K253002 · Esthetic Medical, Inc. · QAI · Feb 19, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253002
Device Name
SkinStylus SteriLock MicroSystem (MP1209SL)
Applicant
Esthetic Medical, Inc.
Product Code
QAI · General, Plastic Surgery
Decision Date
Feb 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SkinStylus SteriLock® MicroSystem is intended to be used as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older. The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, and III in patients aged 22 years and older. The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of periorbital wrinkles in Fitzpatrick skin types I, II, III, IV, V, and VI in adults aged 34 years or older.
Device Story
Handheld, powered microneedling device; uses 5V 1A DC motor to reciprocate 36-needle array (2.5mm max length) to create micro-channels/injuries in skin. Operated by trained professionals in clinical settings. Device includes power source, motor body, depth adjustment, and single-use gamma-sterilized needle cartridge. Procedure involves cleansing skin, applying numbing cream and gliding agent, then performing microneedling at depths (0.5mm to 2.5mm) until pinpoint bleeding occurs. Output is mechanical skin injury intended to stimulate aesthetic improvement. Healthcare providers assess treatment response via standardized photography and Lemperle wrinkle grading. Benefits include improved appearance of periorbital wrinkles, facial acne scars, and abdominal hypertrophic scars.
Clinical Evidence
Single-arm, two-center clinical study (n=51) evaluated periorbital wrinkle improvement. Primary endpoint: mean improvement of one grade on Lemperle scale at 120/150 days. 80% of subjects showed improvement at Day 150 (blinded physician review). No treatment-related adverse events reported; side effects (erythema, discomfort, peeling) were mild/moderate and resolved within 8 days.
Technological Characteristics
Powered microneedling device; 36 stainless steel needles (6x6 array); 2.5mm max depth; 0.1mm depth increments; 6200-8840 RPM reciprocation. Powered by rechargeable Li-ion battery or AC/DC adapter. Single-use gamma-sterilized cartridges (polycarbonate body). Biocompatibility per ISO 10993; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for treatment to improve appearance of periorbital wrinkles in adults aged 34+ with Fitzpatrick skin types I-VI.
Regulatory Classification
Identification
A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.
Special Controls
The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 19, 2026
Esthetic Medical Inc.
Sean Song
Manager Regulatory Affairs
3600 E Burnett St
Long Beach, California 90815
Re: K253002
Trade/Device Name: SkinStylus SteriLock® MicroSystem (MP1209SL)
Regulation Number: 21 CFR 878.4430
Regulation Name: Microneedling Device For Aesthetic Use
Regulatory Class: Class II
Product Code: QAI
Dated: January 16, 2026
Received: January 16, 2026
Dear Sean Song:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253002 - Sean Song
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253002 - Sean Song
Page 3
Sincerely,
JODIE GIORDANO -S
Jodie Giordano, Ph.D.
Acting Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253002 | |
| Device Name SkinStylus SteriLock® MicroSystem | |
| Indications for Use (Describe) The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of periorbital wrinkles in Fitzpatrick skin types I, II, III, IV, V, and VI in adults aged 34 years or older. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
# 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).
| Submitter's Name: | Esthetic Medical Inc. |
| --- | --- |
| Submitter's Address: | 3600 East Burnett Street, Long Beach, California 90815 |
| Contact Person: | Sean Song |
| Telephone: | +1.509.413.6458 |
| Date Prepared: | February 19, 2026 |
| Device Trade Name: | SkinStylus SteriLock® MicroSystem |
| Common Name: | Powered microneedle device |
| Classification Name: | Microneedling device for aesthetic use |
| Regulation Number: | 21 CFR 878.4430 |
| Product Code: | QAI |
| Review Panel: | General & Plastic Surgery |
| Predicate Device: | K180778. Exceed Microneedling Device |
# Intended Use
The SkinStylus SteriLock® MicroSystem is intended to be used as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older.
The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, and III in patients aged 22 years and older.
The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of periorbital wrinkles in Fitzpatrick skin types I, II, III, IV, V, and VI in adults aged 34 years or older.
# Device Description
The SkinStylus SteriLock® MicroSystem (SSM) is a handheld microneedling device that uses an adjustable speed 5V 1A DC motor to rapidly reciprocate an array of 32-gauge microneedles that are no longer than $2.5\mathrm{mm}$ to create channels as well as microinjuries into the skin. The device consists of a power source, a motor body with depth adjustment, a removable nosecone interface, and a disposable, single-use, gamma ray sterilized cartridge, containing an array of microneedles.
1 The SkinStylus SteriLock® MicroSystem is already cleared for abdominal scarring under K200044 and for facial acne scarring under K231073. Facial wrinkles is the only indication that is new or changed under this submission.
{5}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
The SSM receives power from either a rechargeable lithium-ion battery or directly from an external AC/DC power adapter (input: 100-240 VAC/50-60 Hz, output: 5 VDC, 1 A).
## Technical specifications, including needle characteristics
| Characteristic | Specification |
| --- | --- |
| Needle geometry | 36 solid needles arranged in a 6 × 6 square |
| Max needle length | 2.5 mm |
| Needle protrusion setting | 0.0 to 2.5 mm with marked 0.1 mm increments |
| Maximum needle penetration in clinical application | 2.5 mm |
| Accuracy of needle penetration depth | ± 0.25 mm |
| Reciprocation rate | 6200 to 8840 RPM |
## Substantial Equivalence
The SkinStylus SteriLock® MicroSystem is substantially equivalent to the Exceed Microneedling Device (K180778), which is a prescription-only powered microneedling device indicated for the treatment of wrinkles in Fitzpatrick skin types I, II, and/or III in the following facial areas: glabellar frown lines, periorbital lines and cheek folds in adults aged 22 years or older.
Page 2 of 14
{6}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
Substantial Equivalence
| Description | Subject device
SkinStylus SteriLock® MicroSystem | Predicate device
Exceed Microneedling Device
(K180778) | Significant differences |
| --- | --- | --- | --- |
| Manufacturer | Esthetic Medical Inc.
3600 E Burnett St
Long Beach, CA 90815 | MT.DERM Gustav-Krone-Str. 3,
14167 Berlin, Germany | N/A |
| Manufacturing location | Guangzhou Carain Beauty Equipment
Guangzhou, China | Not specified in 510(k) summary | N/A |
| Device trade name | SkinStylus SteriLock® MicroSystem | Exceed Microneedling Device | N/A |
| 510(k) number | - | K180778 | N/A |
| Device classification name | Microneedling device for aesthetic use | Microneedling device for aesthetic use | Same |
| Device product code | QAI | QAI | Same |
| Regulation number | 21 CFR 878.4430 | 21 CFR 878.4430 | Same |
| FDA device classification | Class II (special controls) | Class II (special controls) | Same |
| Use | Prescription only | Prescription only | Same |
| Intended use and indications | The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories that are intended to be used for the following purposes:
• As a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older. | The Exceed is a microneedling device and accessories is intended for the treatment of wrinkles in Fitzpatrick skin types I, II and/or III in the following facial areas: glabellar frown lines, periorbital lines and cheek folds in adults aged 22 years or older. | The subject device and Exceed are indicated for the treatment of certain types of facial wrinkles. |
Page 3 of 14
{7}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
| Description | Subject device
SkinStylus SteriLock® MicroSystem | Predicate device
Exceed Microneedling Device
(K180778) | Significant differences |
| --- | --- | --- | --- |
| | • As a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, and III in patients aged 22 years and older.
• As a treatment to improve the appearance of Periorbital wrinkles in Fitzpatrick skin types I, II, III, IV, V, and VI in adults aged 34 years and older. | | |
| Intended location of use | Face, Abdomen | Face (glabellar frown lines, periorbital lines and cheek folds) | Different. SkinStylus is indicated for periorbital wrinkles |
| Mode of action | Microneedling (using one or more needles to mechanically puncture and injure skin tissue for aesthetic use) | Microneedling (using one or more needles to mechanically puncture and injure skin tissue for aesthetic use) | Same |
| Needle geometry | 36 solid needles
(6 × 6 square) | 6 needles in a squared arrangement | Different. Clinical and non-clinical performance testing demonstrates safety and effectiveness of the subject device for the proposed indication. |
| Needle length | 2.5 mm | 2.5 mm | Same |
| Needle protrusion setting | 0.0 to 2.5 mm with marked 0.1 mm increments | 0 to 1.9 mm | Same |
Page 4 of 14
{8}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
| Description | Subject device
SkinStylus SteriLock® MicroSystem | Predicate device
Exceed Microneedling Device
(K180778) | Significant differences |
| --- | --- | --- | --- |
| Maximum needle penetration in clinical application | 2.5 mm | 1.5 mm | Different. Clinical and non-clinical performance testing demonstrates safety and effectiveness of the subject device for the proposed indication. |
| Accuracy of needle penetration depth | ± 0.25 mm | Not stated in K180778 summary | Information not available. Clinical and non-clinical performance testing demonstrates safety and effectiveness of the subject device for the proposed indication. |
| Motor speed | 6200 to 8840 RPM | 100 to 150 Hz (±10%)
6000 to 9000 RPM | Similar – the motor speed range of the SkinStylus is within the range of the Exceed device. |
| Treatment protocol | Abdominal scarring:
3 treatments, at least 2 weeks between treatments
Facial Acne: 4 treatments, at least 30 days between treatments | 4 treatments spaced 4 weeks apart | Identical with respect to facial wrinkle indication. |
Page 5 of 14
{9}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
| Description | Subject device
SkinStylus SteriLock® MicroSystem | Predicate device
Exceed Microneedling Device
(K180778) | Significant differences |
| --- | --- | --- | --- |
| | Periorbital Wrinkles: 4 treatments, at least 30 days between treatments | | |
| Power source(s) | 5 V DC/1 amp rechargeable lithium-ion battery, or
AC adapter 5VC +/-, 1 A minimum | 15 V DC | Different. Non-clinical performance testing demonstrates safety and effectiveness of the subject device for the proposed indication. |
Page 6 of 14
{10}
SkinStylus SteriLoc®k MicroSystem
510(k) Summary
K253002
# Substantial Equivalency and Comparison of Technological Similarities and Differences
## Similarities between subject device and predicate
The Exceed Microneedling Device (K180778) is a microneedling system cleared for the same intended use and similar indication as proposed for the subject device. The key technological similarities between the SkinStylus SteriLock® MicroSystem and the Exceed Microneedling Device include:
- Both devices are microneedling systems containing >1 needle that uses a powered reciprocating motion to mechanically puncture or injure the skin for aesthetic use.
- Both devices are generally indicated for the treatment of facial wrinkles.
- Both devices use a sterile, single-use disposable needle cartridge with stainless steel microneedles.
- Both devices are intended for prescription use only and have identical treatment protocols.
- The subject device’s motor speed range is within the operating range of the predicate.
## Differences between subject device and predicate
The following differences are noted between the subject device and the Exceed system (K180778):
- Use of the Exceed system is indicated for facial wrinkles in the glabellar frown lines, periorbital lines and cheek folds area. The subject device is indicated for the periorbital region. Clinical performance testing supports the use of the subject device as described.
- The intended patient demographics of the subject device (age 34 years and older, Fitzpatrick skin type I-VI) differ from the patient demographics of the predicate device (age 22 years and older, Fitzpatrick skin type I, II, or III). Clinical performance testing included with this submission supports the targeted patient population of the subject device.
The subject device’s needle cartridge uses 36 microneedles arranged in a 6 × 6 square, while the Exceed predicate device cartridge contains 6 microneedles arranged in a square. Additionally, the subject device has a greater maximum depth setting (2.5 mm vs 1.9 mm) and maximum permissible depth for the indication (2.5 mm vs 1.9 mm). Clinical performance testing demonstrates that the subject device needle cartridge and depth settings are safe and effective for the indicated use, and the safety of the device is further supported by postmarket data referenced via the MAUDE database.
Page 7 of 14
{11}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
The technological characteristics of the proposed device have been addressed by the sponsor through application of relevant safety standards (general controls and mitigation measures), special controls, and both clinical and non-clinical performance testing.
## General controls and mitigation measures
The SkinStylus SteriLock® MicroSystem is subject to the general controls of the FD&C Act and special controls under 21 CFR 878.4430: To demonstrate safety and effectiveness and support substantial equivalence the device has undergone several non-clinical performance tests in line with recognized standards in terms of biocompatibility, electrical safety, software and general requirements.
## Biocompatibility
The critical component of the SSSM device in terms of patient contact and biocompatibility is the SteriLock® cartridge.
The following testing data is provided to demonstrate the biocompatibility of the needles:
- ISO 10993-5:2009 tests for in vitro cytotoxicity
- ISO 10993-10:2010 tests for irritation
- ISO 10993-10:2010 tests for skin sensitization
- ISO 10993-11:2017 tests for systemic toxicity and pyrogens
- Chemical composition/metallographic analysis
The body of the cartridge is composed of polycarbonate. While the polycarbonate does not penetrate the skin surface, it may contact the surface of the skin during treatment. The following biocompatibility test data was provided to demonstrate biocompatibility endpoints for the polycarbonate:
- ISO 10993-5:2009 tests for in vitro cytotoxicity
- ISO 10993-10:2010 tests for irritation
- ISO 10993-10:2010 tests for skin sensitization
- ISO 10993-11:2017 tests for systemic toxicity and pyrogens
## Sterility and shelf-life testing
The single-use SteriLock® needle cartridge is the only component of the system that is provided sterile and has been validated according to:
- ISO 11737-1:2018 Sterilization of health care products – Microbiological methods – Part 1: Determination of a population of microorganisms on products
- ISO 11737-2:2009 Sterilization of health care products – Microbiological methods – Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
Page 8 of 14
{12}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
The single-use cartridge packaging has been validated to maintain sterility over the product's shelf-life according to:
- ASTM F88/F88M-15 Seal Strength of Flexible Barrier Materials
- ASTM F1929-15 Dye Penetration Test
- ASTM F1608 Microbial Ranking Test
## EMC and electrical performance
The SSM has been tested to applicable industry standards for electromagnetic compatibility (EMC) and electrical safety and complies to the following:
- IEC 60601-1:2005 + CORR 1:2006 + CORR 2:2007 + A1:2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
## Battery testing
The Lithium-ion battery for the SSSM system has been tested and complies with the following standards:
- IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes. Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- UN38.3 Recommendations on the Transport of Dangerous Goods, Manual of Tests and Criteria (ST/SG/AC.10.11/Rev.6) Sixth revised edition.
## Software
The SkinStylus SteriLock® MicroSystem does not incorporate any software in the device design.
## Non-Clinical Performance
In addition to the general standards and risk mitigation measures identified above the SSM device has been subjected to several non-clinical safety tests to include.
1. The technical specifications and needle characteristics have been identified, including needle length, geometry and maximum penetration depth.
a. Accuracy of the needle penetration depth and puncture rate
b. Safety features to protect against cross-contamination, including fluid ingress protection
c. Maximum safe needle penetration depth.
2. Validation was performed for reusable components of the device.
Page 9 of 14
{13}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
3. Suitable labelling has been provided to allow safe and effective use of the device to include:
a. Information on how to operate the device and its components and the typical course of treatment.
b. A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and frequency.
c. Validated methods and instructions for reprocessing of any reusable components.
d. Disposal instructions.
4. Maximum Shelf life.
5. Patient labeling included:
a. Information on how the device operates and the typical course of treatment.
b. The probable risks and benefits associated with use of the device; and
c. Post-operative care instructions.
6. Usability/Human factors. The subject device has been previously cleared under K200444 and K231073.
## Clinical Performance
A clinical study was conducted to support the safety and effectiveness of the SkinStylus SteriLock® MicroSystem for the treatment of facial wrinkles.
The single-arm study was conducted at two centers and included treatments on day 0 (baseline), day 30, day 60, and day 90, with follow-up visits on day 120 and day 150. During each visit, standardized photography was taken, study investigators graded subjects' wrinkles, and subjects provided a self-evaluation of their facial wrinkles and associated psychological well-being. If the visit included a microneedling treatment, the treatment was performed after photography and other assessments. After treatment, subjects assessed side effects and received a diary to record their side effects over the following week. During day 90, day 120, and day 150 visits, investigators and subjects both provided additional assessments on the overall effectiveness of the treatment. Subjects reported any adverse events to the study site throughout the study period.
Each of the four treatments was performed by a trained and appropriately qualified operator. Facial skin was cleansed using Cetaphil® wipes before BLT numbing cream (Benzocaine 20%, Lidocaine 4%, Tetracaine 2%) was applied for 30 minutes. After the numbing cream was removed, a gliding agent (Supragel) was applied to the face before starting the microneedling procedure. All treatments were performed using a 36-pin (all @2.5 mm) cartridge. Each subject was treated on the forehead, cheek and nasolabial folds, periorbital area, glabella and jawline, and upper lip, corner of mouth and chin. Treatment of each area started at a depth of 0.5 mm and the depth was increased in increments of 0.1 mm until the clinical endpoint of pinpoint bleeding was observed. The modal depth setting used was Glabella -
{14}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
1mm, Periorbital - 1.5mm, Cheek/ Nasolabial - 2mm, Forehead - 0.8mm, Perioral - 1.5mm and Chin - 1.5mm. After each treatment, the subject's facial skin was cleansed with sterile saline and sterile gauze, then moisturizer and sunblock were applied.
A total of 51 subjects enrolled in the study. The following tables describe the demographics of the study group.
| Age | | Sex | | Fitzpatrick skin type | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mean | Range | Male | Female | I | II | III | IV | V | VI |
| 54.7 | 34-72 | 2 | 49 | 3 | 13 | 18 | 9 | 4 | 4 |
| Race | | |
| --- | --- | --- |
| Asian | 2 | 4% |
| Black or African American | 5 | 10% |
| Native Hawaiian or Other Pacific Islander | 1 | 2% |
| White/Caucasian | 39 | 76% |
| Mixed race | 4 | 8% |
| Ethnicity | | |
| Hispanic or Latino | 13 | 25% |
| Not Hispanic or Latino | 38 | 75% |
The primary endpoint of the study was a mean improvement over baseline of one grade on the Lemperle scale for wrinkle assessment of the periorbital region, cheek folds, and glabellar lines at 120 days and 150 days after the first treatment. The Lemperle Wrinkle Grading Scale is a grading system used to assess the severity of facial wrinkles. The scale divides the face into distinct regions and provides a photonumeric scale to assess each individual area. Each area can be graded from 0 to 5, where $0 =$ no wrinkles and $5 =$ Very deep wrinkles. Patient treatment satisfaction and psychological function measured on the FACE-Q scale at day 0 and day 150. Adverse events were recorded throughout the study period. In addition to the major endpoints Global Aesthetic Improvement Scale (GAIS) and Subject Global Aesthetic Improvement Scale (SGAIS) were used to assess overall improvement of facial appearance as judged by the clinical assessors and subjects respectively.
Standardized photography was taken at each study visit using VISIA systems. The images were randomized and three independent, blinded physicians graded the appearance of wrinkles on the VISIA images using the Lemperle scale.
## Clinical Results
## Reduction in Wrinkles Assessment (Primary Endpoint)
The average baseline grading for periorbital lines was 3.14 (range 1-4) at study site #1, and 2.06 (range 1-3) at study site #2. By day 120 (30 days after completing the fourth treatment),
{15}
510(k) Summary
K253002
the average grading for periorbital lines was 1.8 (range 0-3) at study site #1 and 1.8 (range 1-3) at study site #2, an improvement of 1.3 and 0.2 from baseline, respectively. By day 150 (60 days after completing the fourth treatment), the average grading for periorbital lines was 1.69 (range 0-3) at study site #1 and 1.67 (range 0-3) at study site #2, an improvement of 1.5 and 0.2 from baseline, respectively.
The primary endpoint was achieved for periorbital lines at site #1 (ASIRC), and the primary endpoint was not achieved for periorbital wrinkles at site #2 (WDRC).
Reduction in Wrinkle Assessment: Blinded evaluation of photos by independent physicians
Three independent physicians retrospectively graded subjects' wrinkles at baseline and day 150 using the Lemperle scale using VISIA images that were randomized and blinded.
The periorbital region showed the most consistent improvement, with 40 subjects (80%) rated as improved at Day 150 as agreed upon by at least 2 out of 3 independent photograph reviewers (study site #1: 27 (77%) subjects improved, 95% confidence interval [60%, 90%]; study site #2: 13 (87%) subjects improved, 95% confidence interval [60%, 98%]).
Adverse Event Results
No adverse incidents that were related to the study treatment were reported during the study.
Side Effects
Subjects' self-evaluations of side effects have been pooled for both sites.
Subjects' self-evaluations indicated that the pain was most severe (mean grading of 3.0, representing “moderate” pain) immediately following the treatment. Subjects' pain reduced quickly following treatment, to a mean grading of 1.1 by day 2 (the evening of the day after treatment) and completely resolving by day 6.
Subjects' self-evaluations indicated only mild discomfort, with the peak grading of 1.4 (mild discomfort) occurring the evening following the treatment. Discomfort resolved fully by day 8.
Subjects were asked to grade peeling and flaking beginning on day 3 (approximately 48-60 hours after treatment). Peak grading of peeling/flaking occurred on day 4 (mean grading of 1.9, mild peeling/flaking), and by day 8 the mean grading was 0.5 (very mild).
Subjects' self-evaluations indicated that erythema was most severe immediately following treatment, with all subjects reporting some degree of erythema: 74.5% moderate to severe and 25.5% minor to mild. Erythema improved significantly by day 2, with only 18.5% of subjects reporting moderate to severe erythema, 76% reporting minor to mild erythema, and 5.5% reporting no erythema. By day 8, no subjects reported their erythema was moderate to
{16}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
severe, only 11.5% reported minor to mild erythema, and 88.5% reported that their erythema had resolved completely.
## Additional Results
### Physician grading of treatment response (GAIS)
Mean grading of subjects' overall treatment response by the investigator on day 90 was 2.3 (good improvement) at study site #1 and 1.4 (fair improvement) at study site #2. By day 150 (60 days after the last treatment), mean grading was 2.9 at study site #1 (very good improvement) and 2.5 at study site #2 (good to very good improvement), an increase of 0.6 and 1.1, respectively.
On day 90, investigators graded 48 of 50 subjects as showing some degree of improvement, and only two subjects as showing poor/no improvement. The modal grading at day 90 was 3 (very good) at study site #1 and 1 (fair) at study site #2. By day 150, investigators graded all subjects as showing some improvement, with 32 of 50 subjects showing very good or excellent improvement. The modal grading at day 150 was 3 (very good) study site #1 and 2 (good) at study site #2.
### Subject self-evaluation of treatment response (SGAIS)
Mean self-evaluation by subjects of overall treatment response on day 90 was 2.3 (good improvement) at study site #1 and 1.7 at study site #2. By day 150 (60 days after the last treatment), subjects' mean grading was 2.7 at study site #1 and 2.3 at study site #2, an increase of 0.4 and 0.6 respectively.
On day 90, 49 of 50 subjects reported some degree of improvement, and only one subject graded their appearance as having poor/no improvement. The modal grading at day 90 was 3 at study site #1 and 1 at study site #2. By day 150, 48 of 50 subjects reported some degree of improvement, with 24 reporting very good or excellent improvement, 24 reporting fair or good improvement, and two reporting poor improvement. The modal SGAIS grading at day 150 was 2 at both study sites.
## Statement of Substantial Equivalence
513(i) of the FD&C Act (21 U.S.C. 360c(i)) states that for substantial equivalence a proposed device is required to have the same intended use and similar technological characteristics as the predicate device. Where there are differences in technological characteristics, these can be negated by appropriate clinical or scientific data demonstrating that the proposed device is as safe and effective as the predicate device, and that the proposed device does not raise any different questions of safety and effectiveness than the predicate device for the same intended use.
Esthetic Medical has demonstrated that the SkinStylus SteriLock® MicroSystem has the same generic classification (generic description) and basic technologies as the predicate
Page 13 of 14
{17}
SkinStylus SteriLoc®k MicroSystem
K253002
510(k) Summary
devices. The SkinStylus and both predicates are all prescription micro needling systems for use in a physician/aesthetician clinic setting that use multiple needles to mechanically puncture or injure the skin for aesthetic use. The specific indication for the treatment of facial wrinkles in the periorbital region is shared between the subject device and the Exceed Microneedling Device predicate, and Esthetic Medical has undertaken clinical testing that demonstrates the SkinStylus SteriLock® MicroSystem's effectiveness in treating facial wrinkles in the periorbital region.
Esthetic Medical has conducted both clinical and non-clinical performance testing applicable to those general and special controls deemed necessary by the agency for this product classification and has determined that the SkinStylus SteriLock® MicroSystem raises no new questions relating to safety and therefore has demonstrated that the SkinStylus SteriLock® MicroSystem is substantially equivalent to the Exceed Microneedling Device (K180778).
February 19th 2025
Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.