K252749 · Shanghai Medconn Medical Technology Co., Ltd. · PDJ · Jan 6, 2026 · Clinical Chemistry
Device Facts
Record ID
K252749
Device Name
Medconn 8K Glycated Hemoglobin Test System
Applicant
Shanghai Medconn Medical Technology Co., Ltd.
Product Code
PDJ · Clinical Chemistry
Decision Date
Jan 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1373
Device Class
Class 2
Indications for Use
Medconn 8K Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood or hemolysate with ion-exchange high performance liquid chromatography (HPLC) using the Medconn 8K HbA1c Assay Kit (HPLC) on the Medconn 8K Glycated Hemoglobin Analyzer, models MQ-8000 and MQ-8000PT. Hemoglobin A1c measurements are used as an aid in diagnosis of diabetes, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.
Device Story
System uses ion-exchange HPLC to quantify HbA1c in human whole blood or hemolysate. Input: whole blood samples (K2-EDTA). Process: automatic hemolysis/dilution; buffer gradient delivery to analytical cartridge; separation of hemoglobin species based on ionic interactions; absorbance measurement at 415nm. Output: digital chromatogram, HbA1c concentration (NGSP % and IFCC mmol/mol). Used in clinical laboratory settings by professional operators. Output used by clinicians to diagnose diabetes, identify at-risk patients, and monitor long-term glycemic control. Benefits: provides standardized, precise HbA1c measurements for diabetes management.
Clinical Evidence
Bench testing only. Precision studies (n=720 measurements per sample) showed total CV <1.5%. Linearity confirmed across 3.0-17.0% HbA1c range (R2=0.9999). Interference testing confirmed no significant bias from endogenous substances, exogenous drugs, or hemoglobin variants (HbS, HbC, HbD, HbE, HbA2, HbF). Method comparison with predicate (n=139 samples) demonstrated high correlation (Deming slope 0.9868-0.9855) and acceptable total error.
Indicated for quantitative determination of HbA1c in human whole blood or hemolysate for diagnosis of diabetes, identification of patients at risk for diabetes, and monitoring long-term blood glucose control in individuals with diabetes mellitus.
Regulatory Classification
Identification
A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
Special Controls
(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
# ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K252749
B Applicant
Shanghai Medconn Medical Technology Co., Ltd.
C Proprietary and Established Names
Medconn 8K Glycated Hemoglobin Test System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PDJ | Class II | 21 CFR 862.1373 - Hemoglobin A1c Test System | CH - Clinical Chemistry |
| LCP | Class II | 21 CFR 864.7470 - Glycosylated hemoglobin assay | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Whole Blood Glycosylated Hemoglobin (HbA1c)
C Type of Test:
Quantitative Ion exchange High Performance Liquid Chromatography (HPLC)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K252749 - Page 2 of 14
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
Medconn 8K Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood or hemolysate with ion-exchange high performance liquid chromatography (HPLC) using the Medconn 8K HbA1c Assay Kit (HPLC) on the Medconn 8K Glycated Hemoglobin Analyzer, models MQ-8000 and MQ-8000PT.
Hemoglobin A1c measurements are used as an aid in diagnosis of diabetes, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
- The device should not be used in newborns, pregnant patients, patients with heterozygous sickle cell trait, hemolytic diseases and recent significant or chronic blood loss.
- The device should not be used in patients during pregnancy for diagnosis of gestational diabetes.
- In cases of rapidly evolving type 1 diabetes the increase of HbA1c values might be delayed compared to the acute increase in glucose concentrations. In these conditions diabetes mellitus must be diagnosed based on plasma glucose concentration and/or the typical clinical symptoms.
- The device should not be used to diagnose or monitor patients with conditions that alter the life span of the red blood cells, such as: iron deficiency and hemolytic anemia, various hemoglobinopathies, thalassemia, hereditary spherocytosis, malignancies and severe chronic hepatic and renal disease polycythemia or post-splenectomy. For example samples of patients with hemolytic anemia may present a decreased value of HbA1c due to a shortened life span of their red blood cells. The magnitude of this effect depends on the severity of the anemia. Samples of patients with polycythemia or post-splenectomy patients may present an increased value of HbA1c due to the relatively long life span of their red blood cells.
### D Special Instrument Requirements:
The Medconn 8K Glycated Hemoglobin Analyzer
## IV Device/System Characteristics:
### A Device Description:
The Medconn 8K Glycated Hemoglobin Test System consists of the following components:
{2}
- The Medconn 8K Glycated Hemoglobin Analyzer; models MQ-8000 and MQ-8000PT: automated instrument for running the Medconn 8K HbA1c Assay (HPLC) to measure the content of HbA1c in human whole blood. These analyzers are identical except the sample disc cover colors. The color for model MQ-8000 analyzer is blue, and the color for the MQ-8000PT is black.
- The Medconn 8K HbA1c Assay Kit (HPLC): distributed with 200-test to 10,000-test packages that contains HbA1c reagents A, B, C. These reagents are succinate buffers with different ionic strengths used for elution of different substances in HPLC.
- The Medconn Hemoglobin A1c Calibrator: lyophilized powers containing hemoglobin, 1 vial of Calibrator Level 1 (20 uL), and 1 vial of Calibrator Level 2 (20 uL).
- The Medconn Hemoglobin A1c Control (2 levels): lyophilized powders containing hemoglobin, which are reconstituted with labeled volume of distilled water or are directly reconstituted into diluted samples using haemolyser.
- The Medconn HbA1c Haemolyser (reagent H): used for pretreatment of blood for HbA1c testing, containing deionized water, a surfactant, and less than 0.05% of sodium azide as a preservative.
- The Medconn 8K HbA1c Column Kit (HPLC): assembled HPLC columns filled with hydrophilic polymer for use with the Medconn MQ-8000 and MQ-8000PT Glycated Hemoglobin Analyzers.
## B Principle of Operation:
The device is based on the principle of chromatographic separation of HbA1c in high performance liquid chromatography (HPLC). A high-pressure pumping system delivers a buffer solution to an analytical cartridge and detector. The analyzer has two modes: whole blood mode and manually hemolyzed and pre-diluted sample mode. Whole blood samples undergo an automatic hemolysis and dilution process before being introduced into the analytical flow path. The manually hemolyzed and pre-diluted samples loaded in sample cups at a designated location are directly introduced for analysis.
A programmed buffer gradient of increasing ionic strength delivers the sample to the analytical cartridge where the hemoglobin species are separated based upon their ionic interactions with the cartridge material and the buffer gradient. The separated hemoglobin species then pass through the flow cell where changes in the absorbance are measured at 415nm and recorded as a digital chromatogram.
The software performs an analysis of the hemoglobin peaks in the chromatogram, recording information including retention time, peak area, and relative peak area of the detected substance over the total peak area of all substances. Peaks identified as target analytes are calibrated to generate a report and a chromatogram for each sample.
## C Instrument Description Information:
1. Instrument Name:
Medconn 8K Glycated Hemoglobin Analyzer; models MQ-8000 and MQ-8000PT
2. Specimen Identification:
K252749 - Page 3 of 14
{3}
Sample identification is via barcoding or manual entry.
## 3. Specimen Sampling and Handling:
Venous whole blood samples collected in K2-EDTA can be stored at 2-8°C for 5 days and at -20°C for 30 days.
Hemolysate should be stored at 2-8°C for less than 8 hours.
## 4. Calibration:
Calibration, using the Medconn Hemoglobin A1c Calibrators, should be performed once following the use of a new batch of the Medconn 8K HbA1c assay kit, the installation of a new chromatographic column, instrument repair, maintenance, or replacement of key components. Additional calibration may be performed at the discretion of the laboratory.
## 5. Quality Control:
Quality control of the Medconn 8K Glycated Hemoglobin Test System can be conducted using the Medconn Hemoglobin A1c Controls (2 levels). The labeling for the device includes quality control recommendations and instructions.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
D-100 HbA1c, D-100 HbA1c Calibrator Pack
### B Predicate 510(k) Number(s):
K151321
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | K252749 | K151321 |
| --- | --- | --- |
| Device Trade Name | Medconn 8K Glycated Hemoglobin Test System | D-100 HbA1c |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for the quantitative determination of hemoglobin A1c in human whole blood as an aid in diagnosis of diabetes, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term | Same |
K252749 - Page 4 of 14
{4}
| | blood glucose control in individuals with diabetes mellitus. | |
| --- | --- | --- |
| Test principle | Ion exchange HPLC | Same |
| Sample Types | Human whole blood and hemolysate | Same |
| Standardization/ Traceability | Traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP) | Same |
| General Device Characteristic Differences | | |
| Matrices | K2-EDTA | K2-EDTA, K3-EDTA Potassium Oxalate/ Sodium Fluoride, Sodium Citrate, Sodium Heparin, Lithium Heparin |
| Measuring range | 3.0% to 17.0% (NGSP); 9.3 - 162.3 mmol/mol HbA1c (IFCC) | 3.5 to 20% (NSGP); 15 – 195 mmol/mol HbA1c (IFCC) |
VI Standards/Guidance Documents Referenced:
- CLSI EP05-A3 (Reaffirmed: September 2019): Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP06 2nd Edition: Evaluation of the Linearity of Quantitative Measurement Procedures
- CLSI EP07 3rd Edition: Interference Testing in Clinical Chemistry
- CLSI EP09c 3rd Edition: Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP25-A (Replaces EP25-P): Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
- IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)]
- IEC 61326-1 Edition 3.0 2020-10: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC 61326-2-6 Edition 3.0 2020-10: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
- IEC 81001-5-1 Edition 1.0 2021-12: Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
- AAMI TIR69:2017/(R2020): Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems
- IEEE ANSI USEMCSC C63.27-2021: American National Standard for Evaluation of Wireless Coexistence
K252749 - Page 5 of 14
{5}
- ISO 17511 Second edition 2020-04: In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples
## VII Performance Characteristics (if/when applicable):
### 1. Analytical Performance:
Since the analyzer models are functionally identical and only differ in sample disc colors, the Medconn 8K Glycated Hemoglobin Analyzer model MQ-8000 was used as the representative model for the studies described below to support both analyzer models.
### 1. Precision/Reproducibility:
Precision studies were conducted with the Medconn 8K Glycated Hemoglobin Test System using four K2-EDTA venous whole blood samples at concentrations near 5%, 6.5%, 8%, and 12% HbA1c (S1, S2, S3, and S4 respectively) and two quality control samples at concentrations near 6% and 11% HbA1c (QC1 and QC2 respectively). Samples were either analyzed as whole blood samples or as manually prepared hemolysate samples. Samples were analyzed on three Medconn 8K Glycated Hemoglobin Analyzer instruments with three lots of reagents in duplicates per run, two runs per day, for 20 days. For each sample there were 720 measurements. Results in NGSP and IFCC units are shown in the tables below.
All Analyzers Combined, Whole Blood, % HbA1c (NGSP) units
| Sample | Mean Values (%) | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 5.13 | 0.89 | 0.18 | 0.06 | 0.06 | 0.00 | 0.91 |
| S2 | 6.70 | 0.83 | 0.25 | 0.00 | 0.09 | 0.00 | 0.87 |
| S3 | 8.01 | 0.71 | 0.26 | 0.11 | 0.00 | 0.00 | 0.76 |
| S4 | 12.16 | 0.62 | 0.17 | 0.17 | 0.07 | 0.00 | 0.66 |
| QC1 | 5.50 | 1.17 | 0.23 | 0.00 | 0.12 | 0.09 | 1.20 |
| QC2 | 11.06 | 0.70 | 0.03 | 0.14 | 0.02 | 0.01 | 0.72 |
Note: Negative variance component values have been set to zero
All Analyzers Combined, Whole Blood, mmol/mol (IFCC) units
| Sample | Mean Values mmol/mol | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 32.52 | 1.17 | 0.35 | 0.00 | 0.00 | 0.09 | 1.21 |
| S2 | 49.72 | 0.92 | 0.25 | 0.20 | 0.00 | 0.00 | 0.97 |
| S3 | 63.96 | 0.72 | 0.31 | 0.11 | 0.06 | 0.00 | 0.79 |
| S4 | 109.34 | 0.69 | 0.11 | 0.20 | 0.08 | 0.00 | 0.73 |
| QC1 | 36.56 | 1.44 | 0.42 | 0.00 | 0.07 | 0.13 | 1.49 |
| QC2 | 97.37 | 0.81 | 0.10 | 0.17 | 0.06 | 0.00 | 0.83 |
Note: Negative variance component values have been set to zero
K252749 - Page 6 of 14
{6}
All Analyzers Combined, Hemolysate, % HbA1c (NGSP) units
| Sample | Mean Values (%) | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 5.15 | 0.95 | 0.07 | 0.18 | 0.00 | 0.05 | 0.97 |
| S2 | 6.41 | 0.89 | 0.00 | 0.00 | 0.15 | 0.00 | 0.86 |
| S3 | 7.91 | 0.75 | 0.19 | 0.06 | 0.02 | 0.00 | 0.78 |
| S4 | 12.05 | 0.70 | 0.16 | 0.05 | 0.07 | 0.00 | 0.73 |
| QC1 | 5.50 | 1.10 | 0.26 | 0.21 | 0.23 | 0.00 | 1.17 |
| QC2 | 11.07 | 0.67 | 0.24 | 0.10 | 0.00 | 0.05 | 0.71 |
Note: Negative variance component values have been set to zero
All Analyzers Combined, Hemolysate, mmol/mol (IFCC) units
| Sample | Mean Values mmol/mol | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 32.67 | 1.22 | 0.00 | 0.20 | 0.01 | 0.05 | 1.23 |
| S2 | 46.50 | 1.02 | 0.16 | 0.00 | 0.14 | 0.00 | 1.03 |
| S3 | 62.91 | 0.78 | 0.17 | 0.13 | 0.00 | 0.00 | 0.80 |
| S4 | 108.17 | 0.73 | 0.17 | 0.08 | 0.09 | 0.01 | 0.76 |
| QC1 | 36.58 | 1.39 | 0.40 | 0.32 | 0.29 | 0.00 | 1.49 |
| QC2 | 97.40 | 0.77 | 0.18 | 0.18 | 0.03 | 0.05 | 0.81 |
Note: Negative variance component values have been set to zero
2. Linearity:
Linearity was evaluated for the Medconn 8K Glycated Hemoglobin Test System with three lots of reagents using a dilution series consisting of 11 levels of HbA1c prepared by mixing high and low K2-EDTA whole blood pools. The expected sample concentrations were 3.1%, 4.5%, 5.9%, 7.3%, 8.7%, 10.1%, 11.5%, 12.9%, 14.3%, 15.7%, and 17.1% HbA1c. The samples were either analyzed as whole blood samples or were manually hemolyzed before analyzing. Three replicates were tested at each concentration. The measured mean values were compared to the expected values. The linear regressions for the three lots of reagents are the following:
Whole Blood
NGSP unit:
1st lot of reagent: Y=0.9913X + 0.0056, R2=0.9999
2nd lot of reagent: Y=0.9946X - 0.0393, R2=0.9999
3rd lot of reagent: Y=0.9922X - 0.0031, R2=0.9999
IFCC unit:
1st lot of reagent: Y=0.9913X - 0.1420, R2=0.9999
2nd lot of reagent: Y=0.9946X - 0.5566, R2=0.9999
3rd lot of reagent: Y=0.9922X - 0.2172, R2=0.9999
K252749 - Page 7 of 14
{7}
K252749 - Page 8 of 14
# Hemolysate
## NGSP unit:
- 1st lot of reagent: Y=0.9935X - 0.0374, R2=0.9999
- 2nd lot of reagent: Y=0.9916X - 0.0208, R2=0.9999
- 3rd lot of reagent: Y=0.9924X - 0.0295, R2=0.9999
## IFCC unit:
- 1st lot of reagent: Y=0.9935X - 0.5619, R2=0.9999
- 2nd lot of reagent: Y=0.9916X - 0.4257, R2=0.9999
- 3rd lot of reagent: Y=0.9924X - 0.5010, R2=0.9999
Results of the linearity study support the claimed measuring range of the device of 3.0% to 17% HbA1c (9.3-162.3 mmol/mol).
## 3. Analytical Specificity/Interference:
Interference studies were performed to assess the impact of endogenous substances on the performance of the Medconn 8K Glycated Hemoglobin Test System. Pooled K2-EDTA venous whole blood at two HbA1c levels (~6.5 and ~8.0% HbA1c) spiked with potential interferent was analyzed in replicates of two and compared to the same whole blood pool without interferent (control). The highest concentrations at which no significant interference (a difference of no more than ±6% from the control sample as defined by the sponsor) was observed are summarized below:
### Endogenous Substances
| Interferent | Highest Concentration Level Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Unconjugated Bilirubin | 21.3 |
| Conjugated Bilirubin | 19.2 |
| Rheumatoid Factors | 750 IU/mL |
| Total Protein | 21g/dL |
| Glucose | 2,000 |
| Triglycerides | 6,000 |
### Exogenous Substances
| Interferent | Highest Concentration Level Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Acetaldehyde | 60 |
| Acetaminophen | 200 |
| Acetylcysteine | 166 |
| Ampicillin-Na | 1,000 |
| Ascorbic acid | 100 |
| Cefoxitin | 2,500 |
| Cyclosporine | 5 |
| Doxycyclin | 50 |
| Heparin | 5,000 U/L |
| Ibuprofen | 500 |
| Levodopa | 20 |
| Methyldopa | 20 |
{8}
| Interferent | Highest Concentration Level Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Metronidazole | 200 |
| Phenylbutazone | 400 |
| Rifampicin | 64 |
| Theophylline | 100 |
## Cross Reactivity with Hemoglobin Derivatives:
Potential interference from Labile HbA1c, Carbamylated Hb, Acetylated Hb were evaluated. Two HbA1c concentrations of K2-EDTA pooled venous whole blood (~6.5 and ~8.0% HbA1c) were spiked with the prepared potential interferent. The %HbA1c values were compared to the same control sample with no potential interferent present. Significant interference was defined by the sponsor as ≥6% change in HbA1c value of the mean of the test samples relative to the mean of the control samples. The test results and conclusions are as follows:
- Labile A1c - (up to 2000 mg/dL glucose) does not interfere with this assay.
- Carbamylated Hb - (up to 10 mg/dL Potassium cyanate) does not interfere with this assay.
- Acetylated Hb - (up to 1000 mg/dL acetylsalicylic acid) does not interfere with this assay.
Results demonstrated that there was no cross reactivity with these hemoglobin derivatives at tested levels and the labeling states that at physiologically occurring concentrations there is no interference from labile A1c, carbamylated hemoglobin or acetylated hemoglobin.
## Hemoglobin Variant Interference:
Hemoglobin variant studies were performed using K2-EDTA venous whole blood samples known to contain hemoglobin variants HbS, HbC, HbD, HbE, HbA2, and HbF. The samples containing the hemoglobin variants were tested using the Medconn 8K Glycated Hemoglobin Test System and comparator methods that have been demonstrated to be free from interference by these hemoglobin variants. The sponsor defined significant interference was as ≥7% change in HbA1c values in the presence of the hemoglobin variants relative to the comparator method. The following tables summarize the tested samples and results:
Hemoglobin Variant Samples Tested in Study #1
| Hemoglobin Variant | Number of Samples | Variant Concentration Range (%) | Range of HbA1c Concentration (%) |
| --- | --- | --- | --- |
| HbS | 25 | 36 – 42 | 5.0 – 14.0 |
| HbC | 25 | 30 – 38 | 5.0 – 10.9 |
| HbD | 25 | 39 – 43 | 5.3 – 14.7 |
| HbE | 25 | 23 – 27 | 5.1 – 12.2 |
| HbA2 | 20 | 3.5– 5.6 | 5.2 – 9.0 |
| HbF | 23 | 3.0 – 27 | 5.5 – 16.5 |
K252749 - Page 9 of 14
{9}
Hemoglobin Variant Samples Tested in Study #2
| Hemoglobin Variant | Number of Samples | Variant Concentration Range (%) | Range of HbA1c Concentration (%) |
| --- | --- | --- | --- |
| HbE | 10 | 25.4 – 31.2 | 5.1 – 10.2 |
| HbF | 11 | 25.6 – 34.5 | 5.1 – 11.2 |
Summary of Hemoglobin Variant Test Results for Study #1
| Hemoglobin Variant | %Bias, Percent Relative Bias Observed Relative to the Comparator Method | | | |
| --- | --- | --- | --- | --- |
| | ~ 6.5 % HbA1c | | ~ 8.0% HbA1c | |
| | %Bias | Range of %Bias | %Bias | Range of %Bias |
| HbS | 1.52 | 0.00 to 3.03 | 1.81 | 0.00 to 3.61 |
| HbC | 2.31 | 1.54 to 3.08 | 1.34 | 0.00 to 2.67 |
| HbD | 0.85 | -1.64 to 3.33 | -1.83 | -2.44 to -1.22 |
| HbE | 1.67 | -3.17 to 3.33 | -0.67 | -1.33 to 0.00 |
| HbA2 | 0.02 | -1.49 to 1.52 | -0.63 | -2.53 to 1.28 |
| HbF | -0.08 | -3.33 to 3.08 | 0.05 | -2.44 to 2.53 |
Summary of Hemoglobin Variant Test Results for Study #2
| Hemoglobin Variant | %Bias, Percent Relative Bias Observed Relative to the Comparator Method | | | |
| --- | --- | --- | --- | --- |
| | ~ 6.5 % HbA1c | | ~ 8.0% HbA1c | |
| | %Bias | Range of %Bias | %Bias | Range of %Bias |
| HbE | 1.19 | 0.73 to 1.64 | 1.26 | 0.63 to 1.89 |
| HbF | 0.06 | -1.54 to 1.67 | 0.89 | 0.00 to 1.78 |
The results demonstrate no significant interference for HbS, HbC, HbE, HbD, HbA2, and HbF at the concentrations stated in the table above.
4. Assay Reportable Range:
The reportable range for this device is 3.0% – 17.0% HbA1c (NGSP) or 9.3 – 162.3 mmol/mol HbA1c (IFCC).
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability:
The Medconn 8K Glycated Hemoglobin Test System is traceable to the International Federation of Clinical Chemistry (IFCC) reference calibrators. The assay is certified by the National Glycohemoglobin Standardization Program (NGSP) and IFCC. The NGSP and IFCC certification expires in one year. See the NGSP website for current certification at http://www.NGSP.org.
K252749 - Page 10 of 14
{10}
HbA1c results are provided to users in two different units: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol). These two units are converted using the Master Equation: NGSP (%) = 0.09148 x IFCC (mmol/mol) + 2.152.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
See method comparison study below.
9. Carry-Over:
A Carryover study was performed and found to be acceptable.
## 2. Comparison Studies:
1. Method Comparison with Predicate Device:
Method comparison study was performed using 139 K2-EDTA venous whole blood samples (3.1 to 17.1% HbA1c concentration range) comparing the candidate device results with those from the predicate device (i.e., Bio-Rad D-100 HbA1c Test) performed at NGSP secondary reference laboratories. The study was conducted using both the whole blood and hemolysate modes of the Medconn 8K Glycated Hemoglobin Analyzer. The distribution of samples spanning the measuring interval is as follows:
| HbA1c levels, % | # of samples tested | % of sample tested |
| --- | --- | --- |
| ≤ 5 | 9 | 6.5 |
| 5.0 – 6.0 | 15 | 10.8 |
| 6.0 – 6.5 | 34 | 24.5 |
| 6.5 – 7.0 | 35 | 25.2 |
| 7.0 – 8.0 | 18 | 12.9 |
| 8.0 – 9.0 | 11 | 7.9 |
| > 9.0 | 17 | 12.2 |
| Total | 139 | 100 |
Regression analyses were performed for the candidate device versus the comparator method. Summary of results are as follows:
K252749 - Page 11 of 14
{11}
Whole blood mode, NGSP unit:
| Method | Slope | 95% CI | y-intercept | 95% CI |
| --- | --- | --- | --- | --- |
| Deming | 0.9868 | 0.9617 - 1.0120 | 0.1619 | -0.0062-0.3300 |
| Passing-Bablok | 1.0000 | 1.0000 - 1.0118 | 0.1000 | 0.0106-0.1000 |
Hemolysate mode, NGSP unit:
| Method | Slope | 95% CI | y-intercept | 95% CI |
| --- | --- | --- | --- | --- |
| Deming | 0.9855 | 0.9645 – 1.0066 | 0.1589 | 0.0173 - 0.3004 |
| Passing-Bablok | 1.0000 | 1.0000 - 1.0000 | 0.1000 | 0.1000 - 0.1000 |
Bias Estimation for the whole blood mode:
| HbA1c levels (%) | Deming | | Passing Boblok | |
| --- | --- | --- | --- | --- |
| | Bias (%) | %Bias | Bias (%) | %Bias |
| 5.0 | 0.0959 | 1.92 | 0.1000 | 2.00 |
| 6.5 | 0.0761 | 1.17 | 0.1000 | 1.54 |
| 8.0 | 0.0563 | 0.70 | 0.1000 | 1.25 |
| 12.0 | 0.0035 | 0.03 | 0.1000 | 0.83 |
Bias Estimation for the hemolysate mode:
| HbA1c levels (%) | Deming | | Passing Boblok | |
| --- | --- | --- | --- | --- |
| | Bias (%) | %Bias | Bias (%) | %Bias |
| 5.0 | 0.0864 | 1.73 | 0.1000 | 2.00 |
| 6.5 | 0.0647 | 0.99 | 0.1000 | 1.54 |
| 8.0 | 0.0429 | 0.54 | 0.1000 | 1.25 |
| 12.0 | -0.0151 | -0.13 | 0.1000 | 0.83 |
## Total Error calculations and estimation:
The bias estimation values determined in the method comparison study and the precision estimates determined in the precision study were used to determine the total error at each of the HbA1c levels listed in the tables below. Total error was calculated by the following equation: $\% \mathrm{TE} = |\% \mathrm{Bias}| + 1.96 \times \% \mathrm{CV} \times (1 + \% \mathrm{Bias} / 100)$
Total Error for the whole blood mode:
| Regression | HbA1c levels, % | %Bias | %CV | %TE |
| --- | --- | --- | --- | --- |
| Deming | 5.0 | 1.92 | 0.91 | 3.74 |
| | 6.5 | 1.17 | 0.87 | 2.90 |
| | 8.0 | 0.70 | 0.76 | 2.20 |
| | 12.0 | 0.03 | 0.66 | 1.32 |
| Passing Boblok | 5.0 | 2.00 | 0.91 | 3.82 |
| | 6.5 | 1.54 | 0.87 | 3.27 |
| | 8.0 | 1.25 | 0.76 | 2.76 |
| | 12.0 | 0.83 | 0.66 | 2.13 |
K252749 - Page 12 of 14
{12}
Total Error for the hemolysate mode:
| Regression | HbA1c levels, % | %Bias | %CV | %TE |
| --- | --- | --- | --- | --- |
| Deming | 5.0 | 1.73 | 0.97 | 3.66 |
| | 6.5 | 0.99 | 0.86 | 2.69 |
| | 8.0 | 0.54 | 0.78 | 2.08 |
| | 12.0 | -0.13 | 0.73 | 1.30 |
| Passing Boblok | 5.0 | 2.00 | 0.97 | 3.94 |
| | 6.5 | 1.54 | 0.86 | 3.25 |
| | 8.0 | 1.25 | 0.78 | 2.80 |
| | 12.0 | 0.83 | 0.73 | 2.27 |
2. Matrix Comparison:
Not applicable. Only K2-EDTA venous whole blood samples are to be used with this device.
3. Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
4. Clinical Cut-Off:
Not applicable.
5. Expected Values/Reference Range:
The sponsor provided the following expected values in their labeling: The following HbA1c ranges recommended by the American Diabetes Association (ADA) may be used as an aid in the diagnosis of diabetes mellitus$^{1-3}$
| Hemoglobin A1c | | Suggested Diagnosis |
| --- | --- | --- |
| NGSP % | IFCC mmol/mol | |
| ≥6.5 | ≥48 | Diabetic |
| 5.7–6.4 | 39–47 | Pre-Diabetic |
| <5.7 | <39 | Non-Diabetic |
K252749 - Page 13 of 14
{13}
$^{1}$ American Diabetes Association. Standards of Medical Care in Diabetes - 2013. Diabetes Care 2013; 36 (Suppl.1), S11-66.
$^{2}$ World Health Organization. Abbreviated Report of a WHO Consultation: WHO; 2011. Use of Glycated Haemoglobin (HbA1c) in the Diagnosis of Diabetes Mellitus.
$^{3}$ American Diabetes Association. Standards of Medical Care for Patients with Diabetes Mellitus. Diabetes Care 2002; 25 (Suppl.1), S33-49.
## 6. Other Supportive Instrument Performance Characteristics Data:
Not applicable.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K252749 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.