← Product Code [GCJ](/productcode/GCJ) · K252858

# 4K UHD Endoscopy Camera System (KMS-4K-2088, KMS-4K-2188, KMS-4K-2288) (K252858)

_Kms Medical Technology Co., Ltd. · GCJ · Apr 30, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252858

## Device Facts

- **Applicant:** Kms Medical Technology Co., Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Apr 30, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The 4K UHD Endoscopy Camera System, when used in conjunction with optical endoscope, cold light source and monitor, is intended to convert the optical signals captured by the endoscope into electronic signals and transmits them to the monitor for imaging, and can be used for both diagnostic and therapeutic interventions.

## Device Story

4K UHD Endoscopy Camera System captures optical signals via endoscope; converts to electronic signals using CMOS image sensor in camera head; transmits to Camera Control Unit (CCU). CCU processes, records, and outputs 4K resolution (3840x2160) video to external monitors via DVI, 3G-SDI, or HDMI. Used in professional healthcare facilities by trained medical personnel. Device is reusable; camera head requires cleaning and sterilization between patients. CCU is non-sterile. System aids visualization during diagnostic and therapeutic interventions; provides real-time video feedback to clinicians for procedural guidance.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via IEC 60601-1, 60601-2-18, 60601-1-6, 60601-1-2, and 60601-4-2 standards. Cleaning validation per AAMI ST98:2022; sterilization validation per ISO 17664-1 achieving SAL 10^-6. Repeated use testing (50 cycles) confirmed device integrity.

## Technological Characteristics

System includes CCU and camera head with 1/3 inch CMOS sensor (16:9 aspect ratio). Resolution 3840x2160. Connectivity: DVI, 3G-SDI, HDMI, USB 2.0/3.0. Power: 100-240V AC. Protection: Class 1, Type CF. Camera head IPX8, CCU IPX0. Sterilization: Low-temperature hydrogen peroxide plasma. Software: Basic level of concern.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- LOGIC 4K CAMERA CONTROLLER, LOGIC 4K CAMERA HEAD ([K180583](/device/K180583.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 30, 2026

KMS Medical Technology Co., Ltd.
Yi Dai
Registration Affairs Specialist
Building A1 and A2, No.769 Qingzhuhu Road, Qingzhuhu
Sub-district, Kaifu
Changsha, Hunan 410201
China

Re: K252858
Trade/Device Name: 4K UHD Endoscopy Camera System (KMS-4K-2088, KMS-4K-2188, KMS-4K-2288)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ
Dated: September 9, 2025
Received: September 9, 2025

Dear Yi Dai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}

K252858 - Yi Dai
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K252858 - Yi Dai
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha Hithe
Digitally signed by Tanisha Hithe
Date: 2026.04.30
12:49:57 -04'00"

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252858 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  4K UHD Endoscopy Camera System (KMS-4K-2088, KMS-4K-2188, KMS-4K-2288)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The 4K UHD Endoscopy Camera System, when used in conjunction with optical endoscope, cold light
source and monitor, is intended to convert the optical signals captured by the endoscope into electronic
signals and transmits them to the monitor for imaging, and can be used for both diagnostic and therapeutic
interventions.  |   |   |
|  Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) Summary

# 510(k) Summary K252858

## 1. Submitter

Name: KMS Medical Technology Co., Ltd.

Address: Building A1 and A2, No.769 Qingzhuhu Road, Qingzhuhu Sub-district, Kaifu, Changsha, Hunan, 410201, China

Contact Person: Yi Dai

Position: Registration Affairs Specialist

Phone: 86-731-13588367339

E-mail: 1104173990@qq.com

Preparation date: April 29, 2026

## 2. Device

Common Name/Trade Name: 4K UHD Endoscopy Camera System

Model: KMS-4K-2088, KMS-4K-2188, KMS-4K-2288

Regulation Name: Endoscope and accessories

Regulation Number: 21 CFR 876.1500

Regulatory Class: II

Product Code: GCJ

Review Panel: General &amp; Plastic Surgery

## 3. Predicate Device

|   | Predicate Device  |
| --- | --- |
|  510(k) Number | K180583  |
|  Trade Name | LOGIC 4K CAMERA CONTROLLER, LOGIC 4K CAMERA HEAD  |
|  Manufacture | Richard Wolf Medical Instruments Corporation  |
|  Regulation Name | Endoscope and accessories  |
|  Regulation Number | 21 CFR 876.1500  |
|  Regulatory Class | II  |
|  Product Code | FET  |
|  Review Panel | Gastroenterology-Urology Devices  |

{5}

510(k) Summary

## 4. Device Description

The 4K UHD (Ultra High Definition) Endoscopy Camera System, when used in conjunction with commercially available optical endoscope, cold light source and monitor, is intended to convert the optical signals captured by the endoscope into electronic signals and transmits them to the monitor for imaging, for using in providing endoscopic video during diagnostic and therapeutic procedures.

The device is reusable for multi-patient use, not intended to contact the patient directly or indirectly, delivered non-sterile, and the camera head must be cleaned and sterilized prior to each use.

The device is intended to be used in professional healthcare facility environment, and exclusively used by specialized medical personnel with medically qualified and adequately trained persons.

The device consists of a Camera Control Unit (CCU) and a Camera Head with an integral cable that connects to CCU and with a detachable coupler that connects to endoscope. The device cannot be connected to network.

The Camera Head is designed to convert the optical signals into electronic signals and transmit them to CCU. The CCU is the control center of the system, designed for processing and recording the images of electronic signals and can process a 4K resolution (3840 × 2160 pixels). The CCU equipped with various inputs/outputs including DVI, 3G-SDI, HDMI, HCOM, USB port.

## 5. Indications for use

The 4K UHD Endoscopy Camera System, when used in conjunction with optical endoscope, cold light source and monitor, is intended to convert the optical signals captured by the endoscope into electronic signals and transmits them to the monitor for imaging, and can be used for both diagnostic and therapeutic interventions.

{6}

510(k) Summary

6. Comparison of technological characteristics with the predicate devices

|  Item | Proposed Device | Predicate Device K180583 | Remark  |
| --- | --- | --- | --- |
|  Regulation Number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same  |
|  Product Code | GCJ | FET | Gap  |
|  Device Class | II | II | Same  |
|  Indication for Use | The 4K UHD Endoscopy Camera System, when used in conjunction with optical endoscope, cold light source and monitor, is intended to convert the optical signals captured by the endoscope into electronic signals and transmits them to the monitor for imaging, and can be used for both diagnostic and therapeutic interventions. | The ENDOCAM® Logic 4K Camera System 5525 has been designed for high-definition video endoscopy and can be used for both diagnostic and therapeutic interventions. The ENDOCAM® Logic 4K Camera System 5525 is used in conjunction with other video equipment and endoscopic accessories. | Gap  |
|  Target population | Adult patients | Adult patients | Same  |
|  Anatomical site | Not intended to contact the patient directly or indirectly | Not intended to contact the patient directly or indirectly | Same  |
|  Where used | Professional health care facility | Professional health care facility | Same  |
|  Prescription/OTC use | Prescription | Prescription | Same  |
|  Item | Proposed Device | Predicate Device K180583 | Remark  |
| --- | --- | --- | --- |
|  Principle | Utilize camera head with CMOS image sensor to convert the optical signals captured by endoscope into electronic signals and transmit to control unit. The control unit can process and records image data, and transmit to monitors. | Utilize camera head with CMOS image sensor to convert the optical signals captured by endoscope into electronic signals and transmit to control unit. The control unit can process and records image data, and transmit to monitors. | Same  |
|  Software | Control unit contains software; Camera head does not contain software; The software is classified as Basic Documentation Level of concern, since failures or latent design flaws are unlikely to | Camera controller contains software; Camera head does not contain software; The software is classified as Basic Documentation Level of concern, since failures or latent | Same  |

{7}

510(k) Summary

|  Item | Proposed Device | Predicate Device K180583 | Remark  |
| --- | --- | --- | --- |
|   | cause any injury to the patient or operator. | design flaws are unlikely to cause any injury to the patient or operator. |   |
|  Structure and Composition | Camera control unit and Camera head with a detachable coupler | Camera controller and Camera head with a detachable coupler | Same  |
|   |  1x 1/3 inch CMOS image sensors, aspect ratio 16:9 | 3x 1/3 inch CMOS image sensors, aspect ratio 16:9 | Gap  |
|  4K UHD resolution | 3840 x 2160px | 3840 x 2160px and 4096 x 2160px | Gap  |
|  Weight | Control unit: 4.8kg
Camera head: 0.15kg | Controller: 6.0kg
Camera head: 121g | Gap  |
|  Dimensions Control unit (w x h) | 346 x 300 x 130 mm | 300 x 120 x 416 mm | Gap  |
|  Dimensions Touchscreen (w x h) | 130*70mm | 144 x 78.24 mm | Gap  |
|  Materials of use | Metal housing | Metal housing | Same  |
|  Materials with patient contact | No | No | Same  |
|  Single use / reusable | Reusable for multi-patient use | Reusable for multi-patient use | Same  |
|  Delivered sterile / non-sterile | Control unit:
Delivered non-sterile;
Cannot be sterilized;

Camera head:
Delivered non-sterile;
Require cleaning and sterilization before each use | Controller:
Delivered non-sterile;
Cannot be sterilized;

Camera head:
Delivered non-sterile;
Require cleaning and sterilization before each use | Same  |
|  Operating conditions | +5°C to +40°C | +10°C to +40°C | Gap  |
|  Combinations | Optical endoscope | Optical endoscope | Same  |
|   |  External light source | External light source | Same  |
|   |  Monitors with DVI / 3G-SDI / HDMI interface | Monitors with 3G-SDI / HDMI interface | Gap  |
|   |  USB2.0, USB3.0 (FAT32/NTFS format)) | USB2.0 (FAT32/NTFS format)) | Gap  |
|  File format of saved pictures / videos | JPEG, BMP (max.3840 x2160)
AVI (max.3840 x2160) | JPEG, TIFF (max. 1920 x 1200)
MPEG4 (max. 1920 x 1080/60/P) | Gap  |
|  LAN (Ethernet) Network Connector usage | No | Yes | Gap  |
|  Power supply | Rated voltage: AC100-240V
Rated frequency: 50/60 Hz
Rated power: 120 VA | Rated voltage: AC100-240V
Rated frequency: 50/60 Hz
Rated power: 120 VA | Same  |

{8}

510(k) Summary

|  Item | Proposed Device | Predicate Device K180583 | Remark  |
| --- | --- | --- | --- |
|  Type of protection against electric shock | Class 1 | Class 1 | Same  |
|  Degree of protection against electric shock | CF | CF | Same  |
|  Degree of protection against ingress of water | Control unit: IPX0 | Controller: IP20 (not protected) | Gap  |
|   |  Camera head: IPX8 | Camera head: IPX7  |   |
|  Mode of operation | Continuous | Continuous | Same  |
|  Cooling method | Constant fan colling | Constant fan colling | Same  |
|  Biocompatibility | The device does not contain components that come into direct or indirect contact with patients. | The device does not contain components that come into direct or indirect contact with patients. | Same  |
|  Performance Standards | IEC 60601-1 | IEC 60601-1 | Same  |
|   |  IEC 60601-2-18 | IEC 60601-2-18  |   |
|   |  IEC 60601-1-6 | IEC 60601-1-6  |   |
|   |  IEC 60601-1-2 | IEC 60601-1-2 | Same  |
|  IEC 60601-4-2 | IEC 60601-4-2  |   |   |
|  Focal length | 14-32mm | 13-29mm | Gap  |
|  Minimum illumination | ≤1 Lux / F1.6 | < 0.7 Lux / F1.8 @ 50% Level | Gap  |
|  Signal to noise ratio (SINR) | 56dB | 52 dB | Gap  |

The proposed and predicate device have equivalent Indications for use.

The proposed and predicate device have the same structure and composition.

The proposed and predicate device have different technological characteristics, and these differences do not raise different questions of safety and effectiveness.

## 7. Non-Clinical Testing

### Performance testing

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards

- IEC60601-1 Edition3.2 2020-08 CONSOLIDATED VERSION Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
- IEC60601-2-18 Edition3.0 2009-08 Medical electrical equipment Part 2-18:

{9}

510(k) Summary

Particular requirements for the basic safety and essential performance of endoscopic equipment.

- IEC 60601-1-6 Edition3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance Collateral standard: Usability.
- IEC 60601-1-2 Edition4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic disturbances Requirements and tests.
- IEC TS 60601-4-2 Edition 1.0 2024-03 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.

## Biocompatibility

Not intended to contact the patient directly or indirectly.

Therefore, material biocompatibility according to ISO 10993-1 Biological Evaluation of Medical Devices is not required.

## Cleaning validation:

The cleaning validation demonstrates that no visible soil was seen on the processed devices under normal lighting conditions after cleaning. The cleaning method is acceptable.

The cleaning validation is Reference AAMI ST98:2022 Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices.

## Sterilization validation:

The sterilization validation demonstrates that the sterilization method was effective on the test device and achieve a SAL of $10^{-6}$.

The sterilization validation is reference ISO17664-1: 2021 Processing of health care

{10}

510(k) Summary

products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices.

## Repeated cleaning + sterilization tolerance validation:

After 50 repeated washes and low-temperature hydrogen peroxide plasma sterilization, no abnormalities were observed in the visible parts of the three samples of the instruments, and the device still conforms to product standards.

## 8. Clinical Testing

None

## 9. Conclusion

The 4K UHD Endoscopy Camera System is substantially equivalent to the predicate device. The non-clinical testing demonstrates that the device is as safe and effective as the predicate device.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K252858](https://fda-staging.innolitics.com/device/K252858)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
