← Product Code [ODP](/productcode/ODP) · K252781

# MSFX MIKRON PEEK CAGES (K252781)

_Mikron Makina Sanayi VE Tic. Ltd. Sti. · ODP · Dec 19, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K252781

## Device Facts

- **Applicant:** Mikron Makina Sanayi VE Tic. Ltd. Sti.
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Dec 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

MSFX Mikron Cervical PEEK Cages are made to be implanted into the appropriate vertebral section to help provide stability for spinal fusion after a diseased cervical disc producing neck and/or armpain is removed during spinal decompression for patients who have had six weeks of non-operative treatment.Cervical PEEK Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level(C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.Cervical PEEK Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and implanted with auto graft/autologous bone graft. They are to be used with supplemental fixation. MSFX Mikron Lumbar PEEK Cages are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment.The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.

## Device Story

Intervertebral body fusion devices; implanted into cervical or lumbar spine to maintain disc space height and facilitate fusion. Cervical cages implanted via anterior approach; lumbar cages for 1-2 contiguous levels. Composed of PEEK (ASTM F2026) body with Ti6Al4V ELI (ASTM F136) pins, markers, and screws/expansion mechanisms. Used with autograft/autologous bone graft and supplemental fixation. Intended for surgical use by physicians to treat degenerative disc disease; provides spinal stability; promotes bone fusion.

## Clinical Evidence

Bench testing only. Biomechanical testing performed per ASTM F2077 (static/dynamic axial compression, shear, and torsion) and ASTM F2267 (subsidence). No clinical data presented.

## Technological Characteristics

Materials: PEEK (ASTM F2026) cage body; Ti6Al4V ELI (ASTM F136) pins, markers, screws, and expansion mechanisms. Form factor: Intervertebral cages of various shapes/sizes. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- ARTFX Medical Cervical PEEK Cages ([K240893](/device/K240893.md))
- ARTFX Lumbar PEEK Cages ([K240889](/device/K240889.md))
- Px® PEEK IBF System, Px HA® PEEK IBF System, TxHA™ PEEK IBF System ([K180078](/device/K180078.md))
- Cascadia Lateral, K2M ([K172941](/device/K172941.md))
- FIX-C PEEK Anterior Cervical Interbody System ([K212266](/device/K212266.md))
- ROMERO CERVICAL CAGE ([K212358](/device/K212358.md))
- Lucent® Intervertebral Body Fusion Device ([K152011](/device/K152011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 19, 2025

Mikron Makina Sanayi Ve Tic. Ltd. STI.
Mr. Mustafa Ekiz
Medical Devices Manager
Ivedik OSB Mah. 1372. Sok. No:31 Yenimahalle
Yenimahalle, Ankara 06378
Turkey

Re: K252781
Trade/Device Name: MSFX MIKRON PEEK CAGES
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: ODP, OVE, MAX
Dated: November 26, 2025
Received: November 26, 2025

Dear Mr. Ekiz:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252781 - Mustafa Ekiz
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252781 - Mustafa Ekiz
Page 3

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252781 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  MSFX MIKRON PEEK CAGES  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  MSFX Mikron Cervical PEEK Cages are made to be implanted into the appropriate vertebral section to help provide stability for spinal fusion after a diseased cervical disc producing neck and/or armpain is removed during spinal decompression for patients who have had six weeks of non-operative treatment.Cervical PEEK Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level(C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.Cervical PEEK Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and implanted with auto graft/autologous bone graft. They are to be used with supplemental fixation.

MSFX Mikron Lumbar PEEK Cages are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment.The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) #: K252781
510(k) Summary
Prepared on: 2025-11-26

# Contact Details
21 CFR 807.92(a)(1)

|  Applicant Name | MIKRON MAKINA SAN. VE TIC. LTD. STI.  |
| --- | --- |
|  Applicant Address | Ivedik OSB Mah.1372. sok. No:31 Yenimahalle/Ankara/TURKEY YENIMAHALLE ANKARA 06378 Turkey  |
|  Applicant Contact Telephone | 03128020066  |
|  Applicant Contact | Mr. MUSTAFA EKIZ  |
|  Applicant Contact Email | mustafaekiz@mikronmakina.com  |

# Device Name
21 CFR 807.92(a)(2)

|  Device Trade Name | MSFX MIKRON PEEK CAGES  |
| --- | --- |
|  Common Name | Intervertebral Fusion Device With Bone Graft, Cervical, Intervertebral Fusion Device With Bone Graft, Lumbar  |
|  Classification Name | Intervertebral Body Fusion Device  |
|  Regulation Number | 888.3080  |
|  Product Code(s) | ODP, OVE, MAX  |

# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)

|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
| --- | --- | --- |
|  K240893 | ARTFX Medical Cervical PEEK Cages | ODP  |
|  K240889 | ARTFX Lumbar PEEK Cages | MAX  |
|  K180078 | Px® PEEK IBF System, Px HA® PEEK IBF System, TxHA™ PEEK IBF System | MAX  |
|  K172941 | Cascadia Lateral, K2M | MAX  |
|  K212266 | FIX-C PEEK Anterior Cervical Interbody System | ODP  |
|  K212358 | ROMERO CERVICAL CAGE | ODP  |
|  K152011 | Lucent® Intervertebral Body Fusion Device | MAX  |

# Device Description Summary
21 CFR 807.92(a)(4)

The MSFX Mikron Peek Cages are intervertebral body fusion devices to be implanted into appropriate cervical and lumbar spine. MSFX Mikron Peek Cages are composed of:
- MSFX Mikron Cervical Peek Cages
- MSFX Mikron Lumbar Peek Cages

MSFX Mikron Cervical Peek cages are designed to maintain the height of the intervertebral space. It is used in surgical procedures where

{5}

two or more vertebrae are connected or fused together to achieve fusion in spinal disorders. It supports bone fusion and is available in different shapes and sizes. The cage body is made from PEEK (ASTM F2026), and the pin, marker and screws are produced from Ti6Al4V ELI (ASTM F136). They are to be used with supplemental fixation, placed via anterior approach and implanted with autograft/autologous bone graft.

MSFX Mikron Lumbar Peek cages are designed to maintain the height of the intervertebral space. It is used in surgical procedures where two or more vertebrae are connected or fused together to achieve fusion in spinal disorders. It supports bone fusion and is available in different shapes and sizes. The cage body is made from PEEK (ASTM F2026), and the pin, marker and expansion mechanism are produced from Ti6Al4V ELI (ASTM F136). The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.

## Intended Use/Indications for Use

21 CFR 807.92(a)(5)

MSFX Mikron Cervical PEEK Cages are made to be implanted into the appropriate vertebral section to help provide stability for spinal fusion after a diseased cervical disc producing neck and/or arm pain is removed during spinal decompression for patients who have had six weeks of non-operative treatment. Cervical PEEK Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level(C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Cervical PEEK Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and implanted with auto graft/autologous bone graft. They are to be used with supplemental fixation.

MSFX Mikron Lumbar PEEK Cages are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.

## Indications for Use Comparison

21 CFR 807.92(a)(5)

MSFX Mikron Peek Cages have the same indications for use in comparison to the predicate devices.

## Technological Comparison

21 CFR 807.92(a)(6)

MSFX Mikron Peek Cages have the same technological characteristics in comparison to the predicate devices.

## Non-Clinical and/or Clinical Tests Summary &amp; Conclusions

21 CFR 807.92(b)

MSFX Mikron Cervical/Lumbar Peek Cages were tested biomechanically according to the following standards:

- ASTM F2077 :Static Axial Compression Test, Static Axial Compression Shear Test, Static Torsion Test, Dynamic Axial Compression Test, Dynamic Axial Compression Shear Test, Dynamic Torsion Test and

- ASTM F2267 :Subsidence

Based on the data presented in this submission, the subject device is determined to be substantially equivalent to the predicates in terms of its biomechanical performance. We can conclude that MSFX Mikron Spinal Cages are as safe, as effective, and perform as well as the legally predicate devices identified in this submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K252781](https://fda-staging.innolitics.com/device/K252781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
