STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device; STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
K252743 · ETHICON, Inc. · GAM · Dec 22, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252743
Device Name
STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device; STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
Applicant
ETHICON, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252743 · Dec 22, 2025
STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device; STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control Device
ETHICON, Inc.
Hospital electronic medical records (EMR)
A retrospective chart review was conducted to evaluate clinical outcomes (anastomotic leakage, bleeding, infection, and stricture) in patients undergoing robotically-assisted hand-sewn gastrojejunal anastomosis using the subject device. The findings were used to justify the removal of a warning statement in the device labeling.
Retrospective chart review of hospital electronic medical records; Retrospective chart review; Follow-up/Duration: 30 days post-procedure
Patients undergoing robotically-assisted hand-sewn gastrojejunal anastomosis for Roux-en-Y gastric bypass (RYGB) using STRATAFIX Spiral Knotless Tissue Control Devices; Sample Size: 145
Not applicable for this study
Inpatient mortality, anastomotic leakage, anastomotic bleeding, bleeding of unspecified origin, intra-abdominal infection, reoperations, and anastomotic stricture
Indications for Use
The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.
Device Story
Antibacterial monofilament, synthetic absorbable suture; composed of poly(p-dioxanone) with IRGACARE® MP (triclosan) antibacterial agent. Available in unidirectional (barbed with fixation loop) and bidirectional (barbed with un-barbed center transition zone) configurations; armed with surgical needles. Used by surgeons for soft tissue approximation without requiring surgical knots; barbs anchor tissue. Intended for use in clinical settings; replaces traditional knotted sutures. Clinical benefit includes simplified tissue approximation and removal of previous contraindications for specific gastric bypass procedures.
Clinical Evidence
Retrospective chart review of 145 patients undergoing robotically-assisted hand-sewn gastrojejunal anastomosis for Roux-en-Y gastric bypass. Primary endpoints included mortality, anastomotic leakage, bleeding, and infection within 30 days. Results: 0% mortality, 0% leakage, 0% bleeding, 0% infection, 0% reoperations. 2.1% (n=3) experienced anastomotic stricture, resolved on follow-up.
Indicated for soft tissue approximation in patients where absorbable sutures are appropriate, including those undergoing robotically-assisted hand-sewn gastrojejunal anastomosis for Roux-en-Y gastric bypass.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
STRATAFIX™ Spiral PDS™ Plus Bidirectional Knotless Tissue Control Device (K192144)
STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device (K182873)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
Ethicon, Inc.
Samantha Mecker
Associate Director, Regulatory Affairs
1000 Rte. 202 South
Raritan, New Jersey 08869
Re: K252743
Trade/Device Name: STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable Polydioxanone Surgical Suture
Regulatory Class: Class II
Product Code: NEW
Dated: November 26, 2025
Received: November 26, 2025
Dear Samantha Mecker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252743 - Samantha Mecker
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252743 - Samantha Mecker
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE
-S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252743 | |
| Device Name STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices | |
| Indications for Use (Describe) The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252743 - Page 1 of 5
# 510(k) Summary (K252743)
Submitter: Ethicon Inc. a Johnson & Johnson company
1000 US-202
Raritan, New Jersey 08869
Contact Person: Samantha Mecker
Ethicon, Inc. a Johnson & Johnson company
e-mail: smecker@its.jnj.com
Phone: (732) 837-8599
Date Prepared: December 22, 2025
Device Trade Name: STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices
Device Common Name: Suture, Surgical, Absorbable, Polydioxanone
Class: II
Classification Name: Suture, Surgical, Absorbable, Polydioxanone
(21 CFR 878.4840)
Product Code: NEW
| Legally Marketed Predicate Devices | 510(k) Number | Product Code |
| --- | --- | --- |
| STRATAFIX™ Spiral PDS™ Plus Bidirectional Knotless Tissue Control Device | K192144 | NEW |
| STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Device | K182873 | NEW |
{5}
K252743 - Page 2 of 5
## Device Description:
STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are antibacterial monofilament, synthetic absorbable devices consisting of dyed (violet) polyester, poly(p-dioxanone), the empirical molecular formula of which is (C4H6O3)x. The devices contain IRGACARE®* MP (triclosan), a broad spectrum antibacterial agent, at no more than 2360 µg/m. The pigment for the violet dye is D&C Violet No. 2. Polydioxanone polymer has been found to be nonallergenic, nonpyrogenic and elicits only a slight tissue reaction during absorption.
The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are available in Unidirectional and Bidirectional forms:
- The STRATAFIX™ Spiral PDS™ Plus Unidirectional Knotless Tissue Control Devices consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral PDS™ Plus Unidirectional Knotless Tissue Control Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. Barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
- The STRATAFIX™ Spiral PDS™ Plus Bidirectional Knotless Tissue Control Device consists of barbed suture material, armed with a surgical needle on each end. The device also contains an un-barbed center transition zone that facilitates the initiation of the device use. The barbs allow for tissue approximation without the need to tie surgical knots.
While the formation of barbs in the STRATAFIX™ Spiral PDS™ Plus Devices reduce the tensile strength relative to non-barbed suture material of the same size, tying of knots in non-barbed suture materials also reduces their effective strength. For this reason, the strength of the STRATAFIX™ Spiral PDS™ Plus Devices can be compared to USP knot strength of non-barbed sutures. Additionally, USP designations for diameter are used to describe the STRATAFIX™ Spiral PDS™ Plus Devices suture material after barbing, except for minor variation in suture diameter with a maximum overage of 0.1 mm.
## Indications for Use:
The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.
The indications for use are identical between the subject and predicate devices.
{6}
K252743 - Page 3 of 5
# Comparison of Technological Characteristics with the Predicate Devices:
There are no changes to the technological characteristic of the STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices. The subject devices differ from their respective predicate devices only in the Instructions for Use (IFU). The Instructions for Use of the subject devices are updated with the following change:
Ethicon will remove the warning related to use in gastrojejunal anastomoses for Roux-en-Y gastric bypass in the IFU for STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices.
Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] has been followed and it was determined that subject Devices are substantially equivalent to the predicate in that they share:
a) the same fundamental scientific technology,
b) the same intended use,
c) the same design,
d) the same materials,
e) equivalent packaging materials and configuration,
f) the same labeling components
g) the same sterilization process (Ethylene Oxide)
h) the same sterility assurance level (SAL) is 10-6.
# Performance Data:
A retrospective chart review of hospital electronic medical records was performed to support the update of the STRATAFIX Spiral PDS™ Plus Knotless Tissue Control Devices warning, evaluating robotically-assisted hand-sewn gastrojejunal anastomosis for Roux-en-Y gastric bypass (RYGB) using STRATAFIX Spiral Knotless Tissue Control Devices (KTCD).
A total of 145 patients met the selection criteria for the robotically-assisted hand-sewn gastrojejunal anastomosis cohort. Their demographics are shown in Table 1. There was no inpatient mortality at index and no anastomotic leakage, anastomotic bleeding, bleeding of unspecified origin, or intra-abdominal infection within 30 days post-procedure. There were no cases of reoperations. Three patients (2.1%) experienced anastomotic stricture, which was resolved on follow-up.
Taken together, we believe there is sufficient evidence to justify the removal of the warning statement regarding their use in gastrojejunal anastomoses for Roux-en-y gastric bypass.
{7}
K252743 - Page 4 of 5
Table 1. Demographics of Patient Group Population, n=145
| Age category | |
| --- | --- |
| 18-34 | 17.20% |
| 34-44 | 35.20% |
| 45-54 | 30.30% |
| 55-64 | 13.10% |
| 65-74 | 4.10% |
| 75+ | 0.00% |
| Female | 81.40% |
| Race | |
| White | 49.70% |
| Black | 42.10% |
| Other/unknown | 8.30% |
| Ethnicity | |
| Hispanic | 7.60% |
| Not Hispanic | 87.60% |
| Other/Unknown | 4.80% |
| Diabetes Status | |
| Diabetes without complications | 20.70% |
| Diabetes with complications | 8.30% |
| Smoking status | |
| Former smoker | 30.30% |
| Current smoker | 0.10% |
| Body Mass Index Category | |
| Normal weight: 20 to <25 | 0.00% |
| Overweight: 25 to <30 | 1.40% |
| Obesity: 30 to <35 | 6.20% |
| Obesity: 35.0 to 35.9 | 2.10% |
| Obesity: 36.0 to 36.9 | 2.80% |
| Obesity: 37.0 to 37.9 | 2.80% |
| Obesity: 38.0 to 38.9 | 5.50% |
| Obesity: 39.0 to 39.9 | 3.40% |
| Obesity: 40.0 to 44.9 | 28.30% |
| Obesity: 45.0 to 49.9 | 20.00% |
| Obesity: 50.0 to 59.9 | 21.40% |
| Obesity: 60.0 to 69.9 | 5.50% |
{8}
K252743 - Page 5 of 5
**Substantial Equivalence Conclusion:**
STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.