K252937 · Dosisoft SA · IYE · Jun 12, 2026 · Radiology
Device Facts
Record ID
K252937
Device Name
ThinkQA 3
Applicant
Dosisoft SA
Product Code
IYE · Radiology
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ThinkQA 3 is a software medical device designed for Patient-Specific Quality Assurance (QA) in External Beam Radiation Therapy (RT). ThinkQA 3 is intended to help verify that each patient's treatment is delivered as planned, by identifying discrepancies between treatment plan and actual delivery. ThinkQA 3 consists of a software platform and a suite of independent and complementary verifications. For each patient, ThinkQA 3 imports treatment plan data and treatment machine information, such as delivery logs and EPID (Electronic Portal Image Device) images. ThinkQA 3 verifications are performed either (1) pre-treatment, or (2) post-treatment with data collected during actual treatment delivery (in vivo). ThinkQA 3 is designed for integration into both standard RT and Adaptive RT (ART) workflows. ThinkQA 3 verifications operate independently, offering distinct insights into treatment delivery quality. With these, the clinical users can better assess proper treatment course, identify possible root causes of detected errors, evaluate clinical relevance, and assist in treatment adaptation if necessary. ThinkQA 3 is intended for use only by qualified healthcare professionals trained in radiation therapy QA procedures. ThinkQA 3 does not recommend clinical interventions, nor directly control or modify treatment delivery. All clinical decisions must be made by qualified professionals based on their judgment and full clinical context.
Device Story
ThinkQA 3 is a software platform for patient-specific QA in external beam radiation therapy. It ingests treatment plan data, delivery logs, and EPID images. The system performs independent verifications—Secondary Dose Calculation (SDC), EPID pre-treatment/in vivo, and logfile pre-treatment/in vivo analysis—to detect discrepancies between planned and actual delivery. Used in clinical settings by radiation therapy professionals, it provides a web-based dashboard displaying pass/fail results and insights into treatment quality. It does not control treatment delivery or recommend clinical interventions; it supports clinical decision-making by identifying potential errors and assisting in treatment adaptation. Benefits include improved verification of treatment accuracy and integration into standard and adaptive RT workflows.
Clinical Evidence
Bench testing only. Performance, functional, and algorithmic testing were conducted under clinically representative conditions. Validation included risk management assessment and cybersecurity testing. The device complies with AAPM working group 219 and AAPM Task Group Report 307 recommendations. No clinical trial data was required or presented.
Technological Characteristics
Software-only device; runs on AlmaLinux 9. Architecture includes a central database, web-based interface, and modular verification suite (SDC, EPT, LPT, LIV, EIV). Uses a proprietary dose computation engine. Connectivity is networked within the RT center. No physical materials or energy sources; software-based analysis of external data (treatment plans, logs, EPID images).
Indications for Use
Indicated for patient-specific quality assurance in external beam radiation therapy (RT) to verify treatment delivery accuracy by identifying discrepancies between planned and delivered radiation. Intended for use by qualified healthcare professionals trained in RT QA procedures.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
DOSIsoft SA
Ahmad Khalil
Quality and Regulatory Affairs Responsible
45/47 Avenue Carnot
Cachan, 94230
France
Re: K252937
Trade/Device Name: ThinkQA 3
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: IYE, MUJ
Dated: May 13, 2026
Received: May 13, 2026
Dear Ahmad Khalil:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252937 - Ahmad Khalil
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252937 - Ahmad Khalil
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K252937
?
Please provide the device trade name(s).
?
ThinkQA 3
Please provide your Indications for Use below.
?
ThinkQA 3 is a software medical device designed for Patient-Specific Quality Assurance (QA) in External Beam Radiation Therapy (RT).
ThinkQA 3 is intended to help verify that each patient's treatment is delivered as planned, by identifying discrepancies between treatment plan and actual delivery.
ThinkQA 3 consists of a software platform and a suite of independent and complementary verifications. For each patient, ThinkQA 3 imports treatment plan data and treatment machine information, such as delivery logs and EPID (Electronic Portal Image Device) images. ThinkQA 3 verifications are performed either (1) pre-treatment, or (2) post-treatment with data collected during actual treatment delivery (in vivo). ThinkQA 3 is designed for integration into both standard RT and Adaptive RT (ART) workflows.
ThinkQA 3 verifications operate independently, offering distinct insights into treatment delivery quality. With these, the clinical users can better assess proper treatment course, identify possible root causes of detected errors, evaluate clinical relevance, and assist in treatment adaptation if necessary.
ThinkQA 3 is intended for use only by qualified healthcare professionals trained in radiation therapy QA procedures. ThinkQA 3 does not recommend clinical interventions, nor directly control or modify treatment delivery. All clinical decisions must be made by qualified professionals based on their judgment and full clinical context.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
DOSI soft
CERTIFIED ISO 9001:2015
## Premarket Notification 510(k) Summary
[as required by 21 CFR 807.92]
# #K252937 510(k) Summary
**1. Date the Summary was prepared:** 2026-06-12
# **2. Contact Details**
- Applicant Name: DOSIsoft SA
- Applicant Adress: 45/47 Avenue Carnot
94230 Cachan – FRANCE
- Applicant Contact Telephone: +33 1 41 24 26 26
- Applicant Contact: Mr. Marc USZYNSKI
- Applicant Contact Email: uszynski@dosisoft.com
- Correspondent Name: DOSIsoft SA
- Correspondent Address: 45/47 Avenue Carnot
94230 Cachan – FRANCE
- Correspondent Contact Telephone: +33 1 41 24 26 26
- Correspondent Contact: Mr. Ahmad KHALIL
- Correspondent Contact Email: regulatory@dosisoft.fr
# **3. Device Name**
- Device Trade Name: ThinkQA 3
- Common Name: Medical charged-particle radiation therapy system
- Classification Name: Accelerator, Linear, Medical
- Regulation Number: 892.5050
- Product Code: IYE, MUJ
# **4. Legally Marketed Predicate Devices**
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K231573 | ThinkQA (Edition 2) | IYE |
| Reference # | Reference device Trade Name | Product Code |
| K180106 | ThinkQA | IYE |
| K112723 | EPIgray | MUJ |
| K172534 | QualiForme D LINACWATCH/LINACVIEW | IYE |
DOSISOFT SA
45/47, avenue Carnot
94230 Cachan - France
+33 (0)1 41 24 26 26
+33 (0)1 41 24 26 28
info@dosisoft.com
www.dosisoft.com
RCS Créteil 443 195 433
Page 1/4
{5}
DOSI soft
CERTIFIED ISO 9001:2015
## Executive Summary
### 5. Device Description Summary
ThinkQA 3 consists of a software platform and a suite of software modules, performing various pre-treatment and in vivo verifications.
ThinkQA 3 includes:
- a software platform that collects and receives data, performs calculations, implements workflows, and stores the results in a single database.
- a web application that allows within a homogeneous look & feel to display the computed results from any compatible web browser across the institution (with access control) and pass/fail synthetic information for the active verification's.
- a set of software modules, implementing several verifications related to a patient treatment course, namely:
- a Secondary Dose Calc verification (SDC)
- an EPID Pre-Treatment verification (EPT)
- a Logfiles Pre-Treatment verification (LPT)
- a Logfiles InVivo verification (LIV)
- an EPID InVivo verification (EIV)
- a common administration interface, implementing users settings, verification protocols, system and users logs, licensing, database management, installation & configurations settings.
ThinkQA 3 scales to offer computation capacity from small to large radiation therapy (RT) centres, and the capability to offer a single web-based interface to access all patient treatment courses and patient information from within the RT centre.
ThinkQA 3 can be used in a modular fashion, performing only a subset of available verifications, but also provide a synthetic overview of all verifications performed for a given treatment fraction and overall treatment course for a given patient.
### 6. Intended Use/Indications for Use
ThinkQA 3 is a software medical device designed for Patient-Specific Quality Assurance (QA) in External Beam Radiation Therapy (RT).
ThinkQA 3 is intended to help verify that each patient's treatment is delivered as planned, by identifying discrepancies between treatment plan and actual delivery.
ThinkQA 3 consists of a software platform and a suite of independent and complementary verifications. For each patient, ThinkQA 3 imports treatment plan data and treatment machine information, such as delivery logs and EPID (Electronic Portal Image Device) images. ThinkQA 3 verifications are performed either (1) pre-treatment, or (2) post-treatment with data collected during actual treatment delivery (in vivo). ThinkQA 3 is designed for integration into both standard RT and Adaptive RT (ART) workflows.
ThinkQA 3 verifications operate independently, offering distinct insights into treatment delivery quality. With these, the clinical users can better assess proper treatment course, identify possible root causes of detected errors, evaluate clinical relevance, and assist in treatment adaptation if necessary.
ThinkQA 3 is intended for use only by qualified healthcare professionals trained in radiation therapy QA procedures. ThinkQA 3 does not recommend clinical interventions, nor directly control or modify treatment delivery. All clinical decisions must be made by qualified professionals based on their judgment and full clinical context.
DOSISOFT SA
45/47, avenue Carnot
94230 Cachan - France
+33 (0)1 41 24 26 26
+33 (0)1 41 24 26 28
info@dosisoft.com
www.dosisoft.com
RCS Créteil 443 195 433
Page 2/4
{6}
## 7. Indication for Use Comparison
ThinkQA 3 represents substantial equivalence to ThinkQA Edition 2.
The indications for use of ThinkQA Edition 2 are encompassed within the broader indications for use of ThinkQA 3.
Specifically, ThinkQA Edition 2 is intended for secondary dose verification by independently verifying that the dose distribution calculated by a Treatment Planning System (TPS) for External Beam Radiation Therapy (EBRT) is consistent with treatment plan parameters.
ThinkQA 3 is intended to help verify that each patient's radiation treatment is delivered as planned, by identifying discrepancies between the treatment plan and actual delivery. This intended use includes secondary dose verification functionality as well as additional pretreatment and in vivo independent verification capabilities.
Both devices are software-only medical devices that receive treatment-related data, perform independent calculations and analyses, and store results within a database. ThinkQA 3 is built upon the same underlying software platform as ThinkQA Edition 2. While ThinkQA 3 incorporates additional software modules and expanded verification functionalities, these enhancements do not alter the fundamental intended purpose of independently verifying radiation therapy treatments.
## 8. Technological Comparison
ThinkQA 3 is a new software Edition (Edition 3), built upon successive software Editions which were individually focusing on elementary QA verifications. ThinkQA 3 brings these elementary QA verifications into a single and consistent architecture, database and system, providing to the user an overall, integrated and comprehensive dashboard of Patient-specific QA tasks.
ThinkQA 3 has the same software architecture as ThinkQA Edition 2, already prepared to integrate several verifications. ThinkQA 3 integrates the same single dose computation engine as in ThinkQA 2. This dose computation engine is used by several other verifications inside ThinkQA 3.
ThinkQA 3 integrates also capacity to read and analyze regulatory logfiles generated by the treatment machines. These logfiles are analyzed by ThinkQA 3 as additional verification using the same dose computation engine and used either in a pre-treatment configuration or during a treatment fraction.
ThinkQA 3 also runs on a different and up-to-date operating system (AlmaLinux 9) compared to ThinkQA Edition 2.
## 9. Non-Clinical and/or Clinical Tests Summary
ThinkQA 3 was designed and documented in accordance with the recommendations of FDA'Guidance for the Content of Premarket Submissions for Device Software Functions' for software devices requiring Enhanced Documentation.
ThinkQA 3 was submitted to performance, functional and algorithmic testing, risk management assessment, including cybersecurity and validation activities under clinically representative conditions. The results of performance, functional and algorithmic testing, risk management assessment and validation activities under clinically representative conditions demonstrate the safety and effectiveness of ThinkQA 3 in comparison to the predicate device.
ThinkQA 3 meets the requirements of the device, its user needs and intended use, which are demonstrated to be substantially equivalent to those of the predicate device ThinkQA Edition 2.
Also, ThinkQA 3 complies with applicable recommendations of the AAPM working group 219 regarding independent calculation-based dose/MU verification for IMRT (see [1]) and the AAPM Task Group Report 307: Use of EPIDs for Patient-Specific IMRT and VMAT QA. https://doi.org/10.1002/mp.16536.
The substantial equivalence discussion sustains the claim that ThinkQA 3 intended use, clinical and technical (principles of operation, functionalities and critical performances) characteristics are the same as for the predicate device ThinkQA Edition 2 legally marketed in the U.S.
**DOSISOFT SA**
45/47, avenue Carnot
94230 Cachan - France
![Phone icon]() +33 (0)1 41 24 26 26
![Phone icon]() +33 (0)1 41 24 26 28
info@dosisoft.com
www.dosisoft.com
RCS Créteil 443 195 433
Page 3/4
{7}
DOSIsoft
CERTIFIED ISO 9001:2015 COMPANY
The new verification implementation, included in ThinkQA 3, for EPID in vivo has been validated in terms of performance and functional point of view to its reference device EPIgray. ThinkQA 3 provides enhancements compared to EPIgray in terms of computation and usability in the classical and new adaptative radiotherapy workflows. This approach improves commissioning, usability, training requirements, administration aspects, and support & maintenance procedures for the clinical centers.
ThinkQA 3 also includes a verification based on machine delivery logfiles with a similar functional point of view to its reference device LinacWatch / Linac View. ThinkQA 3 logfiles verification has demonstrated similar error detection capabilities as described in the scientific state of the art literature.
The results of verification and validation activities demonstrate the safety and effectiveness of ThinkQA 3 in comparison to the predicate. ThinkQA 3 meets the requirements of the device, its user needs and intended use, which are demonstrated to be substantially equivalent to those of the predicate.
In summary, DOSIsoft therefore considers that, in its opinion, ThinkQA 3 is substantially equivalent to and is at least as safe and effective as the selected predicate.
## 10 Conclusion
The substantial equivalence discussion sustains the claim that ThinkQA 3 shares the same intended use, clinical and technical characteristics as the predicate ThinkQA Edition 2 legally marketed in the U.S. (K231573).
In conclusion, ThinkQA 3 aligns with the device requirements, user needs, and intended use, showcasing substantial equivalence to the predicate device. Differences in certain features are assessed to have either negligible impact or no significant effect on safety and effectiveness. Thus, DOSIsoft asserts that ThinkQA 3 is substantially equivalent to its predicate device.
DOSISOFT SA
45/47, avenue Carnot
94230 Cachan - France
+33 (0)1 41 24 26 26
+33 (0)1 41 24 26 28
info@dosisoft.com
www.dosisoft.com
RCS Créteil 443 195 433
Page 4/4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.