CL1902A and CL2103F are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
Device Story
LCD monitors (CL1902A, CL2103F) designed for general radiography image display; input via DVI, VGA, or BNC interfaces; utilizes LED backlight panels with brightness stabilization control circuits to maintain luminance consistency; anti-glare screens reduce reflection. Operated by trained medical practitioners in clinical settings; output provides high-definition visual representation of radiological data for diagnostic review. Devices are calibrated to DICOM Part 3.14 standards. CL1902A features physical OSD buttons; CL2103F features capacitive touch OSD. Benefit includes high-precision, stable imaging for clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated per FDA guidance 'Display Devices for Diagnostic Radiology' and AAPM TG18 guidelines. Metrics included spatial resolution, pixel defects, artifact presence, temporal response, luminance (max/min/achievable), DICOM GSDF conformance, and color tracking. No clinical or animal data required.
Technological Characteristics
TFT Color LCD panels (IPS) with LED backlights. CL1902A: 19-inch (1280x1024), DVI/VGA/BNC. CL2103F: 21.3-inch (1600x1200), DVI/VGA. Integrated brightness stabilization control circuit. Anti-glare screen. Power: AC100-240V. Compliance: DICOM Part 3.14, VESA DPMS, EPA power management.
Indications for Use
Indicated for trained medical practitioners to display radiological images for review, analysis, and diagnosis. Not intended for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 19, 2025
Shenzhen Beacon Display Technology Co., Ltd.
Yafei Li
Safety Engineer
15F, Building 6, Hengda Shishang Huigu (East)
Fulong Road, Dalang Subdistrict, Longhua
Shenzhen, Guangdong 518109
China
Re: K252780
Trade/Device Name: LCD Monitor (CL1902A, CL2103F)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: PGY
Dated: November 18, 2025
Received: November 18, 2025
Dear Yafei Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252780 - Yafei Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K252780 - Yafei Li
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252780 | ? |
| Please provide the device trade name(s). | | ? |
| LCD Monitor (CL1902A, CL2103F) | | |
| Please provide your Indications for Use below. | | ? |
| CL1902A and CL2103F are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
LCD Monitors CL1902A, CL2103F
K252780
# 510(k) Summary
## 1. Submitter's Information
Name of Sponsor: Shenzhen Beacon Display Technology Co., Ltd.
Address: 15F, Building 6, Hengda Shishang Huigu(East), Fulong Road, Dalang Subdistrict, Longhua, Shenzhen, 518109 China
Contact Name: Li Yafei
Telephone No.: +86-024-88087610
Fax No.: +86-024-88087629
Email Address: liyf@beacon-display.cn
Date of Prepared: July 24,2025
## 2. Trade Name, Common Name, Classification
- Trade Name/Model: LCD Monitor (CL1902A, CL2103F)
- Common Name: Display, diagnostic radiology
- Classification Name: Medical image management and processing system
- Regulation Number: 21 CFR 892.2050
- Product code: PGY
- Classification Panel: Radiology
- Device Class: II
## 3. Identification of Predicate Device(s)
The identified predicate within this submission is as follows:
EIZO Corporation, RadiForce MX217, K230684;
EIZO Corporation, RadiForce MX194, K180961.
## 4. Description of the Device
1.3MP LCD Monitor CL1902A is a 19-inch TFT LCD monitor, which is specifically designed to provide the high definition image output for general radiography.
This product has been strictly calibrated so that it can meet DICOM Part 3.14 and other standards. It
510(k) Summary
{5}
LCD Monitors CL1902A, CL2103F
K252780
uses the latest generation of LED backlight panel, supporting resolution 1280 x 1024. The built-in brightness stabilization control circuit makes sure the brightness of this monitor is stable, so this product meets the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.
2MP LCD Monitor CL2103F is a 21.3-inch TFT LCD monitor, which is specifically designed to provide the high definition image output for general radiography.
This product has been strictly calibrated so that it can meet DICOM Part 3.14 and other standards. It uses the latest generation of LED backlight panel, supporting resolution 1600 x 1200. The built-in brightness stabilization control circuit makes sure the brightness of this monitor is stable, so this product meets the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.
| Component / Feature | CL1902A | CL2103F |
| --- | --- | --- |
| Main Board Interfaces | DVI + VGA + BNC | DVI + VGA |
| Power Board | Built-in | Built-in |
| OSD Control Board | Physical buttons | Capacitive touch |
# 5. Indications for use
CL1902A and CL2103F are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
# 6. Substantial Equivalence
## 6.1 Comparison table
Table 1 General Comparison of CL1902A
| Comparison Item | Proposed Device | Predicate Device | Explanation of Differences |
| --- | --- | --- | --- |
| | CL1902A | RadiForce MX194 | |
| Display technology | | | |
| | TFT Color
LCD Panel (IPS) | TFT Color
LCD Panel (VA) | Different Panel |
510(k) Summary
{6}
LCD Monitors CL1902A, CL2103F
K252780
| Screen size | | | |
| --- | --- | --- | --- |
| | 48.26cm/19.0"
Aspect ratio: 5:4 | 48.1cm / 19.0"
Aspect ratio: 5: 4 | Different Panel |
| Backlight type | | | |
| | LED | LED | - |
| Frame rate and refresh rate | | | |
| Digital Scanning Frequency (H, V) | 31 - 75 kHz / 59 - 61 Hz | 31 - 64 kHz / 59 - 61 Hz | - |
| Analog Scanning Frequency (H / V) | 15.5 – 75 kHz / 25 - 60 Hz | 24.8 - 80 kHz / 50 - 75 Hz | Different design scheme |
| Pixel pitch | | | |
| | 0.294 x 0.294 mm | 0.294 x 0.294 mm | - |
| Display Interface | | | |
| Input video signals | DVI-D x 1,
D-Sub mini 15 pin x 1
BNC x 3 | DVI-D x 1,
DisplayPort x 1,
D-Sub mini 15 pin x 1 | Different design scheme |
| Output video signals | N/A | N/A | - |
| Video bandwidth | | | |
| | DVI : 162MHz
Analog : 162MHz | DVI : 108MHz
DisplayPort : 108MHz
Analog : 135MHz | Different design scheme |
| Ambient light sensing | | | |
| Ambient light sensor | N/A | Yes | Different design scheme |
| Luminance calibration tools: | | | |
| | Integrated backlight sensor | Integrated optical sensor
External optical sensor
Calibration software: RadiCS | Different design scheme |
| USB Ports | | | |
| | N/A | / | Different design scheme |
| Power Related Specifications | | | |
| Power Requirements | AC100-240V ±10%,
50/60Hz±3Hz, 1.1A-0.6A | AC 100 - 240 V: 50 / 60 Hz | Different design scheme |
| Maximum power consumption | <50W | <28W | Different design scheme |
510(k) Summary
{7}
LCD Monitors CL1902A, CL2103F
K252780
| Power save mode | <3W | <0.6W | Different design scheme |
| --- | --- | --- | --- |
| Power Management | VESA DPMS and EPA power saving management | VESA DPMS and EPA power saving management | - |
| Indications for Use | This product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. This product is not intended for mammography. | This product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. This product is not intended for mammography. | - |
Table 2 General Comparison of CL2103F
| Comparison Item | Proposed Device | Predicate Device | Explanation of Differences |
| --- | --- | --- | --- |
| | CL2103F | RadiForce MX217 | |
| Display technology | | | |
| | TFT Color LCD Panel (IPS) | TFT Color LCD Panel (IPS) | - |
| Screen size | | | |
| | 54.0cm / 21.3" Aspect ratio: 4: 3 | 54.0cm / 21.3" Aspect ratio: 3: 4 | Different Panel |
| Backlight type | | | |
| | LED | LED | - |
| Frame rate and refresh rate | | | |
| Digital Scanning Frequency (H, V) | 31 - 75 kHz / 59 - 61 Hz | 31 - 100 kHz / 59 - 61 Hz | Different design scheme |
| Analog Scanning Frequency (H / V) | 31 - 75 kHz / 59 - 61 Hz | N/A | Different design scheme |
| Pixel pitch | | | |
| | 0.27 x 0.27mm | 0.27 x 0.27mm | - |
| Display Interface | | | |
| Input video signals | DVI-D x 1, D-SUB MINI 15PIN x 1 | DVI-D x 1, DisplayPort x 1 | Different design scheme |
| Output video signals | N/A | DisplayPort x 1 (daisy chain) | Different design scheme |
510(k) Summary
{8}
LCD Monitors CL1902A, CL2103F
K252780
| Video bandwidth | | | |
| --- | --- | --- | --- |
| | DVI: 25-162MHz
VGA: 25-162MHz | DVI : 25-164.5MHz
DisplayPort : 25-164.5MHz | Different design scheme |
| Ambient light sensing | | | |
| Ambient light sensor | Yes | Yes | Different design scheme |
| Luminance calibration tools: | | | |
| | Integrated backlight sensor | Integrated optical sensor
External optical sensor
Calibration software: RadiCS | Different design scheme |
| USB Ports | | | |
| | N/A | / | Different design scheme |
| Power Related Specifications | | | |
| Power Requirements | AC100-240V ±10%, 50/60Hz±3Hz, 1.1A-0.6A | AC 100 - 240 V: 50 / 60 Hz | Different design scheme |
| Maximum power consumption | <50W | <54W | Different design scheme |
| Power save mode | <3W | <0.6W | Different design scheme |
| Power Management | VESA DPMS and EPA power saving management | VESA DPMS and EPA power saving management | - |
| Indications for Use | This product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. This product is not intended for mammography. | This product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. This product is not intended for mammography. | - |
The technological characteristics differences discussed above do not affect the safety and the effectiveness of CL1902A and CL2103F.
510(k) Summary
{9}
LCD Monitors CL1902A, CL2103F
K252780
## 6.2 Performance Testing
According to the instructions in "Guidance for Industry and Food and Drug Administration Staff Display Devices for Diagnostic Radiology", the bench tests were performed on CL1902A and CL2103F as below.
- Measurement of spatial resolution
- The maximum number allowed for each type of pixel defects/faults
- Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
- Measure the temporal response using the typical data provided by the panel manufacturer
- Measure the maximum, minimum, achievable, and recommended luminance
- Verification of the conformance to DICOM GSDF as specified in Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- Measurement of Color tracking (primary colors and color gamut)
The test results showed that CL1902A and CL2103F are with display characteristics equivalent to those of the predicate devices.
No animal or clinical testing is needed for CL1902A and CL2103F.
## 7. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd. concludes that:
- The intended uses of the proposed devices (CL1902A and CL2103F) are equivalent to those of the predicate devices.
- The differences in technological characteristics between the proposed devices (CL1902A and CL2103F) and the predicate devices do not impact safety or effectiveness and do not introduce new risks.
- Bench testing demonstrates that the display characteristics of the proposed devices (CL1902A and CL2103F) are equivalent to those of the predicate devices.
- The data support the devices are substantially equivalent to the predicates.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.