HyperFlex® Bunion Correction System

K252997 · Hyperflex Medical · HRS · Dec 11, 2025 · Orthopedic

Device Facts

Record IDK252997
Device NameHyperFlex® Bunion Correction System
ApplicantHyperflex Medical
Product CodeHRS · Orthopedic
Decision DateDec 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The HyperFlex® Medical HyperFlex® Bunion Correction System is intended for use in reconstruction (correction) of the hallux valgus deformity by holding the reduction of the 1st metatarsal-2nd metatarsal intermetatarsal (IM) angle in adults and pediatric patients aged ≥13 years in which growth plates have fused.

Device Story

Implantable bunion correction system; treats hallux valgus deformity by maintaining reduction of 1st-2nd metatarsal intermetatarsal (IM) angle. System comprises metatarsal attachment components joined by flexible central section. Used by surgeons in clinical settings. Implant holds IM angle without restricting range of motion. Available in multiple sizes to accommodate patient anatomy. Benefits patient by correcting deformity and stabilizing metatarsal alignment.

Clinical Evidence

Bench testing only. Testing included static and fatigue tensile testing, screw torsional strength, insertion torque, and pullout strength per ASTM F543-17, particulate testing per ASTM 1877-16, and additive manufacturing characterization.

Technological Characteristics

Implantable bone fixation system. Components manufactured from titanium alloy. Testing performed per ASTM F543-17 (screw mechanical properties) and ASTM 1877-16 (particulate). Includes flexible central section. Sterilized via validated methods.

Indications for Use

Indicated for hallux valgus deformity reconstruction in adults and pediatric patients aged ≥13 years with fused growth plates.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION December 11, 2025 HyperFlex Medical % Dave McGurl Vice President, Regulatory Affairs - Orthopedics MCRA 803 7th Street NW Washington, District of Columbia 20001 Re: K252997 Trade/Device Name: HyperFlex® Bunion Correction System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: December 5, 2025 Received: December 5, 2025 Dear Dave Mcgurl: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252997 - Dave McGurl Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252997 - Dave McGurl Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252997 | ? | | Please provide the device trade name(s). | | ? | | HyperFlex® Bunion Correction System | | | | Please provide your Indications for Use below. | | ? | | The HyperFlex® Medical HyperFlex® Bunion Correction System is intended for use in reconstruction (correction) of the hallux valgus deformity by holding the reduction of the 1st metatarsal-2nd metatarsal intermetatarsal (IM) angle in adults and pediatric patients aged ≥13 years in which growth plates have fused. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K252997 - Page 1 of 2 # 510(k) Summary Device Trade Name: HyperFlex® Bunion Correction System Manufacturer: HyperFlex® Medical 2929 Lapeer Road Port Huron, MI 48060 (810) 982-7777 Official Correspondent: Dave McGurl MCRA, LLC 803 7th Street, NW, 4th Floor Washington, DC 20001 Office: 202.552.5800 Date Prepared: December 10, 2025 Classifications: 21 CFR 888.3030, Single/multiple component metallic bone fixation appliances and accessories 21 CFR 888.3040, Smooth or threaded metallic bone fixation fastener Common Name: Bunion Correction System Class: II Product Code: HRS, HWC Primary Predicate: HyperFlex Medical (formerly FootBridge Medical) HyperFlex Bunion Correction System (K241995) Reference Device: Arthrex Mini TightRope and the Mini TightRope FT (Arthrex, K090107) # Indications For Use: The HyperFlex® Medical HyperFlex® Bunion Correction System is intended for use in reconstruction (correction) of the hallux valgus deformity by holding the reduction of the 1st metatarsal-2nd metatarsal intermetatarsal (IM) angle in adults and pediatric patients aged ≥13 years in which growth plates have fused. # Device Description: The HyperFlex® Medical HyperFlex Bunion System is an implantable device and accompanying instrumentation used to treat a hallux valgus deformity (bunion). It includes an implant with components that attach to the 1st and 2nd metatarsals which are joined by a flexible central section. Page 1 of 2 {5} K252997 - Page 2 of 2 The system holds the IM angle between the 1st and 2nd metatarsal without restricting range of motion. The implant will be offered in several sizes to fit the anatomical needs of the patient population. ## Predicate Device: HyperFlex® Medical submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA’s regulation of medical devices, HyperFlex System is substantially equivalent in indications, design principles, and performance to the following predicate devices: - Primary Predicate: HyperFlex Medical (formerly FootBridge Medical) HyperFlex Bunion Correction System (K241995). - Reference Device: Arthrex Mini TightRope and the Mini TightRope FT (Arthrex, K090107) ## Performance Testing Summary: The following bench testing has been conducted on the HyperFlex® Bunion Correction System: - Static and Fatigue Tensile Testing - Screw Torsional Strength Testing per ASTM F543-17 - Screw Insertion Torque Testing per ASTM F543-17 - Screw Pullout Strength Testing per ASTM F543-17 - Particulate Testing per ASTM 1877-16 - Additive Manufacturing Characterization ## Substantial Equivalence: The subject device differs from the primary predicate device because the subject metatarsal components are manufactured from titanium alloy, whereas the predicate components are manufactured from stainless steel. Also, there have been changes to the packaging and sterilization methods. The subject devices were demonstrated to be substantially equivalent to the predicates cited above with respect to indications, design materials, function, manufacturing, and performance. The non-clinical tests performed by the company demonstrated that the HyperFlex® Bunion Correction System is able to withstand physiological loading and is substantially equivalent to the predicate devices. ## Conclusion: The subject device and the predicate devices have the same intended use, the same design and similar technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in similar materials and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the identified predicate devices. The HyperFlex® System is as safe, as effective, and performs as well as, or better, than the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...