
Regulatory Class: Class 2
Device Panel: Anesthesiology
Product
Classification Code: CCJ
## III. IDENTIFICATION OF THE PREDICATE
21 CFR 807.92(a)(3)
### Primary Predicate
Predicate Manufacturer: Capnia, Inc.
Predicate Trade Name: CoSense ETCO Monitor
Predicate 510(k): K151107
No secondary predicate or reference device was used in this submission.
510(k) Summary
{5}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
## IV. DESCRIPTION OF THE DEVICE
21 CFR 807.92(a)(4)
The CoSense® ETCO Monitor is a battery-operated carbon monoxide (CO) monitor. It uses an infrared capnometer to detect the end-tidal portion of the breath and an electrochemical carbon monoxide sensor to measure the end-tidal breath CO concentration. The device consists of a portable unit with software-controlled menu (date, time, patient identification, measurement time of monitoring), single-use nasal cannula, replaceable CO Sensor, and a battery charger/power supply.
The front panel of the CoSense device contains the ON/OFF button and a touch screen display which contains the controls and indicators. The front panel includes the menu controls, battery status, time, date, sensor status, and options to select patient information and measurement results.
## V. STATEMENT OF INTENDED USE
21 CFR 807.92(a)(5)
The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end-tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It can be used with the pediatric population. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation.
### INDICATIONS FOR USE COMPARISON
The indications for use for the subject device are expanded to include the use for pediatric population. The difference is the patient population, and it is justified by the clinical data.
## VI. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
21 CFR 807.92(a)(6)
### TECHNOLOGICAL COMPARISON
The subject device has the same technological characteristics (i. e., design, material, chemical composition, principle of operation, energy source etc.) as the predicate device as stated below:
- The dimensions of the subject device are 246mm x 197mm x 68mm that is the same as the predicate device.
- The weight of the subject device is 3.3lbs that is the same as the predicate device.
- The CO sensor cell type is electrochemical for both the subject device and the predicate device.
- The Cannula for both the subject and the predicate devices is made of Non-DEHP PVC.
- The battery type used for both the subject and the predicate devices is Li-Ion.
510(k) Summary
{6}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
## VII. SUBSTANTIAL EQUIVALENCE COMPARISON TABLE
21 CFR 807.92(a)(6)
The subject device CoSense ETCO Monitor is substantially equivalent to the predicate version. The following features are identical to those that previously received 510(k) clearance:
- Have the same intended use
- Use the same operating principle
- Incorporate the same basic design
- Incorporate the same materials
- Have the same shelf life
- Packaged using the same materials and processes
There are no changes in design characteristics from the previous device to the device in this application. The only change is the subject device’s indications for use, expanding to include the pediatric population. The difference is the patient population, and it is justified by the clinical data.
510(k) Summary
{7}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
Table 1: Comparison of CoSense ETCO Monitor (K253119) with CoSense ETCO Monitor (K151107).
| | Capnia CoSense ETCO Monitor (Subject Device) K253119 | Capnia CoSense ETCO Monitor (Predicate Device) K151107 | Justification for Differences |
| --- | --- | --- | --- |
| 510(k) Number | K253119 | K151107 | N/A |
| Manufacturer | Capnia, Inc. | Capnia, Inc. | Same as the predicate device. |
| Classification | Class II | Class II | Same as the predicate device. |
| Product Code | CCJ | CCJ | Same as the predicate device. |
| Regulation | 21 CFR 868.1430 | 21 CFR 868.1430 | Same as the predicate device. |
| Indications for Use | CoSense is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end-tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It can be used with the pediatric population. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation. | The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation. | The indications for use for the subject device are expanded to include the pediatric population. The difference is the patient population, and it is justified by the clinical data. |
| Patient Interface | Nasal cannula | Nasal cannula | Same as the predicate device. |
| Dimensions (LxWxH) | 246mm x 197mm x 68mm | 246mm x 197mm x 68mm | Same as the predicate device. |
| Weight | 3.3lbs | 3.3lbs | Same as the predicate device. |
| Materials | N/A | N/A | N/A |
| CO Sensor Cell Type | Electrochemical | Electrochemical | Same as the predicate device. |
| Cannula | Non-DEHP PVC | Non-DEHP PVC | Same as the predicate device. |
| Battery | Li-Ion | Li-Ion | Same as the predicate device. |
510(k) Summary
{8}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
| Performance Specifications | N/A | N/A | N/A |
| --- | --- | --- | --- |
| Accuracy | +/- 10% or +/-0.5ppm whichever is greater | +/- 10% or +/-0.5ppm whichever is greater | Same as the predicate device. |
| CO Measurement Range | 1.0 – 25.0ppm | 1.0 – 25.0ppm | Same as the predicate device. |
| Resolution | 0.1 ppm | 0.1 ppm | Same as the predicate device. |
| Breaths per Minute | 10 – 50 bpm | 10-50 bpm | Same as the predicate device. |
| Sample Collection Rate | 48 mL/min ± 2.0 mL/min | 48 mL/min ± 2.0 mL/min | Same as the predicate device. |
| Measurement Time | Less than 5 minutes | Less than 5 minutes | Same as the predicate device. |
| Sample collection | Collection of a normal breath using a disposable nasal cannula | Collection of a normal breath using a disposable nasal cannula | Same as the predicate device. |
| Modes | Expired | Expired | Same as the predicate device. |
| Device shelf life | 7 years | 7 years | Same as the predicate device. |
| CO Sensor shelf life | 15 months | 15 months | Same as the predicate device. |
| Cannula shelf life | 5 years / 60 months | 5 years / 60 months | Same as the predicate device. |
| Screen | LCD | LCD | Same as the predicate device. |
| Software/ Hardware | Analog and digital electronics with microprocessor | Analog and digital electronics with microprocessor | Same as the predicate device. |
| Rechargeable Battery | Yes | Yes | Same as the predicate device. |
| Power Source | Rechargeable Battery | Rechargeable Battery | Same as the predicate device. |
510(k) Summary
Page 5 of 7
{9}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
# VIII. BRIEF DISCUSSION OF NONCLINICAL TESTS
21 CFR 807.92(b)(1)
The following nonclinical testing was conducted to support the determination of substantial equivalence.
## Biocompatibility Testing
- Cytotoxicity Testing (MEM Elution) - ISO 10993-5
- Sensitization Testing (Guinea Pig Maximization) - ISO 10993-10
- Irritation Testing - ISO 10993-10
## Sterilization Validation
Not applicable, because the subject device and components are non-sterile and components are not intended to be sterilized by the user.
## Shelf-life
Accelerated age testing in accordance with:
- ASTM F1980-16
- ASTM D4169-09
## Software Validation, Cybersecurity, Wireless Safety, and Interoperability
Software validation was conducted in accordance with:
- IEC 62304
Cybersecurity, wireless safety, and interoperability testing are not applicable to the subject device.
## Electrical Safety & EMC Testing
- IEC 60601-1-2
- IEC 60601-1
## Performance Testing (non-clinical benchtop):
- Non-clinical performance testing was performed to verify CO measurement accuracy—including the revised indications for use.
- Usability Testing was performed to validate that usability risk controls implemented are effective for the revised indications for use.
## Performance Testing (animal studies):
Not Applicable, because the device does not require animal studies to demonstrate performance.
# IX. BRIEF DISCUSSION OF CLINICAL TESTS
21 CFR 807.92(b)(2)
The clinical testing is supported by peer-reviewed studies written by clinical experts.
510(k) Summary
{10}
Capnia, Inc.
CoSense ETCO Monitor 510(k) Submission
## X. CONCLUSIONS FROM NONCLINICAL AND CLINICAL TESTS
21 CFR 807.92(b)(3)
The subject device, CoSense ETCO Monitor (K253119), has equivalent indications for use when compared with the predicate device, CoSense ETCO Monitor (K151107). The subject device and the predicate device have the same technological characteristics. Both devices passed the same safety and performance testing. Non-clinical benchtop testing shows that the subject device is substantially equivalent in safety and effectiveness to the predicate device and performs as well as the predicate device.
510(k) Summary
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.