K253027 · Ortobio S.A. · MBI · May 6, 2026 · Orthopedic
Device Facts
Record ID
K253027
Device Name
Anchor with Fiber Wire and Disposable Inserter
Applicant
Ortobio S.A.
Product Code
MBI · Orthopedic
Decision Date
May 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ortobio Anchor with Fiber Wire and Disposable Inserter is intended to be used for suture or tissue fixation in the elbow, shoulder, hand, wrist, foot, ankle and knee. Specific indications are listed below: • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio- Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
Device Story
Device consists of titanium alloy anchor with external threads and hexagonal head; preloaded with UHMWPE suture; preassembled into disposable applicator (stainless steel shaft, polypropylene handle). Used by surgeons in orthopedic procedures (elbow, shoulder, hand, wrist, foot, ankle, knee) for soft tissue-to-bone fixation. Surgeon inserts anchor into bone; suture secures tissue to anchor. Provides mechanical fixation for ligament/tendon repairs. MR Conditional. Single-use, sterile.
Clinical Evidence
No clinical data included. Bench testing only: mechanical performance (insertion torque, inclined insertion, quasi-static/inclined/post-fatigue pullout) compared to predicate; MR compatibility testing; biocompatibility per ISO 10993-1; sterilization validation per ISO 11135; shelf-life stability testing.
Indicated for suture or tissue fixation in elbow, shoulder, hand, wrist, foot, ankle, and knee for patients requiring ligament or tendon repair/reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Suture Anchors - HTA Headless Titanium Anchor and ZIP Anchors (K223114)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 6, 2026
Ortobio S.A.
Tatiana Jabor Botura
Regulatory Affairs Specialist
PR Serviços Regulatórios Administrativos Ltda
Edifício Pórtico Sul - R. José Jaime Delibo, 160,
Nova Aliança
Ribeirão Preto, SP 14026-563
Brazil
Re: K253027
Trade/Device Name: Anchor with Fiber Wire and Disposable Inserter
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: MBI
Dated: April 07, 2026
Received: April 07, 2026
Dear Tatiana Jabor Botura:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253027 - Tatiana Jabor Botura
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 809; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253027 - Tatiana Jabor Botura
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253027 | ? |
| Please provide the device trade name(s). | | ? |
| Anchor with Fiber Wire and Disposable Inserter | | |
| Please provide your Indications for Use below. | | ? |
| The Ortobio Anchor with Fiber Wire and Disposable Inserter is intended to be used for suture or tissue fixation in the elbow, shoulder, hand, wrist, foot, ankle and knee. Specific indications are listed below: • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio- Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{4}
Traditional510(k)
Anchor with Fiber Wire and Disposable Inserter
510(k) Summary
ADMINISTRATIVE INFORMATION
Sponsor/Manufacturer Name
Ortobio S.A.
Av. Mauá, 1133 - Zona 03, Maringá - PR, 87050-020 Phone:
+55 (44) 3269-5045
Consultants
Tatiana Jabor Botura and Letícia Teixeira
Regulatory Affairs Specialists
PR Serviços Regulatórios Administrativos Ltda
Email: tatiana@passarini.com.br
Telephone +55 (16) 988280027
Date Prepared
August 25, 2025
DEVICE NAME AND CLASSIFICATION
Trade/ Proprietary Name
Anchor with Fiber Wire and Disposable Inserter
Common Name
Bone Anchor
Classification Name
Fastener, Fixation, Nondegradable, Soft Tissue
Product Code
MBI
Classification Regulation
21 CFR 888.3040, Class II
Review Panel
Orthopedic
PREDICATE DEVICE INFORMATION
Predicate Devices
K223114 - Suture Anchors - HTA Headless Titanium Anchor
and ZIP Anchors
Reference Device
K201083 - Parcus V-lox Titanium Suture Anchors, Parcus
Miti Suture Anchors
Indications for use
The Ortobio Anchor with Fiber Wire and Disposable Inserter is intended to be used for suture or tissue fixation in the elbow, shoulder, hand, wrist, foot, ankle and knee. Specific indications are listed below:
- Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Ortobio S.A
{5}
Traditional510(k)
Anchor with Fiber Wire and Disposable Inserter
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
# Subject Device Description
The Fiber Wire Anchor with Disposable Applicator is a sterile, single-use medical device composed of a titanium alloy anchor featuring an external threaded portion and a hexagonal head with a central hole designed for suture thread passage.
The product is packaged with one or two preloaded non-absorbable sutures and is preassembled into a disposable applicator. Optional configurations include disposable suture needles.
The applicator is comprised of a stainless-steel shaft (AISI 304), a polypropylene handle, and an O-ring located at the handle's distal end. The suture thread is made of ultra-high molecular weight polyethylene (UHMWPE), size 2 (0.5 mm). Disposable needles are made of stainless steel (ASTM F899).
The anchor is intended for the fixation of soft tissue to bone and is typically used in ligament and tendon repair procedures in the shoulder, foot, elbow, ankle, hand, and wrist. The anchor is manufactured from titanium alloy in accordance with ISO 5832-3:1997 – Implants for Surgery – Metallic Materials – Part 3: Wrought Titanium 6-Aluminum 4-Vanadium Alloy.
Ortobio S.A
{6}
Traditional510(k)
Anchor with Fiber Wire and Disposable Inserter
# Substantial Equivalence
A comparison between the subject device and predicates is presented in the table below:
Table 1. Substantial equivalence
| Trade Name information | Subject Device | Predicate Device | Reference Predicate | Equivalent Discussion |
| --- | --- | --- | --- | --- |
| | Anchor with Fiber Wire and Disposable Inserter | Suture Anchors - HTA Headless Titanium Anchor and ZIP Anchors | Parcus V-lox Titanium Suture Anchors, Parcus Miti Suture Anchors | |
| | Ortobio | GM dos Reis | Parcus Medical, LLC | - |
| K number | - | K223114 | K201083 | - |
| Regulation Number | 21 CFR 888.3040 | 21 CFR 888.3040 | 21 CFR 888.3040 | Equivalent |
| Product Code | MBI | MBI | MBI | Equivalent |
| Classification Name | Fastener, Fixation, Nondegradable, Soft Tissue | Fastener, Fixation, Nondegradable, Soft Tissue | Fastener, Fixation, Nondegradable, Soft Tissue | Equivalent |
| Class | II | II | II | Equivalent |
| Indications for Use | The Ortobio Anchor with Fiber Wire and Disposable Inserter is intended to be used for suture or tissue fixation in the elbow, shoulder, hand, wrist, foot, ankle and knee. Specific indications are listed below:Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament ReconstructionShoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral ReconstructionHand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstructionKnee: Medial Collateral | The Suture Anchors - HTA Headless Titanium Anchor and ZIP Anchors are indicated to be used for suture or tissue fixation in the elbow, shoulder, hand, wrist, foot, ankle, knee, and hip. Specific indications are listed below:Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament ReconstructionShoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral ReconstructionHand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstructionKnee: Medial Collateral | The Parcus Miti Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament RepairElbow: Tennis Elbow Repair, Biceps Tendon Reattachment. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.The Parcus V-loxTM Titanium Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended | Equivalent |
Ortobio S.A
{7}
Traditional 510(k)
Anchor with Fiber Wire and Disposable Inserter
| | Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis | Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hip: Capsular repair, acetabular labral repair. | for the following indications: Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair. Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC. Hip: Acetabular Labral Repair | |
| --- | --- | --- | --- | --- |
| Anchor Diameter | ø1,7 mm; ø 1,9 mm, 2,0 mm; ø2,8 mm; ø4,0 mm; ø5,0 mm; ø5,5 mm | ø1,7 mm; ø2,2 mm; ø2,7 mm; ø3,5 mm and ø5,0 mm | Miti Suture Anchor: ø 2.0mm, ø 2.5mm and ø 3.5mm V-lox Titanium Suture Anchor: ø4.5mm, ø 5.0mm, ø 5.5mm and ø 6.5mm. | The specifications and dimensions of the device in question are within the range of the predicate and reference devices. Minor differences in dimensions do not raise any safety or efficacy concerns. |
| Suture (USP) | #2; #2-0 | #1; #2; #0; #2-0; #3-0; #4-0 | 3-0 and #2 | Equivalent |
| Anchor Material | Titanium (ASTM F136) | Titanium (ASTM F136) | Titanium (ASTM F136) | Equivalent |
| Suture Material | UHMWPE | UHMWPE | UHMWPE | Equivalent |
| Sterilization | Sterile | Sterile | Sterile | Equivalent |
| Use | Single use | Single use | Single use | Equivalent |
| Principle of operation | Fixation of soft tissue to the bone | Fixation of soft tissue to the bone | Fixation of soft tissue to the bone | Equivalent |
## Comparison of Technological Characteristics with the Predicate Device
The subject device, Ortobio Anchor with Fiber Wire and Disposable Inserter, is substantially equivalent in indications and design principles to the predicate and reference devices. The subject and predicate devices have equivalent intended use and share the same principle of operation for
Ortobio S.A
Page 4 of 4
{8}
Traditional510(k)
Anchor with Fiber Wire and Disposable Inserter
soft tissue fixation to bone. They are all manufactured from titanium alloy (ASTM F136) and use UHMWPE sutures fixed by integrated eyelets, provided sterile and single use. The subject device dimensions and suture sizes are within the range of the predicate and reference devices. These differences do not raise new issues of safety or effectiveness. Any difference in technological characteristics do not raise new issues of safety or efficacy. No clinical data was included in this submission
## Non-clinical performance data
This submission was prepared in accordance with the FDA Guidance "Bone Anchors - Premarket Notification (510(k)) Submissions" and the "Class II Special Controls Guidance Document: Surgical Sutures".
The Ortobio Anchor with Fiber Wire and Disposable Inserter was evaluated for use in the MR environment and determined to be MR Conditional, based on testing performed in accordance with the FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" and applicable ASTM standards. Worst-case devices were tested for magnetically induced force, torque, RF heating, and image artifact, and the results established the MR system conditions suitable for safe use.
Performance testing endpoints included insertion torque, inclined insertion, quasi-static pullout, inclined pullout, and post-fatigue pullout, all performed in accordance with FDA guidance and ASTM standards. Statistical analysis using the TOST method and non-inferiority evaluation demonstrated that the performance of the Ortobio anchors is equivalent or not inferior to the predicate HTA Headless Titanium Anchor (K223114). The results consistently confirmed that no significant differences exist between the subject and predicate devices, supporting the conclusion of substantial equivalence in mechanical performance
Biological evaluation endpoints were selected in accordance with ISO 10993-1, considering the chemical characteristics of the device, as well as the nature, frequency, and duration of body contact associated with the intended use.
Sterilization was validated according to ISO 11135:2014 to achieve a SAL of $10^{-6}$, with routine control and monitoring parameters established. Shelf life was supported by accelerated aging testing, demonstrating stability of the device for five years.
Ortobio S.A
{9}
Traditional510(k)
Anchor with Fiber Wire and Disposable Inserter
## Conclusion
The documentation submitted in this premarket notification demonstrates that Anchor with Fiber Wire and Disposable Inserter has comparable features and performance and, therefore, is substantially equivalent to the predicate and reference device.
Ortobio S.A
Page 6 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.