The HemoSphere Stream™ Module when used with a Smart Pressure Controller (PC1Q) and VitaWave™ Plus Finger Cuff is indicated for use in adult patients to provide continuous, non-invasive arterial pressure waveform output to a compatible multi-parameter patient monitor. The device is designed for use in clinical environments requiring continuous assessment of blood pressure waveform morphology, without the need for an invasive catheter. The HemoSphere Stream™ Module is intended to be used by qualified personnel or trained professionals in a hospital setting. The HemoSphere Stream™ Module is intended for use with compatible VitaWave™ Plus Finger Cuffs. The HemoSphere Stream™ Module is intended to transmit a continuous, non-invasive blood pressure waveform to a compatible patient monitor.
Device Story
HemoSphere Stream Module functions as a component of a non-invasive blood pressure monitoring system. It utilizes a Smart Pressure Controller (PC1Q) and VitaWave™ Plus Finger Cuff to acquire arterial pressure signals from the patient's finger. The module processes these signals to generate a continuous, non-invasive arterial pressure waveform, which is transmitted to a compatible multi-parameter patient monitor. Operated by qualified healthcare professionals in hospital settings, the device provides real-time waveform morphology assessment. Unlike its predicate, the Stream Module is limited to waveform transmission and does not calculate or display derived hemodynamic parameters. This focused functionality aims to provide a cost-effective solution for continuous blood pressure monitoring, benefiting patients by enabling hemodynamic assessment without the risks associated with invasive arterial catheters.
Clinical Evidence
Bench testing only. Verification activities included software and system verification performed in accordance with FDA guidance. Testing confirmed the device performs as intended, with pass/fail criteria based on the predicate device specifications. No clinical data was required or provided.
Technological Characteristics
Non-invasive blood pressure measurement system consisting of a module, Smart Pressure Controller (PC1Q), and finger cuff. Functions as a programmable diagnostic computer (21 CFR 870.1425). Provides continuous arterial pressure waveform output to external monitors. Software-based signal acquisition and transmission. Identical waveform acquisition technology to predicate K243781.
Indications for Use
Indicated for adult patients requiring continuous, non-invasive arterial pressure waveform monitoring in clinical settings, avoiding invasive catheterization.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 21, 2025
Edwards Lifesciences
Varad Raghuwanshi
Director Regulatory Affairs
One Edwards Way
Irvine, California 92614
Re: K253034
Trade/Device Name: HemoSphere Stream Module
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable diagnostic computer
Regulatory Class: Class II
Product Code: DQK, DXN
Dated: September 19, 2025
Received: September 22, 2025
Dear Varad Raghuwanshi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253034 - Varad Raghuwanshi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253034 - Varad Raghuwanshi
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
for
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253034 | ? |
| Please provide the device trade name(s). | | ? |
| HemoSphere Stream Module | | |
| Please provide your Indications for Use below. | | ? |
| The HemoSphere Stream™ Module when used with a Smart Pressure Controller (PC1Q) and VitaWave™ Plus Finger Cuff is indicated for use in adult patients to provide continuous, non-invasive arterial pressure waveform output to a compatible multi-parameter patient monitor. The device is designed for use in clinical environments requiring continuous assessment of blood pressure waveform morphology, without the need for an invasive catheter. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253034
Page 1 of 3
510(k) Summary – HemoSphere Advanced Monitoring Platform
I. Submitter:
Sponsor: Edwards Lifesciences LLC
One Edwards Way
Irvine, CA 92614
Establishment Registration Number: 2015691
Contact Person: Varad Raghuwanshi
Director, Regulatory Affairs
One Edwards Way
Irvine, CA 92614
Date Prepared: September 15th, 2025
October 16, 2025
II. Device Information:
Device Name: HemoSphere Stream Module
Trade Name: HemoSphere Stream Module
Common Name: Noninvasive blood pressure measurement system
Classification Name: Programmable Diagnostic Computer 21 CFR 870.1425
Noninvasive blood pressure measurement system 21 CFR 870.1130
Product Code and Regulatory Class: DQK, Class II
DXN, Class II
III. Predicate Device
Primary HemoSphere ClearSight Module manufactured by Edwards Lifesciences, Frederick Device: K243781 cleared July 23rd, 2025.
{5}
K253034
Page 2 of 3
## IV. Device Description
**Device Description:**
The HemoSphere Stream™ Module when used with the Smart Pressure Controller (PC1Q) and VitaWave™ Plus Finger Cuff is indicated for use in adult patients to provide continuous, non-invasive arterial pressure waveform output to a compatible multi-parameter patient monitor.
## V. Indications for Use:
The HemoSphere Stream™ Module when used with a Smart Pressure Controller (PC1Q) and VitaWave™ Plus Finger Cuff is indicated for use in adult patients to provide continuous, non-invasive arterial pressure waveform output to a compatible multi-parameter patient monitor. The device is designed for use in clinical environments requiring continuous assessment of blood pressure waveform morphology, without the need for an invasive catheter.
Refer to the VitaWave Plus finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
## Intended Use:
The HemoSphere Stream™ Module is intended to be used by qualified personnel or trained professionals in a hospital setting.
The HemoSphere Stream™ Module is intended for use with compatible VitaWave™ Plus Finger Cuffs.
The HemoSphere Stream™ Module is intended to transmit a continuous, non-invasive blood pressure waveform to a compatible patient monitor.
## VI. Comparison of Technological Characteristics with the Predicate Devices:
The purpose of this 510(k) submission is to introduce the following modification to the HemoSphere ClearSight Module (Cleared in K243781 on July 23rd, 2025):
- Introduction of the HemoSphere Stream Module, a configuration of the HemoSphere ClearSight™ Module to provide limited functionality
The existing previously cleared HemoSphere ClearSight Module (K243781, cleared July 23, 2025) is intended for use with the Smart Pressure Controller and a compatible Edwards finger cuff to noninvasively measure blood pressure waveform, blood pressure parameters, and associated hemodynamic parameters. The ClearSight
{6}
K253034
Page 3 of 3
Module acquires the waveform, calculates the parameters, and displays them on the connected monitor.
To make this technology more accessible in a cost-effective manner and support broader patient benefit, Edwards has introduced the HemoSphere Stream Module, which is designed to provide the continuous noninvasive blood pressure waveform to the connected monitor. The HemoSphere Stream Module provides only the continuous blood pressure waveform to the connected monitor and does not calculate or display blood pressure parameters or derived hemodynamic parameters.
## Performance Data:
The following verification activities were conducted to assess the modification included in this submission. Pass/fail criteria were based on the specifications cleared for the predicate device, and the results demonstrated substantial equivalence.
### Software Verification
Software verification was performed in accordance with FDA’s Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. This process included verification of software design, development, and traceability. There were no changes to how the HemoSphere Stream Module acquires and provides continuous blood pressure waveforms compared to the HemoSphere ClearSight Module. The same methods and acceptance criteria as the predicate device (K243781) were applied. All tests passed successfully.
### System Verification
System verification confirmed that the HemoSphere Stream Module performs as intended, providing the blood pressure waveform to the connected monitor without introducing any issues that could impact device safety or effectiveness. The same methods and acceptance criteria as the predicate device (K243781) were used. All tests passed successfully.
## Conclusions
The technological characteristics of the subject and predicate devices are identical with respect to blood pressure waveform functionality. The HemoSphere Stream Module successfully completed functional and performance testing, including software and system verification. These results confirm that the modification does not adversely affect the device’s safety or effectiveness and that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.