DS Core Detect

K253009 · Dentsply Sirona, Inc. · MYN · Jan 7, 2026 · Radiology

Device Facts

Record IDK253009
Device NameDS Core Detect
ApplicantDentsply Sirona, Inc.
Product CodeMYN · Radiology
Decision DateJan 7, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2070
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Periapical radiolucency detection——Sensitivity: 0.78 (High sensitivity), 0.66 (Standard sensitivity); Specificity: 0.93 (High sensitivity), 0.97 (Standard sensitivity)——Standalone study: 306 CBCTs—
Periapical radiolucency detection——Unaided AUC: 0.415; Aided AUC: 0.639; Difference: 0.224——MRMC study: 11 U.S. readers11 (U.S. readers)

Indications for Use

DS Core Detect is a concurrent read, computer assisted detection software intended to aid the dentist in the incidental detection of permanent teeth that may be associated with periapical radiolucencies in CBCT images. The device is not intended as a replacement for a complete dentist’s review or a replacement of their clinical judgment. The device is to be used by licensed dentists. The device is designed for patients who already have CBCT data obtained for other reasons. DS Core Detect does not generate X-ray images and does not provide an indication for taking such images. The device is intended for adults aged 22 years and older with permanent teeth.

Device Story

Cloud-based SaMD; utilizes ML-enabled CADe to analyze existing CBCT images for periapical radiolucency (PARL). Input: CBCT data acquired for other clinical reasons. Processing: identifies teeth potentially associated with PARL; generates a list of flagged teeth and a visual PARL 'map'. Output: displayed within dental image viewer for practitioner review. Workflow: dentist reviews findings; retains full autonomy to confirm or reject device-flagged teeth. Clinical impact: aids incidental detection of PARL; supports diagnostic decision-making; does not replace standard of care or clinical judgment. Used in dental clinics by licensed dentists.

Clinical Evidence

Performance validated via standalone study (n=306 CBCTs) and MRMC reader study (n=11 U.S. dentists). Standalone sensitivity: 0.78 (High) / 0.66 (Standard); specificity: 0.93 (High) / 0.97 (Standard). MRMC study showed significant improvement in AUC (0.415 unaided vs 0.639 aided; difference 0.224, 95% CI 0.124-0.324) and sensitivity (0.421 unaided vs 0.649 aided; difference 0.227, 95% CI 0.124-0.330) with minimal impact on specificity.

Technological Characteristics

Cloud-based SaMD; ML-enabled CADe algorithm. Input: 3D CBCT images. Output: list of teeth with potential PARL and visual PARL map. Operates on existing CBCT data; no X-ray generation. Designed for adult patients (22+) with permanent teeth. Software-based analysis.

Indications for Use

Indicated for licensed dentists to aid in incidental detection of permanent teeth associated with periapical radiolucencies (PARL) in adult patients (aged 22+) using existing CBCT data. Not for use as a replacement for clinical judgment or standard of care diagnosis.

Regulatory Classification

Identification

Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations. (ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable. (iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures ( *e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation ( *e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following: (i) A detailed description of the patient population for which the device is indicated for use. (ii) A detailed description of the intended reading protocol. (iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device. (iv) A detailed description of the device inputs and outputs. (v) A detailed description of compatible imaging hardware and imaging protocols. (vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level ( *e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions. (viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 7, 2026 Dentsply Sirona Inc % Deepthi Paknikar Senior Manager, Regulatory Affairs 221 West Philadelphia St. YORK, PA 17401 Re: K253009 Trade/Device Name: DS Core Detect Regulation Number: 21 CFR 892.2070 Regulation Name: Medical Image Analyzer Regulatory Class: Class II Product Code: MYN Dated: December 5, 2025 Received: December 5, 2025 Dear Deepthi Paknikar: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253009 - Deepthi Paknikar Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253009 - Deepthi Paknikar Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253009 | | | Device Name DS Core Detect | | | Indications for Use (Describe) DS Core Detect is a concurrent read, computer assisted detection software intended to aid the dentist in the incidental detection of permanent teeth that may be associated with periapical radiolucencies in CBCT images. The device is not intended as a replacement for a complete dentist’s review or a replacement of their clinical judgment. The device is to be used by licensed dentists. The device is designed for patients who already have CBCT data obtained for other reasons. DS Core Detect does not generate X-ray images and does not provide an indication for taking such images. The device is intended for adults aged 22 years and older with permanent teeth. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” | | {4} Page 1 of 4 510(k) Summary 510(k) #: K253009 Applicant Name: Dentsply Sirona Inc Applicant Address: 221 West Philadelphia St. York PA 17401 United States Applicant Contact Telephone: 630-201-1612 Applicant Contact: Dr. Deepthi Paknikar Applicant Contact Email: Deepthi.Paknikar@dentsplysirona.com Device Trade Name: DS Core Detect Common Name: Medical image analyzer Classification Name: Medical Image Analyzer Regulation Number: 21 CFR 892.2070 Product Code(s): MYN Predicate Device: Videa Dental Assist (K232384) Common Name: Medical image analyzer Classification Name: Medical Image Analyzer Regulation Number: 21 CFR 892.2070 Product Code(s): MYN Device Description Summary The DS Core Detect device is a cloud-based SaMD that is a Computer Aided Detection "CADe" Machine Learning "ML" enabled device for the aid in incidental detection of permanent teeth that may have periapical radiolucency "PARL" on a CBCT image. The device is designed for patients who already have CBCT data obtained from a CBCT imaging system for other reasons. DS Core Detect does not generate X-ray images and does not provide an indication for taking such images. The device itself is accessed by the dental practitioner through their dental image viewer. From within the dental viewer the user can view the CBCT with results from the subject device. The device will provide a list of teeth that may potentially have PARL and a corresponding PARL "map" to visualize where the PARL is detected. The user has completely autonomy to confirm or reject a tooth that is flagged by the device as potentially PARL positive. The intended users of the device are trained and licensed dentists. The intended patient population for the device is adult patients with permanent teeth at least age 22. Intended Use/Indications for Use DS Core Detect is a concurrent read, computer assisted detection software intended to aid the dentist in the incidental detection of permanent teeth that may be associated with periapical radiolucencies in CBCT images. The device is not intended as a replacement for a complete dentist's review or a replacement of their clinical judgment. The device is to be used by licensed dentists. The device is designed for patients who already have CBCT data obtained for other {5} reasons. DS Core Detect does not generate X-ray images and does not provide an indication for taking such images. The device is intended for adults aged 22 years and older with permanent teeth. ## Indications for Use Comparison The minor differences in indications for use between the subject and predicate do not constitute a new intended use. Both devices are diagnostic aids for detection of PARL. The primary predicate device provides PARL findings for 2D periapical images whereas the subject device detects PARL positive teeth on a CBCT image. The predicate is indicated for bitewing, periapical, and panoramic radiographs and detects multiple pathologies, restorations, and normal anatomy, whereas the subject device is indicated only for incidental detection of PARL on CBCT. The subject device is also specific with regards to being used with CBCTs that have already been taken for other reasons, for aiding with incidental detection of PARL only. Both devices are SaMD and AI/ML enabled. Both devices are diagnostic aids only, not a replacement for standard of care diagnosis. ## Technological Comparison The main technological difference is that the predicate device provides PARL detection, as well as multiple other detections (anatomy, caries, etc.) for 2D dental radiographs, whereas the subject device provides PARL detection on 3D CBCT images. The predicate device spans a large age range and permanent and primary teeth findings, whereas the subject device is limited to permanent teeth for patients 22 and older. The subject device input is CBCT's taken for other reasons for the incidental detection of PARL, whereas the predicate device has 2D bitewing, periapical, and panoramic images as the input. The subject device will provide a list of teeth that may potentially have PARL which the user must confirm or reject and a corresponding PARL "map" to visualize where the PARL is detected. The predicate device outputs a set of bounding boxes for each finding, including PARL. The minor difference in technological characteristics do not raise a concern of substantial equivalence as demonstrated by the performance testing of the subject device. ## Non-Clinical and/or Clinical Tests Summary & Conclusions The DS Core Detect device underwent comprehensive performance testing including standalone testing and a Multi-Reader Multi-Case Reader Study (MRMC). Software verification and validation testing was completed per applicable standards and agency guidance. DS Core Detect indications detections in two operating points: Standard sensitivity and High sensitivity, whereby sensitivity describes the portion of detected teeth with evidence of periapical radiolucency. The standalone study assessed the performance of the device compared to a consensus ground truth. The MRMC study evaluated reader performance with and without the assistance of the device, in a fully crossed study design. Depending on the operating point, these teeth are detected with the following standalone study accuracy with 95% CI's: {6} 1. Sensitivity: High sensitivity 0.78 [0.74; 0.81], Standard sensitivity 0.66 [0.62; 0.70] 2. Specificity: High sensitivity 0.93 [0.92; 0.94], Standard sensitivity 0.97 [0.97; 0.98] Standalone Study Subgroup performance testing subjects included patients 22 and older from various clinical sites across the U.S., with permanent teeth, with CBCT images. Subgroup reporting included lesion size, age, device, gender, and endodontic treatment status, with results as follows: | Operating point | Group | | \( Sensitivity^1 \) | 95% confi-dence interval | \( Specificity^1 \) | 95% confi-dence interval | | --- | --- | --- | --- | --- | --- | --- | | High sensitivity | Lesion | >8 mm3 | 0.82 | [0.79; 0.86] | 0.93 | [0.92; 0.94] | | | | ≤ 8 mm3 | 0.65 | [0.59; 0.71] | 0.93 | [0.92; 0.94] | | | Age | >52 years | 0.76 | [0.71; 0.80] | 0.95 | [0.94; 0.96] | | | | ≤ 52 years | 0.8 | [0.75; 0.84] | 0.92 | [0.91; 0.93] | | | Device | Orthophos S/SL | 0.76 | [0.72; 0.81] | 0.93 | [0.91; 0.94] | | | | Axeos | 0.80 | [0.75;0.84] | 0.94 | [0.91; 0.94] | | | Gender | Male | 0.78 | [0.73;0.83] | 0.93 | [0.92; 0.95] | | | | Female | 0.79 | [0.74; 0.85] | 0.93 | [0.91; 0.95] | | | Endodontic treatment | Tooth treated | 0.82 | [0.78; 0.86] | 0.82 | [0.77; 0.87] | | | | Tooth un-treated | 0.72 | [0.67; 0.78] | 0.94 | [0.93; 0.95] | | Standard sensitivity | Lesion | >8 mm3 | 0.73 | [0.68; 0.77] | 0.97 | [0.97; 0.98] | | | | ≤ 8 mm3 | 0.50 | [0.43; 0.57] | 0.97 | [0.97; 0.98] | | | Age | >52 years | 0.64 | [0.58; 0.70] | 0.98 | [0.97; 0.98] | | | | ≤ 52 years | 0.68 | [0.62; 0.74] | 0.97 | [0.96; 0.98] | | | Device | Orthophos S/SL | 0.64 | [0.58; 0.69] | 0.97 | [0.96; 0.98] | | | | Axeos | 0.70 | [0.64; 0.76] | 0.98 | [0.97; 0.98] | | | Gender | Male | 0.67 | [0.60; 0.73] | 0.97 | [0.96; 0.98] | | | | Female | 0.70 | [0.64; 0.76] | 0.97 | [0.96; 0.98] | | | Endodontic treatment | Tooth treated | 0.75 | [0.70; 0.80] | 0.90 | [0.86; 0.94] | | | | Tooth un-treated | 0.56 | [0.50; 0.62] | 0.98 | [0.97; 0.98] | | 1 Result of a representative standalone study with 306 CBCTs | | | | | | | The MRMC Tooth Level AUC results are as follows: | Number of U.S. readers = 11 | Unaided | Aided | Difference | | --- | --- | --- | --- | | | OR Model-Based AUC Estimates (95% CIs) | | | | | 0.415 (0.279, 0.550) | 0.639 (0.529, 0.749) | 0.224 (0.124, 0.324) | | | Tooth Level Sensitivity (95% CIs) | | | | | 0.421 (0.281, 0.562) | 0.649 (0.535, 0.763) | 0.227 (0.124, 0.330) | | | Tooth Level Specificity (95% CIs) | | | | | 0.962 (0.928, 0.995) | 0.946 (0.915, 0.976) | -0.016 (-0.034, 0.002) | {7} Page 4 of 4 # Conclusion The standalone and MRMC study results collectively demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...