← Product Code [ODC](/productcode/ODC) · K252910

# Orca Air/Water and Suction Valves (K252910)

_Boston Scientific · ODC · Oct 9, 2025 · Gastroenterology, Urology · SESK_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252910

## Device Facts

- **Applicant:** Boston Scientific
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Oct 9, 2025
- **Decision:** SESK
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure. The Orca Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.

## Device Story

Orca Air/Water and Suction Valves are single-use endoscopic channel accessories; they attach to endoscope handle ports to manually control air/water flow and suction during GI endoscopic procedures. The device is operated by a clinician; it requires no energy source. The valves function as mechanical interfaces; they allow the user to modulate fluid/gas delivery or suction pressure by manual depression. The device benefits the patient by enabling precise control of endoscopic functions during diagnostic or therapeutic procedures. The subject device expands compatibility to include both Olympus and Exalt endoscopes.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), and performance testing (insertion/removal/depression force, actuation time, flow rate, leakage, seal pressure, operational life, and backflow/back pressure).

## Technological Characteristics

Materials: Polycarbonate, ABS resin, thermoplastic elastomers. Single-use, sterile (EO, SAL 10^-6). Manual actuation; no energy source. Compatible with Olympus and Exalt endoscopes. Packaging: Tray with Tyvek lid.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Orca Air/Water and Suction Valves ([K160403](/device/K160403.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 9, 2025

Boston Scientific
Claire Paddock
Senior Regulatory Affairs Specialist
100 Boston Scientific Way
Marlborough, Massachusetts 01752

Re: K252910
Trade/Device Name: Orca Air/Water and Suction Valves
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: ODC
Dated: September 11, 2025
Received: September 12, 2025

Dear Claire Paddock:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252910 - Claire Paddock
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252910 - Claire Paddock
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

SIVAKAMI VENKATACHALAM -S

for

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

Orca Single Use Air/Water and Suction Valves
Page 10 of 38

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252910 | ?  |
|  Please provide the device trade name(s). |  | ?  |
|  Orca Air/Water and Suction Valves |  |   |
|  Please provide your Indications for Use below. |  | ?  |
|  The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure. |  |   |
|  The Orca Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure. |  |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

K252910
Page 1 of 4
Special 510(k) Premarket Notification, Orca Air/Water and Suction Valves
Boston Scientific Corporation

# 510(k) Summary

I. Submitter
Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, MA 01750
Phone: (508)683-4000
Contact Person: Claire Paddock
Date Prepared: September 11, 2025

II. Device
Trade Name: Orca Air/Water and Suction Valves
Classification Name: Endoscope Channel Accessory
Regulation Number: 876.1500
Product Code: ODC
Classification: Class II

III. Predicate Device
Orca Air/Water and Suction Valves, K160403
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.

IV. Device Description
Orca Air/Water and Suction Valves are endoscopic channel accessories that attach to the Air/Water and Suction ports on the endoscope handle to control the amount of air/water flow and suction of an endoscope.

V. Indications for Use
The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.
The Orca Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.

510(k) Summary
016_1

{5}

K252910
Page 2 of 4
Special 510(k) Premarket Notification, Orca Air/Water and Suction Valves
Boston Scientific Corporation

VI. Comparison of Technological Characteristics with the Predicate Device

|  Feature | Predicate Device
Orca Air/Water Valve,
Orca Suction Valve
(K160403) | Subject Device
Orca Air/Water Valve,
Orca Suction Valve | Comparison  |
| --- | --- | --- | --- |
|  Product Code | ODC | ODC | Identical  |
|  Regulation Number | 21 CFR 876.1500 | 21 CFR 876.1500 | Identical  |
|  Intended Use | The SmartStart Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. | The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. | Identical  |
|   |  The SmartStart Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure. | The Orca Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.  |   |
|  Sterility | Sterile | Sterile | Identical  |
|  Sterilization Method | EO | EO | Identical  |
|  SAL | 10^{-6} | 10^{-6} | Identical  |
|  Usage | Single Use | Single Use | Identical  |
|  Materials | Stainless Steel, ABS Resin, Thermoplastic Elastomers | Polycarbonate, ABS Resin, Thermoplastic Elastomers | Reason for Submission, Substantially Equivalent  |
|  Target Population | Patients undergoing an endoscopic procedure | Patients undergoing an endoscopic procedure | Identical  |
|  Energy Used / Delivered | None | None | Identical  |
|  Method of Application | Manual Actuation | Manual Actuation | Identical  |
|  Compatible Endoscopes | Olympus Endoscopes | Olympus and Exalt Endoscopes | Substantially Equivalent  |
|   |   |   |  Exalt Compatibility Established in K193202  |
|  Packaging | Supplied as a kit in Tray sealed in a Tyvek/clear polymer peel pack | Supplied as a kit in Tray with Tyvek Lid | Equivalent  |

510(k) Summary

{6}

K252910
Page 3 of 4
Special 510(k) Premarket Notification, Orca Air/Water and Suction Valves
Boston Scientific Corporation

The subject Orca Air/Water and Suction valves have the same mode of operation, intended use, target population, sterility status, and sterilization method, as the predicate devices. The subject devices differ in their material composition, labeled compatibility, and packaging. These differences were evaluated through biocompatibility, packaging, and bench testing, and were found to be equivalent between the subject and predicate devices.

## VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility Testing

The biocompatibility evaluation for the Orca device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 “Use of International Standard ISO-10993, ‘Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,’” May 1, 1995, and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The battery of testing included the following tests:

- Cytotoxicity
- Sensitization
- Irritation

The Orca Air/Water and Suction Vales are considered tissue contacting for a duration of less than 24 hours.

## Non-Clinical Bench Testing

### Air/Water Valve

- Insertion Force
- Removal Force
- Depression Force
- Actuation Time
- Flow Rate
- Air
- CO2
- Water
- Air/Water Leakage
- Seal Pressure Testing
- Operational Life
- Backflow Prevention

Suction Valve:

510(k) Summary
016_3

{7}

K252910
Page 4 of 4
Special 510(k) Premarket Notification, Orca Air/Water and Suction Valves
Boston Scientific Corporation

- Insertion Force
- Removal Force
- Depression Force
- Actuation Time
- Flow Rate
- Operational Life
- Back Pressure

VIII. Conclusions

The Orca Air/Water and Suction Valves have been demonstrated to be substantially equivalent to the predicate devices with respect to indications for use, comparison of technological characteristics with the predicate device, and performance in the specified use conditions. The information submitted within this premarket notification supports the substantially equivalent determination for these devices.

510(k) Summary
016_4

---

**Source:** [https://fda-staging.innolitics.com/device/K252910](https://fda-staging.innolitics.com/device/K252910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
