Retrospective evaluation of archived clinical specimens; Study Period: 2024-2025 respiratory season
Individuals with signs and symptoms of influenza infection during the 2024-2025 North American respiratory season; Sample Size: 118
FDA-cleared, CLIA waived RT-PCR assay
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for RSV
Indications for Use
The Fast PCR Mini Respiratory Panel is an automated, multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B and respiratory syncytial virus (RSV) nucleic acids in anterior nasal swab specimens obtained from individuals with signs and symptoms of respiratory tract infections. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B and RSV infections in humans and is not intended to detect influenza C virus infection. Nucleic acids from the viral organisms identified by this test are generally detectable in anterior nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aid in the diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive results do not rule out co-infection with other organisms. The organism(s) detected by the Fast PCR Mini Respiratory Panel may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infections.
Device Story
Fast PCR Mini Respiratory Panel (MRP) is an automated, multiplexed RT-PCR test for qualitative detection of SARS-CoV-2, influenza A, influenza B, and RSV in anterior nasal swab specimens. The system utilizes a proprietary Hyperbaric Heating (HBH) process (15 seconds) within a single-use consumable Test Disc to lyse viruses, neutralize PCR inhibitors, and protect RNA. Fluidics are centrifugally driven, requiring no pumps. The Fast PCR Instrument, comprising a tablet, base station, and operating modules, performs HBH and ultrafast thermocycling. Results are delivered in approximately 10 minutes. Used in CLIA-waived settings by untrained operators; output is qualitative (positive/negative). Results aid clinicians in differential diagnosis when used with other clinical/epidemiological data. Benefits include rapid turnaround time for point-of-care respiratory infection management.
Clinical Evidence
Clinical study conducted at 9 CLIA-waived sites with 1,906 evaluable subjects (1,788 prospective, 118 archived). Performance compared to FDA-cleared comparator. Prospective PPA/NPA: Influenza A (97.5%/97.1%), Influenza B (100%/99.7%), RSV (91.7%/99.5%), SARS-CoV-2 (97.7%/99.4%). Combined prospective and archived PPA/NPA: Influenza A (97.5%/97.2%), Influenza B (100%/99.7%), RSV (93.0%/99.5%), SARS-CoV-2 (97.7%/99.4%).
Technological Characteristics
Multiplexed RT-PCR; Hyperbaric Heating (HBH) for sample prep; centrifugal fluidics in single-use Test Disc; targets ORF1ab, ORF8, nucleocapsid (SARS-CoV-2) and matrix genes (Flu A/B, RSV); RNase P internal control; standalone Fast PCR instrument; ambient storage.
Indications for Use
Indicated for qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and RSV nucleic acids in anterior nasal swab specimens from symptomatic individuals. For prescription use only.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT**
# **I Background Information:**
# **A 510(k) Number:**
K252932
# **B Applicant:**
Autonomous Medical Devices Incorporated
# **C Proprietary and Established Names:**
Fast PCR Mini Respiratory Panel
# **D Regulatory Information:**
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QOF | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology |
# **II Submission/Device Overview:**
# **A Purpose for Submission:**
To demonstrate that the Fast PCR Mini Respiratory Panel (MRP) is substantially equivalent to the Xpert Xpress CoV-2/Flu/RSV *plus* (K242071) on the GeneXpert Xpress System and to obtain clearance for the Fast PCR MRP.
# **B Measurand:**
- Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) RNA
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
- Influenza A RNA
- Influenza B RNA
- Respiratory Syncytial Virus (RSV) RNA
# C Type of Test:
Qualitative reverse transcriptase polymerase chain reaction (RT-PCR)
# III Intended Use/Indications for Use:
# A Intended Use(s):
See Indications for Use below.
# B Indication(s) for Use:
The Fast PCR Mini Respiratory Panel is an automated, multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B and respiratory syncytial virus (RSV) nucleic acids in anterior nasal swab specimens obtained from individuals with signs and symptoms of respiratory tract infections. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B and RSV infections in humans and is not intended to detect influenza C virus infection.
Nucleic acids from the viral organisms identified by this test are generally detectable in anterior nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aid in the diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
Positive results do not rule out co-infection with other organisms. The organism(s) detected by the Fast PCR Mini Respiratory Panel may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infections.
# C Special Conditions for Use Statement(s):
Rx – For Prescription Use Only
For In Vitro Diagnostic Use Only
K252932 - Page 2 of 26
{2}
# **D Special Instrument Requirements:**
For use with the Fast PCR Instrument
# **IV Device/System Characteristics:**
# **A Device Description:**
The Fast PCR Mini Respiratory Panel (MRP) is a multiplexed RT-PCR test for use with the FAST PCR instrument for the simultaneous, qualitative detection and identification of SARS-CoV-2, influenza A, influenza B, and RSV from nucleic acids in anterior nasal swab (ANS) specimens in approximately 10 minutes.
To perform the test, an anterior nasal swab is collected and added to the AMDI Sample Buffer tube. A provided fixed volume pipette is used to add 400 μL of the eluted sample to the Fast PCR MRP Test Disc sample port. Each specimen is processed individually. An operator logs into the Fast PCR App and scans the Patient ID on the AMDI sample buffer tube containing the collected ANS specimen and the unique barcode on the Fast PCR Mini Respiratory Panel Test Disc. A disposable pipette provided with the Test Disc kit is used to remove a fixed volume of the AMDI sample buffer from the sample buffer tube. The pipette is inserted into the sample port of the Test Disc and the sample buffer is dispensed into the Test Disc. The Test Disc contains all necessary reagents for the detection of SARS-CoV-2, influenza A, influenza B and RSV RNA and the internal control (RNase P). The pipette is removed and the closure for the Test Disc is pressed into place. The disc loading tray on the Operating Module (part of the Fast PCR instrument) is opened, and the Test Disc is inserted into the Operating Module.
The Fast PCR instrument automatically performs the test process, and the Operator has no further hands-on tasks.
# **B Principle of Operation:**
The Fast PCR MRP is performed using the Fast PCR instrument, which automates all test processes from sample processing, through RT-PCR and detection steps, to result interpretation and result reporting.
The Fast PCR MRP uses a sample preparation technology called Hyperbaric Heating (HBH) that is applied to ANS specimens. This process simultaneously lyses the viruses present in the sample, neutralizes PCR inhibitors, and protects single-stranded RNA in part by eliminating RNase activity in the sample. The HBH process is performed in the Fast PCR MRP consumable Test Disc that derives all its liquids from the AMDI Sample Buffer, and therefore requires no on-board storage of liquids. All the fluidics in the Test Disc are then driven by centrifugal forces produced when rotating the Test Disc inside the Operating Module within the Fast PCR instrument.
The assay RT-PCR primers and probes target the ORF1ab non-structural region, ORF8, and the nucleocapsid genes of SARS-CoV-2 and target the matrix genes of influenza A, influenza B, and RSV. The RNase P internal control is included to control for adequate processing of the target
K252932 - Page 3 of 26
{3}
viruses through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.
The Fast PCR instrument automatically analyzes and interprets the results, reporting results directly in the Fast PCR App.
### C Instrument Description Information:
# 1. 1. Instrument Name:
Fast PCR instrument
# 1. 2. Specimen Identification:
Specimen identification is either entered manually or scanned via a barcode scanner.
# 1. 3. Specimen Sampling and Handling:
Anterior nasal swabs in AMDI Sample Buffer.
# 1. 4. Calibration:
Not applicable.
# 1. 5. Quality Control:
# *Internal Controls*
The assay uses RNase P as an internal control (IC) to ensure adequate sample collection and processing. The internal control monitors the overall performance of the RT-PCR reaction for potential sample mediated inhibition or failure of the reagents. If the internal control result fails, all four target analytes (Influenza A, Influenza B, RSV and SARS-CoV-2) will have an invalid result. If the internal control result passes, the four target analytes will have either a positive or a negative result.
# *External Controls*
The recommended Fast PCR MRP External Controls (Positive and Negative) are packaged and sold separately from the assay kit. The positive control is a lyophilized bead containing inactivated whole viruses (influenza A, influenza B, Respiratory Syncytial Virus, and SARS-CoV-2) and armored RNase P RNA. The negative control is a lyophilized bead containing armored RNase P RNA. External Controls (EC) may be performed to conform with internal Quality Control (QC) procedures, or with local, state, or federal regulations. The Fast PCR MRP External Controls were evaluated in the analytical and clinical studies.
K252932 - Page 4 of 26
{4}
# **V Substantial Equivalence Information:**
# **A Predicate Device Name(s):**
Xpert Xpress CoV-2/Flu/RSV *plus*
# **B Predicate 510(k) Number(s):**
K242071
# **C Comparison with Predicate(s):**
| Device & Predicate Device(s): | K252932 | K242071 |
| --- | --- | --- |
| Device Trade Name | Fast PCR Mini Respiratory Panel | Xpert Xpress CoV-2/Flu/RSV *plus* |
| General Device Characteristic Similarities | K252932 | K242071 |
| Technology/Detection | Same | Multiplexed Real-time Reverse Transcription Polymerase Chain Reaction (RT-PCR) |
| Intended Use/Indications For Use | The Fast PCR Mini Respiratory Panel is an automated, multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B and respiratory syncytial virus (RSV) nucleic acids in anterior nasal swab specimens obtained from individuals with signs and symptoms of respiratory tract infections. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, | The Xpert Xpress CoV-2/Flu/RSV *plus* test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous *in vitro* qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. |
K252932 - Page 5 of 26
{5}
| | influenza A, influenza B and RSV infections in humans and is not intended to detect influenza C virus infection. Nucleic acids from the viral organisms identified by this test are generally detectable in anterior nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aid in the diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Positive results do not rule out co-infection with other organisms. The organism(s) detected by the Fast PCR Mini Respiratory Panel may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infections. | The Xpert Xpress CoV-2/Flu/RSV *plus* is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Xpert Xpress CoV-2/Flu/RSV *plus* test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. |
| --- | --- | --- |
| **Assay Targets** | Same | SARS-CoV-2, Influenza A, Influenza B, RSV viral RNA |
| **Test Format** | Same | Single Use |
| **Automation** | Same | Automated Nucleic Acid Extraction, Detection and Results Interpretation |
| **Assay Result** | Same | Qualitative |
| **Test Interpretation** | Same | Automated test interpretation and reporting |
K252932 - Page 6 of 26
{6}
| General Device Characteristic Differences | K252932 | K242071 |
| --- | --- | --- |
| **Specimen Types** | Anterior nasal swab | • Nasopharyngeal swab (NPS) • Anterior nasal swab (ANS) |
| **Transport Media** | AMDI Sample Buffer | • Universal Transport Medium (UTM) /Viral Transport Medium (VTM) • eNAT |
| **Internal Control** | RNase P | • Sample Processing Control (SPC) • Probe Check Control (PCC) |
| **External Controls** | • Fast PCR MRP Positive External Control • Fast PCR MRP Negative External Control | • External Positive Control: NATtrol Flu/RSV/SARS-CoV-2; Cat # NATFRC-6C-IVD • External Negative Control: Coxsackievirus A9; Cat # NATCV9-6C-IVD |
| **Instrument** | Fast PCR Instrument | Cepheid GeneXpert Xpress System |
| **Time to Result** | 10 min for sample preparation and RT-PCR | 36 min or less for sample preparation and RT-PCR |
## VI Standards/Guidance Documents Referenced:
- • 21 CFR Part 866.3981, 'Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test'
- • 21 CFR 862.2570, 'Instrumentation for Clinical Multiplex Test Systems'
- • FDA Guidance: 'Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses' (July 15, 2011)
- • FDA Guidance: 'Respiratory Viral Panel Multiplex Nucleic Acid Assay – Class II Special Controls Guidance for Industry and FDA Staff' (October 9, 2009)
- • FDA Guidance: 'Instrumentation for Clinical Multiplex Test Systems - Class II Special Controls Guidance for Industry and FDA Staff' (March 10, 2005)
- • FDA Guidance: 'Recommendations for CLIA Waiver Applications for Manufacturers of In Vitro Diagnostic Devices' (February 2020)
- • FDA Guidance: 'Recommendations for Dual 510(k) and CLIA Waiver by Application Studies' (February 2020)
- • FDA Guidance: 'Content of Premarket Submissions for Device Software Functions' (June 2023)
- • FDA Guidance: 'Cybersecurity in Medical Devices – Quality System Considerations and Content of Premarket Submission' (February 2026)
- • FDA Guidance: 'Electromagnetic Compatibility (EMC) of Medical Devices' (June 2022)
K252932 - Page 7 of 26
{7}
- • CLSI EP12 3$^{rd}$ Edition, 7-315
- • CLSI EP17-A2, 7-233
- • CLSI EP25 2$^{nd}$ Edition, 7-318
- • CLSI EP07 3$^{rd}$ Edition, 7-275
- • IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION, 19-34
- • IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, 19-36
- • IEC 60825-1 Edition 2.0 2007-03, 12-273
- • IEC 61326-1 Edition 3.0 2020-10, 19-42
- • IEC 61326-2-6 Edition 3.0 2020-10, 19-43
- • ISTA 3A 2018, 5-126
- • CISPR11 ed.5.0 (2024), Class A
- • FCC 47 CFR Part 18 – Industrial, Scientific, and Medical Equipment
- • FCC Part 15B/ICES-003 – Radio Frequency Devices
- • ISO 14971
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
A precision and reproducibility study was conducted to assess the total variability of the Fast PCR Mini Respiratory Panel (MRP) across untrained operators, study sites, testing days, device lots, and instruments. Test panels, consisting of three different concentrations of samples, were prepared by AMDI and were aliquoted, blinded and labeled for distribution to the testing sites.
Variability was assessed using a test panel at three study sites with two untrained operators per site. Testing was carried out over at least five non-consecutive testing days at each site with three Fast PCR instruments per site. Three lots of the Fast PCR MRP kit, three lots of Fast PCR MRP Positive External Controls, three lots of Fast PCR MRP Negative External Controls, and three replicates per test condition were used.
The four target analytes (influenza A, influenza B, RSV, and SARS-CoV-2) were diluted to 2x Limit of Detection (LoD) (weak positive) and 5x LoD (moderate positive) in pooled negative anterior nasal swab matrix. The negative sample consisted of pooled negative anterior nasal swab matrix without analyte. Strains and concentrations are shown in **Table 1**.
**Table 2** presents the qualitative summary for influenza A, influenza B, RSV, and SARS-CoV-2 during panel member testing. The reproducibility of the Fast PCR System for each sample of the test panel stratified by operator is summarized in **Table 3** and the Ct Value analysis is summarized in **Table 4**. The results of the study demonstrate acceptable assay precision and reproducibility.
K252932 - Page 8 of 26
{8}
**Table 1.** Strains used in the Precision and Reproducibility Study
| Target | Strain | Weak Positive (2x LoD) | Moderate Positive (5x LoD) |
| --- | --- | --- | --- |
| Negative | Negative Anterior Nasal Swab in AMDI sample Buffer | N/A | N/A |
| Influenza A | Influenza A/Darwin/9/2021 (H3N2) | 500 copies/mL | 1250 copies/mL |
| Influenza B | Influenza B/Colorado/06/2017 (Victoria lineage) | 1000 copies/mL | 2500 copies/mL |
| RSV | RSV A 2006 Isolate | 500 copies/mL | 1250 copies/mL |
| SARS-CoV-2 | USA/WA1/2020 | 1000 copies/mL | 2500 copies/mL |
**Table 2.** Fast PCR MRP Reproducibility Across Study Sites
| Sample | Analyte | Site | Invalid | Negative | Positive | #Pass | N Valid | %Pass | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Negative | Influenza A | A | 0 | 87 | 2 | 87 | 89 | 97.8% | 92.2%-99.4% |
| | Influenza A | B | 2 | 95 | 0 | 95 | 95 | 100.0% | 96.1%-100.0% |
| | Influenza A | C | 5 | 94 | 0 | 94 | 94 | 100.0% | 96.1%-100.0% |
| | Influenza B | A | 0 | 89 | 0 | 89 | 89 | 100.0% | 95.9%-100.0% |
| | Influenza B | B | 2 | 95 | 0 | 95 | 95 | 100.0% | 96.1%-100.0% |
| | Influenza B | C | 5 | 94 | 0 | 94 | 94 | 100.0% | 96.1%-100.0% |
| | RSV | A | 0 | 89 | 0 | 89 | 89 | 100.0% | 95.9%-100.0% |
| | RSV | B | 2 | 95 | 0 | 95 | 95 | 100.0% | 96.1%-100.0% |
| | RSV | C | 5 | 94 | 0 | 94 | 94 | 100.0% | 96.1%-100.0% |
| | SC2 | A | 0 | 88 | 1 | 89 | 89 | 98.9% | 93.9%-99.8% |
| | SC2 | B | 2 | 95 | 0 | 95 | 95 | 100.0% | 96.1%-100.0% |
| | SC2 | C | 5 | 94 | 0 | 94 | 94 | 100.0% | 96.1%-100.0% |
| Negative Overall | Influenza A | | 7 | 276 | 2 | 276 | 278 | 99.3% | 97.4%-99.8% |
| | Influenza B | | 7 | 278 | 0 | 278 | 278 | 100.0% | 98.6%-100.0% |
| | RSV | | 7 | 278 | 0 | 278 | 278 | 100.0% | 98.6%-100.0% |
K252932 - Page 9 of 26
{9}
| Sample | Analyte | Site | Invalid | Negative | Positive | #Pass | N Valid | %Pass | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SC2 | | 7 | 277 | 1 | 277 | 278 | 99.6% | 98.0%-99.9% |
| 2x LoD | Influenza A | A | 8 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | Influenza A | B | 2 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | Influenza A | C | 11 | 1 | 92 | 92 | 93 | 98.9% | 94.2%-99.8% |
| | Influenza B | A | 8 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| | Influenza B | B | 2 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| | Influenza B | C | 11 | 1 | 92 | 92 | 93 | 98.9% | 94.2%-99.8% |
| | RSV | A | 8 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | RSV | B | 2 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | RSV | C | 11 | 3 | 90 | 90 | 93 | 96.8% | 90.9%-99.8% |
| | SC2 | A | 8 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | SC2 | B | 2 | 1 | 89 | 89 | 90 | 98.9% | 94.0%-99.8% |
| | SC2 | C | 11 | 1 | 92 | 92 | 93 | 98.9% | 94.2%-99.8% |
| 2x LoD Overall | Influenza A | | 21 | 3 | 270 | 270 | 273 | 98.9% | 96.8%-99.6% |
| | Influenza B | | 21 | 1 | 272 | 272 | 273 | 99.6% | 98.0%-99.9% |
| | RSV | | 21 | 5 | 268 | 268 | 273 | 98.2% | 95.8%-99.2% |
| | SC2 | | 21 | 3 | 270 | 270 | 273 | 98.9% | 96.8%-99.6% |
| 5x LoD | Influenza A | A | 4 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | Influenza A | B | 0 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | Influenza A | C | 6 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| | Influenza B | A | 4 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | Influenza B | B | 0 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | Influenza B | C | 6 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| | RSV | A | 4 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
K252932 - Page 10 of 26
{10}
| Sample | Analyte | Site | Invalid | Negative | Positive | #Pass | N Valid | %Pass | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | RSV | B | 0 | 1 | 90 | 90 | 91 | 98.9% | 94.0%-99.8% |
| | RSV | C | 6 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| | SC2 | A | 4 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | SC2 | B | 0 | 0 | 91 | 91 | 91 | 100.0% | 95.9%-100.0% |
| | SC2 | C | 6 | 0 | 90 | 90 | 90 | 100.0% | 95.9%-100.0% |
| **5x LoD Overall** | Influenza A | | 10 | 0 | 272 | 272 | 272 | 100.0% | 98.6%-100.0% |
| | Influenza B | | 10 | 0 | 272 | 272 | 272 | 100.0% | 98.6%-100.0% |
| | RSV | | 10 | 1 | 271 | 271 | 272 | 99.6% | 97.9%-99.9% |
| | SC2 | | 10 | 0 | 272 | 272 | 272 | 100.0% | 98.6%-100.0% |
**Table 3.** Fast PCR Reproducibility Across Operators
| Site: | | Site A | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Operator: | | Operator 1 | | | Operator 2 | | |
| Sample | Analyte | Pass | Valid | %Pass | Pass | Valid | %Pass |
| **Negative** | Influenza A | 44 | 45 | 97.8% | 43 | 44 | 97.7% |
| | Influenza B | 45 | 45 | 100.0% | 44 | 44 | 100.0% |
| | RSV | 45 | 45 | 100.0% | 44 | 44 | 100.0% |
| | SC2 | 44 | 45 | 97.8% | 44 | 44 | 100.0% |
| **2x LoD** | Influenza A | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| | Influenza B | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| | RSV | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| | SC2 | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| **5x LoD** | Influenza A | 46 | 46 | 100.0% | 45 | 45 | 100.0% |
| | Influenza B | 46 | 46 | 100.0% | 45 | 45 | 100.0% |
| | RSV | 46 | 46 | 100.0% | 45 | 45 | 100.0% |
| | SC2 | 46 | 46 | 100.0% | 45 | 45 | 100.0% |
| Site: | | Site B | | | | | |
| Operator: | | Operator 1 | | | Operator 2 | | |
| Sample | Analyte | Pass | Valid | %Pass | Pass | Valid | %Pass |
| **Negative** | Influenza A | 47 | 47 | 100.0% | 48 | 48 | 100.0% |
| | Influenza B | 47 | 47 | 100.0% | 48 | 48 | 100.0% |
| | RSV | 47 | 47 | 100.0% | 48 | 48 | 100.0% |
| | SC2 | 47 | 47 | 100.0% | 48 | 48 | 100.0% |
| **2x LoD** | Influenza A | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| | Influenza B | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| | RSV | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| | SC2 | 44 | 45 | 97.8% | 45 | 45 | 100.0% |
| **5x LoD** | Influenza A | 45 | 45 | 100.0% | 46 | 46 | 100.0% |
| | Influenza B | 45 | 45 | 100.0% | 46 | 46 | 100.0% |
K252932 - Page 11 of 26
{11}
| | RSV | 45 | 45 | 100.0% | 45 | 46 | 97.8% |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | SC2 | 45 | 45 | 100.0% | 46 | 46 | 100.0% |
| Site: | | Site C | | | | | |
| Operator: | | Operator 1 | | | Operator 2 | | |
| Sample | Analyte | Pass | Valid | %Pass | Pass | Valid | %Pass |
| Negative | Influenza A | 49 | 49 | 100.0% | 45 | 45 | 100.0% |
| | Influenza B | 49 | 49 | 100.0% | 45 | 45 | 100.0% |
| | RSV | 49 | 49 | 100.0% | 45 | 45 | 100.0% |
| | SC2 | 49 | 49 | 100.0% | 45 | 45 | 100.0% |
| 2x LoD | Influenza A | 45 | 45 | 100.0% | 47 | 48 | 97.9% |
| | Influenza B | 45 | 45 | 100.0% | 47 | 48 | 97.9% |
| | RSV | 43 | 45 | 95.6% | 47 | 48 | 97.9% |
| | SC2 | 45 | 45 | 100.0% | 47 | 48 | 97.9% |
| 5x LoD | Influenza A | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| | Influenza B | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| | RSV | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| | SC2 | 45 | 45 | 100.0% | 45 | 45 | 100.0% |
| Overall | | | | | | | |
| Sample | Analyte | %Pass | 95% CI | | | | |
| Negative | Influenza A | 99.8% | 97.4 – 99.8 | | | | |
| | Influenza B | 100.0% | 98.6 – 100.0 | | | | |
| | RSV | 100.0% | 98.6 – 100.0 | | | | |
| | SC2 | 99.9% | 98.0 – 99.9 | | | | |
| 2x LoD | Influenza A | 99.6% | 96.8 – 99.6 | | | | |
| | Influenza B | 99.9% | 98.0 – 99.9 | | | | |
| | RSV | 99.2% | 95.8 – 99.2 | | | | |
| | SC2 | 99.6% | 96.8 – 99.6 | | | | |
| 5x LoD | Influenza A | 100.0% | 98.6 – 100.0 | | | | |
| | Influenza B | 100.0% | 98.6 – 100.0 | | | | |
| | RSV | 99.9% | 97.9 – 99.9 | | | | |
| | SC2 | 100.0% | 98.6 – 100.0 | | | | |
Table 4. Precision and Reproducibility of the Fast PCR MRP, Ct Value Analysis
| Panel | Target | N | Mean Ct | Within Run (Repeatability) | | Between Days | | Between Instruments | | Between Lots | | Between Operators | | Between Sites | | Overall (Reproducibility) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| Low Positive (2x LoD) | Influenza A | 269 | 38.68 | 0.99 | 2.57 | 0.38 | 0.97 | 0.35 | 0.91 | 0.25 | 0.66 | 0.17 | 0.45 | 0.13 | 0.35 | 1.17 | 3.02 |
| | Influenza B | 270 | 38.71 | 0.67 | 1.73 | 0.26 | 0.68 | 0.22 | 0.56 | 0.13 | 0.34 | 0.11 | 0.28 | 0.10 | 0.27 | 0.78 | 2.01 |
| | RSV | 268 | 39.28 | 1.07 | 2.73 | 0.21 | 0.54 | 0.20 | 0.50 | 0.18 | 0.46 | 0.15 | 0.37 | 0.28 | 0.70 | 1.17 | 2.97 |
| | SARS-COV-2 | 265 | 37.52 | 0.63 | 1.67 | 0.13 | 0.35 | 0.35 | 0.93 | 0.09 | 0.24 | 0.12 | 0.33 | 0.27 | 0.71 | 0.79 | 2.11 |
| Moderate Positive (5x LoD) | Influenza A | 271 | 37.13 | 0.68 | 1.82 | 0.17 | 0.45 | 0.45 | 1.22 | 0.27 | 0.72 | 0.17 | 0.46 | 0.22 | 0.60 | 0.92 | 2.47 |
| | Influenza B | 272 | 37.24 | 0.50 | 1.36 | 0.11 | 0.29 | 0.16 | 0.42 | 0.10 | 0.27 | 0.02 | 0.05 | 0.09 | 0.23 | 0.56 | 1.49 |
| | RSV | 271 | 37.82 | 0.95 | 2.51 | 0.23 | 0.61 | 0.24 | 0.64 | 0.04 | 0.11 | 0.12 | 0.33 | 0.13 | 0.34 | 1.02 | 2.71 |
| | SARS-COV-2 | 269 | 36.11 | 0.61 | 1.69 | 0.14 | 0.40 | 0.21 | 0.59 | 0.09 | 0.26 | 0.10 | 0.28 | 0.21 | 0.57 | 0.71 | 1.96 |
SD = standard deviation; CV(%) = percent coefficient of variation.; N = total number of samples
K252932 - Page 12 of 26
{12}
# 2. Linearity:
Not applicable. This is a qualitative assay.
# 3. Analytical Specificity/Interference:
# a. Analytical Reactivity (Inclusivity):
Wet-Testing: The inclusivity of the Fast PCR MRP was established through wet testing by evaluating a total of forty-two (42) strains, including 22 influenza A strains, six (6) influenza B strains, six (6) RSV strains, and eight (8) SARS-CoV-2 strains. The Fast PCR MRP was able to detect 100% of the strains tested for all analytes. Inactivated strains were used for SARS-CoV-2 and cultured virus were used for RSV and influenza strains. Each virus was spiked into pooled negative nasal matrix at analyte concentrations of 3x LoD and tested in triplicate. All strains of influenza A (Table 5), influenza B (Table 6), RSV (Table 7), and SARS-CoV-2 (Table 8) tested positive in all three replicates. Additionally, a total of twelve aliquots of anterior negative swab matrix were tested, and all were negative for each of the four target analytes.
**Table 5.** Influenza A Analytical Reactivity Wet-Testing Results
| Subtype | Strain | Concentration | % Positive |
| --- | --- | --- | --- |
| H1N1 pdm09 | A/California/07/2009 | 250 copies/mL | 100% (3/3) |
| | A/NY/01/09 | 27 TCID50/mL | 100% (3/3) |
| | A/NY/02/09 | 27 TCID50/mL | 100% (3/3) |
| | A/NY/03/09 | 27 TCID50/mL | 100% (3/3) |
| | A/Sydney/05/2021 | 250 copies/mL | 100% (3/3) ^{a} |
| | A/Wisconsin/588/2019 | 250 copies/mL | 100% (3/3) |
| | A/Wisconsin/67/2022 | 2.17E4 copies/mL | 100% (3/3) |
| H1N1 | A/Brisbane/59/2007 | 250 copies/mL | 100% (3/3) |
| | A/Michigan/45/15 | 27 TCID50/mL | 100% (3/3) |
| | A/New Caledonia/20/99 | 27 TCID50/mL | 100% (3/3) |
| | A/Taiwan/42/06 | 27 TCID50/mL | 100% (3/3) |
| | A/Victoria/2570/2019 | 250 copies/mL | 100% (3/3) |
| H3N2 | A/Brisbane/10/07 | 2.4 TCID50/mL | 100% (3/3) |
| | A/Cambodia/E0826360/2020 | 750 copies/mL | 100% (3/3) ^{b} |
| | A/Hong Kong/2671/2019 | 750 copies/mL | 100% (3/3) |
| | A/Hong Kong/4801/2014 | 750 copies/mL | 100% (3/3) |
| | A/Massachusetts/18/2022 | 750 copies/mL | 100% (3/3) |
| | A/Perth/16/2009 | 750 copies/mL | 100% (3/3) |
| | A/Singapore/INFIMH-16-0019/2016 | 750 copies/mL | 100% (3/3) |
| | A/Tasmania/503/2020 | 750 copies/mL | 100% (3/3) |
| | A/Texas/50/12 | 2.4 TCID50/mL | 100% (3/3) |
| | A/Thailand/08/2022 | 750 copies/mL | 100% (3/3) |
$^{a}$ RSV false positive result
K252932 - Page 13 of 26
{13}
$^{b}$ Influenza B false positive result
**Table 6.** Influenza B Analytical Reactivity Wet-Testing Results
| Sub-lineage | Strain | Concentration | % Positive |
| --- | --- | --- | --- |
| Victoria | B/Austria/1359417/2021 | 1500 copies/mL | 100% (3/3) |
| | B/Brisbane/60/2008 | 1500 copies/mL | 100% (3/3) ^{a} |
| | B/Maryland/15/2016 | 1500 copies/mL | 100% (3/3) |
| | B/Washington/02/2019 | 1500 copies/mL | 100% (3/3) |
| Yamagata | B/Brisbane/09/2014 | 750 copies/mL | 100% (3/3) |
| | B/Massachusetts/02/2012 | 750 copies/mL | 100% (3/3) |
$^{a}$ Influenza A false positive result
**Table 7.** RSV Analytical Reactivity Wet-Testing Results
| Subtype | Strain | Concentration | % Positive |
| --- | --- | --- | --- |
| RSV A | 2013 Isolate | 0.78 TCID50/mL | 100% (3/3) |
| | 4/2015 Isolate #1 | 0.78 TCID50/mL | 100% (3/3) |
| | A2 | 750 copies/mL | 100% (3/3) ^{a} |
| RSV B | 3/2015 Isolate #1 | 3.27 TCID50/mL | 100% (3/3) |
| | 9320 | 2250 copies/mL | 100% (3/3) |
| | CH93(18)-18 | 3.27 TCID50/mL | 100% (3/3) |
$^{a}$ Influenza A false positive result
**Table 8.** SARS-CoV-2 Analytical Reactivity Wet-Testing Results
| Strain | Concentration | % Positive |
| --- | --- | --- |
| Japan/TY7-503/2021 | 0.72 TCID50/mL | 100% (3/3) |
| USA/CA-Stanford-139_S23/2023 | 1500 copies/mL | 100% (3/3) |
| USA/MDHP20874/2021 | 1500 copies/mL | 100% (3/3) |
| USA/MD-HP47946/2023 | 0.72 TCID50/mL | 100% (3/3) |
| USA/MD-HP49152/2023 | 0.72 TCID50/mL | 100% (3/3) |
| USA/NY-Wadsworth-21006055-01/2021 | 0.72 TCID50/mL | 100% (3/3) |
| USA/NY-Wadsworth-23068107-01/2023 | 0.72 TCID50/mL | 100% (3/3) |
| USA/PHC658/2021 | 0.72 TCID50/mL | 100% (3/3) |
*In Silico Analysis:* The inclusivity of the Fast PCR Mini Respiratory Panel was evaluated using *in silico* analysis of the forward primers, reverse primers, and probes for SARS-CoV-2 against sequences from GISAID (sequences from November 2021 onward) and the NCBI Virus database (sequences from June 2024 onward). The analysis of the three SARS-CoV-2 targets (N2, ORF1ab and ORF8) was conducted with over 1.3 million SARS-CoV-2 genomic RNA sequences. Results of the *in-silico* analysis predict that the Fast PCR Mini Respiratory Panel will detect greater than 99.99% of SARS-CoV-2 variants.
K252932 - Page 14 of 26
{14}
# b. Cross-reactivity / Microbial Interference:
Cross-reactivity (exclusivity) and microbial interference of the Fast PCR Mini Respiratory Panel was evaluated against a panel of 40 non-targeted microorganisms to determine the assay's ability to detect influenza A, influenza B, RSV, or SARS-CoV-2 in the presence of related pathogens, high prevalence disease agents, and normal or pathogenic microorganisms reasonably likely to be present in a clinical ANS sample.
For cross-reactivity studies, each non-targeted microorganism was spiked independently into negative anterior nasal swab matrix without the presence of on-panel targets and tested in triplicate. Exclusivity was determined if 0/3 replicates returned a detected result for each of the four (4) viral targets. None of the evaluated organisms demonstrated cross-reactivity with the Fast PCR MRP at the tested concentrations listed in **Table 9**.
For microbial interference studies, each non-targeted microorganism was spiked into negative anterior nasal swab matrix in the presence of on-panel targets at 3x LoD and tested in triplicate. Absence of microbial interference was determined if 3/3 replicates returned a detected result for each of the four (4) target analytes. With the exception of *Streptococcus pneumoniae*, none of the evaluated microorganisms demonstrated interference with the Fast PCR MRP at the tested concentrations listed in **Table 9**, with each returning 3/3 replicates detected for each of the four viral targets. *Streptococcus pneumoniae* at 1E6 CFU/mL was found to interfere with detection of one of three replicates of influenza B, RSV and SARS-CoV-2. Three additional replicates of *S. pneumoniae* were tested at 1E6 CFU/mL, for a total of six replicates. Three replicates of *S. pneumoniae* were also tested at concentrations of 5E5 CFU/mL and 2.5E5 CFU/mL. No interference (100% positivity) was observed upon retesting of *S. pneumoniae* at 1E6 CFU/mL or with *S. pneumoniae* at 5E5 or 2.5E5 CFU/mL. The overall positivity rate was 83.3% (5/6) for influenza B, RSV, and SARS-CoV-2 and 100% (6/6) for influenza A with *S. pneumoniae* at 1E6 CFU/mL.
Across both cross-reactivity and microbial interference studies, 23 negative external control samples and 23 positive control samples were run alongside contrived test samples. External controls yielded expected results.
**Table 9.** Non-targeted Microorganisms Evaluated in the Cross-reactivity and Microbial Interference Wet-Testing Studies
| Organism | Concentration |
| --- | --- |
| Adenovirus Type 1 | 1E5 TCID50/mL |
| Adenovirus Type 7 | 1E5 TCID50/mL |
| Cytomegalovirus | 4.7E3 TCID50/mL |
| Enterovirus D68 | 1.67E4 TCID50/mL |
| Epstein-Barr virus | 1E5 copies/mL |
| Human coronavirus 229E | 1.39E4 TCID50/mL |
| Human coronavirus NL63 | 1.18E4 TCID50/mL |
| Human coronavirus OC43 | 1E5 TCID50/mL |
| Human Metapneumovirus (hMPV) | 6.27E3 TCID50/mL |
| Influenza A/Victoria/4897/2022 | 1E5 copies/mL |
K252932 - Page 15 of 26
{15}
| Organism | Concentration |
| --- | --- |
| Influenza B/Colorado06/2017 | 1E5 copies/mL |
| Measles virus | 1.68E4 TCID50/mL |
| MERS-CoV | 1E5 genome equivalents/mL |
| Mumps virus | 1E5 TCID50/mL |
| Parainfluenza virus 1 | 1E5 TCID50/mL |
| Parainfluenza virus 2 | 1E5 TCID50/mL |
| Parainfluenza virus 3 | 1E5 TCID50/mL |
| Parainfluenza virus 4 | 1E5 TCID50/mL |
| Rhinovirus 1A | 1.39E4 TCID50/mL |
| RSV A | 1E5 copies/mL |
| RSV B | 1E5 copies/mL |
| SARS-CoV-2 USA/WA1/2020 | 1E5 copies/mL |
| *Bordetella parapertussis* | 1E6 CFU/mL |
| *Bordetella pertussis* | 1E6 CFU/mL |
| *Corynebacterium diphtheriae* | 1E6 CFU/mL |
| *Escherichia coli* | 1E6 CFU/mL |
| *Fusobacterium necrophorum* | 1E6 CFU/mL |
| *Haemophilus influenzae* | 1E6 CFU/mL |
| *Lactobacillus plantarum* | 1E6 CFU/mL |
| *Legionella pneumophila* | 1E6 CFU/mL |
| *Moraxella catarrhalis* | 1E6 CFU/mL |
| *Mycobacterium tuberculosis* | 1E6 CFU/mL |
| *Mycoplasma genitalium* | 1E6 CFU/mL |
| *Mycoplasma pneumoniae* | 1E6 CFU/mL |
| *Neisseria meningitidis* | 1E6 CFU/mL |
| *Neisseria elongata* | 1E6 CFU/mL |
| *Pseudomonas aeruginosa* | 1E6 CFU/mL |
| *Staphylococcus aureus* (coagulase negative) | 1E6 CFU/mL |
| *Staphylococcus aureus* (Protein A producer) | 1E6 CFU/mL |
| *Staphylococcus epidermidis* | 1E6 CFU/mL |
| *Streptococcus pneumoniae*^{a} | 1E6 CFU/mL |
| *Streptococcus pyogenes* | 1E6 CFU/mL |
| *Streptococcus salivarius* | 1E6 CFU/mL |
| *Aspergillus fumigatus* | 1E6 CFU/mL |
| *Candida albicans* | 1E6 CFU/mL |
| **Pneumocystis jirovecii-S. cerevisiae recombinant* | 1E6 CFU/mL |
* This organism is a recombinant and not a valid source of Pneumocystis jirovecii. In silico analysis further supports lack of cross-reactivity against this organism.
† Cross-reactivity testing for one negative sample yielded a false positive result.
a Microbial interference testing for Streptococcus pneumoniae yielded a false negative result for influenza B, RSV, and SARS-CoV-2.
K252932 - Page 16 of 26
{16}
# c. Competitive Interference:
Competitive interference of the Fast PCR Mini Respiratory Panel caused by co-infections of on-panel targets were evaluated by testing a quadspike sample of influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020). Each quadspike sample included one on-panel target spiked at a high concentration of 1E5 copies/mL while the remaining three on-panel targets were each spiked at 3x LoD. Each combination was tested in triplicate. No competitive interference was observed for the potential co-infections evaluated at the concentrations tested.
**Table 10.** Results for Competitive Interference Study
| Test Virus at 3X LoD | Interferent Virus (concentration) | Positive Calls |
| --- | --- | --- |
| Influenza A | Influenza A/Victoria/4897/2022 (1E5 copies/mL) | 3/3 |
| Influenza B | | 3/3 |
| RSV | | 3/3 |
| SARS-CoV-2 | | 3/3 |
| Influenza A | Influenza B/Colorado/06/2017 (1E5 copies/mL) | 3/3 |
| Influenza B | | 3/3 |
| RSV | | 3/3 |
| SARS-CoV-2 | | 3/3 |
| Influenza A | RSV A/2006 Isolate (1E5 copies/mL) | 3/3 |
| Influenza B | | 3/3 |
| RSV | | 3/3 |
| SARS-CoV-2 | | 3/3 |
| Influenza A | RSV B/Washington (1E5 copies/mL) | 3/3 |
| Influenza B | | 3/3 |
| RSV | | 3/3 |
| SARS-CoV-2 | | 3/3 |
| Influenza A | SARS-CoV-2/USA/WA1/2020 (1E5 copies/mL) | 3/3 |
| Influenza B | | 3/3 |
| RSV | | 3/3 |
| SARS-CoV-2 | | 3/3 |
# d. Interfering Substances:
The performance of the Fast PCR MRP was evaluated in the presence of endogenous and exogenous substances that may be commonly found in ANS specimens. Potentially interfering endogenous and exogenous substances (**Table 11**) were tested at or above clinically relevant levels in pooled negative nasal matrix in the presence and absence of test target analytes. Positive samples were prepared by co-spiking the test target analytes, influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020) each at 3x LoD in pooled negative nasal matrix containing an individual exogenous or endogenous substance. Negative samples consisted of
K252932 - Page 17 of 26
{17}
pooled negative nasal matrix containing an individual exogenous or endogenous substance. Each sample was tested in triplicate.
For each substance tested, interference was not observed with the following exceptions: whole blood was found to cause interference with the detection of influenza A at 4% (v/v) but not at 3% (v/v) and nasal spray containing oxymetazoline was found to cause interference with the detection of SARS-CoV-2 at a concentration of 15% (v/v), but not at 12% (v/v). The FluMist nasal vaccine was not tested as cross-reactivity with the influenza test targets is expected.
Qualitative results for test samples containing potentially interfering substances are shown in Table 11.
Table 11. Results for Potentially Interfering Substances
| Potential Interferent | Active Ingredient | Concentration | Percent Positivity | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Influenza A | Influenza B | RSV | SARS-CoV-2 | RNase P |
| Negative Sample | N/A | N/A | 0.0% (0/9) | 0.0% (0/9) | 0.0% (0/9) | 0.0% (0/9) | 100.0% (9/9) |
| Baseline (3x LoD) | N/A | N/A | 100.0% (9/9) | 100.0% (9/9) | 100.0% (9/9) | 100.0% (9/9) | 100.0% (9/9) |
| Chloraseptic | Menthol, Benzocaine | 1.5 mg product/mL | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Nasal Spray | Fluticasone Propionate | 5% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Homeopathic Nasal Wash (Alkalol) | Eucalyptol, menthol, thymol, camphor, benzoin; oils of wintergreen, spearmint, pine and cinnamon | 10% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Human Whole Blood (EDTA collection tube) | N/A | 4% (v/v) | **33.3% (1/3) *** | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 3% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 2% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 1% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Mucin from bovine submaxillary glands | Purified mucin protein | 0.5% (w/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Mupirocin | Mupirocin | 10 mg product/mL | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Nasal Drops | Phenylephrine | 15% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Nasal Spray | Cromolyn | 15% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
K252932 - Page 18 of 26
{18}
| Potential Interferent | Active Ingredient | Concentration | Percent Positivity | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Influenza A | Influenza B | RSV | SARS-CoV-2 | RNase P |
| Nasal Spray | Oxymetazoline | 15% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 66.7% (2/3)^{b} | 100.0% (3/3) |
| | | 12% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 9% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 6% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| | | 3% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Naso GEL (NeilMed) | Sodium hyaluronate, Aloe vera | 5% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Sore Throat Phenol Spray | Phenol | 15% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Tamiflu | Oseltamivir Phosphate | 5 mg/mL | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Tobramycin | Tobramycin | 4 µg/mL | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Zanamivir | Zanamivir | 282 ng/mL | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
| Zicam | Galphimia glauca, Luffa opperculata, Sabadilla | 5% (v/v) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) | 100.0% (3/3) |
$^{a}$ Two false negative results for influenza A
$^{b}$ One false negative result for SARS-CoV-2
#### 4. Assay Reportable Range:
Not applicable, this is a qualitative assay.
#### 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
##### a. Controls – External Control Evaluation:
See Section **IV.C.5. Quality Control**.
##### b. Specimen Stability:
The stability of the viral targets influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020) at 3x LoD in anterior nasal swab matrix collected in AMDI Sample Buffer and stored at either 2-8°C or 30°C was established for the Fast PCR Mini Respiratory Panel.
Results from these studies support the following:
- Anterior nasal swab specimens collected in AMDI Sample Buffer may be stored for up to 72 hours at 2-8°C or for up to four (4) hours at 15-30°C.
K252932 - Page 19 of 26
{19}
# c. Fresh vs. Frozen - Specimen Stability:
The stability of the viral targets influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020) at either 2x or 5x LoD in anterior nasal swab matrix collected in AMDI Sample Buffer and stored frozen ($\leq -15^\circ\text{C}$ or $\leq -70^\circ\text{C}$) was established following one and two freeze/thaw cycles and tested using the Fast PCR Mini Respiratory Panel.
Results from these studies support the following:
- Anterior nasal swab specimens collected in AMDI Sample Buffer may undergo one (1) freeze/thaw cycle.
# d. Kit Shelf-life Stability:
Real-time kit stability data for the Fast PCR MRP supports the recommended shelf-life stability claim of ambient room temperature storage (15-30°C / 59-86°F) for up to six (6) months.
# e. AMDI Fast PCR MRP External Control Kit Shelf-life Stability:
Real-time kit stability data for the AMDI Fast PCR MRP External Control Kit supports the recommended shelf-life stability claim of ambient room temperature storage (15-30°C / 59-86°F) for up to nine (9) months.
# 6. Detection Limit:
# a. Individual Analyte Detection Limits:
Limit of detection (LoD) studies for the Fast PCR MRP were conducted to determine the lowest detectable concentration of influenza A, influenza B, RSV, and SARS-CoV-2 at which $\geq 95\%$ of all replicates test positive. Viral strains tested are shown in **Table 12**. Testing used pooled negative nasal matrix. The preliminary LoD was established as the minimum concentration yielding 3/3 detected results in a limited dilution series for each target viral strain. The confirmatory LoD was determined by testing at the preliminary LoD plus at least one higher and one lower dilution with 20 replicates. Confirmed LoD values are shown in **Table 12**.
**Table 12.** Confirmed LoD for Target Analytes
| Virus | Strain | LoD (copies/mL) |
| --- | --- | --- |
| Influenza A H1N1 | A/Victoria/4897/2022 | 83.3 |
| Influenza A H3N2 | A/Darwin/9/2021 | 250 |
| Influenza B (Victoria) | B/Colorado/06/2017 | 500 |
| Influenza B (Yamagata) | B/Phuket/3073/2013 | 250 |
| RSV A | 2006 isolate | 250 |
K252932 - Page 20 of 26
{20}
| RSV B | Washington | 500 |
| --- | --- | --- |
| SARS CoV-2 | USA/WA1/2020 | 500 |
As part of the influenza A/Victoria/4897/2022 confirmatory LoD testing, one replicate at 41.7 copies/mL resulted in false positive for SARS-CoV-2. As part of the influenza B/Colorado/06/2017 confirmatory LoD testing, two replicates at 500 copies/mL and one replicate at 150 copies/mL resulted in false positive for influenza A.
# b. *Co-spiked Analyte Detection Limit:*
A Co-Spiked LoD study for the Fast PCR MRP was conducted to determine the lowest detectable concentration of influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020) in a co-spiked sample at which $\geq 95\%$ of all replicates test positive. Testing used pooled negative nasal matrix. The preliminary LoD concentrations used were the LoD concentrations determined for individual analytes above. The confirmatory LoD was determined by testing at the preliminary LoD with 20 replicates. This study confirmed that the LoD concentrations for influenza A (A/Darwin/9/2021), influenza B (B/Colorado/06/2017), RSV (A 2006 Isolate), and SARS-CoV-2 (USA/WA1/2020) are the same whether spiked individually or co-spiked in the same contrived sample.
# 7. *Assay Cut-Off:*
The Fast PCR MRP Analysis Algorithm is responsible for converting PCR test data from the Fast PCR Instrument into test results. The cut-off for determining if an amplification curve should be classified as “Amplified” or “Not Amplified” utilizes a set of Amplification Curve Analysis Parameters (including the Ct value). The influenza A, influenza B, RSV, SARS-CoV-2, and RNase P assays each have their own Amplification Curve Analysis Parameters that were validated. For influenza A, influenza B, and RSV, if Ct $\geq 50$, the status is “Not Amplified”, else it is “Amplified”. For SARS-CoV-2, if Ct $\geq 42.2$, the status is “Not Amplified”, else it is “Amplified”. For RNase P, if Ct $\geq 40.8$, the status is “Not Amplified”, else it is “Amplified”.
# 8. *Accuracy (Instrument):*
Not Applicable.
# 9. *Carry-Over:*
A study was conducted to assess potential carryover or cross-contamination in the single-use, self-contained Fast PCR Mini Respiratory Panel Test Disc by testing high positive and no template samples in an alternating fashion on the same Fast PCR Instrument. The negative sample used in this study was negative anterior nasal swab matrix screened to be negative for the four viral targets in the Fast PCR Mini Respiratory Panel. The high positive quadspike sample consisted of influenza A (influenza A/Darwin/9/2021 at 2.0E7 copies/mL), influenza B (influenza B/Colorado/06/2017 at 2.9E8 copies/mL), RSV (RSV A 2006 isolate at 4.8E7 copies/mL), and SARS-CoV-2 (USA/WA1/2020 at 2.9E7 copies/mL) in negative anterior nasal swab matrix. The study consisted of five (5) total runs on three (3) Fast PCR Instruments, with each run consisting of eight (8) replicates of alternating high positive quadspike and negative samples run on the same Fast PCR Instrument. A total of 40 high
K252932 - Page 21 of 26
{21}
positive quadspike samples and 40 negative samples were tested. All 40 high positive quadspike samples correctly reported as positive for all four targets. All negative samples correctly reported as negative for all four targets. No cross-contamination was observed.
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
Not Applicable.
### 2. Matrix Comparison:
The Fast PCR Mini Respiratory Panel is to be used with the AMDI Sample Buffer only. No additional collection devices have been evaluated for use with the device.
## C Clinical Studies:
### 1. Prospective Study:
The clinical performance of the Fast PCR Mini Respiratory Panel assay was evaluated using clinical specimens prospectively collected between December 2024 and May 2025 from nine (9) geographically diverse clinical sites located within the United States with 27 untrained test operators. The clinical study utilized prospective specimens collected from pediatric and adult patients exhibiting clinical signs and symptoms of respiratory tract infection. All collection and assay testing were performed at CLIA-waived facilities, representative of the intended use environment, by untrained operators according to the test QRG. The results of all four viral targets were compared to results from an FDA-cleared, CLIA waived RT-PCR assay (comparative reference method).
Initial enrollment included a total of 1947 unique prospectively collected specimens that met the pre-determined inclusion criteria. Of these, a total of 159 (8.2%) specimens were excluded. 33 (1.7%) specimens were excluded due to an invalid result produced by the Fast PCR MRP and 126 (6.5%) specimens were excluded due to protocol deviations. **Table 13** provides a summary of the demographic information for the remaining 1788 subjects included in the data analysis. Of the 1788 samples included in the data analysis, 34.6% (619/1788) were tested fresh while 65.4% (1169/1788) were tested frozen. Clinical performance of the fast PCR MRP with prospectively collected samples is presented in **Table 14**.
**Table 13.** Subject Demographics - Prospective Symptomatic Population
| Characteristics | Symptomatic Subjects |
| --- | --- |
| Total, N | 1788 |
| Age Group (Years), n (%) | |
| ≤ 5 yr | 94 (5.3%) |
K252932 - Page 22 of 26
{22}
| 6-18 yr | 311 (17.4%) |
| --- | --- |
| 19-40 yr | 664 (37.1%) |
| 41-60 yr | 496 (27.7%) |
| ≥ 61 yr | 223 (12.5%) |
| **Sex, n (%)** | |
| Male | 683 (38.2%) |
| Female | 1104 (61.7%) |
| Unknown | 1 (0.1%) |
| **Ethnicity, n (%)** | |
| Hispanic / Latino | 339 (19.0%) |
| Not Hispanic / Not Latino | 1,384 (77.4%) |
| Note Reported | 58 (3.2%) |
| Unknown | 7 (0.4%) |
| **Race, n (%)** | |
| American Indian / Alaskan Native | 4 (0.2%) |
| Asian | 81 (4.5%) |
| Black / African-American | 283 (15.8%) |
| White | 1301 (72.8%) |
| Other or Mixed Race | 70 (3.9%) |
| Unknown | 20 (1.1%) |
| Not Reported | 29 (1.6%) |
**Table 14.** Fast PCR MRP Prospective Clinical Performance
| Analyte | Sample Category | N | TP | FP | FN | TN | PPA | | NPA | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | % | 95% CI | % | 95% CI |
| **Influenza A** | Prospective Fresh | 619 | 25 | 7 | 1 | 586 | 96.2 | 81.1 – 99.3 | 98.8 | 97.6 – 99.4 |
| | Prospective Frozen | 1169 | 211 | 38 | 5 | 915 | 97.7 | 94.7 – 99.0 | 96.0 | 94.6 – 97.1 |
| | Overall | 1788 | 236 | 45 | 6 | 1501 | 97.5 | 94.7 – 98.9 | 97.1 | 96.1 – 97.8 |
| **Influenza B** | Prospective Fresh | 619 | 52 | 2 | 0 | 565 | 100.0 | 93.1 – 100.0 | 99.6 | 98.7 – 99.9 |
| | Prospective Frozen | 1169 | 25 | 3 | 0 | 1,141 | 100.0 | 86.7 – 100.0 | 99.7 | 99.2 – 99.9 |
| | Overall | 1788 | 77 | 5 | 0 | 1706 | 100.0 | 95.2 – 100.0 | 99.7 | 99.3 – 99.9 |
| **RSV** | Prospective Fresh | 619 | 9 | 4 | 0 | 606 | 100.0 | 70.1 – 100.0 | 99.3 | 98.3 – 99.7 |
| | Prospective Frozen | 1169 | 24 | 5 | 3 | 1,137 | 88.9 | 71.9 – 96.1 | 99.6 | 99.0 – 99.8 |
| | Overall | 1788 | 33 | 9 | 3 | 1743 | 91.7 | 78.2 – 97.1 | 99.5 | 99.0 – 99.7 |
K252932 - Page 23 of 26
{23}
| Analyte | Sample Category | N | TP | FP | FN | TN | PPA | | NPA | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | % | 95% CI | % | 95% CI |
| **SARS-CoV-2** | Prospective Fresh | 619 | 18 | 4 | 0 | 597 | 100.0 | 82.4 – 100.0 | 99.3 | 98.3 – 99.7 |
| | Prospective Frozen | 1169 | 67 | 7 | 2 | 1093 | 97.1 | 90.0 – 99.2 | 99.4 | 98.7 – 99.7 |
| | Overall | 1788 | 85 | 11 | 2 | 1690 | 97.7 | 92.0 – 99.4 | 99.4 | 98.8 – 99.6 |
PPA = Positive Percent Agreement, NPA = Negative Percent Agreement
TP=true positive; FP=false positive; TN=true negative; FN=false negative
Five specimens tested positive for co-infections by the Fast PCR MRP that were also identified as the same co-infections by the comparator. Eight specimens tested positive for co-infections by the Fast PCR MRP that were identified as single infections by the comparator. One specimen tested positive for co-infection by the Fast PCR MRP that was identified as negative by the comparator.
### 2. Retrospective/Archived Study:
To supplement the prospective clinical study results for RSV, retrospective frozen clinical ANS specimens collected from individuals with signs and symptoms of influenza infection during the 2024-2025 North American respiratory season were evaluated. Paired ANS swabs were collected from each enrolled subject, with one swab placed in the AMDI Sample Buffer, the other swab placed in a transport media for comparator testing. Both samples were immediately frozen at -80°C and remained at -80°C until thawed and tested at the clinical sites (integrated into the daily testing workflow) or at the comparator testing site. Of the 132 enrolled archived specimens, 14 were excluded (one due to an invalid Fast PCR MRP result and 13 due to protocol deviations) for a total of 118 archived specimens. Clinical performance for archived samples is shown in Table 15.
Table 15. Fast PCR MRP Retrospective Clinical Performance
| Analyte | N | TP | FP | FN | TN | PPA | | NPA | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | % | 95% CI | % | 95% CI |
| **RSV** | 118 | 33 | 2 | 2 | 83 | 94.3 | 81.4 – 98.4 | 100 | 95.6 – 100.0 |
PPA = Positive Percent Agreement, NPA = Negative Percent Agreement
TP=true positive; FP=false positive; TN=true negative; FN=false negative
One specimen tested positive for co-infection with RSV and influenza A by the Fast PCR MRP, which was only positive for RSV by the comparator.
### 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
### D Clinical Cut-Off:
Not Applicable.
K252932 - Page 24 of 26
{24}
## E Expected Values/Reference Range:
The positivity rate for SARS-CoV-2, influenza A and influenza B, as determined by the Fast PCR MRP, are shown below, stratified by the study site (Table 16), by age group (Table 17), and by sex (Table 18).
**Table 16.** Expected Results – Positive Results and Prevalence by Site
| Clinical Site ID | Site Location (City, State) | SARS-Cov-2 | | | Influenza A | | | Influenza B | | | RSV | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Total No. | No. Pos. | Exp. Value | Total No. | No. Pos. | Exp. Value | Total No. | No. Pos. | Exp. Value | Total No. | No. Pos. | Exp. Value |
| **Overall** | | **1788** | **96** | **5.4%** | **1788** | **281** | **15.7%** | **1788** | **82** | **4.6%** | **1788** | **42** | **2.3%** |
| 1 | Laredo, TX | 214 | 5 | 2.3% | 214 | 11 | 5.1% | 214 | 2 | 0.9% | 214 | 2 | 0.9% |
| 2 | Bellevue, WA | 89 | 0 | 0.0% | 89 | 3 | 3.4% | 89 | 5 | 5.6% | 89 | 2 | 2.2% |
| 3 | Birmingham, AL | 211 | 23 | 10.9% | 211 | 47 | 22.3% | 211 | 3 | 1.4% | 211 | 5 | 2.4% |
| 4 | Forney, TX | 203 | 17 | 8.4% | 203 | 54 | 26.6% | 203 | 1 | 0.5% | 203 | 3 | 1.5% |
| 5 | Riverdale, GA | 165 | 5 | 3.0% | 165 | 10 | 6.1% | 165 | 0 | 0.0% | 165 | 4 | 2.4% |
| 6 | Kansas City, KS | 164 | 6 | 3.7% | 164 | 9 | 5.5% | 164 | 0 | 0.0% | 164 | 2 | 1.2% |
| 7 | Redmond, WA | 256 | 6 | 1.9% | 256 | 33 | 12.9% | 256 | 16 | 6.3% | 256 | 3 | 1.2% |
| 8 | Smithfield, RI | 413 | 28 | 6.8% | 413 | 86 | 20.8% | 413 | 50 | 12.1% | 413 | 20 | 4.8% |
| 9 | Chicago, IL | 73 | 6 | 8.2% | 73 | 28 | 38.4% | 73 | 5 | 6.8% | 73 | 1 | 1.4% |
**Table 17.** Expected Results – Positive Results and Prevalence by Age Group
| Target | ≤ 5 yr | 6-18 yr | 19-40 yr | 41-60 yr | ≥ 61 yr | Total |
| --- | --- | --- | --- | --- | --- | --- |
| Influenza A | 9 (9.6%) | 45 (14.5%) | 127 (19.1%) | 67 (13.5%) | 33 (14.8%) | 281 (15.7%) |
| Influenza B | 3 (3.2%) | 34 (10.9%) | 28 (4.2%) | 17 (3.4%) | 0 (0.0%) | 82 (4.6%) |
| RSV | 2 (2.1%) | 9 (2.9%) | 8 (1.2%) | 13 (2.6%) | 10 (4.5%) | 42 (2.3%) |
| SARS-CoV-2 | 3 (3.2%) | 9 (2.9%) | 28 (4.2%) | 40 (8.1%) | 16 (7.2%) | 96 (5.4%) |
**Table 18.** Expected Results – Positive Results and Prevalence by Sex
| Target | Male | Female | Unknown | Total |
| --- | --- | --- | --- | --- |
| Influenza A | 102 (14.9%) | 178 (16.1%) | 1 (100%) | 281 (15.7%) |
| Influenza B | 34 (5.0%) | 48 (4.3%) | 0 (0.0%) | 82 (4.6%) |
| RSV | 17 (2.5%) | 25 (2.3%) | 0 (0.0%) | 42 (2.3%) |
| SARS-CoV-2 | 37 (5.4%) | 59 (5.3%) | 0 (0.0%) | 96 (5.4%) |
## F Other Supportive Instrument Performance Characteristics Data:
Not Applicable.
K252932 - Page 25 of 26
{25}
# **VIII Proposed Labeling:**
The labeling supports the finding of substantial equivalence for this device.
# **IX Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K252932 - Page 26 of 26
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.