The VEOFIX® snap-off screw is indicated for fixation of bone fractures or for bone reconstruction, including : - Fixation of small bone fragments - Weil osteotomy - Mono-cortical fixation - Osteotomies and fractures fixation in the foot
Device Story
VEOFIX snap-off screw is a metallic bone fixation fastener used for bone fracture fixation and reconstruction in the foot. The device consists of a cartridge and an implantable screw; it is used with a handle and adapter. The screw features a self-drilling and self-tapping design; the head separates from the cartridge upon contact with bone. The device is intended for permanent implantation. Surgeons use the device to stabilize bone fragments to facilitate fusion. It is provided sterile and is MR Conditional.
Clinical Evidence
Bench testing only. Performance testing included pullout strength, driving/removal torques, and torsional resistance compared against the predicate and FDA guidance criteria. Biocompatibility evaluated per ISO 10993-1; sterility validated per ISO 11137-1; endotoxin testing per USP <85>.
Indicated for fixation of bone fractures or bone reconstruction in the foot, including small bone fragments, Weil osteotomy, and mono-cortical fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
June 5, 2026
Steps Ortho
% Christine Scifert
Partner
MRC Global, LLC
9085 East Mineral Circle
Suite 110
Centennial, Colorado 80112
Re: K253108
Trade/Device Name: VEOFIX Snap Off Screw
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC
Dated: September 17, 2025
Received: September 24, 2025
Dear Christine Scifert:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253108 - Christine Scifert
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253108 - Christine Scifert
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253108 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| VEOFIX snap off screw | | |
| Please provide your Indications for Use below. | | ? |
| The VEOFIX® snap-off screw is indicated for fixation of bone fractures or for bone reconstruction, including : - Fixation of small bone fragments - Weil osteotomy - Mono-cortical fixation - Osteotomies and fractures fixation in the foot | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Page 1 of 1
{4}
STEPS ORTHO
VEOFIX® snap-off screw – Traditional 510(k)
## 510(k) Summary
Prepared: June 5, 2026
I. SUBMITTER
STEPS ORTHO
3, allée des Ginkgos
69500 BRON - FRANCE
Phone: +33(0)4 28 89 81 15
Contact: Morgane GRENIER
Quality and Regulatory Affairs Director
Phone: + 33 (0) 4 28 89 81 15
e-mail: m.grenier@stepsortho.com
II. DEVICE
| Name of Device: | VEOFIX® snap-off screw |
| --- | --- |
| Common/Usual Name: | Screw, Fixation, Bone |
| Regulation Numbers: | 21 CFR 888.3040 |
| Regulation Name: | Smooth or threaded metallic bone fixation fastener |
| Regulatory Class: | II |
| Product Codes: | HWC, Screw, fixation, bone |
III. PREDICATE DEVICE
| Primary Predicate | SPIN® snap-off screw K011946 |
| --- | --- |
IV. DEVICE DESCRIPTION
The VEOFIX® snap-off screw is manufactured in TA6V (the composition is compliant with ISO 5832-3 standard).
The VEOFIX® snap-off screw has two distinct parts, the cartridge and the implantable screw. The cartridge can be combined with the adapter using a handle.
The parts will separate when the head of the screw is in contact with the bone
The VEOFIX® Snap-off screw is available in 4 sizes of the same diameter Ø 2mm (11 mm, 12 mm, 13 mm and 14 mm).
Page 1 / 3
K253108
{5}
STEPS ORTHO
VEOFIX® snap-off screw – Traditional 510(k)
V. INDICATIONS FOR USE
The VEOFIX® snap-off screw is indicated for fixation of bone fractures or for bone reconstruction, including :
- Fixation of small bone fragments
- Weil osteotomy
- Mono-cortical fixation
- Osteotomies and fractures fixation in the foot
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
VEOFIX® snap-off screw and SPIN® snap-off screw are both intended to be implanted for fixation of bone fractures or for bone reconstruction. VEOFIX® snap-off screw has the same technological characteristics to the predicate SPIN® snap-off screw. Specifically, both devices have the same design (self-drilling and self-tapping snap-off screw) and are placed permanently in the foot. Both subject device and predicate device are comprised of Titanium Alloy which further support the substantial equivalence. The additional diameter available for the SPIN® snap-off screw do not raise new types of questions of safety or effectiveness either, as both devices are designed to offer the surgeon to accommodate individual patient anatomy. The key questions of safety and effectiveness for both devices are the same, i.e., whether the implant can be correctly implanted, whether the implant can perform the stabilization and attachment of the bones to each other to achieve bone fusion.
In addition, bench testing confirms that the technological characteristics of VEOFIX® snap-off screw is appropriate for its intended use and functions as intended. Thus, a conclusion of substantial equivalence is supported.
VII. PERFORMANCE TESTING
The following performance testing was conducted to support substantial equivalence. Torsional yield strength testing was compared against the predicate device. Pullout strength, driving and removal torque testing was compared against the FDA guidance document, “Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway”.
Performance Testing
Mechanical and functional testing was performed to verify the design of the implant, including:
- Pullout strength
- Driving and removal torques
- Torsional resistance
Biocompatibility
Biocompatibility testing was conducted in accordance with NF EN ISO 10993-1 :2020 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and NF EN ISO 14971 :2019 Medical devices - Application of risk management to medical devices.
Page 2 / 3
K253108
{6}
STEPS ORTHO
VEOFIX® snap-off screw – Traditional 510(k)
# Sterility
Sterilization validation testing was performed in accordance with NF EN ISO 11137-1:2016
“Sterilization of health care products – Radiation – Part 1: Requirements for development,
validation and routine control of a sterilization process for medical devices.”
# MR Compatibility
VEOFIX® snap-off screw has been evaluated for safety in the magnetic resonance environment and
has been determined to be MR Conditional.
STEPS ORTHO implant can be scanned safely after implantation under specific conditions
mentioned in the Instructions for Use.
# Pyrogenicity
Endotoxin testing was completed per USP Chapter <85>.
# VIII. CONCLUSIONS
The differences that exist between VEOFIX® snap-off screw and its predicate do not raise different
questions of safety or effectiveness. The results of performance testing demonstrate that VEOFIX®
snap-off screw will perform safely as intended and support a determination of substantial
equivalence to the predicate device which is marketed for the same intended use.
Page 3 / 3
K253108
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.