AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02)
K253080 · Suzhou and Science & Technology Development Corp. · HSB · Jun 18, 2026 · Orthopedic
Device Facts
Record ID
K253080
Device Name
AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02)
Applicant
Suzhou and Science & Technology Development Corp.
Product Code
HSB · Orthopedic
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AND Medical Femoral Nail System is Suitable for intramedullary fixation of femoral fractures. Indications for use of the PFNA02(Length 170 mm - 240 mm) include: - Pertrochanteric fractures; - Intertrochanteric fractures; - High subtrochanteric fractures. Indications for use of the PFNA01 (Length 320 mm - 420 mm), include: - Low and extended subtrochanteric fractures; - Ipsilateral trochanteric fractures; - Combination fractures (in the proximal femur); - Pathological fractures.
Device Story
Intramedullary fixation system for femoral fractures; consists of titanium alloy (Ti6-Al4-V) nails, lock nails, blade nails, and extended caps. System divided into PFNA01 (320-420 mm) and PFNA02 (170-240 mm) variants based on fracture site and surgical approach. Delivered non-sterile; requires surgical implantation by orthopedic surgeons. Device provides mechanical stabilization of femoral fractures to facilitate bone healing. Performance verified via mechanical bench testing (bending, torsion, pullout, cut-out) per ASTM standards.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance verified through static/dynamic four-point bending, torsion, three-point bending, insertion/removal, pullout, cantilever bending, and cut-out tests per ASTM F1264, F543, and F384 standards.
Technological Characteristics
Material: Ti-6Al-4V (ISO 5832-3:2016). Components: Intramedullary nail, lock nail, blade nail, extended cap. Form factor: Left/right specific, varying lengths/diameters. Mechanical testing per ASTM F1264-16e1, ASTM F543-17, ASTM F384-17. Non-sterile delivery.
Indications for Use
Indicated for patients requiring intramedullary fixation of femoral fractures, including pertrochanteric, intertrochanteric, high subtrochanteric, low/extended subtrochanteric, ipsilateral trochanteric, combination proximal femur, and pathological fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Suzhou And Science & Technology Development Corp.
Limin Zhu
Regulatory Affairs
#2 Ansheng Rd., Hexing Jinfeng Town
Zhangjiagang, Jiangsu 215625
China
Re: K253080
Trade/Device Name: AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02)
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB, HWC
Dated: May 22, 2026
Received: May 22, 2026
Dear Limin Zhu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253080 - Limin Zhu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253080 - Limin Zhu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**FARZANA SHARMIN -S**
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K253080
Device Name
AND Medical Femoral Nail System (PFNA01);
AND Medical Femoral Nail System (PFNA02)
Indications for Use (Describe)
AND Medical Femoral Nail System is Suitable for intramedullary fixation of femoral fractures.
Indications for use of the PFNA02(Length 170 mm - 240 mm) include:
- Pertrochanteric fractures;
- Intertrochanteric fractures;
- High subtrochanteric fractures.
Indications for use of the PFNA01 (Length 320 mm - 420 mm), include:
- Low and extended subtrochanteric fractures;
- Ipsilateral trochanteric fractures;
- Combination fractures (in the proximal femur);
- Pathological fractures.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K253080 (Page 1 of 2)
510(k) #: K253080
# 510(k) Summary
Prepared on: 2026-06-18
## Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Suzhou AND Science & Technology Development Corp. |
| --- | --- |
| Applicant Address | No.2 Ansheng Road,Hexing Jinfeng Town,Zhangjiagang City,Jiangsu,China Zhangjiagang Jiangsu 215625 China |
| Applicant Contact Telephone | +86 15190928088 |
| Applicant Contact | Mr. zhu limin |
| Applicant Contact Email | Zhulimin@andtosi.com |
## Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02) |
| --- | --- |
| Common Name | Intramedullary fixation rod |
| Classification Name | Rod, Fixation, Intramedullary And Accessories |
| Regulation Number | 888.3020 |
| Product Code(s) | HSB, HWC |
## Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K172830 | Double Medical Femoral Nail System | HSB |
## Device Description Summary
21 CFR 807.92(a)(4)
AND Medical Femoral Nail System is made of Ti6-Al4-V material that complies with the ISO 5832-3:2016 standard. Interlocking intramedullary nails are classified into PFNA01 and PFNA02 types based on their structural type, usage site, and surgical approach. Among them, PFNA01 and PFNA02 are composed of nail, lock nail, blade nail, and extended cap. The specifications of the product are diameter x length, and the specifications of PFNA01 and PFNA02 are also divided into left/right. The product is delivered in a non-sterile state.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
AND Medical Femoral Nail System is Suitable for intramedullary fixation of femoral fractures.
Indications for use of the PFNA02(Length 170 mm - 240 mm) include:
- - Pertrochanteric fractures;
- - Intertrochanteric fractures;
- - High subtrochanteric fractures.
Indications for use of the PFNA01 (Length 320 mm - 420 mm), include:
- - Low and extended subtrochanteric fractures;
- - Ipsilateral trochanteric fractures;
- - Combination fractures (in the proximal femur);
- - Pathological fractures.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
{5}
K253080 (Page 2 of 2)
The device has same indications for use and material in comparison to the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
The predict device contains two patient contact materials, Ti-6Al-4V following ISO 5832-3 or Ti-6Al-4V Eli following ASTM F 136. There is only one material of the declared product, Ti-6Al-4V following ISO 5832-3. The titanium alloy used is determined to be biocompatible and will not cause different questions of safety and effectiveness.
The proposed device is mainly different in dimension with the predicate device, but the mechanical test demonstrated the results of both devices are very similar.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Non-Clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device conforms to ASTM F1264-16e1 Standard Specification and Test Methods for Intramedullary Fixation Devices, ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws, ASTM F384-17 Standard Specifications and Test Methods for Metallic Angled Orthopedic Fracture Fixation Devices, including:
- ☑ Static Four-point Bending Test of Nail
- ☑ Dynamic Four-point Bending Test of Nail
- ☑ Static Torsion Test of Nail
- ☑ Static Three-point Bending Test of Locking Screw
- ☑ Dynamic Three-point Bending Test of Locking Screw
- ☑ Insertion/removal Test of Locking Screw
- ☑ Pullout Test of Locking Screw
- ☑ Torsion Test of Locking Screw.
- ☑ Cantilever Bending Test
- ☑ Cut-out Test
Non-Clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
The test results demonstrated that the proposed device conforms to ASTM F1264-16e1 Standard Specification and Test Methods for Intramedullary Fixation Devices, ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws, ASTM F384-17 Standard Specifications and Test Methods for Metallic Angled Orthopedic Fracture Fixation Devices, that demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device (K172830) identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.