K253116 · The Surgeon General, Department of the Army (TSG-DA) · GXN · Jun 18, 2026 · Neurology
Device Facts
Record ID
K253116
Device Name
3D MAC Titanium Cranial Plate (TCP) System
Applicant
The Surgeon General, Department of the Army (TSG-DA)
Product Code
GXN · Neurology
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 3D MAC Titanium Cranial Plate (TCP) System is a patient-specific device that is indicated for use to replace bony voids in the cranial and/or craniofacial skeleton (e.g., frontal bone, temporal bone, occipital bone, nasal bone, parietal bone, supraorbital process, lacrimal bone, zygomatic bone, sphenoid bone, ethmoid process, vomer).
Device Story
System comprises patient-specific Titanium Cranial Plate (TCP) and Skull Anatomical Model (SAM); used in cranioplasty surgery to replace bony voids. Input: patient CT scan data. Process: SAM printed via stereolithography (SLA) from CT data; TCP designed from SAM model and printed via electron beam melting (EBM) using Ti6Al4V powder. Output: custom-fit titanium implant and physical anatomical model. Used in clinical setting by neurosurgeons; SAM serves as surgical reference tool for fit-check and positioning of TCP. TCP fixed to skull using commercially available screws. Benefits: provides patient-specific reconstruction of cranial/craniofacial defects.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-5, -10, -23, -11, -6, -18; ASTM F756), static and dynamic mechanical testing, and MR compatibility. Design validation performed via mock implantation by neurosurgeons using three cases of varying complexity; all successfully completed with acceptable results.
Technological Characteristics
Implant: Ti6Al4V titanium alloy (ASTM F2924) via electron beam melting (EBM). SAM: clear photopolymer resin via stereolithography (SLA). Non-load bearing, non-alterable. Provided non-sterile; user-sterilized via steam. Fixation via commercially available screws.
Indications for Use
Indicated for patients requiring replacement of bony voids in the cranial and/or craniofacial skeleton.
Regulatory Classification
Identification
A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
Predicate Devices
BioArchitects Patient Specific Cranial/Craniofacial Plate (K151692)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
The Surgeon General, Department of the Army (TSG-DA)
India El
Regulatory Affairs Scientist
DHA R&D - Medical Research and Development Command; ATTN:
FCMR-ORA, 1541 Porter Street
Fort Detrick, Maryland 21702-9247
Re: K253116
Trade/Device Name: 3D MAC Titanium Cranial Plate (TCP) System
Regulation Number: 21 CFR 882.5330
Regulation Name: Preformed Nonalterable Cranioplasty Plate
Regulatory Class: Class II
Product Code: GXN
Dated: May 18, 2026
Received: May 19, 2026
Dear India El:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253116 - India El
Page 2
(https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253116 - India El
Page 3
Sincerely,
JULIA E.
SLOCOMB -
S
Digitally signed by
JULIA E. SLOCOMB -S
Date: 2026.06.18
10:05:40 -04'00'
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253116
Device Name
3D MAC Titanium Cranial Plate (TCP) System
Indications for Use (Describe)
The 3D MAC Titanium Cranial Plate (TCP) System is a patient-specific device that is indicated for use to replace bony voids in the cranial and/or craniofacial skeleton (e.g., frontal bone, temporal bone, occipital bone, nasal bone, parietal bone, supraorbital process, lacrimal bone, zygomatic bone, sphenoid bone, ethmoid process, vomer).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) Summary
## Contact Information
| **Applicant Information:** | The Surgeon General, Department of the Army (TSG-DA) |
| --- | --- |
| **Applicant Address:** | Defense Health Agency Research & Development; ATTN: FCMR-ORA 1541 Porter Street Fort Detrick, MD 21702-9247 |
| **Primary Contact:** | India El, MS, BS Regulatory Affairs Scientist Office of Regulated Activities Defense Health Agency Research & Development 301-619-8296 india.o.el.ctr@health.mil |
| **Alternate Contact:** | Peter Liacouras, PhD Director of Services, 3D Medical Applications Center 301-295-8592 peter.c.liacouras.civ@health.mil |
| **Date Prepared:** | June 16, 2026 |
## Device Information
| **Trade Name:** | 3D MAC Titanium Cranial Plate System |
| --- | --- |
| **Common Name:** | Cranial Plate |
| **Classification Name:** | Preformed nonalterable cranioplasty plate |
| **Classification:** | 21 CFR 882.5330 |
| **Product Code:** | GXN |
## Predicate Device Information
| **Predicate Device Name:** | BioArchitects Patient Specific Cranial/Craniofacial Plate (K151692) |
| --- | --- |
| **Classification:** | 21 CFR 882.5330 |
| **Product Code:** | GXN |
## Device Description
The 3D MAC Titanium Cranial Plate (TCP) System includes the TCP Implant and TCP Skull Anatomical Model (SAM). The TCP System is indicated for use in cranioplasty surgery to replace bony voids in the cranial and/or craniofacial skeleton and is patient-specific. The TCP Implant replaces the bony void in the skull and the SAM is provided as a surgical reference tool to aid in the fitting and positioning of the TCP Implant. The TCP Implant is manufactured on an electron beam melting (EBM) 3D printer out of Ti6Al4V titanium alloy powder. The SAM is printed on a stereolithography (SLA) 3D printer out of clear photopolymer resin. The TCP Implant is an implantable, non-load bearing, and non-alterable medical device. The TCP Implant
3D MAC Titanium Cranial Plate System
Page 1 of 4
{5}
is individually-sized and designed specifically for each patient case to replace the bony void of the patient's cranium. The TCP Implant is initially designed from a 3D Model of the SAM. The SAM is designed from the patient's CT scan and utilized as a fit check for the TCP Implant.
### **Indications for Use Statement**
The 3D MAC Titanium Cranial Plate (TCP) System is a patient-specific device that is indicated for use to replace bony voids in the cranial and/or craniofacial skeleton (e.g., frontal bone, temporal bone, occipital bone, nasal bone, parietal bone, supraorbital process, lacrimal bone, zygomatic bone, sphenoid bone, ethmoid process, vomer).
### **Predicate Device Comparison**
| | **Subject Device (K253116) 3D MAC Patient-Specific TCP** | **Predicate Device (K151692) BioArchitects Patient-Specific Cranial/Craniofacial Plate** |
| --- | --- | --- |
| **Classification** | II | II |
| **Regulation** | 882.5330 | 882.5330 |
| **Product Code** | GXN | GXN |
| **Indications for Use Statement** | The 3D MAC Titanium Cranial Plate (TCP) System is a patient-specific device that is indicated for use to replace bony voids in the cranial and/or craniofacial skeleton (e.g., frontal bone, temporal bone, occipital bone, nasal bone, parietal bone, supraorbital process, lacrimal bone, zygomatic bone, sphenoid bone, ethmoid process, vomer). | The BioArchitects Patient Specific Cranial/Craniofacial Plate implant device is intended to replace bony voids in the cranial and/or craniofacial skeleton (e.g., frontal bone, temporal bone, occipital bone, nasal bone, parietal bone, supraorbital process, lacrimal bone, zygomatic bone, sphenoid bone, ethmoid process, vomer). It is a patient specific device. |
| **Material** | Ti6Al4V per ASTM F2924 | Ti6Al4V ELI per ASTM F3001-14 |
| **Manufacturing Technology** | Additive manufacturing: Electron Beam Melting and Vat Photopolymerization | Additive manufacturing: Electron Beam Melting |
| **Anatomical Model Included?** | Yes | Not Specified |
| **Fixation to Skull** | Commercially available screw systems | Commercially available screw systems |
| **Sterilization** | Provided non-sterile, user sterilized | Provided non-sterile, user sterilized |
### **Predicate K151692 Device Comparison**
The intended use of the subject device 3D MAC TCP System and predicate K151692 device are the same (replacing voids in the cranial and craniofacial skeleton). The subject device and predicate K151692 device have the following same technological characteristics:
- Method of manufacturing (Implant): Additive manufacturing using powder bed fusion technology
- Method of design: Patient-specific design that is approved by the surgeon before implantation
- Method of fixation: Commercially available screws
- Sterilization: Provided non-sterile, will be steam sterilized by end user
3D MAC Titanium Cranial Plate System
Page 2 of 4
{6}
The subject and predicate devices have the following different technological characteristics:
- Material: Ti6Al4V per ASTM F2924 (Subject) and Ti6Al4V ELI per ASTM F3001-14 (Predicate)
This difference in technological characteristics does not raise different questions of safety and effectiveness because the material alloy differences can be evaluated using biocompatibility and performance testing.
## Summary of Performance Testing
Multiple nonclinical tests were performed to determine the substantial equivalence of the subject device 3D MAC TCP System. The following tests are listed below:
- Biocompatibility testing
- Static mechanical testing
- Dynamic mechanical testing
- Design validation testing
- Magnetic Resonance (MR) compatibility testing (for labeling)
## Biocompatibility Testing
The 3D MAC TCP System comes in contact with tissue/bone and is implanted. The 3D MAC TCP System was evaluated for biocompatibility endpoints and the following tests and results are shown in the following table:
| Test | Conclusion |
| --- | --- |
| Cytotoxicity (ISO 10993-5) | Non-cytotoxic |
| Sensitization (ISO 10993-10) | Non-sensitizing |
| Irritation (ISO 10993-23) | Non-irritating |
| Acute Systemic Toxicity (ISO 10993-11) | Non-toxic |
| Material-Mediated Pyrogenicity (ISO 10993-11) | Non-pyrogenic |
| Implantation Effects (ISO 10993-6) | No unexpected results |
| Chemical Characterization (ISO 10993-18) | Acceptable margin of safety for all reported extractable substances |
| Hemolysis (ASTM F756) | Non-hemolytic |
## Static and Dynamic Mechanical Testing
Static and dynamic mechanical tests were performed to support the substantial equivalence of the 3D MAC TCP System. A static compression test was performed and the 3D MAC TCP System was shown to be able to withstand loads expected in a clinical setting. A dynamic impact test was performed and all 3D MAC TCP System samples passed the acceptance criteria.
3D MAC Titanium Cranial Plate System
Page 3 of 4
{7}
### Design Validation Testing
A design validation was performed to validate the design and manufacturing of the TCP System by trained neurosurgeons in a mock implantation. Three different engineer designers, three different cases of differing complexities, and three different neurosurgeons were involved in the design validation testing. The neurosurgeons were able to successfully complete the mock implantation design validation, and they found the TCP System acceptable and meets their predefined user needs.
### MR Compatibility Testing (for MR Conditional Labeling)
MR compatibility testing was performed to determine the conditions under which a patient implanted with the TCP System could safely undergo an MRI scan. The MR Conditional language is included in the labeling. This testing was performed based on FDA’s guidance document, “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment.”
### Conclusion
Based on the comparison of the intended use and results of the performance testing, the 3D MAC TCP System is substantially equivalent to the predicate device.
3D MAC Titanium Cranial Plate System
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.