K253047 · Nipro Medical Corporation · FMI · Mar 24, 2026 · General Hospital
Device Facts
Record ID
K253047
Device Name
PRESSONE™
Applicant
Nipro Medical Corporation
Product Code
FMI · General Hospital
Decision Date
Mar 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PRESSONE™ is indicated for use with syringes for general purpose fluid injection/withdrawal.
Device Story
PRESSONE™ is a sterile, single-use, disposable hypodermic needle consisting of a stainless-steel cannula and a polypropylene luer-lock hub. It is used by healthcare professionals to deliver or withdraw fluids from the body. The device operates via manual pressure applied through a connected syringe or infusion set, utilizing pressure differentials to facilitate fluid transfer through the single-lumen cannula. The needle is inserted into the skin and underlying tissue to perform standard medical procedures. The device is provided in a hard plastic protective case to maintain sterility until the point of use.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including performance testing per ISO 80369-7, ISO 9626, ISO 7864, and USP <788>. Biocompatibility was assessed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility). Packaging, sterility, and shelf-life were validated per ISO 11137-1, ASTM D4169, ASTM F1980-16, ASTM F1929-15, and ASTM F88/F88M-21.
Technological Characteristics
Materials: stainless steel cannula, polypropylene hub. Dimensions: 25G x 5/8". Sterilization: gamma radiation. Connectivity: ISO 80369-7 compliant luer-lock hub. Standards: ISO 9626 (tubing), ISO 7864 (sterile needles), ISO 10993-1 (biocompatibility), ISO 11137-1 (sterilization).
Indications for Use
Indicated for use with syringes for general purpose fluid injection/withdrawal in patients requiring fluid administration or aspiration.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 24, 2026
Nipro Medical Corporation
Jessica Oswald-Mcleod
Director, Regulatory Affairs
3150 NW 107th Ave.
Miami, Florida 33172
Re: K253047
Trade/Device Name: PRESSONETM
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: February 26, 2026
Received: February 26, 2026
Dear Jessica Oswald-Mcleod:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253047 - Jessica Oswald-Mcleod
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253047 - Jessica Oswald-Mcleod
Page 3
Sincerely,
Shruti N. Mistry -S
Shruti Mistry
Assistant Director, Injection Devices
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253047 | |
| Device Name PRESSONE™ | |
| Indications for Use (Describe) PRESSONE™ is indicated for use with syringes for general purpose fluid injection/withdrawal. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
NIPRO
K253047 | 510(k) Summary - PRESSONE™ | Page 1
# 510(k) Summary - PRESSONE™
Date Created: March 23, 2026
## Contact Details (21 CFR 807.92(a)(1))
Applicant Name: Nipro Medical Corporation
Applicant Address: 3150 NW 107th Ave. Miami FL 33172 United States
Applicant Contact Telephone: 305-432-6699
Applicant Contact: Mrs. Jessica Oswald-McLeod
Applicant Contact Email: Jessicao@nipromed.com
## Device Name (21 CFR 807.92(a)(2))
Device Trade Name: PRESSONE™
Common Name: Hypodermic single lumen needle
Classification Name: Needle, Hypodermic, Single Lumen
Regulation Number: 880.5570
Product Code: FMI
## Legally marketed predicate device (21 CFR 807.92(a)(3))
K070440, BD Hypoint Needle, Product Code FMI
## Device Description Summary (21 CFR 807.92(a)(4))
PRESSONE™ is a sterile, single-use disposable hypodermic needle designed to deliver or withdraw fluids from the body via a plastic luer-lock hub and single-lumen stainless steel cannula. PRESSONE™ is for prescription use only.
The needle size is 25G x 5/8". It is sterilized by gamma radiation with a shelf-life of 5 years.
Each needle is individually packaged in a hard plastic container and sealed with a label to maintain sterility until the point of use. The case serves as the needle protector.
PRESSONE™ enables fluid transfer through manual pressure applied via a connected syringe or infusion set equipped with an ISO 80369-7 compliant luer-lock hub, for use in standard medical procedures involving fluid administration or withdrawal.
## Intended Use/Indications for Use (21 CFR 807.92(a)(5))
PRESSONE™ is indicated for use with syringes for general purpose fluid injection/withdrawal.
## Indications for Use Comparison (21 CFR 807.92(a)(5))
The indications for use are the different from that of the predicate device; However, the intended use of the devices is the same.
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. I Miami, FL I 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
{5}
NIPRO
K253047 | 510(k) Summary - PRESSONE™ | Page 2
Technological Comparison (21 CFR 807.92(a)(6))
| Technological Characteristic | Subject device: PRESSONE™ | Predicate device: Hypoint | Conclusion |
| --- | --- | --- | --- |
| Indications for Use | PRESSONE™ is indicated for use with syringes for general purpose fluid injection/withdrawal. | The BD Hypoint™ Needle is intended for use with syringes and injection devices for general purpose fluid injection / aspiration. | The indications for use are different; however, the intended use of the devices are the same. |
| Mechanism of Action | Mechanical / manual | Mechanical / manual | Same |
| Technology Overview | The PRESSONE™ achieves its intended use through mechanical penetration of the skin and underlying tissue using a sharpened, beveled stainless-steel tip. Fluid transfer occurs via a single internal lumen, utilizing pressure differentials created by attached compatible devices (e.g., syringe). | The BD Hypoint™ achieves its intended use through mechanical penetration of the skin and underlying tissue using a sharpened, beveled stainless-steel tip. Fluid transfer occurs via a single internal lumen, utilizing pressure differentials created by attached compatible devices (e.g., syringe). | Same |
| Device Components | Stainless Steel cannula attached to a polypropylene hub | Stainless Steel cannula attached to a polypropylene hub | Same |
| Needle Gauge | 25G | 25G | Same |
| Needle Length | 5/8" | 5/8" | Same |
| Gauge color | Orange (ISO 6009) | Orange (ISO 6009) | Same |
| Sterilization Method | Gamma radiation | Gamma radiation | Same |
| Single-Use | yes | yes | Same |
| Shelf Life | 5 years | 5 years | Same |
| Individual packaging | Hard container with peel off label | Hard container with peel off label | Same |
| Performance Testing | ISO 9626, ISO 80369-7, ISO 7864 | ISO 9626, ISO 80369-7, ISO 7864 | Same |
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. | Miami, FL | 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
{6}
NIPRO
K253047 | 510(k) Summary - PRESSONE™ | Page 3
# Non-Clinical and/or Clinical Tests Summary & Conclusions (21 CFR 807.92(b))
The submitted performance testing is representative of the functional testing that is performed at final inspection and testing during routine manufacturing. Bench testing was performed on the subject device and the predicate device. The following performance standards were used.
1) ISO 80369-7:2016, Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications (General I (QS/RM)
2) ISO 9626:2016, Stainless steel needle tubing for the manufacturer of medical devices (General Plastic Surgery/General Hospital)
3) ISO 7864 Fourth Edition 2016-08-01, Sterile Hypodermic Needles for Single Use - Requirements and Test Methods. (General Plastic Surgery/General Hospital)
4) USP <788> "Particulate Matter in Injections"
Biocompatibility was assessed to ISO 10993-1:2018. Endpoints include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity and Hemocompatibility. All requirements were met.
The packaging, sterility and shelf -life are validated to standards
1) ISO 11137-1:2025 Sterilization of health care products - Radiation. Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices and
2) ASTM D4169, Standard Practice for Performance Testing of Shipping Containers and Systems. All requirements were met.
3) ASTM F1980-16, Standard Guide for Accelerated Aging Of Sterile Barrier Systems For Medical Devices
4) ASTM F1929-15, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
5) ASTM F88/F88M-21, Standard Test Method for Seal Strength of Flexible Barrier Materials
No Clinical testing was performed.
Non-clinical testing results confirm that the subject device complies with applicable voluntary consensus standards pertaining to performance, biocompatibility, packaging, transportation, and sterilization. All testing was successfully conducted in accordance with internal protocols, as well as relevant national and international standards.
Internal verification and validation activities demonstrate that the PRESSONE™ meets its design specifications and exhibits functional performance equivalent to the predicate device. These results support the claim of substantial equivalence to the predicate device in both design and technological characteristics.
NIPRO MEDICAL CORPORATION
3150 NW 107th Ave. | Miami, FL | 33172 USA
T: +1 305.599.7174
info@nipromed.com | www.nipro.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.