K252949 · Skeletal Dynamics, Inc. · HRS · Nov 13, 2025 · Orthopedic
Device Facts
Record ID
K252949
Device Name
Geminus Volar Distal Radius Plating System
Applicant
Skeletal Dynamics, Inc.
Product Code
HRS · Orthopedic
Decision Date
Nov 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Skeletal Dynamics Geminus Volar Distal Radius Plating System is intended for fixation of fractures and osteotomies of the distal radius.
Device Story
The Geminus Volar Distal Radius Plating System is a bone fixation system used by orthopedic surgeons for the repair of distal radius fractures and osteotomies. The system consists of medical-grade titanium alloy (Ti 6Al-V4 ELI) bone plates, an ulnar hook plate, radial hook plate extensions, and various bone screws and pegs made of titanium alloy or Cobalt-Chromium-Molybdenum (CoCrMo). It also includes a buttress button assembly designed to capture and hold small dorsal corner fractures. The implants and specialized instrumentation are provided non-sterile and must be sterilized at the healthcare facility prior to use. By providing rigid fixation, the system stabilizes bone fragments to facilitate healing and restore function to the upper extremity.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through static and dynamic mechanical construct testing and MR environment safety testing (ASTM F2052, F2213, F2119, F2182).
Technological Characteristics
Materials: Ti 6Al-V4 ELI (ASTM F136), CoCrMo (ASTM F1537), CP Ti Grade 2 (ASTM F86). Components: Bone plates, locking/non-locking screws, pegs, hook plates, buttress button assembly. Sterilization: Non-sterile, facility-sterilized. MR compatibility: Evaluated for 1.5T and 3T scanners.
Indications for Use
Indicated for patients requiring fixation of fractures and osteotomies of the distal radius.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Geminus Volar Distal Radius Plate System (K182492)
Reference Devices
Sponger Cable System, Modification - Pioneer Surgical Technology Inc. Sternal Cable System (K935481)
Synthes Titanium Alloy Volar Distal Radius Plate System (K963798)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 13, 2025
Skeletal Dynamics Inc
Daniel Gallego
Regulatory Affairs Specialist
7300 North Kendall Drive
Suite 800
Miami, Florida 33156
Re: K252949
Trade/Device Name: Geminus Volar Distal Radius Plating System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: September 15, 2025
Received: September 16, 2025
Dear Daniel Gallego:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252949 - Daniel Gallego
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252949 - Daniel Gallego
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252949 | ? |
| Please provide the device trade name(s). | | ? |
| Geminus Volar Distal Radius Plating System | | |
| Please provide your Indications for Use below. | | ? |
| The Skeletal Dynamics Geminus Volar Distal Radius Plating System is intended for fixation of fractures and osteotomies of the distal radius. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
skeletal dynamics
UNDERSTANDING THE UPPER EXTREMITY
K252949
510(k) SUMMARY
Skeletal Dynamic's
GEMINUS Volar Distal Radius Plating System
## Submitter
Skeletal Dynamics, Inc.
7300 N. Kendall Drive, Suite 800
Miami, FL 33156
Phone: 305-596-7585
E-mail: dgallego@skeletaldynamics.com
Contact Person: Daniel E. Gallego
Date Prepared: November 12th, 2025
## Name and Classification
Trade Name: GEMINUS Volar Distal Radius Plating System
Common Name: Plate, Fixation, Bone.
Classification Name: Single/multiple component metallic bone fixation appliances and
Accessories (Primary), Smooth or threaded metallic bone fixation fastener.
Classification Number: 21 CFR §888.3030 (Primary), 21 CFR §888.3040
Regulatory Class: Class II
Product Code: HRS (Primary), HWC
## Predicate Devices
- Geminus Volar Distal Radius Plate System (K182492)
## Reference Device
- Sponger Cable System, Modification - Pioneer Surgical Technology Inc. Sternal Cable System (K935481)
- Geminus Fossa Specific Plate System (K122310)
- Geminus Fossa Specific Plate System (K122737)
- Proximal Humerus Fixation System (K242436)
- Protean Fragment Plating System (K241815)
- Synthes Titanium Alloy Volar Distal Radius Plate System (K963798)
## Device Description
The current Geminus Volar Distal Radius Plating System includes medical grade titanium alloy (Ti 6Al-V4 ELI) bone plates offered in 12 sizes, an ulnar hook plate, Cobalt-Chromium-Molybdenum (CoCrMo) and titanium alloy (Ti 6Al-V4 ELI) bone screws and pegs, and specialized instrumentation for the repair of distal radius and osteotomies. The system also includes the Radial Hook Plate extensions to buttress a volar marginal fragment, and the Buttress Button assembly to capture and hold small dorsal corner fractures.
Page 1 of 3
{5}
K252949
skeletal dynamics
UNDERSTANDING THE UPPER EXTREMITY
The implants and instruments are provided non-sterile and shall be sterilized at the health care facility.
The Geminus Volar Distal Radius Plating System is comprised of:
| System Components | Material(s) | Standard(s) |
| --- | --- | --- |
| Geminus Volar Distal Radius Plates | Ti 6Al-V4 ELI | ASTM F136 |
| Locking Smooth Pegs | Ti 6Al-V4 ELI | ASTM F136 |
| Locking Threaded Pegs | | |
| High Compression Locking Pegs | | |
| Non-Locking Threaded Pegs | | |
| Cortical Non-Locking Screws | | |
| Cortical Locking Screws | | |
| Washer Button | Ti 6Al-V4 ELI | ASTM F136 |
| Polyaxial Locking Screws | CoCrMo | ASTM F1537 |
| Geminus Hook Plate Screws | Ti 6Al-V4 ELI | ASTM F136 |
| Geminus Ulnar Hook Plate | Ti 6Al-V4 ELI | ASTM F136 |
| Geminus Radial Hook Plate | Ti 6Al-V4 ELI | ASTM F136 |
| Geminus Buttress Button Assembly | Ti 6Al-V4 ELI | ASTM F136 |
| | CP Ti Grade 2 | ASTM F86 |
## Indications for Use
The Skeletal Dynamics Geminus Volar Distal Radius Plating System is intended for fixation of fractures and osteotomies of the distal radius.
## Substantial Equivalence
The Geminus Volar Distal Radius Plating System is substantially equivalent to the Geminus Volar Distal Radius Plate System (K182492). The subject and predicate devices have the same intended use, and only minor differences in the technological characteristics are identified. These differences do not present any concerns related to the safety or effectiveness of the device as demonstrated by the performance testing.
## Comparison of Technological Characteristics with the Predicate Device
The subject Geminus Volar Distal Radius Plating System is substantially equivalent to the predicate Geminus Volar Distal Radius Plate (K182492). The subject device consists of the same types of components (bone plates and bone screws) as the predicate device. Similarities in material, manufacturing process, performance, sterility, and packaging are demonstrated in this premarket notification.
The primary difference between the subject device and its predicate device is the introduction of the buttress button assembly and radial hook plate extensions. However, the addition of these accessories do not raise any concerns of safety or effectiveness as demonstrated by the performance testing results.
{6}
K252949
skeletal dynamics
UNDERSTANDING THE UPPER EXTREMITY
## Performance Testing
Based on the similarities in indications for use, technological characteristics, and the results of engineering analysis and mechanical performance testing, the Skeletal Dynamics Geminus Volar Distal Radius Plating System has been demonstrated to be substantially equivalent to the predicate device. Construct testing, including both static and dynamic testing, confirmed that the subject device and its accessories met the established acceptance criteria. Therefore, the subject device does not raise new questions of safety or effectiveness and is considered as safe and effective as the legally marketed predicate device.
Additionally, Skeletal Dynamics Inc has conducted testing to assess the safety and compatibility of the Skeletal device portfolio in a magnetic resonance (MR) environment and provide labeling recommendations for the 1.5 T and 3 T clinical scanners. Magnetically induced displacement force per ASTM F2052, Magnetically induced torque per ASTM F2213, MR image artifact per ASTM F2119, and radiofrequency (RF) induced heating per ASTM F2182 were evaluated following the FDA Guidance for Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (FDA-2019-D-2837).
## Conclusions
The Skeletal Dynamics’ Geminus Volar Distal Radius Plating System is substantially equivalent to the predicate device identified in this premarket notification.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.