K252909 · Ningbo Runyes Medical Instrument Co., Ltd. · EHD · Feb 3, 2026 · Dental
Device Facts
Record ID
K252909
Device Name
Diagnostic X-Ray Equipment Model POCT22
Applicant
Ningbo Runyes Medical Instrument Co., Ltd.
Product Code
EHD · Dental
Decision Date
Feb 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Intended as extraoral x-ray source to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians.
Device Story
Portable dental X-ray system; generates ionizing radiation for radiographic examination of teeth, jaw, and oral structures. Device contains internal X-ray tube and timer; operator selects exposure parameters via LCD interface (patient size, tooth type) or manual up-down buttons. Used in general dentistry; assists with patients in nursing homes, special needs, or those with limited mobility. Compatible with analog film, digital phosphorous plates, and digital sensors. Output is X-ray exposure to receptor; clinician uses resulting image for diagnostic assessment of oral pathology. Benefits include portability for point-of-care imaging in non-traditional settings.
Clinical Evidence
No clinical data. Performance supported by bench testing, including electrical safety, EMC, software validation, usability, and risk assessment conducted in accordance with recognized IEC and ISO standards.
Technological Characteristics
Portable X-ray system; 70kVp, 2mA output; 0.4mm focal spot; 1.5mm Al total filtration; 1.6mm Al HVL. Constant potential (DC) waveform. Battery-powered. User interface: LCD with preset exposure times. Dimensions: 97 x 247 x 130mm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-65, IEC 62304, IEC 62366, ISO 14971.
Indications for Use
Indicated for use by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors in adult and pediatric patients.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 3, 2026
Ningbo Runyes Medical Instrument Co., Ltd.
% Lee Strong
Regulatory Consultant
510K FDA, Inc.
156 E. Granada Blvd.
ORMOND BEACH, FL 32176
Re: K252909
Trade/Device Name: Diagnostic X-Ray Equipment Model POCT22
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral source x-ray system
Regulatory Class: Class II
Product Code: EHD
Dated: August 28, 2025
Received: January 14, 2026
Dear Lee Strong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252909 - Lee Strong
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252909 - Lee Strong
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252909
Device Name
Diagnostic X-Ray Equipment Model POCT22
Indications for Use (Describe)
The Diagnostic X-Ray Equipment Model POCT22 is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
510k FDA Consulting
Medical Device Clearances
K252909
156 East Granada Blvd.
Ormond Beach, FL 32176
386-506-8711
# 510(k) Summary
## Submitter/Applicant
Ningbo Runyes Medical Instrument Co., Ltd.
No. 456 Tonghui Road
Jiangbei Investment & Pioneering Park C
Ningbo Zhejiang, CN 315000
Phone: +86-574-27709922
Contact: Weiqiong Fang, Registration Dept (xz02@runyes.com)
Date Prepared: August 13, 2025
## Preparer/Consultant
510K FDA Inc.
156 East Granada Blvd.
Ormond Beach, FL 32176
Phone: 386-846-6037
Fax: (386) 675-4621
Primary Contact: Lee Strong, Regulatory Dept. Mgr (lee@510kfda.com)
Secondary Contacts: Claude Berthoin, CEO (claude@510kfda.com).
## Device Classification
Trade/Model Names: Diagnostic X-Ray Equipment Model POCT22
Common Name: Portable X-ray System
Regulation Name: Extra-oral Source X-ray System
Regulation Number: 21 CFR 872.1800
Primary Product Code: EHD
Classification Name: Unit, X-ray, Extraoral with Timer
Regulatory Class: II
510k Review Panel: Dental
{5}
5-2
# Predicate Device
The subject device claims equivalence to the following legally marketed predicate:
510(k) Number: K231449
Applicant: Ningbo Runyes Medical Instrument Co., Ltd.
Date Cleared: July 13, 2023
Trade Name: Portable X-ray System Model Ray98(P)
Regulation Name: Extra-oral Source X-ray System
Regulation Number: 21 CFR 872.1800
Primary Product Code: EHD
Classification Name: Unit, X-ray, Extraoral with Timer
Regulatory Class: II
510k Review Panel: Dental
# Indications for Use
The Diagnostic X-Ray Equipment Model POCT22 is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects.
# Intended Use
Intended as extraoral x-ray source to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians.
# Device Description
The Diagnostic X-Ray Equipment Model POCT22 by Ningbo Runyes Medical Instrument Co., Ltd., is a portable x-ray device. The technology of this device was originally developed in Korea more than a decade ago and global production is still concentrated in that country.
The subject device is designed for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located inside the portable device. All three conventional types of intraoral receptors can be used with this device—analog x-ray film, digital phosphorous plates, and digital x-ray sensors.
This device is used in general dentistry and is supplied with an internal timer to control the duration of the x-ray source to the patient. The portable x-ray device is a choice model to assist doctors with special need patients, nursing home patients and patients in the office that cannot be easily moved, as well as other special situations. The choice of an x-ray generator is a matter of functional utility in the dental operatory and personal preference by the medical professional.
{6}
# Comparison of Technological Characteristics with Predicate
The following table compares technological and other characteristics of the subject and predicate device.
| Device Characteristic | Subject Device Diagnostic X-Ray Equipment Model POCT22 (K252909) | Predicate Device Portable X-ray System Model Ray98(P) (K231449) | Comparison |
| --- | --- | --- | --- |
| 510(k) Owner | Ningbo Runyes Medical Instrument Co., Ltd. (China) | Ningbo Runyes Medical Instrument Co., Ltd. (China) | NA |
| Classification & Product Code | 872.1800; EHD | 872.1800; EHD | Same |
| Intended Use | Intended as extraoral x-ray sources to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians. | Intended as extraoral x-ray sources to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians. | Same |
| Indication for Use | The Diagnostic X-Ray Equipment Model POCT22 is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects. | The device is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral receptors. Its use is intended for both adults and pediatric subjects. | Similar |
| Size | 97 x 247 x 130mm | 146 x 152 x 241mm | Difference of design, size |
| Source to Skin Distance | 20.5 cm ±0.5cm | 20.5 cm ±0.5cm | Same |
| Cone diameter | 6.0 cm ±0.5cm | 6.0 cm ±0.5cm | Same |
5-3
{7}
| Device Characteristic | Subject Device Diagnostic X-Ray Equipment Model POCT22 (K252909) | Predicate Device Portable X-ray System Model Ray98(P) (K231449) | Comparison |
| --- | --- | --- | --- |
| User interface | Up-down buttons for exposure time selection with timer display. Additionally, several user-selectable preset times with patient size, and tooth selection icons on an LCD display. | Up-down buttons for exposure time selection with timer display. Additionally, several user-selectable preset times with patient size, and tooth selection icons on an LCD display. | Same |
| Exposure switch | Exposure button at front cover on right hand side or remote switch. | Exposure button at front cover on right hand side or remote switch. | Same |
| Electrical Information | | | |
| Exposure time | 0.04 ~ 2.0 seconds in fixed increments:
0.04s, 0.05s, 0.06s, 0.08s, 0.10s, 0.12s, 0.16s, 0.20s, 0.25s, 0.32s, 0.40s, 0.50s, 0.63s, 0.80s, 1.00s, 1.25s, 1.60s, 2.00s | 0.04 ~ 2.0 seconds in fixed increments:
0.04s, 0.05s, 0.06s, 0.08s, 0.10s, 0.12s, 0.16s, 0.20s, 0.25s, 0.32s, 0.40s, 0.50s, 0.63s, 0.80s, 1.00s, 1.25s, 1.60s, 2.00s | Same |
| Time accuracy | ± (10% +1 ms) | ± (10% +1 ms) | Same |
| mA | 2mA | 2mA | Same |
| kVp | 70kVp | 70kVp | Same |
| Waveform | Constant Potential (DC) | Constant Potential (DC) | Same |
| Total Filtration | 1.5mmAl | 1.5mmAl | Same |
| Half-Value Layer (HVL) | 1.6mm Al | 1.6mm Al | Same |
| X-ray Focal Spot | 0.4mm | 0.4mm | Same |
| Average Number of 0.5s Exposures with Fully Charged Battery | 500 | 500 | Same |
{8}
| Device Characteristic | Subject Device Diagnostic X-Ray Equipment Model POCT22 (K252909) | Predicate Device Portable X-ray System Model Ray98(P) (K231449) | Comparison |
| --- | --- | --- | --- |
| Performance standards | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-2 | IEC 60601-1-2 | |
| | IEC 60601-1-3 | IEC 60601-1-3 | |
| | IEC 60601-1-6 | IEC 60601-1-6 | |
| | IEC 60601-2-65 | IEC 60601-2-65 | |
| | IEC 62304 | IEC 62304 | |
| | IEC 62366 | IEC 62366 | |
| | ISO 14971 | ISO 14971 | |
The above comparison shows the subject and predicate devices have substantially similar technological characteristics. Differences show up in the shape, size, design of the device and those are in cm and mm measurements of slight difference. The differences of the device are minor and do not raise new issues of safety and effectiveness.
## Non-Clinical Performance Data
The following performance testing was completed on the subject device in support of the substantial equivalence determination of the predicate device.
- Electrical Safety and EMC
- Software Validation
- Usability
- Clinical Evaluation
- Risk Assessment
- All tests were performed in accordance with IEC and ISO standards and tests are recognized by FDA.
- None of the standards were adapted for application to the device under review.
- No deviations from the standards were applied.
- No differences exist between the tested device and the device to be marketed.
- Conformity with all standards was determined by the device manufacturer, Ningbo Runyes Ltd., Korea.
- Electrical test performed by KCTL, Inc. Laboratories, Inc., Korea.
{9}
5-6
# Specific Guidance Document
The Guidance document: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices applies to this device. Details of this guidance are provided within the Software Validation Report.
# Labels
The labels on the device show that this device conforms to the following:
21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products,
21 CFR 1020.30: Diagnostic x-ray systems and their major components,
21 CFR 1020.31: Radiographic Equipment
# Substantial Equivalence
The above comparison chart shows the subject and predicate devices have similar technological characteristics.
Both devices have:
- Similar function and are used in similar environments.
- The same indications for use and the same intended use.
- Similar manufacturing process and similar technological characteristics.
- Both devices have completed similar IEC and ISO standardized testing listed in the comparison chart shown above.
# Conclusion
The subject and predicate device have the same indications for use, the same intended use and similar technological characteristics. The Diagnostic X-Ray Equipment Model POCT22 performs similar functions, in a similar environment as the predicate device. Diagnostic X-Ray Equipment Model POCT22 uses similar technology as the predicate device, based on well-known technology. Diagnostic X-Ray Equipment Model POCT22 is as safe and effective as the predicate device. We believe the subject device does not introduce any new safety concerns and is substantially equivalent to the predicate device. In conclusion, the subject device, Diagnostic X-Ray Equipment Model POCT22, is at least as safe and effective as the predicate device and warrants a finding of substantial equivalence to the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.