The Voyager DSLT device is indicated for use in selective laser trabeculoplasty (SLT).
Device Story
Q-switched, 532 nm frequency-doubled Nd:YAG laser; performs selective laser trabeculoplasty (SLT). Device uses anatomical detection algorithm and eye tracking to locate trabecular meshwork; delivers 120 laser spots in predefined 360° circumferential elliptical pattern through limbus. Non-contact procedure; no gonioscopy lens required. Operator uses joystick to set focal position via overlapping 650 nm ranging beams; confirms treatment location before automated delivery. Eye tracker compensates for movement during sequence. Software includes 'Complete an Incomplete Treatment' feature and updated target definition interface. Used in clinical settings by physicians. Benefits include automated, non-contact, precise laser delivery to trabecular meshwork, reducing procedural complexity.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed alignment accuracy for the 'Complete an Incomplete Treatment' feature. Human factors analysis conducted per FDA guidance. Software verified per IEC 62304:2015 standards. Animal testing from the predicate device (K230722) supports the safety of the updated software features.
Technological Characteristics
Q-switched, 532 nm frequency-doubled Nd:YAG laser; 3 ns pulse duration; 50 Hz repetition rate; 1.1–1.9 mJ pulse energy. 400 μm spot size. 635 nm diode aiming laser; 650 nm ranging beams. LED illumination. Weight 34 kg. Software developed per IEC 62304:2015 (Major level of concern).
Indications for Use
Indicated for selective laser trabeculoplasty (SLT) in patients requiring such treatment.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 7, 2026
BELKIN Vision Ltd. (a wholly owned subsidiary of Alcon Laboratories, Inc.)
% Anne-Marie Ripley
Clinical & Regulatory Affairs Consultant
Regulatory Pathways Group, Inc.
440 N. Barranca Ave., #2471
Covina, California 91723
Re: K252979
Trade/Device Name: Voyager DSLT (430840610)
Regulation Number: 21 CFR 886.4390
Regulation Name: Ophthalmic Laser
Regulatory Class: Class II
Product Code: HQF
Dated: February 26, 2026
Received: February 26, 2026
Dear Anne-Marie Ripley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252979 - Anne-Marie Ripley
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252979 - Anne-Marie Ripley
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CLAUDINE H. KRAWCZYK -S
Claudine Krawczyk
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252979 | ? |
| Please provide the device trade name(s). | | ? |
| Voyager DSLT (430840610) | | |
| Please provide your Indications for Use below. | | ? |
| The Voyager DSLT device is indicated for use in selective laser trabeculoplasty (SLT). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252979
Page 1 of 4
510(k) SUMMARY
K252979
I. Submitter Information
510(k) Owner: BELKIN Vision Ltd. (a wholly owned subsidiary of Alcon Laboratories, Inc.)
13 Gan Rave Blvd.
Yavne, Israel, 8122214
+972-8-857-1619
Contact Person: Anne-Marie Ripley
Regulatory Pathways Group, Inc.
440 N. Barranca Ave., #2471
Covina, CA 91723
aripley@regulatorypathways.com
Date Prepared: April 7, 2026
II. Device Name and Classification
Device Trade Name: Voyager DSLT (430840610)
Common Name: Ophthalmic laser
Classification Name: Ophthalmic laser
Regulation Number: 21 CFR 886.4390
Device Classification: Class 2
Product Code: HQF
III. Predicate Device and Reference Devices
Predicate Device
- BELKIN Eagle device (K230722)
IV. Device Description
The Voyager DSLT device is a Q-switched, 532 nm-wavelength, frequency-doubled Nd:YAG laser that is intended for use in performing selective laser trabeculoplasty. The laser spots produced by the Voyager DSLT device have a 400 μm spot size, a 3 ns pulse duration, and a 50-Hz pulse
{5}
repetition rate. The sequence of laser spots consists of 120 spots in a predefined circumferential elliptical pattern delivered at a pre-defined pulse energy level. The spots are delivered through the limbus to the trabecular meshwork in a non-contact fashion, without the need for the use of a contact gonioscopy lens. The device automatically locates the treatment location. The treatment location may be adjusted slightly by the operator. Once confirmed by the operator, the device then automatically applies the laser treatment sequence to the limbal region of the eye, while the eye tracker compensates for any eye movement. The default energy setting is $1.8\mathrm{mJ} / \mathrm{pulse}$ .
# V. Indications for Use
The Voyager DSLT device is indicated for use in selective laser trabeculoplasty (SLT).
# VI. Comparison of Technological Characteristics with the Predicate Device
The Voyager DSLT has the same intended use, indications for use, target tissue, laser parameters, and laser pattern as the predicate device. It also has the same method of maintaining laser beam targeting and laser delivery as the predicate Eagle device. Refer to the Table below.
| Characteristic | Voyager DSLT (BELKIN Vision) Subject Device | Eagle (BELKIN Vision) Predicate Device K230722 | Comparison of Subject and Predicate Devices |
| --- | --- | --- | --- |
| Device Class | 2 | 2 | Same |
| Classification Product Code | HQF | HQF | Same |
| Regulation Number | 886.4390 | 886.4390 | Same |
| Intended Use | Selective Laser Trabeculoplasty (SLT) | Selective Laser Trabeculoplasty (SLT) | Same |
| Indications for use | Selective laser trabeculoplasty (SLT) | Selective laser trabeculoplasty (SLT) | Same |
| Laser Parameters | | | |
| Laser type | Q-switched, frequency-doubled Nd:YAG | Q-switched, frequency-doubled Nd:YAG | Same |
| Wavelength | 532 nm | 532 nm | Same |
| Laser pulse duration | 3 ns | 3 ns | Same |
| Pulse energy range | 1.1 – 1.9 mJ | 1.1 – 1.9 mJ | Same |
| Pulse repetition rate | 50 Hz | 50 Hz | Same |
| Laser Beam Delivery | | | |
K252979
{6}
| Characteristic | Voyager DSLT (BELKIN Vision) Subject Device | Eagle (BELKIN Vision) Predicate Device K230722 | Comparison of Subject and Predicate Devices |
| --- | --- | --- | --- |
| Method of laser delivery (pattern) | Automated delivery of laser spots in a pre-defined pattern (360° ellipse) through the limbus without the use of a contact gonioscopy lens; the trabecular meshwork (TM) is not directly visualized as laser spots are applied. | Automated delivery of laser spots in a pre-defined pattern (360° ellipse) through the limbus without the use of a contact gonioscopy lens; the trabecular meshwork (TM) is not directly visualized as laser spots are applied. | Same |
| Target tissue | Trabecular meshwork | Trabecular meshwork | Same |
| Method of maintaining laser beam targeting | Anatomical detection algorithm and eye tracking to compensate for eye movements during the procedure. | Anatomical detection algorithm and eye tracking to compensate for eye movements during the procedure. | Same |
| Spot diameter | 400 μm | 400 μm | Same |
| Auxiliary optical characteristics | | | |
| Aiming laser | Diode laser | Diode laser | Same |
| Aiming wavelength | 635 nm | 635 nm | Same |
| Method of maintaining the focal position of treatment beam | Manually set using a joystick to adjust the axial position by visualizing the overlap of two ranging (650 nm) diode laser beams. | Manually set using a joystick to adjust the axial position by visualizing the overlap of two ranging (650 nm) diode laser beams. | Same |
| Illumination | LED illumination ring, visible white light | LED illumination ring, visible white light | Same |
| Physical Dimensions | | | |
| System weight | 34 kg / 74 lbs. | 34 kg / 74 lbs. | Same |
| System dimensions (H x W X D) | 52 x 53 x 63 cm
20.5 x 21.8 x 25 inches | 52 x 53 x 63 cm
20.5 x 21.8 x 25 inches | Same |
The software of the Voyager DSLT device was updated to incorporate 3 changes:
1. Addition of the "Complete an Incomplete Treatment" feature: This feature allows the user to complete a procedure that was interrupted and not fully completed.
2. Modifications to the Target Definition screen to remind the user to ensure proper target definition, and to provide options to the user for how the target definition ring is displayed and for editing the target definition.
3. Automatically add results of the test card daily check to the exported treatment record
K252979
Page 3 of 4
{7}
These changes did not raise new types of questions regarding the safety and effectiveness of the device.
## VII. Summary of Non-Clinical Testing
Based on the risk assessment and design control requirements, the following verification and validation testing/analysis was performed:
- Bench testing was performed to demonstrate the ability of the Voyager DSLT to properly align the target definition ring between the original and new procedure, when completing an incomplete procedure.
- An analysis of the changes on human factors was conducted per FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices"
- Animal testing, performed to support 510(k) clearance of the predicate Eagle device (K230722), also supports the safety of the added "Complete an Incomplete Treatment" feature.
- Software verification and validation testing: BELKIN developed and verified the software in accordance with a major level of concern described in the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and also per the IEC 62304: 2015 Medical Device Software - Software Life Cycle Processes standard.
Results of the non-clinical testing support a substantial equivalence determination. The Voyager DSLT device is substantially equivalent to its predicate device for the indications for use.
## VIII. Conclusions
The Voyager DSLT device has the same indications for use as the predicate device. The differences in technological characteristics of the Voyager device do not raise new or different types of questions of safety or effectiveness. The results of the non-clinical performance testing demonstrate that the Voyager DSLT device functions as intended and is substantially equivalent to the predicate device.
K252979
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.