100 subjects including MS, AD, CI, and healthy controls from Siemens, GE, and Philips scanners
>1 (radiologists)
White matter hyperintensities follow-up
—
—
Dice score 0.59
—
—
165 subjects including MS and AD patients from Siemens, GE, and Philips scanners
>1 (radiologists)
Indications for Use
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images. AI-Rad Companion Brain MR provides the following functionalities. • Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities • Quantitative comparison of each brain structure with normative data from a healthy population • Presentation of results for reporting that includes all numerical values as well as visualization of these results
Device Story
AI-Rad Companion Brain MR is a post-processing software for MR brain images; inputs include T1 MPRAGE and T2-weighted FLAIR datasets. It performs automated segmentation and quantitative analysis of brain structures and white matter hyperintensities (WMH). The device compares structure volumes against normative healthy population data; results are presented as numerical values and 3D overlay maps. Used in clinical settings by healthcare professionals; output assists clinicians in evaluating brain morphometry and WMH, including follow-up analysis to identify new or enlarged lesions. The device supports multi-vendor data (Siemens, GE, Philips) and operates via a cloud/edge architecture. It aids clinical decision-making by providing objective volumetric data and longitudinal change assessment, potentially benefiting patients through improved monitoring of neurodegenerative or white matter conditions.
Clinical Evidence
No clinical data. Bench testing only. Performance validated on 100 subjects for WMH segmentation (Dice 0.60, F1 0.67) and 165 subjects for WMH follow-up (Dice 0.59, F1 0.71). Ground truth established by manual annotation from three radiologists. Pearson correlation with ground truth was 0.96 for WMH volume and 0.76 for new/enlarged WMH volume.
Technological Characteristics
Software-based image processing system. Operates on T1 MPRAGE and T2-weighted FLAIR MR images. Cloud/edge architecture. Complies with IEC 62366-1, ISO 14971, IEC 62304, and DICOM standards. Features automated segmentation and quantification algorithms. No hardware components.
Indications for Use
Indicated for clinicians to assist in viewing, analyzing, and evaluating MR brain images, including automated segmentation and quantification of brain structures and white matter hyperintensities, and comparison against normative healthy population data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 22, 2026
Siemens Healthcare GmbH
Kira Morales
Regulatory Affairs Manager
Henkestrasse 127
Erlangen, 91052
Germany
Re: K253057
Trade/Device Name: AI-Rad Companion Brain MR
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: December 5, 2025
Received: December 5, 2025
Dear Kira Morales:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253057 - Kira Morales
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K253057 - Kira Morales
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253057 | ? |
| Please provide the device trade name(s). | | ? |
| Al-Rad Companion Brain MR | | |
| Please provide your Indications for Use below. | | ? |
| Al-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing,
analyzing, and evaluating MR brain images.
Al-Rad Companion Brain MR provides the following functionalities.
• Automated segmentation and quantitative analysis of individual brain structures and white matter
hyperintensities
• Quantitative comparison of each brain structure with normative data from a healthy population
• Presentation of results for reporting that includes all numerical values as well as visualization of these
results | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
SIEMENS Healthineers
K253057
# 510(k) SUMMARY FOR AI-Rad Companion Brain MR
Submitted by:
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Date Prepared: December 5, 2025
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.
# 1. Submitter
Importer/Distributor
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 193552
Registration Number: 2240869
Manufacturing Site
Siemens Healthcare GmbH
Henkestrasse 127
Erlangen, Germany 91052
Registration Number: 3002808157
# 2. Contact Person
Kira Morales
Regulatory Affairs Manager
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19335
Phone: +1 (484) 901 - 9471
Email: kira.morales@siemens-healthineers.com
# 3. Device Name and Classification
Product Name: AI-Rad Companion Brain MR
Trade Name: AI-Rad Companion Brain MR
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{5}
SIEMENS Healthineers
Device Class: Class II
Product Code: QIH
# 4. Predicate Device
Product Name: AI-Rad Companion Brain MR
Propriety Trade Name: AI-Rad Companion Brain MR
510(k) Number: K232305
Clearance Date: October 23, 2023
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Primary Product Code: QIH
Recall Information: N/A
# 5. Reference Device
Product Name: icobrain
510(k) Number: K192130
Clearance Date: December 13, 2019
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Primary Product Code: LLZ
Recall Information: N/A
# 6. Indications for Use
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.
AI-Rad Companion Brain MR provides the following functionalities.
- Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
- Quantitative comparison of each brain structure with normative data from a healthy population
- Presentation of results for reporting that includes all numerical values as well as visualization of these results
# 7. Device Description
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{6}
SIEMENS Healthineers
AI-Rad Companion Brain MR runs two distinct and independent algorithms for Brain Morphometry analysis and White Matter Hyperintensities (WMH) segmentation, respectively. In overall, comprises four main algorithmic features:
- Brain Morphometry
- Brain Morphometry follow-up
- White Matter Hyperintensities (WMH)
- White Matter Hyperintensities (WMH) follow-up
The feature for Brain Morphometry is available since the first version of the device (VA2x), while segmentation of White Matter Hyperintensities was added since VA4x and the follow-up analysis for both is available since VA5x. The brain morphometry and brain morphometry follow-up feature have not been modified and remain identical to previous VA5x mainline version.
AI-Rad Companion Brain MR VA60 is an enhancement to the predicate, AI-Rad Companion Brain MR VA50 (K232305). Just as in the predicate, the brain morphometry feature of AI-Rad Companion Brain MR addresses the automatic quantification and visual assessment of the volumetric properties of various brain structures based on T1 MPRAGE datasets. From a predefined list of brain structures (e.g. Hippocampus, Caudate, Left Frontal Gray Matter, etc.) volumetric properties are calculated as absolute and normalized volumes with respect to the total intracranial volume. The normalized values are compared against age-matched mean and standard deviations obtained from a population of healthy reference subjects. The deviation from this reference population can be visualized as 3D overlay map or out-of-range flag next to the quantitative values.
Additionally, identical to the predicate, the white matter hyperintensities feature addresses the automatic quantification and visual assessment of white matter hyperintensities on the basis of T1 MPRAGE and T2 weighted FLAIR datasets. The detected WMH can be visualized as a 3D overlay map and the quantification in count and volume as per 4 brain regions in the report.
# 8. Substantially Equivalent (SE) and Technological Characteristics
The intended use of the predicate device and the subject device are equivalent. The main difference is that AI-Rad Companion Brain MR VA60 adds improvements to the White matter hyperintensities and White Matter Hyperintensities Follow-up as compared to the predicate, AI-Rad Companion Brain MR VA50.
The subject device, AI-Rad Companion Brain MR VA60 is substantially equivalent with regard to the intended use and technical characteristics compared to the predicate device, AI-Rad Companion Brain MR VA50 (K232305) with respect to the software features, functionalities, and core algorithms. The additional enhancements and improvements provided in AI-Rad Companion Brain MR VA60 increase the usability and reduce the complexity of the imaging workflow for the clinical user.
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{7}
SIEMENS Healthineers
Icobrain serves as a reference device within this submission and a dedicated comparison of technological characteristics is provided. Siemens Healthineers has determined that AI-Rad Companion Brain MR VA60 is comparable to icobrain (K192130) as it has similar technological and performance characteristics with respect to the white matter hyperintensities & follow-up feature improvements (algorithms cleared in K232305). AI-Rad Companion Brain MR VA60 used equivalent validation methodology to analyze the performance of the white matter hyperintensities follow-up feature compared to icobrain (K192130).
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{8}
SIEMENS Healthineers
The risk analysis and non-clinical data support that both devices perform equivalently and do not raise different questions of the safety and effectiveness.
| Feature | Subject Device: AI-Rad Companion Brain MR VA60 | Predicate Device: AI-Rad Companion Brain MR VA50 (K232305) | Reference Device: icobrain (K192130) | Comparison Results |
| --- | --- | --- | --- | --- |
| Brain Morphometry Segmentation | Pre-processing functionality for automatic segmentation and volumetry of MPRAGE data. | Pre-processing functionality for automatic segmentation and volumetry of MPRAGE data. | Image processing for automatic segmentation and volumetry of MPRAGE data. | Same as predicate |
| Brain Morphometry Quantification | Calculation of label maps (display of brain segmentation) and partially combined label maps (fused with the processed MPRAGE data). | Calculation of label maps (display of brain segmentation) and partially combined label maps (fused with the processed MPRAGE data). | Normalized and unnormalized volume and volume changes of different brain structures. | Same as predicate |
| Brain Morphometry: Deviation Map | Calculation of deviation map (representation of brain status in relation to reference data) and partially combined deviation maps (fused with the processed MPRAGE data) User customizable color labels for the overlay map. | Calculation of deviation map (representation of brain status in relation to reference data) and partially combined deviation maps (fused with the processed MPRAGE data) User customizable color labels for the overlay map. | Not available | Same as predicate |
| Brain Morphometry Follow-Up | Automatic calculation of the atrophy range in | Automatic calculation of the atrophy range in | Not available | Same as predicate |
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{9}
SIEMENS Healthineers
| | percentage for each segmented brain structure | percentage for each segmented brain structure | | |
| --- | --- | --- | --- | --- |
| **Brain Morphometry Follow-Up Time Between Studies** | Configurable between 14-180 days | Configurable time interval between the current and the prior scan should be ≥ 180 days and < the retention period. | Not available | **Enhanced from the predicate** |
| **Brain White Matter Hyperintensities Segmentation** | Pre-processing functionality for automatic segmentation and volumetry of MPRAGE and FLAIR data. | Pre-processing functionality for automatic segmentation and volumetry of MPRAGE and FLAIR data. | Image processing for automatic segmentation and volumetry of FLAIR data. | **Same as predicate** |
| **Brain White Matter Hyperintensities Quantification** | Calculation of white matter hyperintensities count and volume as per 4 brain regions. | Calculation of white matter hyperintensities count and volume as per 4 brain regions. | Unnormalized volume and volume changes of FLAIR white matter hyperintensities as per 4 brain regions | **Same as predicate** |
| **Brain White Matter Hyperintensities Map** | Calculation of white matter hyperintensities map fused with the processed FLAIR data User customizable color labels for the overlay map. | Calculation of white matter hyperintensities map fused with the processed FLAIR data User customizable color labels for the overlay map. | Calculation of white matter hyperintensities map overlaid with the FLAIR data | **Same as predicate** |
| **Brain White Matter Hyperintensities Input Data** | T2-weighted 3D FLAIR image series. | T1-weighted MPRAGE and T2-weighted 3D FLAIR image series | T1-weighted MPRAGE and T2-weighted 3D FLAIR image series | **Streamlined from the predicate** |
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{10}
SIEMENS Healthineers
| White Matter Hyperintensities Follow-Up | MR images from two time points to identify new or enlarged areas. The results of follow-up scan are then refined by using morphological operations to generate WMH changed areas between baseline and follow-up scan | MR images from two time points to identify new or enlarged areas. The results of follow-up scan are then refined by using morphological operations to generate WMH changed areas between baseline and follow-up scan | Assessment of New/Enlarging Lesion count | Same as predicate |
| --- | --- | --- | --- | --- |
| White Matter Hyperintensities Follow Up Input Data | WMH follow-up supports 1.5T and 3T input data in both prior and current study | WMH follow-up is only supported for 3T input data in both prior and current study. | Both 1.5T and 3T is supported | Enhanced from the predicate |
| White Matter Hyperintensities Follow-Up Data Criteria | Not mandatory to have the same Slice Thickness, Pixel Spacing, and Magnetic Field Strength consistent between prior and current studies. *In case if the prior and current Magnetic Field Strength (0018, 0087) are not same, the system will process the input data with an information message. | Mandatory to have the same Slice Thickness, Pixel Spacing, and Magnetic Field Strength consistent between prior and current studies. | Not available | Enhanced from the predicate |
| White Matter Hyperintensities & Follow-Up Multi-Vendor Support | Validated with data from Siemens Healthineers, GE and Philips | Only supports data acquired on 3 Tesla Siemens Healthineers MR scanners | Multi-vendor support | Enhanced from the predicate |
| Distribution & Archiving | Creation of an image series for a morphometry report. Automatic transfer of | Creation of an image series for a morphometry report. Automatic transfer of generated image | Automatic transfer of generated image | Same as predicate |
Al-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{11}
SIEMENS Healthineers
| | generated maps and morphometry report to a PACS system. | generated maps and morphometry report to a PACS system. | series and report to a PACS system. | |
| --- | --- | --- | --- | --- |
| User Interface Confirmation | Confirmation UI with basic visualization functionality | Confirmation UI with basic visualization functionality | Not available. | Same as predicate |
| User Interface Configuration | Configuration UI | Configuration UI | Not available | Same as predicate |
| Layouts | Simplified layout dedicated for confirmation of results | Simplified layout dedicated for confirmation of results | Not available | Same as predicate |
| Architecture | Cloud solution and Edge components deployed on customer premise. | Cloud solution and Edge components deployed on customer premise. | Cloud only solution with no components deployed on customer premise. | Same as predicate |
| DICOM SR | DICOM structured report representation of a natural language report | DICOM structured report representation of a natural language report | DICOM structured report | Same as predicate |
Table 1: Comparison table for AI-Rad Companion Brain MR VA60, predicate device AI-Rad Companion Brain MR VA50 (K232305) and reference device icobrain (K192130)
The conclusions from all verification and validation data suggest that these enhancements are equivalent with respect to safety and effectiveness of the predicate device. These modifications do not change the intended use of the product. Siemens is of the opinion that AI-Rad Companion Brain MR VA60 is substantially equivalent to the currently marketed device, AI-Rad Companion Brain MR VA50.
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{12}
SIEMENS Healthineers
# 9. Nonclinical Tests
Non-clinical tests were conducted to test the functionality of AI-Rad Companion Brain MR. Software validation and bench testing have been conducted to assess the performance claims as well as the claim of substantial equivalence to the predicate device.
AI-Rad Companion has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrates that AI-Rad Companion Organs RT complies with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Device Software Functions" (June 2023) as well as with the following voluntary FDA recognized Consensus Standards listed in Table 2.
| Recognition Number | Product Area | Title of Standard | Reference Number and Date | Standards Development Organization |
| --- | --- | --- | --- | --- |
| 5-129 | General | Medical Devices – Application of usability engineering to medical devices [including Corrigendum 1 (2016)] | IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | IEC |
| 5-125 | General | Medical Devices – application of risk management to medical devices | ISO 14971 Third Edition 2019-12 | ISO |
| 13-79 | Software/ Informatics | Medical device software – software life cycle processes [Including Amendment 1 (2016)] | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | AAMI ANSI IEC |
| 12-349 | Radiology | Digital Imaging and Communications in Medicine (DICOM) Set | PS 3.1 – 3.20 2021e | NEMA |
| 5-134 | General | Medical devices – symbols to be used with information to be supplied by the manufacturer – Part 1: General Requirements | 15223-1 Fourth edition 2021-07 | ISO IEC |
| 13-97 | Software/ Informatics | Health software – Part 1: General requirements for product safety | 82304-1 Edition 1.0 2016-10 | IEC |
Table 2: List of recognized standards
# Verification and Validation
Software documentation level, per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023, is also included as part of this submission. The performance data demonstrates continued
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{13}
SIEMENS Healthineers
conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the subject device during product development.
Software bench testing in the form of Unit, System and Integration tests were performed to evaluate the performance and functionality of the new features and software updates. All testable requirements in the Requirement Specifications and the Risk Analysis have been successfully verified and traced in accordance with the Siemens Healthineers DH product development process. Human factor usability validation is addressed in system testing and usability validation test records. Software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.
Siemens Healthineers adheres to the cybersecurity recommendations as defined the FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 2023) by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.
# 10. Performance Software Validation
AI-Rad Companion Brain MR VA60A brain morphometry and brain morphometry follow-up features are identical to the predicate device AI-Rad Companion Brain MR VA50A.
# White Matter Hyperintensities Features
Performance testing for AI-Rad Companion Brain MR WMH was performed on test data from 100 subjects, which included Multiple Sclerosis patients (MS), Alzheimer’s patients (AD), cognitive impaired (CI) and healthy controls (HC) from Siemens, GE, and Philips scanners. Testing data has balanced distribution with respect to gender and age of the patient according to target patient population and field strength (1.5T and 3T) of the MR scanner used.
Accuracy was validated by comparing the results of the device to manual annotated ground truth from three radiologists. Acceptance criteria for above experiments were defined based on a literature review. In all validation experiments, AI-Rad Companion Brain MR WMH feature passed the acceptance criteria, the Pearson correlation coefficient between the WMH volumes estimated by our software and ground truth annotation was 0.96. The interclass correlation coefficient between the WMH volumes estimated by our software and ground truth annotation was 0.94. The segmentation accuracy of the WMH reaches a Dice score of 0.60. The F1-score of WMH detection is 0.67.
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{14}
SIEMENS Healthineers
| | Subject Device | | | | | Icobrain (K192130) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | All | Gender | | Field Strength | | |
| | | | M | F | 1.5 T | | 3.0 T |
| # Data | | 100 | 39 | 61 | 38 | 62 | 51 |
| Dice | Mean | 0.60 | 0.60 | 0.59 | 0.62 | 0.58 | 0.58 |
| | Med | 0.62 | 0.60 | 0.63 | 0.64 | 0.60 | N.A. |
| | STD | 0.14 | 0.12 | 0.17 | 0.15 | 0.16 | N.A. |
| | 95% CI | [0.57, 0.63] | [0.56, 0.64] | [0.54, 0.63] | [0.57, 0.66] | [0.54, 0.62] | N.A. |
| ASSD | Mean | 0.05 | 0.04 | 0.05 | 0.08 | 0.03 | N.A. |
| | Med | 0.00 | 0.00 | 0.00 | 0.01 | 0.00 | N.A. |
| | STD | 0.15 | 0.11 | 0.17 | 0.21 | 0.08 | N.A. |
| | 95% CI | [0.02, 0.08] | [0.01, 0.08] | [0.02, 0.10] | [0.03, 0.16] | [0.01, 0.05] | N.A. |
| | Subject Device | | | |
| --- | --- | --- | --- | --- |
| | | Siemens | GE | Philips |
| # Data | | 40 | 30 | 30 |
| Dice | Mean | 0.64 | 0.56 | 0.55 |
| | Med | 0.67 | 0.60 | 0.59 |
| | STD | 0.15 | 0.14 | 0.16 |
| | 95% CI | [0.60, 0.69] | [0.51, 0.61] | [0.50, 0.61] |
| ASSD | Mean | 0.02 | 0.09 | 0.04 |
| | Med | 0.00 | 0.01 | 0.00 |
| | STD | 0.06 | 0.23 | 0.11 |
| | 95% CI | [0.00, 0.04] | [0.03, 0.19] | [0.00, 0.08] |
DICE and ASSD Results for the AI-Rad Companion Brain MR White Matter Hyperintensities Algorithm
# White Matter Hyperintensities Follow-up Features
Performance testing for AI-Rad Companion Brain MR WMH follow-up feature was performed on 1.5T and 3T test datasets from 165 subjects, which included Multiple Sclerosis patients (MS) and Alzheimer's patients (AD) scanned using Siemens, GE and Philips scanners. Testing data has balanced distribution with respect to gender and age of the patient according to target patient population.
Accuracy was validated by comparing the results of the subject device to manual annotated ground truth from three radiologists. Acceptance criteria for above experiments were defined based on a literature review. In all validation experiments, AI-Rad Companion Brain MR WMH follow-up feature passed the acceptance criteria, the Pearson correlation coefficient between the new or enlarged WMH volumes estimated by our software and ground truth annotation was 0.76. The segmentation accuracy of the new or enlarged WMH reaches an average Dice score of 0.59. The average F1-score of new or enlarged WMH detection is 0.71.
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{15}
SIEMENS Healthineers
For each dataset, three sets of ground truth of white matter hyperintensity changes between two time points are annotated manually. Each set is annotated by a disjoint group of annotator, reviewer and clinical expert, with the expert randomly assigned per case. For each test dataset, the three initial annotations are annotated by three different in-house annotators, then each initial annotation is reviewed by the in-house reviewer. Afterwards, each initial annotation is reviewed by the referred clinical expert.
## Standard Annotation Process:
For each dataset, three sets of ground truth of white matter hyperintensity changes between two time points are annotated manually. Each set is annotated by a disjoint group of annotator, reviewer, and clinical expert, with the expert randomly assigned per case to minimize annotation bias. For each test dataset, the three initial annotations are annotated by three different in-house annotators. Then, each initial annotation is reviewed by the in-house reviewer. Afterwards, each initial annotation is reviewed by the referred clinical expert. The clinical expert reviews and corrects the initial annotation of the changed WMH areas according to the annotation protocol. If the corrections are significant and time-consuming, the corrections are communicated to the annotator for correction and then re-reviewed.
## Testing & Training Data Independence:
WMH follow-up algorithm does not include any machine learning/ deep learning component. The training data used for the fine tuning the hyper parameters of WMH follow-up algorithm is independent of the data used to test the white matter hyperintensity algorithm follow up algorithm.
## 11. Summary of Nonclinical Tests
Based on the nonclinical performance documented within the Scientific Evaluation, AI-Rad Companion Brain MR VA60 was found to have a safety and effectiveness profile that is similar to the predicate. Since the predicate device was cleared based on the results of the prior conducted scientific evaluation, the same methodology was required to support the substantial equivalence. The nonclinical data and verification and validation results supports the safety and effectiveness of the subject device in that it should perform comparable to the predicate device that is currently marketed.
## 12. Summary of Clinical Tests
The predicate (K232305) was not validated using clinical tests and therefore no clinical tests were conducted to test the performance and functionality of the modifications introduced within AI-Rad Companion Brain MR. Verification and validation of the enhancements and improvements have been performed and these modifications have been validated for their intended use. No animal testing has been performed on the subject device.
## 13. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk management is ensured via ISO 14971:2019 compliance to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
{16}
SIEMENS Healthineers
development of the product. These risks are controlled via measures realized during software development, testing and product labeling.
Furthermore, the device is intended for healthcare professionals familiar with the post processing of magnetic resonance images.
## 14. Conclusion
Based on the discussion and validation testing and performance data above, the proposed device is determined to be as safe and effective as its predicate device, AI-Rad Companion Brain MR VA50 (K232305).
AI-Rad Companion Brain MR
Traditional 510(k)
Siemens Medical Solutions USA, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.