K253058 · Shantou Institute of Ultrasonic Instruments Co., Ltd. · IZL · Jun 18, 2026 · Radiology
Device Facts
Record ID
K253058
Device Name
Portable DR Imaging System
Applicant
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Product Code
IZL · Radiology
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Indications for Use
The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's extremities. This device is not intended for mammography. This device is only intended for adults, not for paediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms.
Device Story
Portable X-ray system for extremity imaging; consists of X-ray unit, flat panel detector, and PIE-5 workstation. System control software (ARM platform, C language) manages high-voltage generation, X-ray tube excitation, collimator, and battery monitoring. PIE-5 workstation (Windows-based) handles patient registration, image acquisition, processing, review, diagnosis, and PACS transmission. Operated by qualified physicians or technicians in hospital shielded rooms. Output consists of digital radiographic images used for clinical diagnosis. Benefits include portable, high-quality extremity imaging for adult patients.
Clinical Evidence
Comparative clinical image quality assessment performed. Images of extremities acquired using both subject and predicate devices were evaluated by an ABR Board-Certified Radiologist. Results demonstrated substantial equivalence in image quality and clinical utility for diagnostic requirements.
Technological Characteristics
Portable X-ray system; 150-560W generator; 40-90kV range. Detector: TFT-amorphous Silicon, 150µm pixel pitch, 14"x17" max size. Connectivity: Wired/Wireless IEEE 802.11a/g/n (2.4/5GHz) with WEP/WPA/WPA2. Software: Embedded system control (ARM/C) and PIE-5 workstation (Windows/C). Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, ISO 10993, and IEC 62304.
Indications for Use
Indicated for adult patients requiring X-ray imaging of extremities. Contraindicated for pediatric patients and mammography. Must be used in hospital shielded rooms with stand-mounted configuration.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Shantou Institute of Ultrasonic Instruments Co., Ltd.
% Guang Chen
Primary Correspondent
#77, Jinsha Road
Shantou, CN 515041
CHINA
Re: K253058
Trade/Device Name: Portable DR Imaging System
Regulation Number: 21 CFR 892.1720
Regulation Name: Mobile X-Ray System
Regulatory Class: Class II
Product Code: IZL, MQB, LLZ
Dated: May 20, 2026
Received: September 22, 2025
Dear Guang Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253058 - Guang Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253058 - Guang Chen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253058
Device Name
Portable DR Imaging System
Indications for Use (Describe)
The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's extremities.
This device is not intended for mammography.
This device is only intended for adults, not for paediatric application.
This device is only suitable for stand-mounted use in hospital shielded rooms.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
SIUI Portable DR Imaging System---
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is provided in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).
The assigned 510(k) number is: K253058
## 1. Submitter
Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI)
77 Jinsha Road, Shantou, Guangdong 515041, China
Tel: 86-754-88250151
Fax: 86-754-86603664
Contact Person: Chen Guang
Shantou Institute of Ultrasonic Instruments Co., Ltd.
77 Jinsha Road, Shantou, Guangdong 515041, China
Date Prepared: June 17, 2026
## 2. Subject Device
Name of Device: Portable DR Imaging System
510(k) submitter: SIUI
Regulation Name: Mobile x-ray system (21 CFR 892.1720)
Regulatory Class: II
Product Code: IZL, MQB, LLZ
## 3. Predicate Device
Name of Device: SR-2300 Portable DR Imaging System (K202353)
510(k) holder: SIUI
Regulation Name: Mobile x-ray system (21 CFR 892.1720)
Regulatory Class: II
1/7
{5}
SIUI Portable DR Imaging System---
Product Code: IZL, MQB, LLZ
Classification Name:
Mobile x-ray system IZL (per 21 CFR 892.1720)
Stationary x-ray system MQB (per 21 CFR 892.1680)
Medical image management and processing system LLZ (per 21 CFR 892.2050)
#### 4. Device Description
The Portable DR Imaging System is a portable digital device developed, designed and manufactured by SIUI. The device consists of the following major components: a portable X-ray unit, a flat panel detector, and workstation.
This system is a Track 3 device. The Portable DR Imaging System contains two pieces of software, one is the embedded software in the X-ray unit: system control software, and the other is PIE-5 DR Workstation (hereinafter referred to as “PIE-5”). The system control software of the Portable DR Imaging System is software designed by SIUI. The software adopts development languages of Keil, the program language is C language, and the hardware platform is ARM. The system control software is for real-time interaction and control with various circuit modules inside the Portable DR Imaging System. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters. The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board, control the high-voltage tank to generate high-voltage and to excite the X-ray tube inside to emit X-rays, control the switch of the collimator indicator, monitor the working status of the device and the battery power status, and control the display of the status indicators.
PIE-5 uses the C programming language, and the operating platform is Microsoft Windows. The software is a medical image processing system, consisting of image acquisition, image review and diagnosis, image processing and digital storage functions, which mainly address digital image workflow in radiographic department of hospitals. The functions of the software include patient registration, image acquisition and processing, image review and diagnosis, image printing, report editing and PACS transmission, digital image storage and recovery.
2/7
{6}
SIUI Portable DR Imaging System
# **5. Comparisons of SR-300Pro, SR-200Pro, SR-100Pro, SR-300, SR-200, SR-100, SR-2300Pro, SR-1100Pro, SR-1000Pro, SR-2300, SR-2000, SR-1100, and SR-1000 devices**
The Portable DR Imaging System includes the models SR-300Pro /SR-2300 / SR-1000 / SR-2000 / SR-2300Pro / SR-1000Pro / SR-1100 /SR-1100Pro / SR-300 / SR-200 / SR-100 /SR-200Pro /SR-100Pro. These models used the same safe critical components, with the same circuit modules and working principles, and the same appearance. Except for the differences in functional configuration, these models are completely the same.
The mechanical structure is designed for market requirements of the devices according to international general safety standard (IEC 60601-1). The hardware design is identical to the following SR series models: SR-300Pro, SR-200Pro, SR-100Pro, SR-300, SR-200, SR-100, SR-2300Pro, SR-1100Pro, SR-1000Pro, SR-2300, SR-2000, SR-1100, and SR-1000. A list of the differences between various models intended for sale in the market is shown below, which are identical except for the differences in settings:
Table 1
| Model | DICOM Setting | Report Setting | Process Setting | Custom Setting |
| --- | --- | --- | --- | --- |
| SR-300Pro | Standard | Standard | Standard | Standard |
| SR-300 | Standard | Standard | Standard | Optional |
| SR-200 | Standard | Standard | Optional | Standard |
| SR-100 | Standard | Optional | Standard | Standard |
| SR-200Pro | Optional | Standard | Standard | Standard |
| SR-100Pro | Standard | Standard | Optional | Optional |
| SR-2300 | Standard | Optional | Standard | Optional |
| SR-1000 | Standard | Optional | Optional | Standard |
| SR-2000 | Optional | Standard | Standard | Optional |
| SR-2300Pro | Optional | Standard | Optional | Standard |
| SR-1000Pro | Optional | Optional | Standard | Standard |
| SR-1100 | Standard | Optional | Optional | Optional |
| SR-1100Pro | Optional | Standard | Optional | Optional |
3/7
{7}
SIUI Portable DR Imaging System
## 6. Indications for Use
The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s extremities.
This device is not intended for mammography.
This device is only intended for adults, not for pediatric application.
This device is only suitable for stand-mounted use in hospital shielded rooms.
## 7. Comparison of Technological Characteristics with the Predicate Device
The comparison between the overall specifications of the predicate device (K202353) and the subject device is shown in Table 2. Any differences between the predicate and the subject device have no impact on safety or efficacy of the subject device and do not raise any new potential or increased safety risks, and the subject device is equivalent in performance to existing legally marketed devices.
### Subject Device : Portable DR Imaging System (K253058)
#### Predicate Device: SIUI SR-2300 Portable DR Imaging System (K202353)
| Description | | Subject Device : Portable DR Imaging System (K253058) | Predicate Device: SIUI SR-2300 Portable DR Imaging System (K202353) |
| --- | --- | --- | --- |
| **Indications for use** | | The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s extremities. This device is not intended for mammography. This device is only intended for adults, not for pediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms. | Intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). This device is not intended for mammography. |
| **Basic Genera-** | **Peak generator** | 150-560W | 5kW |
4/7
{8}
SIUI Portable DR Imaging System
| **tor Charac- teristics** | **Tube current** | 1.6-8mA | 10-100mA |
| --- | --- | --- | --- |
| | **Tube voltage adjustable range** | 40-90kV, step value 1kV | 40-125kV, step value 1kV |
| | **mAs range** | 0.16mAs-8mAs | 0.4mAs-200mAs |
| **Collimator** | | Built in | Built in |
| **X-ray Generator** | | One model | One model |
| **Operator console** | | Touch Control or Touch Screen | Button Control or Touch Screen |
| **Digital X-Ray Detectors** | | SFD-1X/ SFD-2X/ SFD-3X | SFD-1X |
| **Panel Shape** | | Rectangular Panel | Rectangular Panel |
| **Detector Size** | | Max: 14” X 17” | 14” X 17” |
| **Pixel Pitch** | | 150µm | 150µm |
| **Materials Scintillator** | | TFT-amorphous Silicon | TFT-amorphous Silicon |
| **Communication Method** | | Wire Wireless IEEE 802.11a//g/n(2.4GHz/ 5GHz) Security: WEP/WPA/WPA2 | Wire Wireless IEEE 802.11a//g/n(2.4GHz/ 5GHz) Security: WEP/WPA/WPA2 |
| **Acquisition Software** | | PIE-5 | PIE-5 |
| **Software Function** | | Image viewing Image search Image storage Image annotation Image measurement Image processing Image stitch | Image viewing Image search Image storage Image annotation Image measurement Image processing Image stitch |
| **DICOM 3.0 Compatibility** | | Yes | Yes |
| **Power Source** | | AC Line or rechargeable batteries (Generator only) | AC Line or rechargeable batteries |
## 8. Non-clinical Testing Summary
The Portable DR Imaging System complies with and/or were tested in accordance with the following FDA guidance and International Standards:
- ● IEC 60601-1:2005+ AMD1:2012+AMD2:2020 Medical electrical equipment -
5/7
{9}
SIUI Portable DR Imaging System---
Part 1: General requirements for basic safety and essential performance
- ● IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests.
- ● IEC 60601-1-3:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- ● IEC 60601-2-54:2022 Medical electrical equipment - Part 2-54:Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- ● ISO 14971:2019 Medical devices - Application of risk management to medical devices
- ● ISO 10993-1:2018 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process
- ● ISO 10993-5:2009 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity
- ● ISO 10993-10:2021 Biological evaluation of medical devices - Part 10:Tests for skin sensitization
- ● ISO 10993-23:2021 Biological evaluation of medical devices - Part 23:Tests for irritation
- ● IEC 62304:2006+A1:2015 Medical device software – Software life cycle processes
- ● IEC 60601-1-6:2010+ AMD1:2013+AMD2:2020 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- ● IEC 62366-1:2015+AMD1:2020 Medical devices – Part 1: Application of usability engineering to medical devices
- ● ISO 15223-1:2021 Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
- ● ISO 13485:2016 Medical devices - Quality management systems - Requirements for regulatory purposes
- ● Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- ● Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices
6/7
{10}
SIUI Portable DR Imaging System
- Content of Premarket Submissions for Device Software Functions
- Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions
- Electromagnetic Compatibility (EMC) of Medical Devices
The subject device and the predicate device are comparable in terms of technical features, general functions, applications and intended uses. The test results showed compliance with the above standards.
Non-clinical Detector Performance testing as prescribed in the standard of IEC 60601-2-54: 2022.
9. Clinical Testing
SIUI performed Image Quality and Performance Testing by acquiring images of the extremities using both the Portable DR Imaging System (subject device) and the predicate device. The images were assessed for image quality and clinical utility by an ABR Board-Certified Radiologist. Comparative analysis demonstrated that the image quality of the subject device was substantially equivalent to that of the predicate device. The system exhibited sufficient clinical utility and fulfilled diagnostic requirements for extremities.
10. Conclusion
The subject device Portable DR Imaging System and the predicate device SIUI SR-2300 Portable DR Imaging System (K202353) are comparable in terms of technical features, general functions, applications and indications for use.
7/7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.