K252840 · Global Instrumentation, LLC · MHX · May 29, 2026 · Cardiovascular
Device Facts
Record ID
K252840
Device Name
M12 Telemetry System
Applicant
Global Instrumentation, LLC
Product Code
MHX · Cardiovascular
Decision Date
May 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The M12 Telemetry System is intended for use by trained clinical research healthcare professionals for the purpose of centralized monitoring of adult and adolescent patients (11 to 17 years) participating in clinical studies within clinical research facilities. Centralized monitoring involves the display and management of data from networked patient telemetry transmitters including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. The M12 Telemetry System with resting 12-lead ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and adolescent patients when connected to a transmitter with 12-Lead ECG monitoring enabled. The M12RT Telemetry Transmitter is intended for use with the M12A Telemetry Central Station to monitor ECG on ambulatory and non-ambulatory adult and adolescent patients using wireless communication over the M12 telemetry network. The M12RT Telemetry Transmitter can deliver ECG data using a 10-lead cable. The M12 Telemetry System is not intended for general patient monitoring (e.g., hospitals, intensive care units, operating room, and other healthcare facilities except for clinical research).
Device Story
M12 Telemetry System provides centralized monitoring for clinical research. System components: M12A Central Station, Review & Admin Station, Telemetry Server, and M12RT Telemetry Transmitter. Transmitter captures ECG via 10-lead cable; transmits wirelessly to central station. Central station displays real-time ECG, performs arrhythmia analysis on three selectable channels, manages alarms, and stores up to 96 hours of data. Operated by trained clinical research professionals in research facilities. Output includes visual/audible alarms, waveform displays, and printed reports. Enables continuous monitoring of ambulatory/non-ambulatory patients; supports data review for research analysis. Benefits include centralized oversight of cardiac physiological data during clinical trials.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless performance, and device-specific standards (IEC 60601-2-27, ANSI/AAMI EC53, ANSI/AAMI EC57, EN 60601-1-8, IEC 60601-2-25). Biocompatibility testing (ISO 10933-5, -10, -23) and software validation (IEC 62304) were performed. No clinical data provided.
Technological Characteristics
System includes M12RT transmitter (10-lead cable) and M12A central station. Wireless network connectivity. Battery: AA (LR6) disposable with internal lithium polymer for swap. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62133, UL 1642, AAMI TIR 69, IEC 60601-2-27, ANSI/AAMI EC53, ANSI/AAMI EC57, EN 60601-1-8, IEC 60601-2-25, IEC 62304, ISO 10933-5/10/23, ASTM D4169-22, IEC 60601-1-6, IEC 62366-1.
Indications for Use
Indicated for centralized monitoring of adult and adolescent patients (11-17 years) participating in clinical studies within clinical research facilities. Includes resting 12-lead ECG production and interpretation. Not for general patient monitoring in hospitals, ICUs, or ORs.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Infinity Central Station Wide, Infinity M300, Infinity M300+ (K2314774)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
May 29, 2026
Global Instrumentation, LLC
Doug Linquest
Vice President Marketing & Business Development
7010 Fly Road STE 5
East Syracuse, New York 13057
Re: K252840
Trade/Device Name: M12 Telemetry System
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: Class II
Product Code: MHX
Dated: April 30, 2026
Received: April 30, 2026
Dear Doug Linquest:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252840 – Doug Linquest
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K252840 – Doug Linquest
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252840 | |
| Device Name M12 Telemetry System | |
| Indications for Use (Describe) The M12 Telemetry System is intended for use by trained clinical research healthcare professionals for the purpose of centralized monitoring of adult and adolescent patients (11 to 17 years) participating in clinical studies within clinical research facilities. Centralized monitoring involves the display and management of data from networked patient telemetry transmitters including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation. The M12 Telemetry System with resting 12-lead ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and adolescent patients when connected to a transmitter with 12-Lead ECG monitoring enabled. The M12RT Telemetry Transmitter is intended for use with the M12A Telemetry Central Station to monitor ECG on ambulatory and non-ambulatory adult and adolescent patients using wireless communication over the M12 telemetry network. The M12RT Telemetry Transmitter can deliver ECG data using a 10-lead cable. The M12 Telemetry System is not intended for general patient monitoring (e.g., hospitals, intensive care units, operating room, and other healthcare facilities except for clinical research). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Global Instrumentation, LLC
M12 Telemetry System
K252840
# 510(k) SUMMARY
Prepared in accordance with 21 CFR 807.92
## 1. Submitter Information
Submitter Name: Global Instrumentation, LLC
Address: 7010 Fly Road STE 5 East Syracuse NY 13057 United States
Contact Person: Doug Linquest; Vice President Marketing & Business Development
Phone Number: 760-960-0869
Email Address: doug.linquest@gi-med.com
Date Prepared: May 29, 2026
## 2. Device Identification
Proprietary Name: M12 Telemetry System
Common/Usual Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Classification Name: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms
Regulation Number: 21 CFR 870.1025
Product Code: MHX
Device Class: Class II
Review Panel: Cardiovascular
## 3. Predicate Device(s)
Predicate Device Name: Infinity Central Station Wide, Infinity M300, Infinity M300+
Manufacturer: Draeger Medical Systems
510(k) Number: K2314774
{5}
Global Instrumentation, LLC
M12 Telemetry System
K252840
# 4. Device Description
The M12 Telemetry System is a central monitoring system capable of real-time display and storage of ECG physiologic patient data and alarm annunciation for networked devices including ambulatory wireless telemetry transmitters. The system is comprised of the M12A Central Station, M12A Review & Admin Station, M12A Telemetry Server, and the M12RT Telemetry Transmitter.
The M12A Central Station can perform telemetry monitoring using the M12RT Telemetry Transmitter integrated with a wireless network infrastructure installed in the clinical research healthcare organization's facility.
The systems intended functions include:
- Real-time 10-wire (12-lead) ECG signal display
- Real-time analysis program runs on three selectable ECG channels
- Automatic monitoring of arrhythmias including display of a single channel waveform for each patient
- Visual and audible alarms of events, arrhythmias, and set parameters with adjustable priorities
- Graphical trend review
- ECG signal data storage with full-disclosure storage and review
Through telemetry, patients using ambulatory transmitters are monitored while moving in a designated area. To guarantee proper signal transmission in various situations, install an antenna network as per facility requirements.
M12A Central Station may be set up in a variety of system configurations to support the acquisition and storage of real-time patient monitoring data.
Optional data storage is available for on or off premise storage. This increases storage capacity as well as allows storage to duplicate data disks when continuous and complete ECG capture is critical. This is valuable for institutions that collect ECG data for research purposes.
Networked printers are used for automatic or manual printouts of 12-lead ECG, ECG rhythm, and reports.
The M12A Central Station provides the means for users to manage patient records with associated patient demographics and assign each to a particular M12R Telemetry Transmitter for obtaining cardiac physiological information. Physiological data obtained from the telemetry devices are viewable and stored continuously up to 96 hours. Alarm monitoring and management, including definition of alarm limits, as well as data trending and reporting are available. Full disclosure supports review of the data on a particular patient including review of 12-lead ECG waveforms and other data and waveforms for obtained parameters.
Page | 2
{6}
Global Instrumentation, LLC
M12 Telemetry System
K252840
# 5. Intended Use / Indications for Use
The M12 Telemetry System is intended for use by trained clinical research healthcare professionals for the purpose of centralized monitoring of adult and adolescent patients (11 to 17 years) participating in clinical studies within clinical research facilities. Centralized monitoring involves the display and management of data from networked patient telemetry transmitters including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.
The M12 Telemetry System with resting 12-lead ECG is intended for the production and interpretation of diagnostic electrocardiograms for adult and adolescent patients when connected to a transmitter with 12-Lead ECG monitoring enabled.
The M12RT Telemetry Transmitter is intended for use with the M12A Telemetry Central Station to monitor ECG on ambulatory and non-ambulatory adult and adolescent patients using wireless communication over the M12 telemetry network.
The M12RT Telemetry Transmitter can deliver ECG data using a 10-lead cable.
The M12 Telemetry System is not intended for general patient monitoring (e.g., hospitals, intensive care units, operating room, and other healthcare facilities except for clinical research).
Page | 3
{7}
Global Instrumentation, LLC
M12 Telemetry System
K252840
# 6. Comparison of Technological Characteristics
The intended use, principles of operation, and technological characteristics are substantially equivalent to the referenced predicate device.
Technological differences between the subject and predicate device include:
Physiological Parameters Monitored
- Subject device includes ECG
- Predicate includes ECG & pulse oximetry (SpO2)
ECG lead configuration options
- Subject device utilizes a 10-leadwire patient cable to provide 12-lead ECG with leads - I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
- Predicate offers leadwire patient cable configurations of 3-wire (leads I, II, III), 5-wire (leads I, II, III, aVL, aVR, aVF, V), 6-wire (I, II, III, aVL, aVR, aVF, V, V+), and 6-wire TrustST for 12L ECG (I, II, III, aVL, aVR, aVF, dV1, V2, dV3, dV4, V5, dV6)
Monitoring capacity
- The subject system supports up to 64 patients on continuous telemetry monitoring at central station
- Predicate device supports up to 32 patients on continuous telemetry monitoring at central station
Battery Type
- Subject device uses AA (LR6) disposable batteries with lithium polymer internal battery to support battery swap
- Predicate uses a custom lithium-ion battery pack
# 7. Performance Testing
The following non-clinical testing was performed to demonstrate the safety and effectiveness of the subject device as compared to the predicate.
Electrical Safety, EMC, Wireless Testing
- IEC 60601-1 Electrical Safety
- IEC 60601-1-2 EMC Testing
- IEC 62133
- UL 1642
- AAMI TIR 69
Page | 4
{8}
Global Instrumentation, LLC M12 Telemetry System
K252840
The technological characteristics of the subject device have been tested for function and performance characterization in accordance with the following standards:
- IEC 60601-2-27
- ANSI/AAMI EC53
- ANSI/AAMI EC57
- EN 60601-1-8
- IEC 60601-2-25
Software Validation
- IEC 62304
Biocompatibility Testing
- Cytotoxicity Testing (MEM Elution) – ISO 10933-5
- Sensitization Testing (Guinea Pig Maximization) – ISO 10933-10
- Irritation Testing (Intracutaneous Irritation) – ISO 10933-23
Packaging and Transit
- ASTM D4169-22
Safety & Essential Performance - Usability
- IEC 60601-1-6
- IEC 62366-1
Conclusion
To establish substantial equivalence of the M12 Telemetry System, Global Instrumentation conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its safety, performance, and accuracy specification, and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.