JAZBI™ Resorbable Bone Void Filler is indicated for use in bony voids or defects that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults), may be surgically created osseous defects, or osseous defects created by traumatic injury to the bone. JAZBI™ Resorbable Bone Void Filler is intended to be gently packed or placed into bony voids or defects of the extremities and pelvis. Following placement into the bony void, JAZBI™ Resorbable Bone Void Filler resorbs and is replaced with bone during the healing process.
Device Story
Porous, osteoconductive bone void filler; composed of beta-tricalcium phosphate, hydroxyapatite, poly(propylene carbonate) (PPC), and 3-Glycidyloxypropyltrimethoxysilane (GPTMS). Supplied as sterile, single-use granules. Intended for surgical implantation; packed into osseous defects of extremities and pelvis. Device resorbs over time, facilitating natural bone ingrowth and replacement. Used by surgeons to fill non-structural voids. Benefits include support for bone healing process.
Clinical Evidence
No clinical data. Performance demonstrated via rabbit critical-sized defect model comparing radiographic, histologic, and histomorphometric outcomes to predicate. Biocompatibility testing performed per ISO 10993 series (parts 5, 10, 11, 12, 18, 23) and toxicological risk assessment per ISO 10993-17.
Technological Characteristics
Porous calcium phosphate (beta-tricalcium phosphate and hydroxyapatite) in a polymeric matrix (PPC and GPTMS). Granule form (1.4-5.6 mm). Porosity 48.5%; pore size 113-228 μm. Sterilized via gamma irradiation (17.5 kGy). Biocompatibility per ISO 10993. Sterile barrier shelf-life validated per ASTM F1980, F1886/F1886M, F1929, F88/F88M, and F2096.
Indications for Use
Indicated for adults with bony voids or defects not intrinsic to skeletal stability, including benign bone cysts/tumors, surgically created defects, or traumatic injuries in the extremities and pelvis.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 26, 2026
SDIP Innovations Pty Ltd
% Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA
Director, Regulatory Affairs
MCRA, LLC
803 7th Street NW, Floor 3
Washington, DC 20001
Re: K252797
Trade/Device Name: JAZBI™ Resorbable Bone Void Filler
Regulation Number: 21 CFR 888.3045
Regulation Name: Resorbable Calcium Salt Bone Void Filler Device
Regulatory Class: Class II
Product Code: MQV
Dated: April 17, 2026
Received: April 17, 2026
Dear Dr. Kazemzadeh-Narbat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252797 - Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252797 - Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE MUIR - Digitally signed by JESSE
MUIR -S
Date: 2026.05.26 09:14:19
-04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
JAZBI™ Resorbable Bone Void Filler
Page 11 of 51
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252797 | ? |
| Please provide the device trade name(s). | | ? |
| JAZBI™ Resorbable Bone Void Filler | | |
| Please provide your Indications for Use below. | | ? |
| JAZBI™ Resorbable Bone Void Filler is indicated for use in bony voids or defects that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults), may be surgically created osseous defects, or osseous defects created by traumatic injury to the bone. JAZBI™ Resorbable Bone Void Filler is intended to be gently packed or placed into bony voids or defects of the extremities and pelvis. Following placement into the bony void, JAZBI™ Resorbable Bone Void Filler resorbs and is replaced with bone during the healing process. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252797
# 510(k) Summary
Device Trade Name: JAZBI™ Resorbable Bone Void Filler
Submitter: SDIP Innovations Pty Ltd
17/34-36 Leighton Place
Hornsby, New South Wales 2077, Australia
Telephone +61 404108500
Contact: Maryam Parviz, PhD
Co-founder and Chief Executive Officer
SDIP Innovations Pty Ltd
Consultant: Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA
Director, Regulatory Affairs,
MCRA, LLC
803 7th St NW
Washington, DC 20001
Office: 202.552.6011
mkazemzadeh@mcra.com
Date Prepared: May 22, 2026
Classifications: Filler, bone void, calcium compound
Class: Class II
Product Code: MQV
Primary Predicate: Synthes (USA) chronOS™ (K043045)
# Indications For Use:
JAZBI™ Resorbable Bone Void Filler is indicated for use in bony voids or defects that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors, are surgically created or the result of traumatic injury to the bone.
JAZBI™ Resorbable Bone Void Filler is intended to be gently packed or placed into bony voids or defects of the extremities and pelvis. Following placement into the bony void, JAZBI™ Resorbable Bone Void Filler resorbs and is replaced with bone during the healing process.
{5}
K252797
# Device Description:
JAZBI™ Resorbable Bone Void Filler is composed of porous calcium phosphate in a polymeric matrix (poly(propylene carbonate) (PPC), and 3-Glycidyloxypropyltrimethoxysilane (GPTMS)). It is an osteoconductive, porous implant that allows for bony ingrowth across the graft site. The device is slowly resorbed and replaced by new bone tissue during the natural healing process. The bi-phasic calcium phosphate component of the device is made of beta-tricalcium phosphate and hydroxyapatite. The product is available in granules of different sizes and is provided sterile, non-pyrogenic, for single use only.
# Performance Testing Summary:
Non-clinical data submitted to demonstrate substantial equivalence included:
- a scientific rationale for labeling the device as MR Safe
- sterilization gamma irradiation sterilization validation to a sterility assurance level of $10^{-6}$ by selecting and substantiating a $17.5\mathrm{kGy}$ dose according to ISO 13004, ISO 11137-1, and ISO 11137-2
- bacterial endotoxin testing including Limulus amebocyte lysate (LAL) test according to USP $< 85>$ to demonstrate that the sterile product meets a limit of $\leq 20$ EU/device
- sterile barrier shelf-life testing of samples after accelerated aging to the equivalent of 24 months of real time aging according to ASTM F1980, with testing of the packaging sterile barrier according to ASTM F1886/F1886M, ASTM F1929, ASTM F88/F88M, and ASTM F2096
- subject device shelf-life testing of samples after real time aging to 18 months
- biocompatibility testing according to ISO 10993-5, ISO 10993-12, ISO 10993-10, ISO 10993-23, ISO 10993-11, ISO 10993-18, and a toxicological risk assessment according to ISO 10993-17
- the radiographic, histologic, and histomorphometric performance of the subject device were demonstrated to be equivalent to that of the primary predicate device K043045 in a rabbit critical-sized defect model
No clinical data were included in this submission.
# Substantial Equivalence:
JAZBI™ Resorbable Bone Void Filler is substantially equivalent in technological characteristics including formulation, design, performance, route of administration, and intended use to the predicate Synthes (USA) chronOS™. The subject device and the primary predicate device are irregularly-shaped granules that incorporate similar calcium phosphate materials (a mixture of $\beta$-tricalcium phosphate and hydroxyapatite, or $\beta$-tricalcium phosphate only). The addition of polymeric binder constitutes of JAZBI™ Resorbable Bone Void Filler raises no new issues of safety or effectiveness as shown through the biocompatibility studies, the animal study in the critical-sized defect model, and the toxicological risk assessment. Performance data demonstrates that the subject device is substantially equivalent to the primary predicate device in K043045.
{6}
K252797
# SUMMARY OF TECHNOLOGICAL SIMULARITIES AND DIFFERENCES
The subject device and the primary predicate device are irregularly-shaped granules that incorporate similar calcium phosphate materials (a mixture of $\beta$ -tricalcium phosphate and hydroxyapatite, or $\beta$ -tricalcium phosphate only). The subject device includes additional material of PPC and GPTMS as a portion of the material composition which is not present in the predicate device. The subject device and the primary predicate device are provided in the same ranges of granule sizes and are provided packaged in similar amounts as the primary predicate device. The subject device and the primary predicate device have similar porosity and similar ranges of pore sizes, and both are provided to the end user sterilized by gamma irradiation.
The radiographic, histologic, and histomorphometric performance of the subject device was compared to that of the primary predicate device in a rabbit critical-sized defect model study. The results of the study demonstrated the performance of the subject device was equivalent to that of the primary predicate device in the critical-sized defect model. Additionally, similar to the predicate device, the subject device successfully completed a biocompatibility and toxicological risk assessment per ISO 10993, proving that the subject device is safe to use as intended.
The similarities described, combined with animal performance testing and biocompatibility studies, support substantial equivalence of the subject device to the predicate device. The technological differences between the subject device and the primary predicate do not raise new questions of safety or effectiveness, and do not impact substantial equivalence.
Table 1: Summary of Substantial Equivalence
| Device Characteristics | Subject Device | Primary Predicate Device | Comparison with the Primary Predicate |
| --- | --- | --- | --- |
| | JAZBITM Resorbable Bone Void Filler SDIP Innovations Pty Ltd | Synthes (USA) chronOSTM Synthes (USA) | |
| 510(k) Number | K252792 | K043045 | N/A |
| Common Name | Filler, Bone Void, Calcium Compound | Filler, Bone Void, Calcium Compound | Same |
| Regulation Information | 21 CFR § 888.3045 Class II | 21 CFR § 888.3045 Class II | Same |
| Product Code | MQV | MQV, FMF | Same |
| Intended Use | Bone void filler for use in the skeletal system: Extremities and pelvis | Bone void filler for use in the skeletal system: Extremities, pelvis, posterolateral spine | Same |
{7}
K252797
| Device Characteristics | Subject Device | Primary Predicate Device | Comparison with the Primary Predicate |
| --- | --- | --- | --- |
| Indications for Use | JAZBI™ Resorbable Bone Void Filler SDIP Innovations Pty Ltd | Synthes (USA) chronOS™ Synthes (USA) | Same |
| | JAZBI™ Resorbable Bone Void Filler is indicated for use in bony voids or defects that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors, are surgically created or the result of traumatic injury to the bone. JAZBI™ Resorbable Bone Void Filler is intended to be gently packed or placed into bony voids or defects of the extremities and pelvis. Following placement into the bony void, JAZBI™ Resorbable Bone Void Filler resorbs and is replaced with bone during the healing process. | Synthes chronOS™ is indicated for use in bony voids or gaps that are not intrinsic to the stability of the bony structure. Synthes chronOS™ is indicated for use in the treatment of bony defects created surgically or through traumatic injury. Synthes chronOS™ is intended to be gently packed or placed into bony voids or gaps of the skeletal system (i.e. the extremities, spine, and pelvis) and may be combined with autogenous blood and/or bone marrow. Following placement into the bony void, chronOS™ resorbs and is replaced with bone during the healing process. | |
| Mixing Prior to Use | Not indicated for mixing with autogenous blood, bone marrow, or allograft. | May be combined with autogenous blood and/or bone marrow. | Same |
| Form | Granules | Granules, blocks, wedges, cylinders | Same |
| Size | Granules 1.4-2.8 mm granules 2.8-5.6 mm granules | Granules 0.7-1.4 mm granules 1.4-2.8 mm granules 2.8-5.6 mm granules | Same |
| Porosity | Granules – 48.5% | Granules – 60% | Same |
| Pore Size | 113 μm – 228 μm | 100 μm – 500 μm | Same |
| Material | • Calcium phosphate – composed of β-tricalcium phosphate (TCP) and Hydroxyapatite (HA) • PPC • GPTMS | • β-tricalcium phosphate (TCP) | Differences do not raise new questions of safety and efficacy. Testing demonstrates that the differences do not adversely impact performance. |
| Sterility | Provided sterile to end user | Provided sterile to end user | Same |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Same |
| Usage | Single patient, single use | Single patient, single use | Same |
## Conclusion:
JAZBI™ Resorbable Bone Void Filler and the predicate device has the same intended use, has similar technological characteristics, and are made of similar materials. The data included in this submission demonstrates substantial equivalence to the predicate device listed above. JAZBI™ Resorbable Bone Void Filler is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.