These sensors are intended to be used by clinicians that have been appropriately trained, such as nurses, physicians, and EMS personnel. These sensors provide continuous, non-invasive measurement of arterial oxygen saturation (pulse rate signal and plethysmograph wave) to any SpO2 device that has passed validation testing. Either sensor can be comfortably clipped onto the finger of patients weighing > 40kg (typically adult patients).
Device Story
Reusable clip-style SpO2 sensors (M1196A, M1196T) for use with compatible Philips monitor/defibrillator systems. Sensors contain red and infrared light sources and a photodiode; measure arterial oxygen saturation via transmissive spectrophotometry. Input: red/IR light absorption signals; Output: SpO2 and pulse rate data. Used in hospital environments and during patient transport by trained clinicians (nurses, physicians, EMS). Output displayed on host monitor to assist clinical decision-making regarding patient oxygenation status. Benefits include continuous, non-invasive monitoring of patient vital signs.
Clinical Evidence
Single-center, non-randomized, desaturation study in healthy adult volunteers. Validated SpO2 accuracy against SaO2 from arterial blood samples analyzed via CO-Oximetry. Results confirmed an overall ARMS of ≤3%.
Technological Characteristics
Clip-style sensor with hard shell and soft inner material. Transmissive pulse oximetry using red/IR LEDs and photodiode. Non-sterile, reusable. Complies with IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, IEC 60601-1-12, IEC 62366-1, and ISO 10993-1.
Indications for Use
Indicated for non-invasive measurement of oxygen saturation (SpO2) and pulse rate in patients > 40 kg (typically adult patients).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Philips SpO2 Reusable Sensor, Models M1196A and M1196T (K062605)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
Philips Medizin Systeme Böblingen GmbH
Nicole Söder
Regulatory Affairs Specialist
Hewlett-Packard-Strasse 2
Böblingen, BW 71034
Germany
Re: K252821
Trade/Device Name: Reusable Adult SpO2 Clip Sensor (3m) (M1196A); Reusable Adult SpO2 Clip Sensor (0.9m) (M1196T)
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: DQA
Dated: April 30, 2026
Received: April 30, 2026
Dear Nicole Söder:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252821 - Nicole Söder
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K252821 - Nicole Söder
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252821 | ? |
| Please provide the device trade name(s). | | ? |
| Reusable Adult SpO2 Clip Sensor (3m) (M1196A) Reusable Adult SpO2 Clip Sensor (0.9m) (M1196T) | | |
| Please provide your Indications for Use below. | | ? |
| The Philips Reusable SpO2 Sensors are intended for non-invasive measurement of oxygen saturation (SpO2) and pulse rate. M1196A and M1196T are indicated for patients > 40 kg (typically adult patients). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PHILIPS
Philips Reusable SpO2 Sensors M1196A/M1196T
Traditional 510(k) 510(k) Summary
| 510(k) Summary | | | | |
| --- | --- | --- | --- | --- |
| Submitter | | | | |
| **Date Prepared** | May 28, 2026 | | | |
| **Submitter/Owner** | Philips Medizin Systeme Böblingen GmbH FDA Establishment Number 9610816 Hewlett-Packard-Str. 2 71034 Böblingen Germany | | | |
| **Key Contact** | Nicole Söder Regulatory Affairs Specialist nicole.soeder@philips.com Phone: +49 (0)7031/463-1492 | | | |
| **510(k) Submission Type** | Traditional 510(k) | | | |
| Device | | | | |
| **Model Numbers (#)** | M1196A & M1196T | | | |
| **Trade Name** | Reusable Adult SpO2 Clip Sensor | | | |
| **Common Name** | SpO2 Pulse Oximeter Sensor | | | |
| **Classification Name** | Panel & Name: Cardiovascular Devices Subpart & Division: 21 CFR §870.2700 Oximeter Regulatory Class: II Product Code: DQA | | | |
| Predicate Device | | | | |
| | **510(k) No.** | **Company** | **Device Name** | **Product Code** |
| **Predicate Device** | K062605 | Philips Medical Systems 3000 Minuteman RD. Andover, MA 01810 | Philips SpO2 Reusable Sensor, Models M1196A and M1196T | DQA |
| The subject devices are substantially equivalent to the legally marketed predicate devices. | | | | |
| Device Description | | | | |
PHILIPS
Philips Hospital Patient Monitoring
Page 1 of 5
{5}
PHILIPS
Philips Reusable SpO2 Sensors M1196A/M1196T
Traditional 510(k) 510(k) Summary
| Philips Reusable SpO2 Sensors M1196A/M1196T – description of the device per 21 CFR 807.92(a) (4) |
| --- |
| The Philips Reusable SpO2 Sensors M1196A/M1196T are used as accessories with compatible Philips Monitor/Defibrillator product lines as part of a System. |
| These Sensors measure, non-invasively, the arterial oxygen saturation of blood. The measurement method is based on the red and infrared light absorption of hemoglobin and oxyhemoglobin. The Sensors contain a red and infrared light source and a photodiode receiving the non-absorbed red and infrared light. The received signals are forwarded to a measurement device that amplifies the acquired signal and an algorithm that calculates the ratio and converts via a validated calibration table the ratio to a saturation value. |
| The Philips Reusable SpO2 Sensors M1196A/M1196T fall into the clip style category. Clip style Sensors follow the traditional SpO2 sensor design of a hard shell, to house the optical components, with a soft, contoured inner material meant to distribute the clamping pressure of the shell. |
| **Intended Use and Indication for Use** |
| **Intended Use as required per 21 CFR 807.92(a)(5)** |
| **Intended Use** These sensors are intended to be used by clinicians that have been appropriately trained, such as nurses, physicians, and EMS personnel. These sensors provide continuous, non-invasive measurement of arterial oxygen saturation (pulse rate signal and plethysmograph wave) to any SpO2 device that has passed validation testing. Either sensor can be comfortably clipped onto the finger of patients weighing > 40kg (typically adult patients). |
| **Indications for Use:** The Philips Reusable SpO2 Sensors are intended for non-invasive measurement of oxygen saturation (SpO2) and pulse rate. M1196A and M1196T are indicated for patients > 40 kg (typically adult patients). |
Philips Hospital Patient Monitoring
Page 2 of 5
{6}
PHILIPS
Philips Reusable SpO2 Sensors M1196A/M1196T
Traditional 510(k) 510(k) Summary
| Comparison of Technological Characteristics with Predicate Device | |
| --- | --- |
| Similarities | |
| Item of Comparison | Description/Rationale |
| Intended Use/Indications for Use/Patient Population | The subject and predicate device do have the same intended use and indications for use. Intended Use/Indications for Use: Subject and predicate device provide continuous, non-invasive measurement of arterial oxygen saturation (pulse rate signal and plethysmograph wave) to any SpO2 device that has passed validation testing. Either sensor can be comfortably clipped onto the finger of patients weighing > 40kg (typically adult patients). |
| Principle of Operation | The subject and predicate device do have the same Principle of Operation. Spectrophotometric measurement of functional arterial oxygen saturation by transmissive mode pulse oximetry. |
| Anatomical Sites | The subject and predicate device are applied to any finger except thumb. |
| Environment of Use | The subject and predicate device are intended for use in hospital environments and for transport to a professional healthcare facility or between professional healthcare facilities. |
| Sterility | The subject and predicate device are non-sterile |
| Reusability | The subject and predicate device are reusable devices |
| Prescription Use | The subject and predicate device are prescription use only devices |
| Differences | |
| Material and Color | The subject device is made out of a more matte and opaque material. SpO2 Sensor housing, connector and cable colors have been changed. |
| LED and Photodiode (PD) | The positioning of the LED and photodiode (PD) was slightly changed for the subject device to optimize the light path. |
| Calibration Curve | Compared to the predicate device, the subject device applies a different calibration curve. This calibration curve is an existing and known calibration curve, already used in other SpO2 Sensors. |
| Clinical Study | The 510(k) included clinical data from Philips initiated laboratory tests with the intent to demonstrate the accuracy performance of the subject device. The clinical investigation provides supporting evidence that the proposed design changes tested with the subject device meet the existing SpO2 accuracy (ARMS) claims when validated in a large and diverse patient population. |
PHILIPS
Philips Hospital Patient Monitoring
Page 3 of 5
{7}
PHILIPS
Philips Reusable SpO2 Sensors M1196A/M1196T
Traditional 510(k) 510(k) Summary
| **Substantial Equivalence Summary** | | |
| --- | --- | --- |
| In accordance with 21 CFR Part 807, the Philips Reusable SpO2 Sensors M1196A/M1196T are substantially equivalent to the predicate devices. The differences identified between the subject devices and predicate devices do not raise new questions of safety and effectiveness. Performance testing confirms that the Philips Reusable SpO2 Sensors M1196A/M1196T are substantially equivalent to the predicate devices for Philips Monitor/Defibrillator product lines. The Philips SpO2 Reusable Sensor Models M1196A and M1196T (K062605) serve as the predicate devices. | | |
| **Performance Data** | | |
| **Recognized Standards** | | |
| The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below: | | |
| Standard | FDA Recognition # | Title # |
| IEC 60601-1:2005+A1:2012+A2:2020 | 19-49 | Medical electrical equipment – Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | 19-36 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) | 1-139 | Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment |
| IEC 60601-1-12 Edition 1.1 2020-07 CONSOLIDATED VERSION | 19-39 | Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment |
| IEC 62366-1:2015+AMD1:2020: | 5-129 | Medical devices - Part 1: Application of usability engineering to medical devices |
| ISO 10993-1 Fifth edition 2018- | 2-258 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process |
| **Non-Clinical Bench Testing** | | |
| The following Bench tests were performed to demonstrate the correct performance of the Philips Reusable SpO2 Sensors M1196A/M1196T: | | |
PHILIPS
Philips Hospital Patient Monitoring
Page 4 of 5
{8}
PHILIPS
Philips Reusable SpO2 Sensors M1196A/M1196T
Traditional 510(k) 510(k) Summary
- Interoperability testing, covering the performance between the Philips Reusable SpO2 Sensors and the host monitor device
- Material compatibility testing, covering material resistance to intended use and cleaning and disinfection
- Electronic Hardware testing, covering verification activities related to the applicable calibration curve, testing of optical requirements and testing of ranges, limits and tolerances
- Mechanical testing covering insertion and removal force testing
- Reliability testing covering bending, pull force and stress cycles
- Environmental testing
All above listed testing have met the specified acceptance criteria.
The data from the non-clinical testing supports the substantial equivalence of the subject device.
## Clinical Testing
SpO2 accuracy has been demonstrated through a clinical study performed with the Philips FAST Pulse Oximetry technology and the Philips Reusable SpO2 Sensors.
The performed clinical study is a single-center, non-randomized, desaturation study in healthy adult volunteers. The purpose was to validate the SpO2 accuracy of the Philips FAST Pulse Oximetry System against SaO2 obtained from arterial blood samples and analyzed by CO-Oximetry with consideration to anticipated FDA Guidance for Pulse Oximeters and ISO 80601-2-61 updates.
The testing confirmed and overall ARMS of ≤3%.
The data from clinical testing supports the substantial equivalence of the subject device.
## Conclusion
Substantial equivalence assessment, based on performance testing and taken together with the clinical study demonstrate that the subject devices do not raise different questions of safety and effectiveness when compared to the predicate device. The Philips Reusable SpO2 Sensors M1196A/M1196T perform as intended and have performance characteristics that are substantially equivalent to the predicate device.
PHILIPS
Philips Hospital Patient Monitoring
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.