K252853 · Xiamen Inno Medical Technology Co., Ltd. · QNP · Aug 5, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K252853
Device Name
CAD - Lower GI (EC07-01)
Applicant
Xiamen Inno Medical Technology Co., Ltd.
Product Code
QNP · Gastroenterology, Urology
Decision Date
Aug 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1520
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252853 · Aug 5, 2026
CAD - Lower GI (EC07-01)
Xiamen Inno Medical Technology Co., Ltd.
Public third-party clinical research dataset (REAL-Colon)
The sponsor used the REAL-Colon dataset to conduct standalone performance testing, assessing the device's lesion-based sensitivity, false positive rates, and ROC/FROC curves to demonstrate algorithmic performance and support substantial equivalence.
Standalone performance testing; Public dataset; Algorithmic validation; REAL-Colon
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
REAL-Colon dataset evaluation; Retrospective analysis of public third-party clinical research data; Follow-up/Duration: Not applicable
Adult patients undergoing colonoscopy (derived from public dataset); Sample Size: 2,393,916 images; 132 positive lesions; 60 cases; Number of Sites: Not applicable (public dataset)
Not applicable for this study
Frame-level TPR/FPR, lesion-level TPR, FPs per patient, ROC/FROC curves
CAD - Lower GI is a computer-assisted reading tool intended to aid qualified endoscopists in detecting suspected colonic mucosal lesions, such as polyps and adenomas, in real-time endoscopic images obtained during colorectal endoscopy. The device is intended for use in adult patients undergoing colorectal cancer screening or surveillance colonoscopy in medical institutions. The device detects and displays areas of suspected lesions to assist endoscopists during colorectal endoscopy. The device is not intended to characterize lesions or replace clinical judgment. Physicians shall evaluate the device output together with the endoscopic images and other clinical information and shall not make clinical decisions solely based on the device output. The product is intended to be used only with specified compatible image processors and video endoscopes under white light and NBI modes.
Device Story
CAD - Lower GI is an AI-based standalone software that processes live endoscopic video streams to detect suspected colorectal polyps in real-time. It connects between an endoscopic video processor (Olympus CV-190/CV-1500) and a monitor. The software uses a deep learning algorithm to analyze video frames and overlays colored box markers on the display to highlight suspected lesions without altering the original video feed. It is operated by endoscopists in medical institutions during screening or surveillance colonoscopies. The output serves as a visual aid to enhance physician awareness; it does not characterize lesions or replace clinical judgment. Physicians must evaluate the device output alongside endoscopic images and clinical information. The device benefits patients by increasing the detection of small, flat, or proximal adenomas that might otherwise be missed during standard procedures.
Clinical Evidence
A prospective, multicenter, randomized controlled trial (n=840) compared CAD-assisted colonoscopy to standard colonoscopy. Primary endpoints were Adenoma Detection Rate (ADR) and Positive Predictive Agreement (PPA). ADR was 51.2% (Experimental) vs 36.4% (Control), demonstrating superiority (+14.8 pp, 95% CI 8.1, 21.5). PPA was 81.2% (Experimental) vs 82.5% (Control), demonstrating non-inferiority (-1.3 pp, 95% CI -5.8, 3.2). Secondary endpoint APC was 1.027 (Experimental) vs 0.609 (Control), demonstrating superiority (+0.418, 95% CI 0.26, 0.57). No device-related SAEs occurred.
Technological Characteristics
Standalone SaMD; deep learning-based detection algorithm. Compatible with Olympus CV-190 and CV-1500 processors. Supports WLI and NBI modes. Operates on user-provided PC (Ubuntu 22.04 LTS, Intel i7-7700k, NVIDIA RTX 3070, 16GB RAM). Video input via HD-SDI/3G-SDI; output via DP/DVI. Annotation latency ≤ 40 ms; real-time video delay ≤ 30 ms.
Indications for Use
Indicated for adult patients undergoing colorectal cancer screening or surveillance colonoscopy in medical institutions to assist qualified endoscopists in detecting suspected colonic mucosal lesions (polyps/adenomas).
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
Predicate Devices
GI Genius™ Module 100 (GGM100.US); GI Genius™ Module 200 (GGM200.US); ColonPRO™ 4.0 (CPRO40.US) (K233964)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 5, 2026
Xiamen Inno Medical Technology Co.,Ltd.
% Rachel Yu
Regulatory Manager
ZHIHE Info-Tech(Suzhou) Co., Ltd.
Room 616, Building 1, No. 1 Huayun Road, Industrial Park
Suzhou, Jiangsu 215134
China
Re: K252853
Trade/Device Name: CAD - Lower GI (EC07-01)
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulatory Class: Class II
Product Code: QNP
Dated: July 4, 2026
Received: July 6, 2026
Dear Rachel Yu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252853 - Rachel Yu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252853 - Rachel Yu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# SHANIL P. HAUGEN -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252853 | ? |
| Please provide the device trade name(s). | | ? |
| CAD - Lower GI | | |
| Please provide your Indications for Use below. | | ? |
| CAD - Lower GI is a computer-assisted reading tool intended to aid qualified endoscopists in detecting suspected colonic mucosal lesions, such as polyps and adenomas, in real-time endoscopic images obtained during colorectal endoscopy. The device is intended for use in adult patients undergoing colorectal cancer screening or surveillance colonoscopy in medical institutions. The device detects and displays areas of suspected lesions to assist endoscopists during colorectal endoscopy. The device is not intended to characterize lesions or replace clinical judgment. Physicians shall evaluate the device output together with the endoscopic images and other clinical information and shall not make clinical decisions solely based on the device output. The product is intended to be used only with specified compatible image processors and video endoscopes under white light and NBI modes. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252853
Page 1 of 18
## 510(k) Summary
### I Submitter
Device Submitter: Xiamen Inno Medical Technology Co., Ltd.
Unit 203-2, No. 55, Chengyi North Street, Phase 3 of software park, Xiamen, Fujian Province, China
Contact Person: Jianqing Zhao
Title: Manager
Phone: +86 173 5002 8269
Email: zhaojq@xm-innovision.com
Date Prepared: September 5, 2025
### II Device
Trade Name of Device: CAD - Lower GI
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulation Class: II
Product code: QNP
Review Panel: Gastroenterology/Urology
### III Predicate Device 1
Trade/Device Name: GI Genius™ Module 100 (GGM100.US); GI Genius™ Module 200 (GGM200.US); ColonPRO™ 4.0 (CPRO40.US)
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulatory Class: Class II
Product Code: QNP
Premarket Notification: K233964
Manufacturer: Cosmo Artificial Intelligence - AI Ltd
### IV Predicate Device 2
Trade/Device Name: GI Genius
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulatory Class: Class II
Product Code: QNP
Submission Number: DEN200055
Manufacturer: Cosmo Artificial Intelligence - AI, LTD
1 / 18
{5}
K252853
Page 2 of 18
## V Device Description
CAD-Lower GI is an artificial intelligence-based standalone medical device software designed to process colonoscopy images and assist in the real-time detection of suspected colorectal polyps during procedures in adult patients. It is intended to aid physicians in medical institutions and is not meant to replace full patient evaluation or confirm diagnoses.
The device is delivered as a software installation package via email, comprising modules for colorectal polyp detection, video display, recording, system settings, and user management. The polyp detection module uses a deep learning algorithm to analyze endoscopic images and highlight suspected polyp locations in real-time.
CAD-Lower GI is compatible with endoscopic image processors (CV-190 and CV-1500, Olympus Medical Corporation) and their compatible video colorectal endoscopes, supporting white light(WLI) and Narrow Band Imaging (NBI) modes. It connects between the processor's SDI-OUT (HD-SDI/3G-SDI) and a computer system's video capture card, outputting processed images to a monitor via DP-OUT to DVI-IN/DP-IN.
The software operates on a user-provided computer system running Ubuntu 22.04 LTS, requiring an Intel i7-7700k or higher CPU, \(\geqslant\)16 GB memory, NVIDIA RTX 3070 or higher graphics card, a video capture card supporting 1080I@50Hz or 1080P@60Hz with \(\geqslant\)400 MB/s bandwidth, and a 1920x1080 resolution monitor. An Ethernet connection (RJ45, \(\geqslant\)100 Mbps) with a firewall is needed for license activation.
During colonoscopy, CAD-Lower GI processes live video streams, overlaying colored box markers on the display monitor to highlight suspected polyps in real-time, without altering the original video feed. It supports standard endoscopic workflows, enhancing physician awareness through AI-driven image analysis.
## VI Indications for Use
CAD - Lower GI is a computer-assisted reading tool intended to aid qualified endoscopists in detecting suspected colonic mucosal lesions, such as polyps and adenomas, in real-time endoscopic images obtained during colorectal endoscopy.
The device is intended for use in adult patients undergoing colorectal cancer screening or surveillance colonoscopy in medical institutions.
The device detects and displays areas of suspected lesions to assist endoscopists during colorectal endoscopy. The device is not intended to characterize lesions or replace clinical judgment. Physicians shall evaluate the device output together with the endoscopic images and other clinical information and shall not make clinical decisions solely based on the device output.
The product is intended to be used only with specified compatible image processors and video endoscopes under white light and NBI modes.
2 / 18
{6}
K252853
Page 3 of 18
## VII Substantial Equivalence Discussion
The comparison and discussion between the subject device and the predicate devices are listed in below tables:
Table 1: Comparison of CAD - Lower GI to GI Genius
| Items | Subject Device CAD - Lower GI | Predicate Device 1 GI Genius™ Module 100; GI Genius™ Module 200; ColonPRO™ 4.0. (K233964) | Predicate Device 2 GI Genius (DEN200055) |
| --- | --- | --- | --- |
| FDA Reg name | Gastrointestinal lesion software detection system | Gastrointestinal lesion software detection system | Gastrointestinal lesion software detection system |
| FDA Reg # | 21 CFR 876.1520 | 21 CFR 876.1520 | 21 CFR 876.1520 |
| Product Code | QNP | QNP | QNP |
| Indications for Use | CAD - Lower GI is a computer-assisted reading tool intended to aid qualified endoscopists in detecting suspected colonic mucosal lesions, such as polyps and adenomas, in real-time endoscopic images obtained during colorectal endoscopy. The device is intended for use in adult patients undergoing colorectal cancer screening or surveillance colonoscopy in medical institutions. The device detects and displays areas of suspected lesions to assist endoscopists during colorectal endoscopy. The device is not intended to characterize lesions or replace clinical judgment. Physicians shall evaluate the device output together with the endoscopic images and other clinical information and shall not make clinical decisions solely based on the device output. The product is intended to be used only with specified compatible image processors and video endoscopes under white light and NBI modes. | The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making. | The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making. |
| Video delay, signal in to signal out | The maximum real-time video delay is 30ms | GI Genius™ Module 100 with ColonPRO™ 4.0: 1.52 μs GI Genius™ Module 200 with ColonPRO™ 4.0: 0.74 μs | 1.52 μs |
| Annotation delay | The maximum annotation delay is 38ms | Time is less than or equal to 120 milliseconds | Time is less than or equal to 120 milliseconds |
| Video quality integrity test | The maximum color error of image quality degradation caused by device is 0.777427, which is only slightly recognizable to human eyes | There should be no degradation in image quality, meaning that all corresponding pixels (from the endoscopic video processor and from the GI Genius) are identical in the three color channels | There should be no degradation in image quality, meaning that all corresponding pixels (from the endoscopic video processor and from the GI Genius) are identical in the |
3 / 18
{7}
K252853
Page 4 of 18
| Items | Subject Device CAD - Lower GI | Predicate Device 1GI GeniusTM Module 100;GI GeniusTM Module 200;ColonPROTM 4.0.(K233964) | Predicate Device 2GI Genius(DEN200055) |
| --- | --- | --- | --- |
| | | | three color channels |
| Video processor | Olympus CV-190,Olympus CV-1500 | Olympus CV-180,Olympus CV-190,Pentax EPK-i7000,Fujifilm VP-4450HDFujifilm VP-7000Olympus CV-1500 | Olympus CV-180,Olympus CV-190,Pentax EPK-i7000,Fujifilm VP-4450HD |
| Accessories (optional) | None | FootswitchUSB K-switch | N/A |
| Software other function | Recording and file management function | Procedure Highlights function | N/A |
| Electrical safety | The CAD - Lower GI is a standalone SaMD (Software as a Medical Device) without hardware. | IEC 60601-1 | IEC 60601-1 |
| Electromagnetic compatibility | The CAD - Lower GI is a standalone SaMD (Software as a Medical Device) without hardware. | IEC 60601-1-2 | IEC 60601-1-2 |
| LAN port | Yes, non-functional to user | Yes, non-functional to user | No |
CAD - Lower GI has the same intended use and similar technological characteristics as the predicate devices. The identified differences, including support for NBI imaging and differences in video-processing architecture, were evaluated through nonclinical and clinical performance testing. These differences do not raise different questions of safety or effectiveness and do not affect the substantial equivalence determination.
### VIII Non-Clinical Tests
The following non-clinical verification and validation activities have been completed for the CAD - Lower GI device:
- System-level software verification was conducted against the Software Requirements Specification (SRS). All specified requirements were tested and met, covering multiple functional units and items.
- User-level software validation was performed in accordance with the SRS, confirming that the device performs as intended in its clinical context of use.
- Risk control measures identified during the risk management process were verified or validated, as appropriate, to ensure risk mitigations are effective.
- Standalone performance testing was conducted, including evaluation on static image datasets, dynamic video datasets, and public third-party clinical research data. Key parameters assessed included lesion-based sensitivity, true positive and false positive rates per frame, ROC curve (AUC), and false positive clusters per patient. Results demonstrated performance within or exceeding the range of predicate devices.
• Video processing tests verified signal acquisition, annotation latency ( \( \leq \) 40 ms), and frame synchronization.
- Cybersecurity measures related to Ethernet communication with the User Activation Management Platform were implemented and assessed in accordance with FDA guidance documents (Cybersecurity in Medical
4 / 18
{8}
K252853
Page 5 of 18
Devices: Quality System Considerations and Content of Premarket Submissions).
The results of these activities demonstrate that the CAD – Lower GI meets its specifications and support a finding of substantial equivalence to the predicate devices.
### Performance Data
The following testing was conducted for the CAD - Lower GI system with data included in the 510(k) document.
- Software Verification and Validation
The Software verification and validation was conducted for the CAD - Lower GI software to validate it for its intended use per the design documentation in line with recommendations outlined in General Principles of Software Validation, Guidance for Industry and FDA Staff. The CAD - Lower GI software demonstrated passing results on all applicable unit, integration, and requirements testing.
- Standalone Performance Testing
The purpose of the standalone performance testing is to demonstrate that overall algorithmic performance is sufficient to fulfill the indications for use of the CAD - Lower GI. The standalone performance evaluation was carried out using the static image datasets, dynamic video datasets, and public third-party data which are shown in Tables 2-8. The Static Image Dataset is used exclusively as an internal validation dataset. The public third-party data sourced from open-source dataset REAL-Colon. This evaluation was performed in both WLI and NBI modes.
Table 2: Detailed Static Image Dataset
| Items | Classification Criteria | Number of Images | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| Age | <45 | 9405 | 10433 | 19838 | 58464 |
| | ≥45 | 21382 | 17244 | 38626 | |
| Gender | Male | 17254 | 15522 | 32776 | 58464 |
| | Female | 13533 | 12155 | 25688 | |
| Image Processor | Olympus CV-1500 | 6699 | 5571 | 12270 | 12270 |
| Light Source | WLI | 25594 | 26222 | 51816 | 58464 |
| | NBI | 5193 | 1455 | 6648 | |
| Lesion Nature | Inflammatory/Hyperplastic Polyp | 17227 | - | 17227 | 31185 |
| | Adenoma | 13958 | - | 13958 | |
| Lesion Morphology | Yamada Type I | 15944 | - | 15944 | 31504 |
| | Yamada Type II | 12677 | - | 12677 | |
| | Yamada Type III | 2399 | - | 2399 | |
| | Yamada Type IV | 484 | - | 484 | |
Table 3: Detailed Image-Based Dynamic Video Dataset
| Items | Classification Criteria | Number of Images | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| Age | <45 | 243134 | 2021403 | 2264537 | 5161320 |
| | ≥45 | 284219 | 2612564 | 2896783 | |
| Gender | Female | 242005 | 2016984 | 2258989 | 5161320 |
| | Male | 285348 | 2616983 | 2902331 | |
| Image Processor | Olympus CV-1500 | 150337 | 1676445 | 1826782 | 2876012 |
| | Olympus CV-190 | 129204 | 920026 | 1049230 | |
| Light Source | WLI | 423252 | 4487397 | 4910649 | 5161320 |
| | NBI | 104101 | 146570 | 250671 | |
| Lesion | Inflammatory/Hyperplastic | 322869 | - | 322869 | |
5 / 18
{9}
K252853
Page 6 of 18
| Nature | Polyp | | | | 528766 |
| --- | --- | --- | --- | --- | --- |
| | Adenoma | 205897 | - | 205897 | |
| Lesion Morphology | Yamada Type I | 397935 | - | 397935 | 528545 |
| | Yamada Type II | 92928 | - | 92928 | |
| | Yamada Type III | 35386 | - | 35386 | |
| | Yamada Type IV | 2296 | - | 2296 | |
Table 4: Detailed Lesion-Based Dynamic Video Dataset
| Item | Classification Criteria | Number of Positive Lesions | Total |
| --- | --- | --- | --- |
| Age | <45 | 265 | 507 |
| | ≥45 | 242 | |
| Gender | Male | 273 | 507 |
| | Female | 234 | |
| Image Processor | Olympus CV-1500 | 130 | 246 |
| | Olympus CV-190 | 116 | |
| Light Source | WLI | 496 | 725 |
| | NBI | 229 | |
| Lesion Nature | Inflammatory/Hyperplastic Polyp | 377 | 507 |
| | Adenoma | 130 | |
| Lesion Morphology | Yamada Type I | 442 | 507 |
| | Yamada Type II | 53 | |
| | Yamada Type III | 10 | |
| | Yamada Type IV | 2 | |
Table 5: Detailed Case-Based Dynamic Video Dataset
| Items | Classification Criteria | Number of Cases | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Positive Cases | Negative Cases | Total Cases | |
| Age | <45 | 94 | 2 | 96 | 209 |
| | ≥45 | 106 | 7 | 113 | |
| Gender | Male | 107 | 3 | 110 | 209 |
| | Female | 93 | 6 | 99 | |
| Image Processor | Olympus CV-1500 | 46 | 9 | 55 | 104 |
| | Olympus CV-190 | 49 | 0 | 49 | |
Table 6: Detailed Image-Based Third-party Data
| Items | Classification Criteria | Number of Images | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| Age | < 45 | 3512 | 44439 | 47951 | 2393916 |
| | ≥45 | 206153 | 2139812 | 2345965 | |
| Gender | Male | 127544 | 1233158 | 1360702 | 2393916 |
| | Female | 82121 | 951093 | 1033214 | |
| Endoscope Brands | Olympus | 195771 | 2000039 | 2195810 | 2393916 |
| | Fujifilm | 13894 | 184212 | 198106 | |
| Lesion Nature | Hyperplastic/Inflammatory Polyp | 112017 | - | 112017 | 209833 |
| | Adenoma | 97816 | - | 97816 | |
| Lesion Size | 0-5mm | 165992 | - | 165992 | 209665 |
| | 6-9mm | 20096 | - | 20096 | |
6 / 18
{10}
K252853
Page 7 of 18
| | ≥10mm | 23577 | - | 23577 | |
| --- | --- | --- | --- | --- | --- |
Table 7: Detailed Lesion-Based Third-party Data
| Items | Classification Criteria | Number of Positive Lesions | Total |
| --- | --- | --- | --- |
| Age | <45 | 3 | 132 |
| | ≥45 | 129 | |
| Gender | Male | 79 | 132 |
| | Female | 53 | |
| Endoscope Brands | Olympus | 117 | 132 |
| | Fujifilm | 15 | |
| Lesion Nature | Hyperplastic/Inflammatory Polyp | 76 | 132 |
| | Adenoma | 56 | |
| Lesion Size | 0-5mm | 117 | 132 |
| | 6-9mm | 12 | |
| | ≥10mm | 3 | |
Table 8: Detailed Case-Based Third-party Data
| Items | Classification Criteria | Number of Cases | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Positive Cases | Negative Cases | Total Cases | |
| Age | < 45 | 1 | 1 | 2 | 60 |
| | ≥45 | 45 | 13 | 58 | |
| Gender | Male | 27 | 6 | 33 | 60 |
| | Female | 19 | 8 | 27 | |
| Endoscope Brands | Olympus | 42 | 11 | 53 | 60 |
| | Fujifilm | 4 | 3 | 7 | |
| Region | 001 | 9 | 6 | 15 | 60 |
| | 002 | 11 | 4 | 15 | |
| | 003 | 14 | 1 | 15 | |
| | 004 | 12 | 3 | 15 | |
*In Table 2, Table 3 and Table 6, since a single positive image may contain multiple positive lesions with different lesion nature, lesion size and lesion morphologies, there are differences in the statistics of lesion nature, lesion morphology and the total number of positives. In Table 4, since a single lesion is counted multiple times under different light sources, the number of light source lesions is different from other subgroup. In Table 8, 001 represents the United States, the United Kingdom, and Italy; 002 represents Italy; 003 represents Austria; 004 represents Japan.
In static image dataset, sensitivity, specificity and precision were calculated as the evaluation metrics. The evaluation results of static image dataset are show in Table 9 below.
Table 9: Evaluation Results for Static Image Dataset
| Items | | Sensitivity(%)[95% CI] | Specificity(%)[95% CI] | Precision(%) [95% CI] |
| --- | --- | --- | --- | --- |
| Overall | | 98.68 [98.56-98.81] | 98.34 [98.19-98.49] | 98.51 [98.38-98.65] |
| Light Source | WLI | 98.68 [98.54-98.82] | 98.34 [98.19-98.50] | 98.31 [98.15-98.47] |
| | NBI | 98.69 [98.38-99.00] | 98.28 [97.61-98.95] | 99.51 [99.32-99.70] |
| Image Processor | Olympus CV-1500 | 98.52 [98.23-98.81] | 96.46 [95.98-96.95] | 97.10 [96.70-97.50] |
In dynamic video dataset and third-party data, frame level TPR(True Positive Rate) and FPR(False Positive Rate), lesion level TPR and FPs(False Positive clusters) per patient are designed for evaluation metrics. In lesion level TPR, a polyp is considered detected if the CAD - Lower GI bounding box adequately overlaps with the ground truth in at least one frame. The FPs per patient represent the average number of false positives per patient. It specifically refers to false-positive events detected on negative frames that contain no ground-truth(GT) polyps The testing
7 / 18
{11}
K252853
Page 8 of 18
below considers repeated marking overlays of the same false positives as a single statistical event, instead of considering only markings in individual frames as a single statistical event. The evaluation results of dynamic video dataset and third-party data are shown in Table 10-12.
Table 10: Evaluation Results for Dynamic Video Dataset (IoU=0.5, persistence>0)
| Items | | Frame Level TPR(%) [95% CI] | Frame Level FPR(%) [95% CI] | Lesion Level TPR(%) [95% CI] | FPs per patient |
| --- | --- | --- | --- | --- | --- |
| Overall | | 65.04 [64.91-65.17] | 0.84 [0.83-0.84] | 100 [100-100] | 9.87 [8.97-10.76] |
| Light Source | WLI | 64.34 [64.19-64.48] | 0.83 [0.82-0.84] | 100 [100-100] | - |
| | NBI | 67.89 [67.61-68.18] | 1.03 [0.97-1.08] | 100 [100-100] | - |
| Image Processor | Olympus CV-1500 | 63.62 [63.38-63.87] | 1.16 [1.14-1.18] | 100 [100-100] | - |
| | Olympus CV-190 | 65.87 [65.61-66.13] | 0.7 [0.68-0.72] | 100 [100-100] | - |
| Overall Frame Level Performance | | True positive: 342,978 True negative: 4,595,256False positive: 38,711 False negative: 184,375(209 videos/ 507 polyps) | | | |
Table 11: Evaluation Results for Third-party Data (IoU=0.1, persistence>0)
| Items | | Frame Level TPR(%) [95% CI] | Frame Level FPR(%) [95% CI] | Lesion Level TPR(%) [95% CI] | FPs per patient |
| --- | --- | --- | --- | --- | --- |
| Overall | | 60.3[60.09-60.51] | 1.77[1.75-1.79] | 100[100-100] | 61.30[50.12-72.48] |
| Endoscope Brands | Olympus | 60.93[60.72-61.15] | 1.82[1.80-1.84] | 100[100-100] | - |
| | Fujifilm | 51.41[50.58-52.24] | 1.25[1.20-1.30] | 100[100-100] | - |
| Overall Frame Level Performance | | True positive:126,429 True negative: 2,145,564False positive: 38,687 False negative: 83,236(60 videos/ 132 polyps) | | | |
Figures 1-2 showed ROC curve with confidence as threshold for the dynamic video dataset and the third-party data. The AUC values of ROC curves were respectively 0.8715(dynamic video dataset) and 0.8297(third-party data). The results showed that the CAD - Lower GI has high accuracy.

Figure 1: ROC Curve for Dynamic Video Dataset
8 / 18
{12}
K252853
Page 9 of 18

Figure 2: ROC Curve for Third-party Data
Figures 3-4 showed FROC curve with confidence as threshold for the dynamic video dataset and the third-party data.

Figure 3: FROC Curve with Confidence as Threshold (dynamic video dataset)

Figure 4: FROC Curve with Confidence as Threshold (third-party data)
9 / 18
{13}
K252853
Page 10 of 18
Figures 5–6 and Tables 12–13 show the FROC curves and supporting data using IoU as the threshold for the dynamic video dataset and third-party dataset.
The results show that FPC remained unchanged across the evaluated IoU thresholds, while lesion-level sensitivity decreased at higher IoU thresholds due to the more stringent localization criterion.

Figure 5: FROC Curve with IoU as Threshold (dynamic video dataset)
Table 12: FROC Data source table with IoU as Threshold (dynamic video dataset)
| IoU | FPC | 95% CI-L | 95% CI-U | Sensitivity per Lesion | 95% CI-L | 95% CI-U |
| --- | --- | --- | --- | --- | --- | --- |
| 0.1 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| 0.2 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| 0.3 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| 0.4 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| 0.5 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| 0.6 | 9.87 | 8.97 | 10.76 | 99.80% | 99.42% | 100.00% |
| 0.7 | 9.87 | 8.97 | 10.76 | 99.80% | 99.42% | 100.00% |
| 0.8 | 9.87 | 8.97 | 10.76 | 99.61% | 99.06% | 100.00% |
| 0.9 | 9.87 | 8.97 | 10.76 | 95.66% | 93.89% | 97.43% |

10 / 18
{14}
K252853
Page 11 of 18
Figure 6: FROC Curve with IoU as Threshold (third-party data)
Table 13: FROC Data source table with IoU as Threshold (third-party data)
| IoU | FPC | 95% CI-L | 95% CI-U | Sensitivity per Lesion | 95% CI-L | 95% CI-U |
| --- | --- | --- | --- | --- | --- | --- |
| 0.1 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.2 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.3 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.4 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.5 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.6 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.7 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| 0.8 | 61.30 | 50.12 | 72.48 | 98.48% | 96.40% | 100.00% |
| 0.9 | 61.30 | 50.12 | 72.48 | 79.55% | 72.66% | 86.43% |
Figures7-8 and Table14-15 showed FROC curve and data source with detection persistence in time(the duration of a time a mark persists on the same target) as threshold for the dynamic video dataset and the third-party data.

Figure 7: FROC Curve with Detection Persistence in Time as Threshold (dynamic video dataset)
Table 14: FROC Data source table with Detection Persistence in Time as Threshold (dynamic video dataset)
| Detection persistence in time(ms) | FPC | 95% CI-L | 95% CI-U | Sensitivity per Lesion | 95% CI-L | 95% CI-U |
| --- | --- | --- | --- | --- | --- | --- |
| >0 | 9.87 | 8.97 | 10.76 | 100.00% | 100.00% | 100.00% |
| >100 | 4.53 | 4.01 | 5.04 | 99.41% | 98.74% | 100.00% |
| >200 | 3.25 | 2.87 | 3.63 | 98.62% | 97.60% | 99.64% |
| >300 | 2.63 | 2.30 | 2.96 | 97.63% | 96.31% | 98.96% |
| >400 | 2.25 | 1.95 | 2.55 | 96.25% | 94.60% | 97.91% |
| >500 | 1.99 | 1.71 | 2.26 | 94.87% | 92.95% | 96.79% |
| >1000 | 1.16 | 0.98 | 1.35 | 91.72% | 89.32% | 94.12% |
| >1500 | 0.74 | 0.60 | 0.88 | 87.77% | 84.92% | 90.62% |
| >2000 | 0.56 | 0.45 | 0.67 | 85.21% | 82.12% | 88.30% |
| >10000 | 0.06 | 0.03 | 0.09 | 57.40% | 53.09% | 61.70% |
| >20000 | 0.02 | 0.00 | 0.04 | 37.28% | 33.07% | 41.49% |
| >100000 | 0 | 0 | 0 | 1.97% | 0.76% | 3.18% |
11 / 18
{15}
K252853
Page 12 of 18

Figure 8: FROC Curve with Detection Persistence in Time as Threshold(third-party data)
Table 15: FROC Data source table with Detection Persistence in Time as Threshold(third-party data)
| Detection persistence in time(ms) | FPC | 95% CI-L | 95% CI-U | Sensitivity per Lesion | 95% CI-L | 95% CI-U |
| --- | --- | --- | --- | --- | --- | --- |
| >0 | 61.30 | 50.12 | 72.48 | 100.00% | 100.00% | 100.00% |
| >100 | 26.15 | 20.97 | 31.33 | 100.00% | 100.00% | 100.00% |
| >200 | 15.77 | 12.52 | 19.01 | 100.00% | 100.00% | 100.00% |
| >300 | 12.22 | 9.47 | 14.97 | 100.00% | 100.00% | 100.00% |
| >400 | 9.48 | 7.38 | 11.58 | 100.00% | 100.00% | 100.00% |
| >500 | 8.02 | 6.20 | 9.83 | 100.00% | 100.00% | 100.00% |
| >1000 | 4.50 | 3.43 | 5.57 | 99.24% | 97.76% | 100.00% |
| >1500 | 3.12 | 2.30 | 3.93 | 98.48% | 96.40% | 100.00% |
| >2000 | 2.17 | 1.56 | 2.77 | 97.73% | 95.18% | 100.00% |
| >10000 | 0.32 | 0.15 | 0.49 | 86.36% | 80.51% | 92.22% |
| >20000 | 0.07 | 0.00 | 0.13 | 68.94% | 61.05% | 76.83% |
| >100000 | 0 | 0 | 0 | 3.03% | 0.11% | 5.95% |
An activation-time criterion consistent with the predicate device was incorporated into the TP definition: only lesions detected within less than 300 ms were classified as TP. Lesions with an activation time of 300 ms or longer were classified as FN for this analysis
Table 16: Detection Performance for Dynamic Video Dataset and Third-Party Data
| Dataset | Mean Activation Time [95% CI] | FPC(False Positives per Case) [95% CI] | Sensitivity [95% CI] |
| --- | --- | --- | --- |
| Dynamic Video Dataset | 109ms [101ms, 118ms] | 9.87[8.97, 10.76] | 89.94%[87.32%, 92.56%] |
| Third-Party Data | 117ms [98ms, 134ms] | 61.30[50.12, 72.48] | 87.88%[82.31T, 93.45%] |
#### - Standalone Performance Conclusions
Based on the above results, CAD - Lower GI met the pre-defined performance criteria in both modes and demonstrated high values across all subgroups. The test results were observed to be as expected and support that the device has similar performance to the predicate device.
12 / 18
{16}
K252853
Page 13 of 18
#### - Special Control Testing
▶ Pixel-level comparison of degradation of image quality due to the device: The maximum color error of image quality degradation caused by device is 0.78, which is only slightly recognizable to human eyes.
▶ Assessment of video delay due to marker annotation: The maximum annotation delay is 38ms.
▶ Assessment of real-time endoscopic video delay due to the device: The maximum real-time video delay is 30ms. The video delay was determined to be acceptable based on the results of the clinical study.
#### - Human Factors
Human factors validation was performed following the FDA Guidance document Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and FDA Staff recommendations. The human factors validation demonstrated that the device functioned as intended, use-related risk has been mitigated, and the CAD - Lower GI system is safe for its intended use.
### IX Clinical Tests
#### - Study Design and Population
A prospective, multicenter, randomized, parallel two-arm controlled trial was conducted at four academic medical centers in China (Peking Union Medical College Hospital, Nanfang Hospital, Huadong Hospital, and the First Affiliated Hospital of Bengbu Medical College) to evaluate the clinical performance of the CAD - Lower GI system. A total of 840 subjects were randomized in a 1:1 ratio via center-stratified block randomization (block size = 4) to either the Experimental arm (colonoscopy with real-time CAD - Lower GI assistance) or the Control arm (standard colonoscopy without CAD assistance). To eliminate learning bias, all participating endoscopists were pre-assigned to a fixed single arm for the entire study duration. The study enrolled adult patients ( \( \geqslant \) 18 years) undergoing screening or surveillance colonoscopy. Key exclusion criteria included prior colorectal cancer, known polyposis syndromes, active inflammatory bowel disease, inadequate bowel preparation (Boston Bowel Preparation Scale <6), and withdrawal time <8 minutes.
#### - Reference Standard (Sole Gold Standard)
Histopathology, independently read by site panels of three blinded expert gastrointestinal pathologists with central adjudication of discordant cases by a senior gastrointestinal pathologist, served as the reference standard for efficacy evaluations. Diagnostic histopathology was evaluable for 916 of 923 resected or biopsied lesions: 568 of 571 Experimental lesions and 348 of 352 Control lesions. The seven missing histopathology results were handled using the prespecified multiple-imputation and sensitivity analyses. Evaluable lesions were classified according to the Revised Vienna Classification, and the reference-standard procedures were applied uniformly across study arms.
#### - Analysis Populations
The following analysis populations were defined per ICH E9:
▶ Full Analysis Set (FAS) : 824 subjects (412 Experimental, 412 Control) — all randomized subjects who underwent the allocated colonoscopy procedure.
▶ Per-Protocol Set (PPS) : 798 subjects (399 per arm) — FAS subjects without major protocol deviations.
▶ Safety Set (SS) : 840 subjects (420 per arm) — all randomized subjects who underwent the procedure.
▶ NBI Sub-population : 477 subjects (245 Experimental, 232 Control) — subjects in whom Narrow Band Imaging (NBI) mode was utilized during the procedure.
#### - Demographic and Baseline Characteristics (FAS, n=824)
The randomization achieved successful balance across arms, with all standardized differences \(< 0.10\). The mean age was 55.5 years (SD 11.9); \(52.4\%\) were male; \(100\%\) were Asian (Han Chinese); mean BMI was \(25.7~\mathrm{kg / m^2}\).
13 / 18
{17}
K252853
Page 14 of 18
Indication distribution was 64.4% screening and 35.6% surveillance. Image processor distribution was 58.5% Olympus CV-190 and 41.5% Olympus CV-1500, balanced across arms. Other baseline characteristics (ASA class, smoking status, family history of colorectal cancer, medication use, and sedation type) were also well balanced.
#### - Procedure Quality Metrics
Procedure quality met or exceeded ASGE/ACG 2024 benchmarks in both arms and was balanced, supporting the internal validity of the comparison:
▶ Cecal intubation rate: 99.3% (Experimental) vs. 99.0% (Control).
▶ Mean withdrawal time: 11.2 min (SD 2.1) vs. 10.8 min (SD 2.0); 98.8% vs. 99.0% of procedures met the \( \geqslant \) 8 min standard.
▶ Mean BBPS score: 7.7 (SD 1.1) vs. 7.6 (SD 1.0); adequate preparation (BBPS \( \geqslant \) 6) rate: 99.3% in both arms.
▶ Rectal retroflexion performed: 96.1% vs. 95.4%.
#### - Efficacy Endpoints and Definitions
The study employed an Intersection-Union Test (IUT) framework with two co-primary endpoints, both of which were required to independently succeed for trial success (familywise error rate controlled at alpha = 0.05; no multiplicity adjustment per FDA Multiple Endpoints Guidance, 2022).
#### - Co-Primary Endpoint 1 — Adenoma Detection Rate (ADR, Case-Level)
Defined as the proportion of patients with at least one histopathologically confirmed adenoma (Vienna classification categories 3 or 4.1). Superiority design with a prespecified margin of +5 percentage points (i.e., lower bound of the 95% confidence interval for the difference must exceed +5 pp). Statistical analysis used the Cochran-Mantel-Haenszel (CMH) method stratified by study site.
#### - Co-Primary Endpoint 2 — Positive Predictive Agreement (PPA, Lesion-Level)
Defined as the proportion of clinically significant lesions, including adenomas, adenocarcinomas, sessile serrated lesions (SSLs), and proximal hyperplastic polyps \(> = 5\mathrm{mm}\), among all biopsied or resected lesions. Non-inferiority was assessed against a -10 percentage-point margin using patient-cluster bootstrap confidence intervals \((\mathrm{B} = 2,000)\). Missing lesion histopathology was addressed by multiple imputation \((\mathrm{M} = 5)\), with complete-case and actual-arm worst-case analyses used to assess robustness.
#### - Secondary Endpoint — Adenomas Per Colonoscopy (APC)
Defined as the total number of histopathologically confirmed adenomas divided by the total number of procedures. Superiority design with a margin of 0 (i.e., lower bound of the \(95\%\) confidence interval for the difference must exceed 0). Statistical analysis used a mixed-effects Poisson regression model with study site as a random effect; negative binomial was prespecified as a sensitivity analysis if overdispersion was observed (Pearson \(\mathrm{chi}^2 / \mathrm{df} > 1.5\)).
#### - Exploratory Endpoints (not formally hypothesis-tested)
Additional exploratory endpoints included Polyp Detection Rate (PDR), Neoplastic Lesion Detection Rate, Serrated Lesion Detection Rate, and false positive rates, all based on histopathological confirmation.
#### - Co-Primary Endpoint Results (FAS, n = 824)
#### ADR (Superiority) :
▶ Experimental: 51.2% (211/412 patients with \( \geqslant \) 1 adenoma).
▶ Control: 36.4% (150/412 patients with \( \geqslant \) 1 adenoma).
▶ CMH-adjusted difference: +14.8 percentage points (95% CI: 8.1, 21.5).
The lower bound of the 95% CI (8.1 pp) exceeded the prespecified superiority margin of +5.0 pp →
14 / 18
{18}
K252853
Page 15 of 18
#### SUPERIORITY ACHIEVED.
#### PPA (Non-inferiority) :
▶ Experimental: 81.2% (461 clinically significant lesions out of 568 total biopsied/resected lesions).
▶ Control: 82.5% (287 out of 348).
▶ Experimental-minus-Control difference: -1.3 percentage points.
▶ Cluster-bootstrap 95% CI (percentile): -5.8 to 3.2.
The lower confidence limit (-5.8 pp) exceeded the prespecified non-inferiority margin of -10.0 pp -> NON-INFERIORITY ACHIEVED.
Sensitivity (missing data) analyses were consistent: the multiple imputation (M=5) result was difference -1.4 pp (95% CI -6.0, 3.3); the actual-arm worst-case difference was -1.9 pp (95% CI -7.1, 3.1).
#### - Secondary Endpoint Results (FAS, n = 824)
#### APC (Superiority) :
▶ Experimental: 1.027 adenomas per procedure (423 total adenomas / 412 procedures).
▶ Control: 0.609 (251 total adenomas / 412 procedures).
▶ Mixed-Poisson difference: +0.418 (95% CI: 0.26, 0.57).
The lower bound of the 95% CI (0.26) exceeded the superiority margin of 0 → SUPERIORITY ACHIEVED.
Overdispersion parameter (chi \( ^{2} \) /df) was 1.21 (Experimental) and 1.14 (Control), both below the prespecified threshold of 1.5, confirming Poisson model appropriateness.
#### - Lesion Histopathology Composition
The histopathological breakdown of the 916 histopathology-evaluable biopsied/resected lesions (568 Experimental and 348 Control) was as follows. Seven additional lesions had missing histopathology results (3 Experimental and 4 Control).
Table 16: PPA: Histopathology-Evaluable Lesion Composition by Arm (FAS)
| Lesion Category | Exp (n=568) | Ctrl (n=348) |
| --- | --- | --- |
| Adenoma (Vienna 3, 4.1) | 423 (74.5%) | 251 (72.1%) |
| Adenocarcinoma (Vienna 4.2-5) | 4 (0.7%) | 3 (0.9%) |
| Sessile serrated lesion (SSL) | 21 (3.7%) | 17 (4.9%) |
| Proximal HP >=5 mm | 13 (2.3%) | 16 (4.6%) |
| Clinically significant subtotal | 461 (81.2%) | 287 (82.5%) |
| Hyperplastic (distal/<5 mm proximal) | 79 (13.9%) | 44 (12.6%) |
| Normal / inflammatory | 21 (3.7%) | 14 (4.0%) |
| Other benign | 7 (1.2%) | 3 (0.9%) |
#### - Characteristics of Detected Adenomas
The incremental adenomas detected with CAD assistance were predominantly small, flat, and proximal — the categories most prone to being missed during standard white-light colonoscopy:
▶ Size <5 mm: Experimental 65.5% vs. Control 59.4% (+128 additional adenomas).
▶ Size 5-10 mm: 26.5% vs. 29.1% (+39).
▶ Size >10 mm: 8.0% vs. 11.6% (+5).
▶ Paris IIa (flat-elevated): 171 vs. 95 (+76).
▶ Proximal location: 54.6% vs. 51.0% (+103).
▶ Paris IIb/Ic (flat/depressed): 41 vs. 17 (+24).
15 / 18
{19}
K252853
Page 16 of 18
#### - CAD Detection Attribution (Experimental Arm Only)
Within the Experimental arm, detection attribution was prospectively recorded per lesion:
▶ Identified by both endoscopist and CAD: 295 adenomas (69.7%).
▶ CAD-first (endoscopist had not yet flagged the lesion): 89 adenomas (21.0%) — representing the incremental detection mechanism.
▶ Endoscopist-only (CAD did not flag): 39 adenomas (9.2%).
The arm-level incremental effect was +172 additional adenomas detected in the Experimental arm versus Control (+68% relative increase).
#### - Performance by Compatible Image Processor (CV-190 vs. CV-1500)
Per the Indications for Use claim for compatibility with both Olympus processors, a prespecified subgroup analysis was performed:
Table 17: Co-Primary and Secondary Endpoints by Image Processor (FAS)
| Processor / Endpoint | Exp | Ctrl | Difference | 95% CI / note |
| --- | --- | --- | --- | --- |
| CV-190 (n=482: 239/243) | | | | |
| ADR | 51.0% | 36.2% | +14.8 pp | (7.2, 22.4) superior |
| PPA (lesion-level) | 81.2% | 82.8% | -1.6 pp | (-7.3, 4.1) non-inferior |
| APC | 1.02 | 0.61 | +0.41 | (0.22, 0.60) superior |
| CV-1500 (n=342: 173/169) | | | | |
| ADR | 51.4% | 36.7% | +14.7 pp | (5.6, 23.8) superior |
| PPA (lesion-level) | 81.1% | 82.1% | -1.0 pp | (-7.9, 5.9) non-inferior |
| APC | 1.03 | 0.61 | +0.42 | (0.19, 0.65) superior |
| Processor-by-treatment interaction (ADR) | | | | p = 0.99 (no difference) |
Processor-by-treatment interaction for ADR: p = 0.99, confirming consistent device performance across both compatible processors.
#### - NBI Sub-study Results
Narrow Band Imaging (NBI) was used in a prespecified NBI-exposed subpopulation of 477 subjects in the Full Analysis Set, including 245 subjects in the Experimental arm and 232 subjects in the Control arm. NBI was activated when a suspected lesion was identified during withdrawal. The results in the NBI-exposed subpopulation were directionally consistent with the overall FAS results, as summarized below:
Table 18: NBI Sub-Study Results
| Endpoint (NBI sub-pop.) | Exp (n=245) | Ctrl (n=232) | Diff (95% CI) |
| --- | --- | --- | --- |
| ADR | 54.3% | 38.4% | +15.9 (7.9, 23.9) |
| PPA (lesion-level) | 80.7% | 82.1% | -1.4 (-6.9, 4.1) |
| APC | 1.09 | 0.66 | +0.43 (0.21, 0.65) |
These results provide supportive clinical evidence for the use of CAD - Lower GI during colonoscopy procedures in which NBI is used. Because NBI use was clinically triggered rather than randomized by imaging modality, the NBI results are interpreted as supportive subgroup evidence.
#### - Pre-specified Subgroup Analyses (ADR)
Treatment effects were directionally consistent across the evaluable prespecified subgroups. No significant
16 / 18
{20}
K252853
Page 17 of 18
treatment-by-subgroup interaction was observed for the evaluable factors (all interaction p > 0.10), and no individual clinical site was identified as an outlier (Breslow-Day p = 0.78).
#### - Sensitivity Analyses
All primary and secondary endpoint conclusions were robust across multiple sensitivity analyses:
Per-Protocol Set (PPS, n = 798): ADR difference +14.3(95% CI: 7.4, 21.2); PPA difference -1.3(95% CI: -5.9, 3.3); APC difference +0.41 (95% CI: 0.25, 0.57).
PPA missing-histopathology analyses (7 lesions: 3 Experimental and 4 Control): the prespecified M=5 result was -1.4 pp (95% CI -6.0, 3.3), and the completed actual-arm worst-case result was -1.9 pp (95% CI -7.1, 3.1); both confirmed non-inferiority.
▶ ADR missing-case assessment: all 412 FAS patients in each arm had a determinable ADR outcome; no ADR outcomes were imputed. CTR Table 5 is the observed FAS analysis, not an all-randomized worst-case analysis. An extreme all-randomized sensitivity analysis gave a site-stratified CMH difference of 12.6 pp (95% CI 6.0, 19.3), maintaining superiority.
Additional supportive results were consistent: APC complete-case difference +0.42 (95% CI 0.26, 0.58), and complete-case PPA BCa bootstrap difference -1.3 pp (95% CI -5.7, 3.1).
#### - Safety Evaluation
All 840 randomized subjects were included in the Safety Set. Day-30 follow-up was completed for 836/840 subjects (99.5%).
#### - Adverse Events (AEs)
▶ Any AE: Experimental 5.7% (24/420) vs. Control 5.2% (22/420) — low and balanced.
▶ Most common AEs: abdominal discomfort/cramping, post-polypectomy bleeding (minor, all managed conservatively or with clipping), nausea/vomiting, and vasovagal reactions. All AEs were consistent with routine colonoscopy-related events.
#### - Serious Adverse Events (SAEs)
▶ No deaths occurred during the study.
Five SAEs were reported (3 Experimental, 2 Control), all CTCAE Grade 3 and all resolved without sequelae. None were assessed as device-related; all were attributed to routine colonoscopy complications (post-polypectomy bleeding, sedation-related vasovagal syncope, or cardiac arrhythmia).
#### - Device Deficiencies
Five minor, transient device deficiencies were recorded in the Experimental arm (e.g., transient false-positive bursts lasting \(\sim 1.5\) seconds, momentary frame freezes lasting \(\sim 2\) seconds, bounding-box drift on repositioning). All events self-resolved within 3 seconds without any harm to subjects. No algorithm modifications were made during the trial.
Safety Conclusion: CAD - Lower GI demonstrated a favorable safety profile with no device-related AEs or SAEs and no evidence of increased procedural risk.
#### - Clinical Conclusions
In this histopathology-referenced randomized controlled trial, CAD - Lower GI met all pre-specified co-primary and secondary efficacy endpoints:
17 / 18
{21}
K252853
Page 18 of 18
▶ ADR superiority was demonstrated with a +14.8 pp absolute improvement (95% CI 8.1, 21.5), exceeding the +5 pp margin. No FAS ADR outcomes were missing or imputed; the all-randomized extreme sensitivity analysis was also superior at +12.6 pp (95% CI 6.0, 19.3).
▶ PPA non-inferiority was demonstrated in the prespecified M=5 analysis with a difference of -1.4 pp (95% CI -6.0, 3.3). The complete-case result [-1.3 pp (95% CI -5.8, 3.2)] and actual-arm worst-case result [-1.9 pp (95% CI -7.1, 3.1)] were consistent with this conclusion.
▶ APC superiority was demonstrated with a difference of +0.418 (95% CI: 0.26, 0.57), exceeding the 0 margin.
The primary and secondary efficacy conclusions were consistent across both compatible Olympus image processors, CV-190 and CV-1500. Results in the NBI-exposed subpopulation were directionally consistent with the overall FAS results and provide supportive evidence for use of the device during procedures in which NBI is used.
The safety profile was favorable, with no device-related adverse events or serious adverse events and only minor transient device deficiencies without subject harm.
Collectively, these clinical data support the conclusion that CAD - Lower GI is clinically effective and safe for its intended use as a real-time computer-assisted detection tool to aid qualified endoscopists in detecting suspected colonic mucosal lesions during colorectal endoscopy under both white-light and NBI imaging modalities, with histopathology as the definitive reference standard.
## X Conclusion
Based on the intended use, technological characteristics, performance results, and comparison to the predicate, the subject CAD - Lower GI has been shown to be substantially equivalent to the predicate device identified in this submission and does not present any new issues of safety or effectiveness.
18 / 18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.