Retrospective patient clinical data (images and treatment plans)
Retrospective clinical data was used to validate the performance of the device's image registration, synthetic CT generation, contour propagation, and dose calculation algorithms against predefined acceptance criteria.
ART.1-US is a software as a medical device intended to support the treatment re-plan process for cancer patients receiving radiation therapy. ART.1-US calculates delivered dose and contours based on cancer patients’ daily volumetric images, in addition to displaying and reporting relevant metrics. ART.1-US allows for the comparison of dosimetry between the delivery of each fraction and the treatment plan in radiation oncology. ART.1-US is intended for use by trained medical professionals as a decision support tool to aid the treatment re-plan process. ART.1-US is indicated for adult Head & Neck cancer patients undergoing radiation therapy.
Device Story
ART.1-US is a SaMD for adaptive radiation therapy; supports treatment re-planning by calculating delivered dose and contours using daily patient volumetric images. Inputs: DICOM CT (planning), structure sets, dose, radiation plan, CBCT (on-treatment), and SRO (spatial transformation) files. Operation: Backend image registration module calculates Deformable Vector Field (DVF) to map planning CT to daily CBCT, creating synthetic CT; propagates clinician-drawn contours via DVF. Dose calculation module uses Collapsed Cone Convolution on synthetic CT to compute delivered dose. Outputs: Delivered dose, daily contours, dosimetric analytics, dose deviation indicators, and summary reports. Used in clinical radiation oncology settings by trained professionals (physicians, physicists, dosimetrists). Output serves as decision support for offline re-planning; helps clinicians assess dose delivery accuracy and potential need for plan adaptation.
Clinical Evidence
No prospective clinical studies. Evidence based on bench testing using phantoms and retrospective clinical data. Performance metrics: Median Target Registration Error (TRE) < 3mm; median gamma pass rate ≥ 90% (3%/2mm criteria); mean DSC ≥ 0.8 or within 0.05 of inter-rater; mean ASSD ≤ 3mm or within 1mm of inter-rater. Retrospective analysis confirmed dosimetric differences within 5% of prescription dose for ≥ 80% of cases. Usability testing per IEC 62366-1 passed.
Technological Characteristics
SaMD; backend image registration (DVF-based) and dose calculation (Collapsed Cone Convolution). Inputs: DICOM CT, CBCT, RT structure sets, dose, and plan. Outputs: Synthetic CT, propagated contours, dosimetric analytics. Software developed per IEC 62304; cybersecurity per AAMI TIR57. No AI/ML-based segmentation used. Standalone software package.
Indications for Use
Indicated for adult Head & Neck cancer patients undergoing radiation therapy. Not for pediatric patients or patients outside the specified cohort. Incompatible with non-CT planning data, non-CBCT daily treatment data, and non-photon treatments.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2026
SeeTreat Pty Ltd
Eric Qin
VP of Regulatory and Clinical Affairs
320 Pitt Street
Sydney, NSW 2000
Australia
Re: K253078
Trade/Device Name: ART.1-US
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: MUJ
Dated: April 1, 2026
Received: April 1, 2026
Dear Eric Qin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253078 - Eric Qin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253078 - Eric Qin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253078 | |
| Device Name ART.1-US | |
| Indications for Use (Describe) ART.1-US is a software as a medical device intended to support the treatment re-plan process for cancer patients receiving radiation therapy. ART.1-US calculates delivered dose and contours based on cancer patients’ daily volumetric images, in addition to displaying and reporting relevant metrics. ART.1-US allows for the comparison of dosimetry between the delivery of each fraction and the treatment plan in radiation oncology. ART.1-US is intended for use by trained medical professionals as a decision support tool to aid the treatment re-plan process. ART.1-US is indicated for adult Head & Neck cancer patients undergoing radiation therapy. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SeeTreat
# 510(k) summary
# K253078
# 1.0 Submitter details
| Submitter name: | SeeTreat Pty Ltd |
| --- | --- |
| Submitter address: | 320 Pitt Street, Sydney, NSW, 2000, Australia |
| Contact person: | Eric Qin |
| Date summary prepared: | 1 April 2026 |
| Contact number: | +61 413 142 856 |
# 2.0 Device information
# 2.1 Regulatory information
| Regulation number | Device type | Device Class | Product code | Classification panel |
| --- | --- | --- | --- | --- |
| 21 CFR 892.5050 | Medical Charged-Particle Radiation therapy System | Class II | MUJ | Radiology |
# 2.2 Device name
ART.1-US
# 2.3 Intended use
ART.1-US is a software as a medical device intended to support the treatment re-plan process for cancer patients receiving radiation therapy.
ART.1-US calculates delivered dose and contours based on cancer patients' daily volumetric images, in addition to displaying and reporting relevant metrics. ART.1-US allows for the comparison of dosimetry between the delivery of each fraction and the treatment plan in radiation oncology.
ART.1-US is intended for use by trained medical professionals as a decision support tool to aid the treatment re-plan process.
# 2.4 Indications for use
ART.1-US is a software as a medical device intended to support the treatment re-plan process for cancer patients receiving radiation therapy.
ART.1-US calculates delivered dose and contours based on cancer patients' daily volumetric images, in addition to displaying and reporting relevant metrics. ART.1-US allows for the comparison of dosimetry between the delivery of each fraction and the treatment plan in radiation oncology.
ART.1-US is intended for use by trained medical professionals as a decision support tool to aid the treatment re-plan process.
ART.1-US is indicated for adult Head & Neck cancer patients undergoing radiation therapy.
Page 1 of 8
{5}
SeeTreat
## 2.5 Contraindications
ART.1-US is not intended to be used:
- for patients outside the specified cohorts (non-Head & Neck cancer patients)
- Paediatric Patients
ART.1-US is incompatible with:
- Non-CT planning data
- Non-CBCT (Cone Beam Computed Tomography) daily treatment data
- Non-photon treatments
## 2.6 Device description
ART.1-US is a software as a medical device (SaMD) that streamlines adaptive radiation therapy by automatically calculating the delivered dose using the patient's daily anatomical images.
ART.1-US consists of a single software package with two main subcomponents:
- User interface
- Backend.
The backend houses key algorithms which are responsible for image registration and dose calculations.
### 2.6.1 Image registration module
Image registration module in ART.1-US calculates a Deformable Vector Field (DVF) that maps the planning CT image to the CBCT image taken on the treatment day to create a synthetic CT.
The synthetic CT will have the resolution and Field of View of the CT whilst representing the anatomy of the CBCT image.
ART.1-US propagates the contours of the relevant structures by mapping the contours manually drawn by physicians and stored in the DICOM structure set using the same DVF that maps from CT to CBCT. These propagated contours can be overlaid upon the dose to facilitate subsequent dose metrics calculation for defined structures.
### 2.6.2 Dose calculation module
The dose calculation module is responsible for calculating the dose delivered to the patient on the day using the Collapsed Cone Convolution technique. It uses the synthetic CT generated from image registration module as well as the treatment plan parameters specified in the DICOM radiation therapy plan.
The dose-volume histogram of each relevant structure and dose metrics as specified by the clinic are calculated by ART.1-US using the propagated contours and the delivered total dose.
### 2.6.3 Device inputs and outputs
Device inputs:
- DICOM CT image (planning image): the patient CT used for planning
- DICOM Radiotherapy Structure sets: contours used for planning
Page 2 of 8
{6}
Page 3 of 8
# SeeTreat
- DICOM Radiotherapy Dose: planned dose
- DICOM Radiotherapy Plan: radiation parameters to deliver the planned Dose
- DICOM Cone Beam CTs: patient on-treatment reconstructed cone beam CTs
- DICOM SRO: an IGRT file representative of spatial transformations for the patient position at treatment delivery
## Device outputs:
- Delivered dose
- Daily structure contours
- Dosimetric analytics for configured structures
- Dose deviation indicator of each structure’s result based on configuration settings (comparison between delivered and planned dose)
- Reports summary of all the above
## 3.0 Predicate device
| Predicate device name: | ART-Plan |
| --- | --- |
| Manufacturer: | TheraPanacea |
| Regulation number: | 21 CFR 892.5050 |
| Classification: | Class II |
| Product code: | MUJ |
| Classification panel: | Radiology |
| S10(k) clearance: | K232479 |
## 4.0 Summary of substantial equivalence
| | Proposed device (ART.1-US) | Predicate device (ART-Plan) | Comparison |
| --- | --- | --- | --- |
| Regulation name | Medical Charged-Particle Radiation Therapy System | Medical Charged-Particle Radiation Therapy System | Equivalent |
| Regulation Number | 21 CFR 892.5050 | 21 CFR 892.5050 | Equivalent |
| Device Classification | Class II | Class II | Equivalent |
| Classification Product Code | MUJ | MUJ | Equivalent |
| Intended use & Indications for use | ART.1-US is a software as a medical device intended to support the treatment re-plan process for cancer patients receiving radiation therapy. ART.1-US calculates delivered dose and contours based on cancer patients’ daily volumetric images, in addition to displaying and reporting relevant metrics. ART.1-US allows for the comparison of dosimetry | ART-Plan is a software intended to be used by trained clinicians who are familiar with radiation therapy, such as medical physicists, medical dosimetrists and radiation oncologists. The software consists of different applications, each used for specific purposes at a different phase of radiation treatment planning. ART-Plan offers the following tools to aid in the | Similar: both the proposed device and the predicate are intended to provide outputs which can be used to support offline treatment re-plan decisions. There are minor differences in |
{7}
SeeTreat
| | between the delivery of each fraction and the treatment plan in radiation oncology. ART.1-US is intended for use by trained medical professionals as a decision support tool to aid the treatment re-plan process. ART.1-US is suitable for adult Head & Neck cancer patients undergoing radiation therapy. | workflow of radiotherapy treatment: • Multi-modal visualization and rigid- and deformable registration of anatomical and functional images such as CT, MR, PET-CT, 4D-CT, CBCT and synthetic-CT generated from CBCT • Display of fused and non-fused images to facilitate the comparison and delineation of image data by the user • Manual generation, modification and semi-automatic generation of contours for the regions of interest • Automatic generation of contours for organs at risk and healthy lymph nodes, based on medical practices, on medical images such as CT and MR images • Generation of synthetic-CT from MR images for supported anatomies • Generation of synthetic-CT from CBCT images for supported anatomies • Dose computation on CT and/or synthetic-CT images for external beam irradiation with photon beams • Assisted CBCT-based off-line adaptation decision-making for supported anatomies The device is intended to be used in a radiation therapy clinical setting, by trained | the supported image formats. Both devices are intended for professional use only. |
| --- | --- | --- | --- |
Page 4 of 8
{8}
SeeTreat
| | | professionals only. | |
| --- | --- | --- | --- |
| Target population | Adult cancer patients | Adult cancer patients | Equivalent |
| Intended User | Trained clinicians | Trained clinicians | Equivalent |
| Key device inputs | CT, CBCT | CT, CBCT, MR, PET-CT, 4D-CT | Similar: both devices support CT and CBCT as inputs. There are some minor differences in the support of other image file formats. |
| Technological characteristics | Image registration/contouring: Propagates existing clinician-drawn contours using Rigid and Deformable image registration for the purposes of replanning/recontouring. Dose calculation: Dose computation on synthetic-CT images for external beam irradiation with photon beams. | Image registration/contouring: Rigid and deformable automatic and manual initialization registration for the purposes of replanning/recontouring. AI/Deep-learning based automatic segmentation of OARs and healthy lymph nodes. Dose calculation: Dose computation on CT and/or synthetic-CT images for external beam irradiation with photon beams. | Similar: both devices can perform image registration using deformable image registration techniques to propagate existing contours. Note that the proposed device does not generate contours or perform image segmentation using AI. Both devices calculate dose from synthetic-CTs. These minor technological differences do not increase the safety risks for the proposed device. |
| Off-line re-plan decision support | Supported | Supported | Equivalent |
Page 5 of 8
{9}
SeeTreat
Based on the above comparison and the Appendix A of the FDA guidance titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" July 28 2014, it can be concluded that the proposed device is substantially equivalent to the predicate device.
# 5.0 Performance tests
The following key performance tests were performed to evaluate the safety and effectiveness of the proposed device:
| Test | Description | Results |
| --- | --- | --- |
| Usability summary report | This document summarizes all usability testing activities conducted per IEC62366-1, including both formative and summative tests. | Pass |
| Algorithm performance verification | This is a bench test utilizing both phantom and preliminary retrospective clinical data to evaluate the performance of the dose engine, image registration, analysis time and compatibility of the proposed device. | All test results passed the following acceptance criteria: • Median ≥ 90% of voxels passing gamma criteria of 3%/2mm • >95% of voxels passing gamma criteria of 3%/2mm for water phantom cases • >90% of voxels passing gamma criteria of 3%/2mm for air-slab phantom cases • Median Target Registration Error between CBCT and CT < 3 mm |
| Algorithm performance verification – quantitative contour evaluation | This is a bench test which evaluates the proposed device's ability to propagate contours using quantitative metrics such as DSC and ASSD. | All test results passed the following acceptance criteria: • Mean DSC ≥ 0.8 or Mean DSC ≥ Inter-rater DSC – 0.05 • Mean ASSD ≤ 3mm or Mean ASSD ≤ Inter-rater ASSD + 1mm |
| Retrospective Clinical Data Analysis | A study which uses retrospective patient data to evaluate all key processes of the proposed device, including image registration, synthetic CT generation, contour | All test results passed the following acceptance criteria: Image registration: |
{10}
SeeTreat
| | propagation, dose calculation as well as end-to-end device output. | • Median TRE < max voxel dimension sCT quality: • Median y pass rate of ≥90% using 3%/2mm tolerance • The mean Jacobian determinant of the DVF used in the DIR between the pCT and the sCT shall be 1 ± 5% • DVH parameters comparison between sCT and pCT shall be within 5% of the prescription dose for at least 80% of cases Contour evaluation: • ≥80% of cases shall be deemed by clinicians as acceptable or acceptable with minor modifications Dose calculation: • Median gamma pass rate of ≥90% using 3%/2mm tolerance • The dosimetric difference between the proposed device and a reference TPS shall be within 5% of the prescription dose for at least 80% of cases End-to-end testing: • Median gamma pass rate of ≥90% using 3%/2mm tolerance. • The dosimetric difference between the proposed device and a reference TPS shall be within 5% of the prescription dose for at least 80% of cases |
| --- | --- | --- |
Page 7 of 8
{11}
SeeTreat
| | | |
| --- | --- | --- |
| Software verification and validation | Unit, integration, and system-level verification and validation conducted in accordance with IEC 62304 and the FDA guidance “Content of Premarket Submissions for Device Software Functions”. Documentation is at an enhanced level. | Pass |
| Cybersecurity | Cybersecurity risk assessment and penetration testing have been conducted in accordance with AAMI TIR57 Principles for medical device security – risk management and the FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” | Pass |
## 6.0 Animal studies
No animal studies were conducted.
## 7.0 Clinical studies
No prospective clinical studies were conducted.
## 8.0 Conclusion
Based on the test results and comparative analysis, the SeeTreat ART.1-US device demonstrates substantial equivalence to the identified predicate device.
The performance testing demonstrates that SeeTreat ART.1-US meets all applicable safety and effectiveness requirements. The device performs within its specified acceptance criteria and maintains the same risk–benefit profile as the predicate.
Therefore, the SeeTreat ART.1-US device is substantially equivalent to the predicate device and is as safe and effective for its intended use when operated according to the labeling instructions.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.