Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging Processor (Model HV-200)
Applicant
Anhui Happiness Workshop Instruments Co., Ltd.
Product Code
FGB · Gastroenterology, Urology
Decision Date
Apr 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Single Use Digital Flexible Ureteroscope (7.5F) is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract. The Videoscope Imaging Processor, when used with the endoscope, provides imaging for the examination, diagnosis, or treatment of urinary tract conditions.
Device Story
System comprises single-use flexible digital ureteroscope and reusable imaging processor (HV-200). Ureteroscope features CMOS sensor and LED lighting; connects to processor for power, control, and video signal processing. Physician operates ureteroscope manually to access urinary tract; processor outputs real-time video to external monitor via HDMI/DVI/SDI. Processor controls include photo/video capture, zoom, LED brightness, and white balance. Used in clinical settings for visualization, diagnosis, and treatment of urinary tract conditions. Enables minimally invasive procedures; provides high-resolution imaging for clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility (ISO 10993), electrical/EMC safety (IEC 60601 series), mechanical/optical performance (deflection, flow rate, resolution, field of view, distortion), photobiological safety (IEC 62471), and software validation.
Technological Characteristics
Flexible ureteroscope (7.5Fr) with CMOS sensor and LED illumination. Reusable imaging processor (HV-200) provides power and video processing. Connectivity: HDMI, DVI, SDI. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Biocompatibility per ISO 10993-1. Luer taper per ISO 80369-7. Shelf life 3 years per ASTM F1980-21.
Indications for Use
Indicated for adult patients requiring visualization of the urinary tract (urethra, bladder, ureter, calyces, renal papillae) and performance of diagnostic or therapeutic procedures via transurethral or percutaneous access.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Pusen Single Use Flexible Video Ureteroscope (Model: PU3033H, PU3033AH) and Pusen Single Use Suction Access Ureteroscope (Model: PU400A, PU411A) (K233778)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 9, 2026
Anhui Happiness Workshop Instruments Co., Ltd.
% Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lujiazui Rd., Pudong
Shanghai, 200120
CHINA
Re: K252929
Trade/Device Name: Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging Processor (Model HV-200)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FGB, FET
Dated: September 15, 2025
Received: September 15, 2025
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252929 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252929 - Boyle Wang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark R. Kreitz -S
for Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252929 | ? |
| Please provide the device trade name(s). | | ? |
| Single Use Digital Flexible Ureteroscope (7.5F);
Videoscope Imaging Processor (Model HV-200) | | |
| Please provide your Indications for Use below. | | ? |
| Single Use Digital Flexible Ureteroscope (7.5F) is intended to visualize organs, cavities and canals in the
urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access
routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and
therapeutic procedures in the urinary tract.
The Videoscope Imaging Processor, when used with the endoscope, provides imaging for the examination,
diagnosis, or treatment of urinary tract conditions. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252929
Page 1 of 7
510(k) Summary
# 510(k) Summary
# K252929
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Anhui Happiness Workshop Instruments Co., Ltd.
Address: 1/F, 2/F, 3/F, Building 2, Wisdom Industrial Park, 1300 Jiangqiao Road, Bengshan District, 233000 Bengbu, Anhui Province, China
Tel: +86-18895690140
Contact: Ms. Jamie Zhang
Registration Number: N/A
## Designated Submission Correspondent
Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China
Tel: +86-21-50313932
Email: Info@truthful.com.cn
Date of Preparation: April.7th,2026
## 2.0 Device Information
Trade name: Single Use Digital Flexible Ureteroscope (7.5F)
Videoscope Imaging Processor (Model HV-200)
Common name: Endoscope and accessories
Classification name: Ureteroscope And Accessories, Flexible/Rigid
Production code: FGB, FET
Regulation number: 21 CFR 876.1500
Classification: Class II
Panel: Gastroenterology/Urology
## 3.0 Predicate Device Information
### Predicate#
{5}
K252929
Page 2 of 7
510(k) Summary
Manufacturer: Zhuhai Pusen Medical Technology Co, Ltd.
Trade name: Pusen Single Use Flexible Video Ureteroscope (Model: PU3033H, PU3033AH) and Pusen Single Use Suction Access Ureteroscope (Model: PU400A, PU411A)
510(k) number: K233778
Product code: FGB
This predicate device has not been subject to a design-related recall.
## 4.0 Indication for Use Statement
Single Use Digital Flexible Ureteroscope (7.5F) is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Videoscope Imaging Processor, when used with the endoscope, provides imaging for the examination, diagnosis, or treatment of urinary tract conditions.
## 5.0 Device Description
The Digital Endoscope System is consists of the Single Use Digital Flexible Ureteroscope (7.5F) and a Videoscope Imaging Processor (HV-200). The Single Use Digital Flexible Ureteroscope (7.5F) is powered and controlled through connection to the Videoscope Imaging Processor (HV-200), which provides illumination, processes the video signal, and outputs images to an external monitor. Digital Endoscope System is intended to be used by physicians to access, visualize, and perform procedures in the urinary tract and the kidney.
The Single Use Digital Flexible Ureteroscope (7.5F) (Model HUS-T101R, HUS-T102R, HUS-T101S, HUS-T102S) and Videoscope Imaging Processor (Model HV-200) have been designed to be used with each other for physicians to access, visualize, and perform procedures in the urinary tract for diagnosis and treatment.
The Single Use Digital Flexible Ureteroscope (7.5F) is a sterile, single-use, flexible, digital video ureteroscope. It is comprised of a control body with articulation controls and accessory access ports, and a flexible insertion tube with an on-tip camera module (CMOS) and LED lighting source. The diameter of the Insertion portion is 7.5Fr, which is thinner than conventional endoscopes.
Single Use Digital Flexible Ureteroscope (7.5F) is the accessory and detachable part of Videoscope Imaging Processor. Single Use Digital Flexible Ureteroscope (7.5F) has 4 models, HUS-T101R, HUS-T102R, HUS-T101S and HUS-T102S, and can be operated by either left or right hand. The model information is shown as Table 1
Page 2 of 7
{6}
K252929
Page 3 of 7
510(k) Summary
Product Model:
Table 1 Product Models
| Product Model | Description | Allowance |
| --- | --- | --- |
| HUS-T101R | Working length 670mm
with reverse deflection | +10mm |
| HUS-T102R | Working length 660mm
with reverse deflection | +10mm |
| HUS-T101S | Working length 670mm
with standard deflection | +10mm |
| HUS-T102S | Working length 660mm
with standard deflection | +10mm |
| Note: ‘S’ means standard deflection; ‘R’ means reverse deflection. | | |

Figure 1 Standard Deflection

Figure 2 Reverse Deflection
Page 3 of 7
{7}
K252929
Page 4 of 7
510(k) Summary
The Videoscope Imaging Processor is composed of processor, power cord and HDMI video cables. The Processor processes the signal from the endoscope and outputs video signals to a monitor.
The processor provides the necessary power and control to the ureteroscope's LED light source and CMOS image sensor, processes the captured video signals, and outputs real-time images to external monitors via HDMI, DVI, or SDI connections. Front panel controls allow power switching, menu navigation, photo capture (JPG), video recording (AVI), image freezing, zoom, LED brightness adjustment, and white balance calibration.
The HV-200 Videoscope Imaging Processor is a reusable device and do not require sterilization before use.
Together, the reusable Videoscope Imaging Processor and the Single Use Digital Flexible Ureteroscope (7.5F) together complete the Digital Endoscope System enabling a Flexible Ureteroscopy procedure to take place.
# 6.0 Technological Characteristic Comparison Table
Table1-General Comparison
| Item | Subject Device K252929 | Predicate Device K233778 |
| --- | --- | --- |
| Product Name | Single Use Digital Flexible Ureteroscope (7.5F)
Videoscope Imaging Processor (Model HV-200) | Single Use Flexible Video Ureteroscope
Single Use Suction Access Ureteroscope |
| Product Code | FGB, FET | FGB |
| Regulation No. | 21 CFR 876.1500 | 21 CFR 876.1500 |
| Class | Class II | Class II |
| Model | Single Use Digital Flexible Ureteroscope (7.5F):
HUS-T101R; HUS-T102R; HUS-T101S;
HUS-T102S
Videoscope Imaging Processor: HV-200 | PU400A; PU411A
PU3033H; PU3033AH |
| Intended Use/Indication for Use | Single Use Digital Flexible Ureteroscope (7.5F) is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Videoscope Imaging Processor, when used with the endoscope, provides imaging for the examination, | This instrument has been designed to be used with Endo-Therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney. This instrument is also used as a suction catheter that establishes a conduit used for irrigation and aspiration of kidney stones and stone dust during ureteral lithotripsy |
Page 4 of 7
{8}
K252929
Page 5 of 7
510(k) Summary
| | diagnosis, or treatment of urinary tract conditions. | |
| --- | --- | --- |
| Target population | Adults | Adults |
| **Single Use Digital Flexible Ureteroscope** | | |
| Scope type | Flexible | Flexible |
| Scope reusability | Single-use | Single-use |
| Energy used | Powered by line power | Powered by chargeable battery or line power. |
| Digital video technology | CMOS | CMOS |
| Illumination source | LED | LED |
| Field of view | 110° | 120° |
| Direction of view | 0° | 0° |
| Depth of field | 2-50 mm | 3~50 mm |
| Maximum insertion portion width | 2.85mm | PU3033AH, PU3033H: 2.7 mm
PU400A, PU411A: 3.2 |
| Working length | HUS-T101R, HUS-T101S: 670mm
HUS-T102R,HUS-T102S: 660mm | PU3033AH, PU3033H: 650 mm;
PU400A: 650 mm;
PU411A: 680 mm |
| Working channel size | ≥3.6Fr (1.2mm) | PU3033AH, PU3033H: ≥1.2 mm;
PU400A, PU411A: ≥5.1 Fr |
| Up/down deflection | Up:285°
Down:285° | Up: 270°
Down: 270° |
| Irrigation | Provided | Provided |
| Suction | Not Provided | Provided |
| Sterility | Ethylene Oxide (EO)
SAL: 10^{-6} | Ethylene Oxide (EO)
SAL: 10^{-6} |
| Image system | Videoscope Imaging Processor
(Model HV-200) | Processor PV300(Cleared under K222602) |
## 7.0 Non-clinical Testing Summary
The following performance data have been conducted to verify that the subject device meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the targeted device complies with the following standards:
## Conclusions for Biocompatibility Testing
The biocompatibility evaluation for the Single Use Digital Flexible Ureteroscope (7.5F) was conducted in accordance with the FDA's Biocompatibility Guidance "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process'" and FDA recognize international standard ISO 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process". The worst case of the Ureteroscope is considered breached/compromised or compromised surfaces, contacting tissue for a limited duration (< 24 hours).
And the testing included the following tests, results of which demonstrate the
Page 5 of 7
{9}
K252929
Page 6 of 7
510(k) Summary
biological safety of the subject device:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation (ISO 10993-23)
- Material-Mediated Pyrogenicity (ISO 10993-11)
- Acute Systemic Toxicity (ISO 10993-11)
## Electrical and EMC Safety:
The electrical safety and EMC safety testing was performed to, and passed, the following standards:
- IEC 60601-1:2005/AMD2:2020, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- IEC 60601-1-2:2020, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests
- IEC TS 60601-4-2:2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment.
## Summary of Bench Testing
Bench testing was conducted and the results show that the subject device complies with the below standard:
- Mechanical and Optical Performance
- Deflection performance
- Working channel performance
- Flow rate
- Tensile and torsional strength
- Field of view
- Direction of view
- Resolution
- Noise and dynamic range
- Geometric distortion
- Image intensity uniformity
- Image Quality
Comparative testing related to image quality performances including color performance (color reproduction and color contrast enhancement), optical performance (resolution, depth of field, image intensity uniformity and distortion) tests were performed for the subject device and the predicate device to support substantial equivalence.
{10}
K252929
Page 7 of 7
510(k) Summary
> Photobiological Safety
The LEDs in the subject device were tested according to the following FDA recognized standards:
- IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems.
> Luer taper
Luer taper in the subject device was tested according to ISO 80369-7: 2021, Small bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications.
## Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance: Content of Premarket Submissions for Device Software Functions.
The Software Validation is in compliance with FDA Guidance.
## Summary of Sterilization and Shelf Life
-- Sterilization Process has been validated accordance with ISO 11135:2014;
-- EO/ECH residual test was performed according to ISO 10993-7:2008
-- Environmental conditioning and simulated shipping distribution according to ASTM D4169-23
-- The shelf life is determined based on optical testing and product performance testing after accelerated aging test according to ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. The Single Use Digital Flexible Ureteroscope (7.5F) is provided sterile and Package integrity and device performance testing to support a three-year shelf-life for it.
-- Package validation was conducted according to ISO 11607-1:2019 and ISO 11607 2:2019
## 8.0 Clinical Test Summary
No clinical study is included in this submission.
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicated device.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.