The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment. The AffloVest may be used for the pediatric population (≥6 months of age) to geriatric population.
Device Story
AffloVest is a high-frequency chest wall oscillation (HFCWO) system; utilizes synthetic polymer vest with integrated oscillating vibratory motors in four zones. Powered by rechargeable lithium-ion battery; controlled by wired controller to adjust oscillation intensity, frequency, and mode (Drainage, Vibration, Percussion). Used in home, hospital, or clinic settings; operated by patient or caregiver. Controller connects via Bluetooth/Wi-Fi to cloud for usage data upload and OTA updates; optional mobile app retrieves usage data. Device promotes airway clearance by mobilizing bronchial secretions; provides non-invasive therapy. Benefits include improved bronchial drainage and secretion mobilization. Does not contact skin directly; worn over clothing.
Clinical Evidence
Bench testing only. Performance verification included weight, battery, cable life, motor speed, comfort, distribution/environmental, keypad life, and motor life testing. Electrical safety (IEC 60601-1 series) and electromagnetic compatibility (IEC 60601-1-2, 60601-4-1) testing performed. Software and cybersecurity verification and validation completed.
Technological Characteristics
Materials: urethane-coated nylon, nylon plastic, DC motors. Sensing/Actuation: electromechanical actuators (vibratory motors). Energy: rechargeable lithium-ion battery, A/C power supply. Connectivity: Bluetooth/Wi-Fi for data/OTA updates. Form factor: wearable vest (16"-75" sizes). Modes: Drainage (sequential), Vibration (continuous), Percussion (pulsed). Frequencies: 24Hz, 26Hz, 32Hz.
Indications for Use
Indicated for pediatric (≥6 months) to geriatric patients requiring airway clearance and bronchial drainage via external thoracic manipulation.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
Predicate Devices
International Biophysics Corporation AffloVest (K122480)
Reference Devices
Hill-rom's Monarch™ Airway Clearance System (K163378)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 30, 2026
Tactile Medical
% Jessica Buell
Regulatory, Quality, and Compliance Consultant
DuVal & Associates, P.A.
825 Nicollet Mall
Suite 1820, Medical Arts Building
Minneapolis, Minnesota 55402
Re: K252849
Trade/Device Name: AffloVest® Mobile Airway Clearance Therapy
Regulation Number: 21 CFR 868.5665
Regulation Name: Powered Percussor
Regulatory Class: Class II
Product Code: BYI
Dated: September 4, 2025
Received: March 30, 2026
Dear Jessica Buell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252849 - Jessica Buell
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K252849 - Jessica Buell
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252849 | ? |
| Please provide the device trade name(s). | | ? |
| AffloVest® Mobile Airway Clearance Therapy | | |
| Please provide your Indications for Use below. | | ? |
| The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment.
The AffloVest may be used for the pediatric population (≥6 months of age) to geriatric population. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Tactile MEDICAL
HEALING RIGHT AT HOME
# K252849 - 510(k) Summary
## 510(k) Owner:
Tactile Medical
3701 Wayzata Blvd, Suite 300
Minneapolis, MN 55416
Contact: Sunday Hoy
612.355.5121
Date prepared: April 28, 2026
## Device Name:
Trade Name: AffloVest® Mobile Airway Clearance Therapy
Common Name: Chest-percussion airway secretion-clearing system
Classification Name: Powered Percussor
Regulation: 21 CFR §868.5665
Regulatory
Classification: 2
Product Code: BYI
## Predicate Device:
**Predicate:** International Biophysics Corporation AffloVest (K122480)
**Reference 1:** Hill-rom's Monarch™ Airway Clearance System (K163378)
**Reference 2:** Baxter's Vest APX System (PVAPX1) (K233441)
## Device Description
The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment.
The AffloVest is a high frequency chest wall oscillation (HFCWO) Vibratory Vest comprised of a vest of synthetic polymer fabric with installed oscillating vibratory motors organized into four zones. It has buckles and straps on the front and over each shoulder that are used to provide the wearer with a snug and evenly distributed fit. Additionally, there is an included charger for the batteries and a controller that is wired into the vest to adjust the intensity of the oscillation on the patient as necessary and as prescribed by a physician.
The AffloVest is constructed of commonly available materials, including urethane-coated nylon, nylon plastic components, DC motors and wires. The vest is powered with a rechargeable lithium-ion battery, which is charged with an A/C power supply. When the vest is powered ON, low voltage DC energy is provided to the vibratory motors, which are in contact with the patient's chest and back, promoting airway clearance.
Tactile Medical
CONFIDENTIAL
Page 1 of 5
{5}
The AffloVest is intended to be worn over clothing and is made of commonly used fabrics. The product is not intended to directly contact the skin. The vest is the only applied part in the AffloVest system. None of the other parts including the controller and battery that might come into contact with the patient are considered applied parts.
The controller is able to connect to the cloud application over Bluetooth or Wi-Fi to upload usage data and receive over-the-air (OTA) updates. The mobile application connects to the controller via Bluetooth or Wi-Fi to retrieve usage data from the controller. Therapy delivery is not supported by the mobile application. The mobile application is treated as an optional accessory for users that want to access more information about their treatments and other usage data.
## Indications for Use
The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment.
The AffloVest may be used for the pediatric population (≥6 months of age) to geriatric population.
## Comparison of Technological Characteristics with the Predicate Devices
The Subject device, the AffloVest Mobile Airway Clearance Therapy, has the same intended use, indications for use, regulatory classification and product code as the Predicate device. The Subject device differs from the Predicate in the connectivity (Wi-Fi and Bluetooth), which technological characteristic is shared with Reference Device 1. The Subject device also differs from the Predicate in the range of garment sizes (16"-75"), which technological characteristic is shared with Reference Device 2. Testing has demonstrated that these minor differences in technological characteristics do not create a new intended use or raise different questions of safety or effectiveness of the Subject device compared to the Predicate device. Thus, the Subject device meets FDA's criteria for substantial equivalence to the Predicate device. See Table 1 below for a more detailed comparison of the Subject and Predicate devices.
Table 1: Comparison of Technological Characteristics of Subject and Predicate Devices
| Product Feature | Subject Device (K252849)
AffloVest Mobile Airway Clearance Therapy | Predicate Device (K122480)
International Biophysics Corporation AffloVest | Substantial Equivalence Determination |
| --- | --- | --- | --- |
| Regulatory Classification | 21 CFR 868.5665
(Powered Percussor) | Same | Substantially Equivalent |
| Product Code | BYI
(Percussor, Powered-Electric) | Same | Substantially Equivalent |
Tactile Medical
CONFIDENTIAL
{6}
Tactile Medical
CONFIDENTIAL
Page 3 of 5
| Product Feature | Subject Device (K252849)
AffloVest Mobile Airway Clearance Therapy | Predictae Device (K122480)
International Biophysics Corporation AffloVest | Substantial Equivalence Determination |
| --- | --- | --- | --- |
| **Intended Use** | The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment. It can be used in a home, hospital, or clinic setting. Not intended to be used in the emergency medical services (EMS) environment. | The International Biophysics Corporation AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment. | Substantially Equivalent |
| **Indications for Use** | The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment.
The AffloVest may be used for the pediatric population (≥6 months of age) to geriatric population. | The International Biophysics Corporation AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external manipulation of the thorax is the physician's choice of treatment. | Substantially Equivalent
Same as Predicate, with additional clarification of patient population. |
| **Technology of Oscillations** | Electromechanical Actuators | Same | Substantially Equivalent |
| **Treatment Modes Available** | (1) Treatment type (Drainage, Vibration, or Percussion,
(2) Intensity (low/24Hz, medium/26Hz, and high/32Hz), and
(3) Duration (5-45 minutes)
Percussion: all motors operate in a pulsed fashion
Vibration: All motors operate continuously
Drainage: All motors operate in a preprogrammed sequential fashion | (1) Treatment type (Drainage, Vibration, or Percussion)
(2) Intensity (low/22Hz, medium/25Hz, and high/27Hz).
Percussion: all motors operate in a pulsed fashion
Vibration: All motors operate continuously
Drainage: All motors operate in a preprogrammed sequential fashion | Substantially Equivalent
Although the proposed AffloVest has higher frequencies than the predicate and reference devices, similar and even lower forces result during treatment. This is due to the difference in design of the motors. Since the applied forces are similar or lower, there no new questions for safety or effectiveness are raised. |
{7}
| Product Feature | Subject Device (K252849)
AffloVest Mobile Airway Clearance Therapy | Predictae Device (K122480)
International Biophysics Corporation AffloVest | Substantial Equivalence Determination |
| --- | --- | --- | --- |
| Therapy Peak Force Across All Settings | Peak Force (N): 5.7-12.2 (36mm motors)
3.9-4.5 (24mm motors) | Peak Force (N): 4.5 to 7.2 | Substantially Equivalent
The Subject Device's peak force range is slightly broader than the Predicate Device's force range. |
| Garment Sizes | 16-75 inches | 33-53 inches | Substantially Equivalent
The Subject Device's vest sizes are a broader range than the Predicate Device to match availability from competitive devices. Performance testing equivalent to what was done for the Predicate was repeated for the Subject Device and raised no new questions of safety or effectiveness. |
## Performance Data
The AffloVest completed performance testing, including:
- Weight Verification
- Battery Verification
- Carrying Case Life Verification
- Battery Cable Life Testing
- Controller Cable Life Testing
- Vest Functional Testing
- Motor Speed Verification
- Comfort Testing
- Distribution and Environmental Testing
- Keypad Button Life Testing
- Motor Life Testing
- Electrical Safety Testing per IEC 60601-1, 60601-1-6, and 60601-1-11
- Electromagnetic Compatibility Testing per IEC 60601-1-2 and 60601-4-1
- Software Verification and Validation
- Cybersecurity Verification and Validation
## Conclusion
The AffloVest Mobile Airway Clearance Therapy continues to meet all its established acceptance criteria that were based upon the performance of the Predicate device. This demonstrates that the device function is as safe, as effective, and performs as well as the Predicate device.
Tactile Medical
CONFIDENTIAL
Page 4 of 5
{8}
A determination of substantial equivalence requires: 1) the same intended use, and 2) either the same technological characteristics or differences in technological characteristics that do not raise different questions of safety or effectiveness. Both of these conditions have been met and support a determination of substantial equivalence.
Tactile Medical
CONFIDENTIAL
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.