K252987 · Epic Medical Pte. , Ltd. · LHI · Jun 12, 2026 · General Hospital
Device Facts
Record ID
K252987
Device Name
ADRx™ Admixture Device
Applicant
Epic Medical Pte. , Ltd.
Product Code
LHI · General Hospital
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ADRx™ Admixture Device is intended to connect an IV bag, a drug vial, and a standard external IV administration set. It facilitates the safe reconstitution of drug vials using diluents from the IV bag and enables the transfer of vial contents into the IV bag prior to patient administration.
Device Story
Sterile, single-use adaptor; connects IV bag, drug vial, and IV administration set. Facilitates safe reconstitution/admixture of parenteral drugs at point-of-care. Manual operation: vial adaptor attaches to drug vial; IV spike connects to IV bag port. Enables fluid transfer from bag to vial for reconstitution, then back to bag for administration. Used by healthcare professionals in pharmacies, clinics, emergency departments, hospital wards, and pre-hospital settings. Benefits include secure, efficient drug preparation and reduced contamination risk via dry connections.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical performance (tensile, torque, drop, spike penetration), leakage integrity, functional transfer efficiency, biocompatibility (ISO 10993), and sterile barrier/shelf-life validation.
Technological Characteristics
Sterile, single-use, manual fluid transfer adaptor. Materials biocompatible per ISO 10993-1. Features ISO 8536-4:2019 compliant bag spike, 13mm/20mm vial adaptors, and Luer lock connections per ISO 80369-7:2021. Sterilized via Ethylene Oxide (EO) to SAL 10^-6. Shelf-life 3 years.
Indications for Use
Indicated for use by healthcare professionals in clinical settings to facilitate the reconstitution and transfer of parenteral drugs between IV bags and drug vials (13mm or 20mm closures) prior to patient administration.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
Epic Medical Pte. , Ltd.
Freddie LEE
CEO/MD
105 Cecil St. #20-04
The Octagon
Singapore, SG 069534
Singapore
Re: K252987
Trade/Device Name: ADRx™ Admixture Device
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular administration set
Regulatory Class: Class II
Product Code: LHI
Dated: May 13, 2026
Received: May 13, 2026
Dear Freddie LEE:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252987 - Freddie LEE
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252987 - Freddie LEE
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
DAVID WOLLOSCHECK -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K252987
Device Name
ADRx™ Admixture Device
Indications for Use (Describe)
The ADRx™ Admixture Device is intended to connect an IV bag, a drug vial, and a standard external IV administration set. It facilitates the safe reconstitution of drug vials using diluents from the IV bag and enables the transfer of vial contents into the IV bag prior to patient administration.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
![EPIC logo]()
# K252987 – 510(k) Summary
I. Submitter
Epic Medical Pte. Ltd.
105 Cecil Street #20-04
The Octagon
Singapore 069534
Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/Managing Director
Date Prepared: June 11, 2026
Content and Format: Prepared in accordance with 21 CFR 807.92
Type of Submission: Traditional 510(k)
II. Subject Device
| 510(k) Number: | K252987 |
| --- | --- |
| Trade/ Device Name: | ADRx™ *Admixture Device* |
| Common/ Usual Name: | Set, I.V. Fluid Transfer |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set. |
| Regulatory Class: | Class: II |
| Product Code: | LHI |
III. Predicate
| 510(k) Number: | K201415 |
| --- | --- |
| Trade/ Device Name: | Vial2Bag Advance™ *20mm Admixture Device* |
| Common/ Usual Name: | Set, I.V. Fluid Transfer |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set. |
| Regulatory Class: | Class: II |
| Product Code: | LHI |
Page 1 of 7
{5}
EPIC
IV. Device Description
The ADRx™ Admixture Device is a sterile, single-use adaptor designed to connect standard I.V. bag and components for the safe reconstitution and transfer of fluids. It enables secure and efficient preparation and transfer of I.V. fluids and drug solutions, supporting point-of-care mixing and direct bedside administration.
Device Components and Models:
- Component Subgroups:
- ADRx™ SingleDose:
(I.V. bag spike with integrated vial adaptor)
- An integrated assembly comprising an IV bag spike with spiking port and an attached vial adaptor for drug reconstitution and transfer
- Designed to support direct reconstitution and transfer of a single drug dose from a 13 mm or a 20 mm vial into I.V. bag, without the need for additional external bag spikes or adaptor components
- ADRx™ UniqDose:
(Standalone vial adaptor derived from the ADRx™ SingleDose, 20 mm configuration)
- The ADRx™ UniqDose consists of the vial adaptor and vial spike subassembly of the ADRx™ SingleDose 20 mm vial configuration, without the integrated IV bag spike and spiking port
- Designed for use as a standalone vial adaptor, enabling drug reconstitution and transfer when directly connected to an IV bag featuring a compatible Luer access port, or when used in combination with a separate IV bag spike assembly
- The ADRx™ UniqDose retains the same vial interface design, materials, and functional characteristics as the vial adaptor used in the ADRx™ SingleDose 20 mm model
- Compliance:
- ISO 8536-4:2019
- Compatibility:
- ADRx™ SingleDose is compatible with 13 mm or 20 mm drug vials that fit its vial adaptors, and ADRx™ UniqDose, with 20 mm drug vials
- Compatible with standard I.V. bags and I.V. administration sets with ISO 8536-4:2019-compliant bag spikes
- Safety:
- Sterile and single-use, single-patient use
- Dry connections to minimize exposure to contaminants
- Intended Users:
- Health care professionals in clinical settings
Page 2 of 7
{6}
EPIC
# V. *Indications for Use Statement*
The **ADRx™ Admixture Device** is intended to connect an IV bag, a drug vial, and a standard external IV administration set. It facilitates the safe reconstitution of drug vials using diluents from the IV bag and enables the transfer of vial contents into the IV bag prior to patient administration.
# VI. *Comparison of Intended Use & Technological Characteristics*
The Subject device, **ADRx™ Admixture Device**, and the Predicate device, **K201415 – Vial2Bag Advanced™ 20 mm Admixture Device** – share the same intended use and similar technological characteristics. Differences identified have been evaluated through verification testing performed to FDA-recognized standards and do not raise new or different questions of safety and effectiveness.
# *Intended Use comparison*
- *Indications for use* statements (similar)
- Prescription use
- Intended user population/intended use environment
- Intended drug type
# *Technological Characteristics comparison*
The **ADRx™ Admixture Device** and the Predicate device share key design and performance characteristics. Evaluation confirmed that no differences raise new or different questions of safety or effectiveness:
- Principles of operation
- Bag spike design
- Number of access points
- Biocompatibility of fluid path materials
- Sterile barrier packaging
- Sterilization process
- Shelf-life
- Single use
- Labeling specifications
All differences, including the vial adaptor dimensional variation for 13 mm and 20 mm vial closures, were thoroughly evaluated. A side-by-side comparison of key characteristics between the Subject and the Predicate device is provided in the table below:
Page 3 of 7
{7}
EPIC
Table: Side-by-side comparison of intended use and technology & design
| Characteristic compared | Predicate Device (K201415) Vial2Bag Advanced™ 20 mm Admixture Device | Subject Device (K252987) ADRx™ Admixture Device | Comment/Discussion |
| --- | --- | --- | --- |
| Intended use and Indications for Use statement | The Vial2Bag Advanced™ 20mm Admixture Device is indicated to serve as a connecting part between a 50, 100 or 250mL IV bag, vial with 20mm closure, and an external IV administration set. The integrated Vial Adaptor makes it possible to reconstitute and/or admix drugs prior to administration to the patient. | The ADRx™ Admixture Device is intended to connect an IV bag, a drug vial, and a standard external IV administration set. It facilitates the safe reconstitution of drug vials using diluents from the IV bag and enables the transfer of vial contents into the IV bag prior to patient administration. | Different see Comment#1 |
| Prescription use | R only | | Same |
| Intended user population/intended use environment | Adequately trained health care professionals including pharmacists/clinical setting | | Same |
| Intended drug type | Parenteral drugs | | Same |
| Principles of operation (except vial closure size) | The Vial2Bag Advanced® 20mm Admixture Device is operated by manual process. The Vial Adapter is first attached to the drug vial, and after removing the Protector, the IV spike is then connected to the administration port of the IV bag. Fluid is transferred from the IV bag to the drug vial to reconstitute/dilute the drug prior to being transferred back to the IV bag. The IV administration set is then connected to the device's IV Port followed by administration to patient. | The ADRx™ SingleDose and UniqDose are intended for health care professionals in various settings such as pharmacies, clinics, emergency departments, hospital wards, ambulances, and pre-hospital care. It serves as a connector between the I.V. bag and an external I.V. line, facilitating the safe and easy reconstitution/dilution and transfer of drugs/diluents. | Same |
Page 4 of 7
{8}
EPIC
Table: Side-by-side comparison of intended use and technology & design
| Characteristic compared | Predicate Device (K201415) Vial2Bag Advanced™ 20 mm Admixture Device | Subject Device (K252987) ADRx™ Admixture Device | Comment/Discussion |
| --- | --- | --- | --- |
| Vial closure size | 20 mm | 13 mm and 20 mm | Different See Comment #2 |
| Bag spike design | Dual-channel bag spike | | Same |
| | ISO 8536-4:2019 compliant bag spike | | Same |
| Number of access points | Device has three (3) access points: 1. I.V. bag piercing spike, 2. Injection port/medication access port, 3. Spiking port/administration port | | Same |
| Biocompatibility of fluid path materials | Materials that were tested in accordance with the ISO 10993-1 standard and/or USP VI requirements and were determined suitable for the Indications for Use of the product (respective devices) | | Same |
| | Particulate testing was conducted | | Same |
| Sterile barrier packaging | Medical grade paper and medical plastic film, heat sealed | | Same |
| Sterilization process | Ethylene Oxide (EO), SAL 10^{-6} | | Same |
| Shelf-life validation | 3 years (36 months) | | Same |
| Single use or reusable | Single use only | | Same |
| Labeling specs | Met the requirements specified in 21 CFR 801 | | Same |
# Submitter's Comments
# Comment #1 on Indications for Use
Both the Subject device (ADRx™ SingleDose & ADRx™ UniqDose) and the Predicate device (Vial2Bag Advanced™) are intended to function as connectors between an I.V. bag and an external I.V. administration set, enabling the admixture of drugs into infusion solutions. Each device establishes a sterile fluid pathway between vials and I.V. bags, and vice versa to support safe drug/fluid transfer. Both devices are compatible with standard I.V. bags and administration sets as specified in ISO 8536-4:2019, and operates on the same manual principles of use
# Comment #2 on Vial Sizes
All differences, including the vial adaptor dimensional variation for 13 mm and 20 mm vial closures, were thoroughly evaluated through verification testing conducted on twice-EO-sterilized and aged devices and their protective packaging, over the full labeled shelf life. These tests included mechanical performance (e.g., attachment/detachment forces, leakage), functional transfer efficiency, and packaging integrity, performed in accordance with FDA-recognized standards (e.g., ISO 8536-4:2019, USP <788>, and ASTM F1980). Results confirmed that these differences do not introduce new or different questions regarding safety and effectiveness
Page 5 of 7
{9}
EPIC
## VII. Performance Data Supporting Substantial Equivalence
### A. Performance Testing
The Subject device was evaluated to be in conformance with the following ISO and FDA recognized standards, and FDA guidance document:
- ISO 8362-2:2024, Injection containers and accessories — Part 2: Closures for injection vials
- ISO 8536-2:2023, Infusion equipment for medical use — Part 2: Closures for infusion bottles
- ISO 8536-4:2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- ISO 8871-5:2016, Elastomeric parts for parenterals and for devices for pharmaceutical use Part 5: Functional requirements and testing
- ISO 15747:2018, Plastic containers for intravenous injections
- ISO 22413:2021, Transfer sets for pharmaceutical preparations — Requirements and test methods
- ISO 80369-7:2021, Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- Intravascular-Administration-Sets-Premarket-Notification-Submissions-[510(k)]---Guidance-for-Industry-and-FDA-Staff
The following performance testing executed on twice-EO sterilized, aged devices confirmed operational integrity:
#### Mechanical Tests
- 1m Drop Durability
- Attachment Force of Vial Adapter to Vial
- Detachment Force of Vial Adapter from Vial
- IV Port Tensile Strength
- Bag Spike Penetration Force Test (functional) – per ISO 22413:2021, Section 6.6, Annex A.7
- IV Spike to IV Port Attachment Force
- IV Spike from IV Port Detachment Force
- IV Spike Protector Detachment Force
- Protective Cap Tests (bag spike, spiking port & vial adaptor) (functional)– ISO 8536-4:2019 Section 7.13
- Residual Volume
- Vial Adaptor Spike Penetration Force Test (vial adaptor to vial closure) (functional) – per ISO 8536-2:2023 Section 7.2, Annex B
- Vial Adaptor Tensile Detachment Force
- Tensile Strength Test (functional)– per ISO 8536-4:2019, paragraph 7.3 and Annex A.4
- Vial Adaptor Torque Test and Vial Adaptor Torsional Breakage
- Vial Adaptor Wings Breakage
- Vial Attachment Force
- Vial Detachment Force
- Visual Inspection for Product Damage
- Visual Inspection of Device
- Coring
- Fragmentation
- Mass Transfer
- Luer Lock Connection Test – per ISO 80369-7:2021, Clause 6.1.3, 6.2, 6.3, 6.4, 6.5 and 6.6
#### Leakage Tests
- Airtightness & Fluid Leakage Integrity Test (including vial adaptor–vial stopper/closure interface) (functional) – per ISO 8536-4:2019, para. 7.2 & Annex A.3
- Leakage after Removal of Twist Off
- Leakage at Vial Adaptor and Stopper/Closure Interface
- Leakage Test (Device)
- Leakage Test (IV spike to IV port)
- Leakage Testing, Internal Diameter of Upper Skirt
#### Functional Tests
- Dose Concentration (K240940) and Dose Concentration of Delivery Profile
- Uniform Mixing and Transfer Performance (functional) – in-house test method
- Short Circuit Test
- Transfer of Vial Contents to Administration Set
- Transfer of Vial Contents to Bag
Page 6 of 7
{10}
EPIC
## B. Biocompatibility
In accordance with ISO 10993-1:2018, the Subject device just like the below listed existing cleared devices, is classified as: *Externally Communicating Device, Blood Path Indirect, Prolonged Contact* (>24hr to 30d). The following testing were performed on relevant devices cleared under K151650, K241988, K223674/S001 and the Subject device:
- **Cytotoxicity** to ISO 10993-5:2009
- **Sensitization** to ISO 10993-10:2010 & ISO 10993-10:2021
- **Intracutaneous Reactivity** to ISO 10993-23:2021
- **Acute Systemic Toxicity** to ISO 10993-11:2017
- **14-day Subacute/ Subchronic Acute Systemic Toxicity** to ISO 10993-11:2017
- **In-vitro Hemolysis Assessment** to ISO 10993-4:2017 & ASTM F756-17
- **Material Mediated Pyrogenicity** to ISO 10993-11:2017 & USP <151>
- **Chemical Requirements** to ISO 15747:2018, Annex B
- **Particulate matter testing** was conducted on the Subject device, in accordance with ISO 8536-4: 2019, *Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed and USP <788> Particulate Matter in Injections*
- **EO and ECH residues testing** were conducted on the Subject device, in accordance with ISO 10993-7:2008/Amd.1:2019 for neonate patient population
## C. Sterility, Shipping, and Shelf-Life
The Subject device complies with sterilization requirements of ISO 11135:2014, *Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices* and the following testing/ evaluations conducted on the Subject device:
- **Simulated shipping testing** per ASTM D 4169-16, *Standard Practice for Performance Testing of Shipping Containers and Systems* conducted under K252987
- **Package integrity tests** per ASTM F1980-21, *Standard guide for accelerated aging of sterile barrier systems for medical devices and Sterile Barrier Packaging Testing* performed on the proposed device: Seal strength – ASTM F88/F88M-21, *Standard test method for seal strength of flexible barrier materials*; Dye Penetration – ASTM F1929-23, *Standard test method for detecting seal leaks in porous medical device packaging by dye penetration*; EN 868-5:2009, *Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods*
- **Pyrogen tests** per ANSI/AAMI ST72/2019, *Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing*, USP 40 <151>, *Pyrogen test (USP rabbit test)*, USP-NF <161>, *Medical Devices-Bacterial Endotoxin and Pyrogen Tests*, USP-NF <85>, *Bacterial Endotoxins Test* and testing will be conducted on every lot
- **Shelf-life** of 3 years has been validated using the FDA recognized standard, ASTM 1980-21, *Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices*
## VIII. Clinical Tests
Not applicable
## IX. Conclusion
Based on the comparative analysis and verification testing performed to FDA-recognized standards, **ADRx™ Admixture Device** does not raise new questions of safety or effectiveness. Therefore it is substantially equivalent to the Predicate, **Vial2Bag Advanced™ 20 mm Admixture Device (K201415)** in intended use and technological characteristics, per US FDA 510(k) requirements
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.