>2,000 slices (banding artifact reduction test dataset)
—
Indications for Use
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. (Note: For MAGNETOM Flow.Elite, Flow.Neo, and Flow.Rise, the display capability is dependent on optional local coils configured with the system.)
Device Story
MAGNETOM series (Vida, Lumina, Sola, Altea, Viato.Mobile, Flow platform) are MR diagnostic devices. Input: RF signals from patient anatomy via local coils. Operation: System uses Syngo MR XB10 software to perform image reconstruction, spectroscopic analysis, and automated workflows (myExam Companion, Autopilot). Features include AI-based image enhancement (Deep Resolve Boost/Sharp/Swift), motion sensing, and cardiac scan assistance. Output: Cross-sectional images, spectra, and derived physical parameters displayed for physician interpretation. Used in clinical settings by trained healthcare professionals. Benefits: Improved image quality, faster acquisition, and standardized workflows via AI-assisted reconstruction and scan planning. Supports interventional procedures with MR-safe equipment.
Clinical Evidence
No clinical trials conducted. Evidence consists of bench testing, software verification/validation, and sample clinical image quality assessments. AI features (Deep Resolve suite) validated using in-house and collaboration partner datasets (e.g., 29,740 slices for Swift Brain; 27,679 patches for Boost). Metrics included PSNR, SSIM, NMSE, and perceptual loss. Visual inspection by radiologists confirmed diagnostic suitability and artifact reduction compared to conventional imaging.
Technological Characteristics
MR diagnostic system; 1.5T/3T field strengths. Materials: Biocompatible applied parts (ISO 10993-1). Connectivity: DICOM (PS 3.1-3.20). Software: Syngo MR XB10, IEC 62304 compliant. Features: Phased array coils (NEMA MS 9), gradient systems, RF transmitters. AI/ML: Deep learning-based image reconstruction (Deep Resolve) and automated scan planning (myExam Companion).
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities to visualize internal structure and function. Used for diagnostic purposes and interventional procedures with MR-compatible devices. No specific age or gender contraindications stated beyond standard MR safety requirements.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 19, 2025
Siemens Healthcare GmbH
% Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, Pennsylvania 19355
Re: K252838
Trade/Device Name: MAGNETOM Sola; MAGNETOM Altea; MAGNETOM Sola Fit; MAGNETOM Viato.Mobile; MAGNETOM Vida; MAGNETOM Lumina; MAGNETOM Vida Fit; MAGNETOM Flow.Elite; MAGNETOM Flow.Neo; MAGNETOM Flow.Rise
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, LNI, MOS
Dated: November 20, 2025
Received: November 20, 2025
Dear Alina Goodman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252838 - Alina Goodman
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252838 - Alina Goodman
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, PhD
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252838 | |
| Device Name MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise | |
| Indications for Use (Describe) Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile: The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise: The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SIEMENS Healthineers
K252838
510(k) Summary
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
## 1. General Information
**Establishment:** Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Registration Number: 2240869
**Date Prepared:** September 29, 2025
**Manufacturer:** Siemens Healthineers AG
Magnetic Resonance (MR)
Allee am Röthelheimpark 2
91052 Erlangen
Germany
Registration Number: 3002808157
Siemens Shenzhen Magnetic Resonance LTD.
Siemens MRI Center
Hi-Tech Industrial Park (middle)
Gaoxin C. Ave., 2nd
Shenzhen 518057
P.R. CHINA
Registration Number: 3004754211
## 2. Contact Information
Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Phone: +1(317)371-8593
E-mail: alina.goodman@siemens-healthineers.com
## 3. Device Name and Classification
**Device/ Trade name:** MAGNETOM Vida
**Classification Name:** Magnetic Resonance Diagnostic Device (MRDD)
**Classification Panel:** Radiology
**CFR Code:** 21 CFR § 892.1000
**Classification:** II
**Product Code:** Primary: LNH
Secondary: LNI, MOS
**Device/ Trade name:** MAGNETOM Lumina
{5}
SIEMENS Healthineers
510(k) Summary
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Vida Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Sola
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Altea
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Sola Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Viato.Mobile
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Elite
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
{6}
SIEMENS Healthineers
510(k) Summary
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Neo
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Flow.Rise
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
## 4. Legally Marketed Predicate and Reference Device
## 4.1. Predicate Device
Trade name: MAGNETOM Vida
510(k) Number: K231560
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Lumina
510(k) Number: K231560
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola
510(k) Number: K232535
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
{7}
SIEMENS Healthineers
510(k) Summary
Trade name: MAGNETOM Altea
510(k) Number: K232535
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Viato.Mobile
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola Fit
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
4.2. Reference Device
Trade name: MAGNETOM Vida Fit
510(k) Number: K220939
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Flow.Ace
510(k) Number: K250436
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Skyra Fit
510(k) Number: K250443
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
{8}
SIEMENS Healthineers
510(k) Summary
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Free.Max
510(k) Number: K231617
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: MOS
Trade name: syngo.via VB40A
510(k) Number: K191040
Product Code: Primary: LLZ
# 5. Intended Use / Indications for Use
Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
{9}
SIEMENS Healthineers
510(k) Summary
## 6. Device Description
The subject device, MAGNETOM Vida with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
### New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
### Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
### Software
### New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Open Workflow
### Modified features and applications:
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
{10}
SIEMENS Healthineers
510(k) Summary
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- myExam 3D Camera
- BM Contour XL Coil
## Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
## Software
### New Features and Applications:
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- Complex Averaging
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Compressed Sensing Cardiac Cine
- Open Workflow
## Modified Features and Applications:
- GRE_PC
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
{11}
SIEMENS Healthineers
510(k) Summary
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Vida Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- myExam 3D Camera
- Beat Sensor
- BM Contour XL Coil
## Modified Hardware:
- RF Transmitter TBX3 3T (TX Box 3)
- MaRS (Measurement and reconstruction system)
- Host computers
## Software
## New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- myExam Implant Suite
- DANTE blood suppression
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- BioMatrix Motion Sensor
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- GRE_PC
- Open Recon 2.0
- 3D Whole Heart
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- AutoMate Cardiac (Cardiac AI Scan Companion)
- myExam Autopilot Spine
- myExam Autopilot Brain and myExam Autopilot Knee
- Deep Resolve for EPI
- Deep Resolve for HASTE
- Physiologging
{12}
SIEMENS Healthineers
510(k) Summary
- Complex Averaging
- Open Workflow
## Modified features and applications:
- myExam RT Assist workflow improvements
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- myExam Angio Advanced Assist (Test Bolus)
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- BM Contour XL Coil
## Modified Hardware:
- MaRS (Measurement and reconstruction system)
## Software
## New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- 3D Whole Heart
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
{13}
SIEMENS Healthineers
510(k) Summary
## Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Altea with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Altea with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- BM Contour XL Coil
## Modified Hardware:
- MaRS (Measurement and reconstruction system)
## Software
### New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Compressed Sensing Cardiac Cine
- Open Workflow
## Modified features and applications:
- myExam Implant Suite
- GRE_PC
{14}
SIEMENS Healthineers
510(k) Summary
- myExam RT Assist workflow improvements
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- BM Contour XL Coil
## Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
## Software
## New Features and Applications:
- Brachytherapy Support for use with MR conditional applicators
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
## Modified features and applications:
- myExam RT Assist workflow improvements
- myExam Implant Suite
- Deep Resolve Boost for TSE
{15}
SIEMENS Healthineers
510(k) Summary
- “MTC Mode” for SPACE
The subject device, MAGNETOM Viato. Mobile with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato. Mobile with syngo MR XA70A (K250443).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- BM Contour XL Coil
## Modified Hardware:
- MaRS (Measurement and reconstruction system)
- Host computers
## Software
### New Features and Applications:
- CS Vibe
- DANTE blood suppression
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR “Abdomen&Pelvis” mode and SPAIR Breast mode
- RF pulse optimization with VERSE
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- ASNR recommended protocols for imaging of ARIA
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PH5
- EP_SEG_FID_PH5
- myExam Implant Suite
- GRE_PC
- Open Recon 2.0
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Deep Resolve Swift Brain
- myExam Autopilot Spine
- Open Workflow
### Modified features and applications:
- myExam Implant Suite
- Deep Resolve Boost for TSE
- “MTC Mode” for SPACE
With the subject software version, Syngo MR XB10, we are also introducing the following new 1.5T devices, which are part of our MAGNETOM Flow. Platform:
MAGNETOM Flow.Elite
MAGNETOM Flow.Neo
MAGNETOM Flow.Rise
{16}
SIEMENS Healthineers
510(k) Summary
The subject device, MAGNETOM Flow-Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).
A high-level summary of the new and modified hardware and software is provided below:
## New Hardware:
- Magnet
- MREF (Magnet Refrigerator)
- Gradient system
- Gradient Coil
- RF System
- System Cover
- Patient Table
- MaRS (Measurement and Reconstruction System)
- Select&GO Display (TPAN_3G) and Control Panel (CPAN_2G)
- Body Coil
- Head/Neck Coil
- BM Head/Neck Coil (with ComfortSound)
- BM Contour S Coil
- BM Contour M Coil
- BM Contour L Coil
- BM Contour XL Coil
- Foot/Ankle Coil
- BM Spine Coil
- iTx Extremity 18 Flare
- Multi-Index MR-RT Positioning (a part of "RT Pro Edition" marketing bundle) (not available for MAGNETOM Flow.Rise)
## Modified Hardware:
- Gradient Power Amplifier (GPA)
- SAR Monitoring
- In-Vivo Shim
## Software
### New Features and Applications:
- CS Vibe
- BioMatrix Motion Sensor
- SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
- Deep Resolve Boost for FL3D_VIBE and SPACE
- Deep Resolve Sharp for FL3D_VIBE and SPACE
- Preview functionality for Deep Resolve Boost
- EP2D_FID_PHS
- EP_SEG_FID_PHS
- AutoMate Cardiac (Cardiac AI Scan Companion)
- DANTE blood suppression
{17}
SIEMENS Healthineers
510(k) Summary
- SMS Averaging for TSE
- SMS Averaging for TSE_DIXON
- SMS for BLADE without diffusion function
- Ghost reduction (Dual polarity Grappa (DPG))
- Fleet Reference Scan
- Deep Resolve Swift Brain
- Quick Protocols
- myExam Autopilot Spine
- Open Workflow
## Modified features and applications:
- myExam Implant Suite
- GRE_PC
- myExam RT Assist workflow improvements (not available for MAGNETOM Flow.Rise)
- Open Recon 2.0
- Deep Resolve Boost for TSE
- "MTC Mode" for SPACE
- SPACE Improvement: high bandwidth IR pulse
- SPACE Improvement: increase gradient spoiling
## New (general) Software / Platform / Workflow:
- Select&GO extension (coil-based Iso Centering, Patient Registration at the touch display, Start Scan at the touch display)
- New Startup-Timer
- myExam RT Assist (not available for MAGNETOM Flow.Rise)
- myExam Brain RT-Autopilot (not available for MAGNETOM Flow.Rise)
- Eco Power Mode Pro
## Modified (general) Software / Platform:
- Improved Gradient ECO Mode Settings
Furthermore, the following minor updates and changes were conducted for the subject devices MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Sola Fit and MAGNETOM Viato.Mobile:
- Off-Center Planning Support
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
{18}
SIEMENS Healthineers
510(k) Summary
- Select&Go Display (Touch Display (TPAN))
- Marketing bundle "myExam Companion"
The following minor updates and changes were conducted for the subject devices MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:
- Off-Center Planning Support
- Flip Angle Optimization (Lock TR and FA)
- Inline Image Filter
- Automatic System Shutdown (ASS) sensor (Smoke Detector)
- ID Gain (re-naming)
- Marketing bundle "myExam Companion"
- Marketing Bundle "RT Pro Edition"(not available for MAGNETOM Flow.Rise)
# 7. Substantial Equivalence
MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10 are substantially equivalent to the following devices:
| Predicate Devices | FDA Clearance Number and Date | Product Code | Manufacturer |
| --- | --- | --- | --- |
| MAGNETOM Vida with syngo MR XA60A | K231560, cleared on October 23, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Lumina with syngo MR XA60A | K231560, cleared on October 23, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Sola with syngo MR XA61A | K232535, cleared on December 22, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Altea with syngo MR XA61A | K232535, cleared on December 22, 2023 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Viato.Mobile with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Sola Fit with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| Reference Devices | FDA Clearance Number and Date | Product Code | Manufacturer |
| --- | --- | --- | --- |
| MAGNETOM Flow.Ace with syngo MR XA70A | K250436, cleared on June 16, 2024 | LNH, LNI, MOS | Siemens Shenzhen Magnetic Resonance Ltd. |
| MAGNETOM Free.Max with syngo MR XA60A | K231617, cleared on September 11, 2023 | LNH, MOS | Siemens Shenzhen Magnetic Resonance |
{19}
SIEMENS Healthineers
510(k) Summary
| MAGNETOM Skyra Fit with syngo MR XA70A | K250443, cleared on June 16, 2024 | LNH, LNI, MOS | Ltd. |
| --- | --- | --- | --- |
| MAGNETOM Vida Fit with syngo MR XA50A | K220939, cleared on April 29, 2022 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| syngo.via VB40A | K191040 | LLZ | Siemens Healthcare GmbH |
# 8. Technological Characteristics
The subject devices, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10, are substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance.
The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.
While there are some differences in technological characteristics between the subject devices and predicate device, including new and modified hardware and software, these differences have been tested and the conclusions from the non-clinical data suggest that the features bear an equivalent safety and performance profile to that of the predicate device.
# 9. Nonclinical Tests
The following performance testing was conducted on the subject devices:
| Performance Test | Tested Hardware or Software | Source/Rationale for test |
| --- | --- | --- |
| Software verification and validation | New or modified software features | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices |
| Sample clinical images | Coils, new or modified software features | Guidance for submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices |
| Image quality assessment by sample clinical images | - new / modified pulse sequence types.
- comparison images between the new / modified features and the predicate device features | |
| Performance bench test | New and modified hardware | |
| Biocompatibility | surface of applied parts | ISO 10993-1 |
| Electrical safety and electromagnetic compatibility (EMC) | Complete systems MAGNETOM Flow. Platform | IEC 60601-1-2 |
The results from each set of tests demonstrate that the subject devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.
Below table shows an executive summary of training and validation dataset of new AI features in the subject devices.
{20}
SIEMENS Healthineers
510(k) Summary
- Deep Resolve Swift Brain:
| Test result summary | The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and normalized mean squared error (NMSE). Additionally, images were inspected visually to ensure that potential artefacts are detected that are not well captured by the metrics.
After successful passing of the quality metrics tests, work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners. |
| --- | --- |
| Test setup | Equipment: 3T and 1.5T MRI scanners.
Protocols: representative protocols (T1, T2*, T2, T2 FLAIR).
Body region: brain. |
| | Sample size: 29,740 2D slices. |
| | Sample source: in-house measurement.
Note: Due to the network architecture, attributes like gender, age and ethnicity are not relevant to the training data. |
| | Dataset split:
Training:
20,076 slices
Validation:
1.5T Validation: 3,616 slices (1.5T validation);
3T Validation: 6,048 slices (3T validation).
Note: Data split maintained similar data distribution (e.g. contrast, orientation, field strength, ...) in both training and validation datasets. |
| Patient Characteristics | No clinical subgroups have been defined for the datasets. |
| Reference standard | The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition of Gaussian noise to k-space data and uniformly-random cropping of the training data along the readout direction. |
| Data independency | Datasets for training and testing were split prior to training (see Test Setup field above). |
- Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE:
| Test result summary | Quantitative evaluations of structural similarity index (SSIM), peak signal-to-noise ratio (PSNR) and mean squared error (MSE) metrics showed a convergence of the training and improvements compared to conventional parallel imaging. An inspection of the test images did not reveal any negative impact to the image quality. The function has been used either to acquire images faster or to improve image quality. |
| --- | --- |
| Test setup | Equipment: 0.55T, 1.5T and 3T scanners |
{21}
SIEMENS Healthineers
510(k) Summary
| | Protocols: representative measurement protocols (T1, T2 and PD with and without fat saturation) which have been altered for training (e.g. to increase SNR, increase resolution or reduce acceleration). |
| --- | --- |
| | Body regions: broad range of body regions |
| | Used coils: broad range of coils to cover the dedicated body regions |
| | Sample size: 27,679 3D patches from 1265 measurements |
| | Dataset split: Training: 81% of the 1265 measurements
Validation: 19% of the 1265 measurements |
| | Note: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, ...) in both training and validation datasets. |
| | Sample source: in-house measurements (training and validation) and collaboration partners (testing) |
| | |
| Patient Characteristics | Gender distribution:
- Male: 53%
- Female 47%
Age: for training and validation.
- 19 - 45: 14%
- 46 - 65: 43%
- 66 - 89: 43%
Clinical subgroups: No clinical subgroups have been defined for the datasets. |
| Reference standard | The acquired datasets (as described above) represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines as well as creating sub-volumes of the acquired data. |
| Data independency | Datasets determined for training and validation were split prior to training along individual acquisitions to ensure that there is no mixture of sub-volumes stemming from the same acquisition. |
- Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE
| Test result summary | The impact of the Deep Resolve Sharp network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. The tests include rating and an evaluation of image sharpness by intensity profile comparisons of reconstruction with and without Deep Resolve Sharp. Both tests show increased edge sharpness and reduced Gibb’s artifacts. |
| --- | --- |
| Test setup | Equipment: 0.55T, 1.5T and 3T MRI scanners
Protocols: representative measurement protocols (T1, T2 and PD with and without fat saturation) which have been altered (e.g. to increase SNR, increase resolution or reduce acceleration)
Body regions: broad range of different body regions.
Used coils: broad range of coils to cover the dedicated body regions
Sample size: approx. 13,000 high resolution 3D patches from 500 measurements.
Dataset split: Training: 70% of the 500 measurements. |
{22}
SIEMENS Healthineers
510(k) Summary
| Validation: 30% of the 500 measurements |
| --- |
| Note: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, ...) in both training and validation datasets. |
| Sample source: in-house measurements |
| Patient Characteristics |
| Gender distribution: - Male: 66.6% - Female 33.4%
Age: for training and validation. - 19 - 45: 8.4% - 46 - 65: 40.2% - 66 - 89: 51.4%
Clinical subgroups: No clinical subgroups have been defined for the datasets. |
| Reference standard |
| The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation. k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation. |
| Data independency |
| The high-resolution datasets were split to 70% training and 30% validation datasets before training to ensure independence of them. The input and output variables of the network have been derived from the same dataset so that no confounders exist for the training methodology. |
- Deep Resolve Boost for TSE
| Test result summary | The evaluation on the test dataset confirmed very similar metrics in terms of peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and learned perceptual image patch similarity metrics (LPIPS) for the predicate and the modified network with both outperforming conventional GRAPPA as the reference. Visual evaluations confirmed statistically significant reduction of banding artifacts with no significant changes in sharpness and detail visibility. In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics between updated and cleared predicate network.
The function as on the predicate devices was modified to the subject devices but the training and testing from the predicate devices still fits. |
| --- | --- |
| Test setup | Equipment: 0.5ST, 1.5T and 3T MRI scanners
Protocols: representative protocols (T1, T2 and PD with and without fat saturation)
Body regions: broad range of different body regions
Used coils: broad range of coils to cover the dedicated body regions
Testing of the Deep Resolve Boost network has been described in the reference and predicate device submissions. Additional tests have been performed to evaluate the banding artifact reduction capabilities of the updated network.
Dataset split: Training: more than 23250 slices (93%) |
{23}
SIEMENS Healthineers
510(k) Summary
| | Validation: more than 1750 slices (7%) Additional test dataset for banding artifact reduction: more than 2000 slices For training and validation of the network, the identical data was used as for the initial device submission (K213693). Note: Data split maintained similar data distribution (e.g., contrast, orientation, field strength, ...) in both training and validation datasets. Sample source: in-house measurements and collaboration partners |
| --- | --- |
| Patient Characteristics | Due to reasons of data privacy, gender, age and ethnicity during data collection have not been recorded. Due to the network architecture, attributes like gender, age and ethnicity are not relevant to the training data. No clinical subgroups have been defined for the collected dataset |
| Reference standard | The acquired training/validation datasets (identical to the initial submission K213693) represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data. |
| Data independency | Training and validation datasets were kept independent from each other during training and validation. The acquired datasets (one dataset consists of a group of multiple slices) were split into 93% training and 7% validation data prior to the training. A similar distribution was maintained for training and validation data. The test dataset for banding artifact reduction was acquired after the release of the predicate network and is therefore independent of the training/validation data. |
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
# 10. Clinical Tests
No clinical tests were conducted to support substantial equivalence for the subject devices.
Sample clinical images were provided to support substantial equivalence for the subject devices.
# 11. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
The subject devices with Syngo MR XB10 conform to the following FDA recognized and international IEC, ISO and NEMA standards:
{24}
SIEMENS Healthineers
510(k) Summary
| Recognition Number | Product Area | Title of Standard | Reference Number and date | Standards Development Organization |
| --- | --- | --- | --- | --- |
| 19-46 | General II (ES/ EMC) | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | ES60601-1:2005 /(R)2012 & A1:2012, C1:2009/(R)2012 &A2:2010/(R)2012 (Cons. Text) [Incl.AMD2:2021] | ANSI AAMI |
| 19-36 | General | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 60601-1-2 Edition 4.1 2020-09 | IEC |
| 12-347 | Radiology | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis | 60601-2-33 Edition 4.0 2022-08 | IEC |
| 5-125 | General I (QS/ RM) | Medical devices - Application of risk management to medical devices | 14971 Third edition 2019-12 | ISO |
| 5-129 | General I (QS/ RM) | Medical devices - Part 1: Application of usability engineering to medical devices | 62366-1: 2015 + AMD1:2020 | IEC |
| 13-79 | Software/ Informatics | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] | IEC 62304:2006 + AMD1:2015 | IEC |
| 12-232 | Radiology | Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices | MS 4-2010 | NEMA |
| 12-288 | Radiology | Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images | MS 9-2008 (R2020) | NEMA |
| 12-352 | Radiology | Digital Imaging and Communications in Medicine (DICOM) | PS 3.1 - 3.20 (2023e) | NEMA |
| 2-258 | Biocompatibility | Biological evaluation of medical devices - part 1: evaluation and testing within a risk management process. (Biocompatibility) | 10993-1: 2018 | ANSI AAMI ISO |
{25}
SIEMENS Healthineers
510(k) Summary
# 12. Conclusion as to Substantial Equivalence
MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10 have the same intended use and same basic technological characteristics than the respective predicate devices, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate devices, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that the subject devices MAGNETOM Flow.Neo, MAGNETOM Flow.Elite, MAGNETOM Flow.Rise, MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile, with software Syngo MR XB10 are substantially equivalent to the currently marketed devices MAGNETOM Vida and MAGNETOM Lumina with syngo MR XA60A (K231560, cleared on October 23, 2023), MAGNETOM Sola and MAGNETOM Altea with syngo MR XA61A (K232535, cleared on December 22, 2023) and MAGNETOM Sola Fit and MAGNETOM Viato.Mobile with syngo MR XA70A (K250443, cleared on June 16, 2024).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.