K252802 · Shenzhen Bi-Rich Medical Devices Co., Ltd. · BZH · May 29, 2026 · Anesthesiology
Device Facts
Record ID
K252802
Device Name
Peak Flow Meter (PF-10(BR-PEF231))
Applicant
Shenzhen Bi-Rich Medical Devices Co., Ltd.
Product Code
BZH · Anesthesiology
Decision Date
May 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1860
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ResScan Essentials is intended to augment the standard follow-up care of patients by transferring and analyzing machine and therapeutic information from ResMed compatible therapy devices to the ResScan Essentials application. ResScan Essentials is intended to be used by clinicians.
Device Story
ResScan Essentials is an offline desktop application for clinicians to view and analyze therapy data from ResMed compatible devices. Data is transferred from therapy devices to the application via SD card or USB drive. The software displays usage metrics (daily hours, sessions), sleep quality indices (AHI, AI, RERA, CSR, Flow Limitation, Snore Index), pressure metrics (IPAP/EPAP, graphs), mask leak, and oximetry data (SpO2, pulse). Clinicians use these reports to manage patient therapy and support clinical decision-making. Unlike its predicate, this device does not allow modification of therapy settings. It operates on Windows 10/11 and includes authenticated user access and encrypted data storage.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through non-clinical verification and validation testing, including software verification against non-functional requirements and end-to-end functional testing in accordance with FDA guidance on software validation and cybersecurity.
Technological Characteristics
PC-based desktop application; compatible with Windows 10 and 11. Data input via SD card or USB memory stick. Features authenticated user access and encrypted data storage. Software-only device; no physical hardware components. Complies with FDA guidance for software validation and cybersecurity.
Indications for Use
Indicated for patients using ResMed compatible therapy devices requiring clinical follow-up care and analysis of machine and therapeutic data.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
5/27/2026
Resmed Corp
Marzia Signoroni
Senior Specialist Regulatory Affairs
9001 Spectrum Center Blvd.
San Diego, California 92123
Re: K254104
Trade/Device Name: ResScan Essentials
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD, MNS, MNT, CBK
Dated: April 27, 2026
Received: April 28, 2026
Dear Marzia Signoroni:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254104 - Marzia Signoroni
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K254104 - Marzia Signoroni
Page 3
Sincerely,
**Binoy J. Mathews -S**
Digitally signed by
Binoy J. Mathews -S
Date: 2026.05.27
14:47:10 -04'00"
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254104 | |
| Device Name ResScan Essentials | |
| Indications for Use (Describe) ResScan Essentials is intended to augment the standard follow-up care of patients by transferring and analyzing machine and therapeutic information from ResMed compatible therapy devices to the ResScan Essentials application. ResScan Essentials is intended to be used by clinicians. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Resmed
ResScan Essentials
Traditional 510(k)
# 1 510(k) Summary
[As required by 21 CFR 807.921(c)]
Date of Submission: 5th Dec 2025
Company Name/Owner: ResMed Corp
9001 Spectrum Center Boulevard
San Diego, CA 92123, US
Official Contact: Marzia Signoroni
Senior Regulatory Affairs Specialist
ResMed Corp
9001 Spectrum Center Boulevard
San Diego, CA 92123, US
marzia.signoroni@resmed.com.au
Device Trade Name: ResScan Essentials
Device Common Name: Ventilator, Non-Continuous
(Respirator) (BZD)
Classification, Classification Name: 21 CFR 868.5905 (Class II)
Non continuous ventilator (IPPB)
Product codes: BZD, MNS, MNT, CBK
510(k) Summary, Page 1
{5}
Resmed
ResScan Essentials
Traditional 510(k)
Legally Marketed ResScan (K140054)
Predicate
Device:
Device
Description:
ResScan Essentials is an offline desktop application specifically developed to function as a patient data viewer for therapy information collected by compatible ResMed therapy devices. Its primary purpose is to assist clinicians in the analysis of patient therapy data, which can be transferred from supported ResMed devices using SD cards or USB drives. By importing this data into ResScan Essentials, clinicians are able to analyze and manage patient information effectively, supporting the processes of diagnosis and ongoing patient management.
Indication For Use: ResScan Essentials is intended to augment the standard follow-up care of patients by transferring and analyzing machine and therapeutic information from ResMed compatible therapy devices to the ResScan Essentials application. ResScan Essentials is intended to be used by clinicians.
Intended Use
Comparison
| ResScan Essentials (K254104) | ResScan (K140054) | Substantial Equivalence |
| --- | --- | --- |
| ResScan Essentials is intended to augment the standard follow-up care of patients by transferring and analyzing machine and therapeutic information from ResMed compatible therapy devices to the ResScan Essentials application. ResScan Essentials is intended to be used by clinicians. | ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information. This includes the ability to remotely change settings in non-life support devices only. | Same. Devices are intended to support the standard follow-up care of patients by transferring machines and therapeutic information. Unlike the predicate device, ResScan Essentials is not intended to modify therapeutic settings. This modification eliminates the risks associated with changing the compatible therapy device settings while ensuring that no new concerns regarding safety or effectiveness are introduced. |
510(k) Summary, Page 2
{6}
Resmed
ResScan Essentials
Traditional 510(k)
# Technological
# Characteristics Comparison
| Characteristic | Subject Device: ResScan Essentials Manufacturer: Resmed Corp 510(k) Number: K254104 | Predicate Device: ResScan Manufacturer: Resmed Corp 510(k) Number: K140054 | Substantial Equivalence |
| --- | --- | --- | --- |
| Regulation Number | 21 CFR §868.5905 | 21 CFR §868.5905 | Same |
| Classification Name | Ventilator, Non-Continuous (Respirator) (BZD) | Ventilator, Non-Continuous (Respirator) (BZD) | Same |
| Compatible flow generators | • BZD • MNS • MNT • CBK | • BZD • MNS • MNT • CBK | Same |
| Prescription Use | • Yes | • Yes | Same |
| Functionality | • PC Application • Reports (Compliance and Therapy) • Patient management • Display of therapy and usage data | • PC Application • Display of therapy and usage data • Reports (Compliance and Therapy) • Settings management (non-life support devices only) • Patient management | Similar: Settings management is not available on ResScan Essentials. This modification eliminates the risks associated with changing the compatible therapy device settings while ensuring that no new concerns regarding safety or effectiveness are introduced. |
| Operating System | Windows 10, 11 Pro, Home, Enterprise | Microsoft Windows XP Service Pack 2 (or later); Microsoft Windows Vista. Microsoft Windows 7, Windows 10 | Same: obsoleted Operating Systems (OS) removed, Windows API remains identical in latest OS. |
510(k) Summary, Page 3
{7}
Resmed
ResScan Essentials
Traditional 510(k)
| Characteristic | Subject Device: ResScan Essentials Manufacturer: Resmed Corp 510(k) Number: K254104 | Predicate Device: ResScan Manufacturer: Resmed Corp 510(k) Number: K140054 | Substantial Equivalence |
| --- | --- | --- | --- |
| Data transfer technology | • SD Card • USB Memory Stick | • Serial connection (RS232) • SD Card • USB Stick • Smart Media Card • Smart Card | Similar. Data transfer technology reduced to align with user needs. |
| Cybersecurity Features | • Authenticated user access • Encrypted data storage | • Authenticated user access | Similar. Subject adds encrypted data storage |
| Patient information | • Usage: daily hours, sessions, no-use days • Event & Sleep Quality Indices: AHI, AI, RERA, CSR, Flow Limitation Index, Snore Index • Pressure Metrics: Set/Delivered Pressure, IPAP/EPAP, Min/Max, PS, Ramp Start / End Pressure, Pressure Graphs • Mask Leak • Oximetry: SpO₂, Pulse, ODI Therapy & Comfort Settings: Mode, Pressure, Ramp, EPR, Climate Control, Tube Temp, Humidifier, SmartStart/Stop, RR, Vt, MV, I:E Ratio | • Usage: daily hours, sessions, no-use days • Event & Sleep Quality Indices: AHI, AI, RERA, CSR, Flow Limitation Index, Snore Index • Pressure Metrics: Set/Delivered Pressure, IPAP/EPAP, Min/Max, PS, Ramp Start / End Pressure, Pressure Graphs • Mask Leak • Oximetry: SpO₂, Pulse, ODI Therapy & Comfort Settings: Mode, Pressure, Ramp, EPR, Climate Control, Tube Temp, Humidifier, SmartStart/Stop, RR, Vt, MV, I:E Ratio | Same |
510(k) Summary, Page 4
{8}
Resmed
ResScan Essentials
Traditional 510(k)
## Non-clinical testing:
Non-clinical verification and validation testing for ResScan Essentials demonstrated that the device met all intended performance requirements. This testing was conducted in accordance with applicable FDA guidance documents, including General Principles of Software Validation; Final Guidance for Industry and FDA Staff (11 January 2002), Content of Premarket Submissions for Device Software Functions (14 June 2023), and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (27 September 2023).
The verification testing performed with ResScan Essentials was software verification and validation against the non-functional requirements and end-to-end functional testing.
510(k) Summary, Page 5
{9}
Resmed
ResScan Essentials
Traditional 510(k)
## Clinical Testing
Clinical tests were not required to demonstrate safety and effectiveness of ResScan Essentials.
## Substantial Equivalence
The subject device and predicate devices have similar intended use, technological characteristics, and operating principles.
Similar technological characteristics include:
- Both the subject and predicate devices are software that are intended to provide transfer of machine and therapeutic information from compatible ResMed PAP therapy devices.
The technological differences between the predicate device and subject device are:
- The subject device is unable to modify therapy settings on a compatible ResMed PAP therapy device.
510(k) Summary, Page 6
{10}
Resmed
ResScan Essentials
Traditional 510(k)
## Substantial Equivalence Conclusion:
ResScan Essentials is substantially equivalent to the predicate ResScan K140054 because:
- It has similar intended use
- It has similar technological characteristics
- It has similar operating principles
- The differences do not raise any new questions of safety or effectiveness
- It is at least as safe and as effective as the predicate device
510(k) Summary, Page 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.