Venti-Trach Tracheostomy Tube is intended for the airway management of tracheostomized adult patients. The device is operated by health professionals in medical institutions for single patient use. It may be continuous use by replacement and each one should be used within 29 days.
Device Story
Respiratory device inserted into trachea via stoma; provides artificial airway for ventilation and oxygenation; bypasses upper airway obstructions. Consists of cannula, 15mm connector, neck plate, pilot balloon, and optional cuff. Cuff seals trachea for positive pressure ventilation; fenestration (if present) allows airflow for speech. Operated by health professionals in critical/acute care or long-term facilities. Secured via neck strap and locking mechanisms. Introducer assists insertion. Output is a patent airway for mechanical ventilation or spontaneous breathing. Benefits include restored breathing pathway for patients with airway injury or obstruction.
Clinical Evidence
Bench testing only. Verification and validation performed per ISO 5366:2016, ISO 5356-1:2015, ISO 10993-1:2018, and ASTM F1980. Biocompatibility evaluation (RDE-23090001) conducted for materials.
Technological Characteristics
Materials: PVC (cannula, cuff, pilot balloon, neck plate), PP (15mm connector), HDPE (introducer), cotton (neck strap). Energy: None (passive). Connectivity: None. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 5366:2016, ISO 18190:2016, ISO 5356-1:2015, ISO 10993-1:2018. Form factor: Fixed 105° curvature tube.
Indications for Use
Indicated for airway maintenance of tracheostomized adult patients. Operated by health professionals in medical institutions for single patient use; continuous use by replacement within 29 days.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
Rota-Trach Disposable Standard Tracheostomy Tube (K242921)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 11, 2026
Vitaltec Corporation
% Ming-Yie Jan
Principal Consultant
RUSCERT Technology Co., LTD.
8F., No. 187, Lequn 2nd Rd., Zhongshan Dist.
Taipei City, 10462
Taiwan
Re: K253063
Trade/Device Name: Venti-Trach Tracheostomy Tube (570 Series); Venti-Trach Tracheostomy Tube (571 Series); Venti-Trach Tracheostomy Tube (580 Series); Venti-Trach Tracheostomy Tube (581 Series)
Regulation Number: 21 CFR 868.5800
Regulation Name: Tracheostomy Tube And Tube Cuff
Regulatory Class: Class II
Product Code: BTO
Dated: August 14, 2026
Received: August 14, 2026
Dear Ming-Yie Jan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253063 - Ming-Yie Jan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K253063 - Ming-Yie Jan
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253063 | ? |
| Please provide the device trade name(s). | | ? |
| Venti-Trach Tracheostomy Tube (570 series); Venti-Trach Tracheostomy Tube (571 series); Venti-Trach Tracheostomy Tube (580 series); Venti-Trach Tracheostomy Tube (581 series) | | |
| Please provide your Indications for Use below. | | ? |
| Venti-Trach tracheostomy tube is indicated for airway maintenance of tracheostomized patients. The device is operated by health professionals in medical institutions for single patient use. It may be continuous use by replacement and each one should be used within 29 days. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# 510(k) Summary
## Venti-Trach Tracheostomy Tube
### 1. General Information
| 510(k) Owner | Vitaltec Corporation |
| --- | --- |
| Address | No. 12, Lane 4-30, Chyuan-Zhou Rd., Hou-Li Dist. |
| Applicant | Joseph Chang / Managing Director |
| Contact Information | Phone: 886-4-2558 0886 Email: sales@vitaltec.com.tw |
| Date prepared | September 17, 2025 |
### 2. Subject Device Description
| Proprietary Name | Venti-Trach Tracheostomy Tube |
| --- | --- |
| Regulation Number | 21 CFR 868.5800 |
| Regulation Name | Tracheostomy tube and tube cuff |
| Regulatory Class | Class II |
| Product Code | BTO |
| Common Name | Tracheostomy Tube |
### 3. Predicate Device
| Proprietary Name | Rota-Trach Disposable Standard Tracheostomy Tube |
| --- | --- |
| Premarket Notification | K242921 |
| Regulation Number | 21 CFR 868.5800 |
| Regulation Name | Tracheostomy tube and tube cuff |
| Regulatory Class | Class II |
| Product Code | BTO |
| Common Name | Tracheostomy Tube |
### 4. Device Description
Venti-Trach Tracheostomy Tube is a respiratory device inserted into the patient's trachea through a stoma, making direct contact with the tracheal mucosal membrane
5-1
{5}
tissue. It may be with or without a cuff attaching around the tube of the device. The design of the cuff could seal the space between the tube and the patient's trachea from inhaling undesired or foreign matters and for assistance of positive pressure ventilation.
# 5. Indications for Use:
Venti-Trach tracheostomy tube is indicated for airway maintenance of tracheostomized patients. The device is operated by health professionals in medical institutions for single patient use. It may be continuous use by replacement and each one should be used within 29 days.
# 6. Substantial Equivalence
Table below provides a comparison of the indications for use and key technological characteristics of Venti-Trach Tracheostomy Tube with that of the Predicate, Rota-Trach Disposable Standard Tracheostomy Tube (K242921).
| Model Particular | Subject Device Venti-Trach Tracheostomy Tube | Predicate Device Rota-Trach Disposable Standard Tracheostomy Tube | Comparison (Same/ Similar/ Different) |
| --- | --- | --- | --- |
| Intended use defined under 21 CFR 872.4200 | Venti-Trach Tracheostomy Tube is intended for the airway management of tracheostomized adult patients. The device is operated by health professionals in medical institutions for single patient use. It may be continuous use by replacement and each one should be used within 29 days. | Rota-Trach Disposable Standard Tracheostomy Tube is intended for the airway management of tracheostomized adult patients. The device is operated by health professionals in medical institutions for single patient use. It may be continuous use by replacement and each one should be used within 29 days. | Same |
| Indications for use | Venti-Trach Disposable Standard Tracheostomy Tube is indicated for airway maintenance of tracheostomized patients. | Rota-Trach Disposable Standard Tracheostomy Tube is indicated for airway maintenance of tracheostomized patients. | Same |
5-2
{6}
| **Patient Population** | Adults with average height, weight, and anthropometrics | Adults with average height, weight, and anthropometrics | Same |
| --- | --- | --- | --- |
| **Principle of operation** | The tracheostomy tube functions as an artificial airway by being inserted into the trachea through a tracheostomy stoma. It serves as a conduit for ventilation and oxygenation while allowing connection to respiratory care devices. The tube is secured in place using integrated locking mechanisms and a neck strap. If present, a fenestration permits partial airflow through the vocal cords to facilitate speech. An introducer is used to assist with insertion and is removed afterward. | The tracheostomy tube functions as an artificial airway by being inserted into the trachea through a tracheostomy stoma. It serves as a conduit for ventilation and oxygenation while allowing connection to respiratory care devices. The tube is secured in place using integrated locking mechanisms and a neck strap. If present, a fenestration permits partial airflow through the vocal cords to facilitate speech. An introducer is used to assist with insertion and is removed afterward. | Same |
| **Invasive or Noninvasive** | Surgically Invasive | Surgically Invasive | Same |
| **Maximum Use** | Recommended 29 Days | Recommended 29 Days | Same |
| **Functionality** | Venti-Trach Tracheostomy Tube provides an alternate pathway for breathing when the natural airway is blocked or injured. It creates direct route to the trachea, bypassing upper airway obstructions. It can be connected to respiratory equipment, like ventilators, to assist with breathing and provide mechanical ventilation. | Rota-Trach Tracheostomy Tube provides an alternate pathway for breathing when the natural airway is blocked or injured. It creates direct route to the trachea, bypassing upper airway obstructions. It can be connected to respiratory equipment, like ventilators, to assist with breathing and provide mechanical ventilation. | Same |
5-3
{7}
| Dimension | The tube has a fixed curvature of 105°. Dimensions for each size are specified as follows: ID (Inside Diameter): Cannula internal diameter, also expressed as the tube size. OD (Outside Diameter): Cannula external diameter. TL (Tube Length): Overall length of the tube. CRD (Cuff Resting Diameter, Ø mm): For cuffed models only. | The tube has a fixed curvature of 100°. Dimensions for each size are specified as follows: ID (Inside Diameter): Cannula internal diameter, also expressed as the tube size. OD (Outside Diameter): Cannula external diameter. TL (Tube Length): Overall length of the tube. CRD (Cuff Resting Diameter, Ø mm): For cuffed models only. | Similar (Analysis 1) |
| --- | --- | --- | --- |
| Size/ID (mm) | OD (mm) | TL (mm) | CRD (mm) | Size/ID (mm) | OD (mm) | TL (mm) | CRD (mm) |
| 6.0 | 9.0 | 64 | 21 | 6.0 | 8.3 | 55 | 21 |
| 6.5 | 9.5 | 67 | 22 | 6.5 | 9.0 | 60 | 22 |
| 7.0 | 10.0 | 70 | 24 | 7.0 | 9.6 | 65 | 23 |
| 7.5 | 10.5 | 74 | 26 | 7.5 | 10.3 | 70 | 24 |
| 8.0 | 11.0 | 77 | 28 | 8.0 | 11.0 | 76 | 26 |
| 8.5 | 11.5 | 78 | 30 | 8.5 | 11.6 | 82 | 28 |
| 9.0 | 12.0 | 81 | 30 | 9.0 | 12.3 | 89 | 30 |
| 10.0 | 13.0 | 87 | 33 | 10.0 | 13.3 | 97 | 33 |
| Material Composition | Cannula | PVC | Cannula | PVC | Similar (Analysis 2) |
| 15mm connector | PP | 15mm connector | PP |
| Cuff | PVC | Cuff | PVC |
| Neck plate | PVC | Neck plate | TPR |
| Pilot ballon w/t inflating tube | PVC | Pilot ballon w/t inflating tube | PVC |
| Introducer | HDPE | Introducer | HDPE |
| Neck strap | 100% Cotton | Neck strap | 100% Cotton |
| |
| Component | Cannula; 15mm connector; cuff; neck plate; | Cannula; 15mm connector; cuff; neck plate; | Similar (Analysis 3) |
5-4
{8}
| | pilot ballon w/t inflating tube; introducer; neck strap. | pilot ballon w/t inflating tube; introducer; neck strap; disconnect wedge | |
| --- | --- | --- | --- |
| **Sterilization** | Ethylene Oxide (EO) Sterile SAL 10^{-6} to End User | Ethylene Oxide (EO) Sterile SAL 10^{-6} to End User | Same |
| **Shelf Life** | 3-year shelf life intended | 3-year shelf life intended | Same |
| **Single Use Single Patient Use** | Yes | Yes | Same |
| **Environment of Use** | Critical care settings, Acute care settings, Long term care facilities | Critical care settings, Acute care settings, Long term care facilities | Same |
| **Biocompatibility** | Compatibility materials ISO 10993-1: 2018 | Compatibility materials ISO 10993-1: 2009 | Similar Subject device conforms to ISO 10993-1:2018 (latest version, superseding ISO 10993-1:2009 used for predicate). No new biocompatibility risk is introduced. |
| **Biocompatibility evaluation endpoints** | Cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, subchronic toxicity, genotoxicity, implantation and chemical characterization. | Cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, subchronic toxicity, genotoxicity, implantation and chemical characterization. | Same |
| **Device standard** | ISO 5366:2016 ISO 18190:2016 ISO 5356-1:2015 | ISO 5366:2016 ISO 18190:2016 ISO 5356-1:2015 | Same |
### Substantial Equivalence Discussion
#### Analysis 1:
The subject device incorporates a 105° tube curvature, compared with the 100° curvature of the predicate device, leading to minor variations in tube length, outer
5-5
{9}
diameter and cuff diameter. For devices of the same size specification, the total tube length differs by less than 10 mm, the outer diameter by less than 0.7 mm, and the cuff diameter by less than 2 mm.
These geometric differences have been verified through applicable tests in accordance with ISO 5366:2016 (see Design Control Activities Summary), and are disclosed in the Instructions for Use (IFU) and product labeling.
# **Analysis 2:**
The subject device differs from the predicate device by a material change in the neck plate, from TPR to PVC. No new biocompatibility testing is needed because the same PVC material is already used in the predicate device's tube, which has direct airway contact. Since the subject device's neck plate only touches intact skin, the risk of a new biological response is considered lower than what has already been proven safe, thus justifying the change without additional testing. Please refer to Biocompatibility Evaluation (RDE-23090001) for the detail justification of the biological risk.
# **Analysis 3:**
The subject device does not include a disconnect wedge. According to the following, the absence of a wedge does not impact device safety or performance.
(1) The neck plate material was changed from TPR to PVC, providing higher rigidity and stability so that removal does not require auxiliary tools.
(2) The 15 mm connector conforms with ISO 5366 and ISO 5356-1, allowing safe removal by standard handling.
(3) The wedge is not a standard requirement. Some marketed models, such as the marketed device K173384 (IFU #10001029-004), do not include this accessory.
# **Conclusion**
According to the discussion above, the indications for use of subject device, Venti-Trach Tracheostomy Tube is same as predicate device, Rota-Trach Disposable Standard Tracheostomy Tube, and the other differences in non-significant characteristics are clarified, the substantial equivalent is claimed. Since the differences of the devices do not raise different questions of safety and effectiveness, Venti-Trach Tracheostomy Tube is as safe and effective as legally marketed Rota-Trach Disposable Standard Tracheostomy Tube.
# **7. Design Control Activities**
The risks arisen from the design modifications activities have been identified and evaluated are controlled through a failure modes and effects analysis (FMEA) that described in FDA-recognized standard, ISO 14971:2019. The safety and effectiveness of Venti-Trach Tracheostomy Tube are also verified and validated according to device-specific, FDA-recognized standard, ISO 5366:2016, ISO 5356-1:2015, ISO 10993-1:2018 and ASTM F1980.
5-6
{10}
## **8. Conclusion**
Based on the information provided in this premarket notification, the same indications for use with small modifications, Venti-Trach Tracheostomy Tube raises no new questions of safety and effectiveness and subject device is substantially equivalent to predicate device in terms of safety, effectiveness, and performance.
5-7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.