1927nm Thulium Laser System (PZ-DJG75-01, PZ-DJG75-02)
K252732 · Zhengzhou PZ Laser Slim Technology Co., Ltd. · GEX · Oct 23, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252732
Device Name
1927nm Thulium Laser System (PZ-DJG75-01, PZ-DJG75-02)
Applicant
Zhengzhou PZ Laser Slim Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1927nm thulium laser system is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).
Device Story
1927nm Thulium Laser System (models PZ-DJG75-01, PZ-DJG75-02) uses thulium-doped fiber laser module to deliver 1927nm pulsed laser energy. Energy is transmitted via optical fiber and handpiece to treatment area. Principle of operation relies on selective light absorption by water in skin to achieve soft tissue coagulation. Operator controls pulse width, dot matrix spacing, and treatment mode (Slide or Dot matrix) to optimize effects. Used in clinical settings by trained professionals. Output allows for precise dermatological treatment of pigmented lesions and actinic keratosis. Benefits include controlled tissue coagulation for dermatological conditions.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified through testing for spot size accuracy, energy output accuracy, and compliance with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, and ISO 10993 (cytotoxicity, sensitization, irritation).
Technological Characteristics
Thulium-doped fiber laser; 1927nm wavelength; 10W power; 0.2-20ms pulse duration; 45Hz-445Hz repetition rate; 300μm spot size. Fiber and handpiece delivery. Foot switch emission control. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2. Biocompatible per ISO 10993-5, 10993-10, 10993-23.
Indications for Use
Indicated for dermatological procedures requiring soft tissue coagulation, treatment of actinic keratosis, and benign pigmented lesions (lentigos, solar lentigos, ephelides) in patients requiring such dermatological care.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 23, 2025
Zhengzhou PZ Laser Slim Technology Co., Ltd
Yueming Li
Regulatory Affairs Specialist
3rd Floor, Building 1, No. 101 Jinbai Road, High-tech
Dev. Zone, 450001 Zhengzhou City, Henan Province
Zhengzhou,
China
Re: K252732
Trade/Device Name: 1927nm Thulium Laser System (PZ-DJG75-01, PZ-DJG75-02)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: August 28, 2025
Received: August 28, 2025
Dear Yueming Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252732 - Yueming Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252732 - Yueming Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU -S
Digitally signed by YAN FU
Date: 2025.10.23 12:52:49
-04'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252732 | ? |
| Please provide the device trade name(s). | | ? |
| 1927nm Thulium Laser System (PZ-DJG75-01, PZ-DJG75-02) | | |
| Please provide your Indications for Use below. | | ? |
| 1927nm Thulium Laser System is indicated for use in dermatological procedures requiring coagulation of
soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited
to, lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
The assigned 510(k) Number: K252732
# 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
1. Date of preparation: 08/28/2025
2. Sponsor Identification:
Zhengzhou PZ Laser Slim Technology Co., Ltd.
3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001
Zhengzhou City, Henan Province, People's Republic of China.
Contact Person: Shi Xiaojie
Position: RA
Tel: +86-18703657355
Fax: +86-371-55677886
Email: shixiaojie1225@163.com
3. Designated Submission Correspondent
Zhengzhou PZ Laser Slim Technology Co., Ltd.
3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001
Zhengzhou City, Henan Province, People's Republic of China.
Contact Person: Li Yueming
Tel: +86-18860362551
Fax: +86-371-55677886
Email: boluobao1010@qq.com
4. Identification of Proposed Device
Trade Name: 1927nm Thulium Laser System
Models: PZ-DJG75-01, PZ-DJG75-02
Common Name: Powered Laser Surgical Instrument
Regulatory Information
Classification Name: Powered Laser Surgical Instrument
Classification: II
{5}
510(k) Summary
Product Code: GEX
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Review Panel: General& Plastic Surgery
Intended Use: 1927nm thulium laser system is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos(age spots), solar lentigos (sun spots)and ephelides (freckles).
## 5. Device Description
1927nm Thulium Laser System contains a thulium-doped fiber laser module that uses a 1927nm pulsed laser output energy. The beam is directly irradiated to the treatment area through the handle and optical fiber. According to the selective light absorption theory, the 1927nm laser can be preferentially absorbed by the water in the skin, thus achieving the purpose of coagulation of soft tissue.
The operator can optimize the treatment effect of each application by controlling the pulse width, distance between the dot matrices and treatment mode of the pulsed laser.
There are 2 models included, PZ-DJG75-01, PZ-DJG75-02, the two models have the same intended use, mechanism of action and principle, only the step of pulse width under the dot matrix mode are different. The detailed differences shown as following:
Table 1 The differences between two models
| Model | Wavelength | Mode | Pulse width (±20%) | Step |
| --- | --- | --- | --- | --- |
| PZ-DJG75-01 | 1927nm±10nm | Slide | 0.2-2.0ms | 0.1ms |
| | | Dot matrix | 0.2-20.0ms | 0.1ms |
| PZ-DJG75-02 | 1927nm±10nm | Slide | 0.2-2.0ms | 0.1ms |
| | | Dot matrix | 0.2-20.0ms | 0.2ms |
## 6. Identification of Predicate Device
{6}
510(k) Summary
510(k) Number: K241406
Product Name: Lavieen
Manufacturer: WON TECH Co., Ltd
## 7. Identification of Reference Device
510(k) Number: K182173
Product Name: JOULE 1927nm Laser System
Manufacturer: Sciton, Inc
## 8. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards.
IEC 60601-1:2020 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
IEC 60601-2-22:2019, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.
IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements.
IEC 60601-1-2:2020, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic disturbances- Requirements and tests.
ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in Vitro cytotoxicity.
ISO 10993-10:2021, Biological Evaluation of Medical Devices, Part 10-Tests for skin sensitization.
ISO 10993-23:2021, Biological Evaluation of Medical Devices-Part 23: Tests for irritation.
Performance Testing for Spot Size Accuracy and Energy Output Accuracy.
{7}
510(k) Summary
## 9. Clinical Test Conclusion
No clinical study is included in this submission.
## 10. Substantially Equivalent (SE) Comparison
Table 1 General Comparison
| Item | Proposed Device | Predicate Device | Reference Device | |
| --- | --- | --- | --- | --- |
| Device Name | 1927nm Thulium Laser System | Lavieen | JOULE 1927nm Laser System | / |
| Classification regulation | 21CFR 878.4810 | 21CFR 878.4810 | 21CFR 878.4810 | Same |
| Classification Panel | General&Plastic Surgery | General&Plastic Surgery | General&Plastic Surgery | Same |
| class | II | II | II | Same |
| Product Code | GEX | GEX | GEX | Same |
| Common name | Powered laser surgical instrument | Powered laser surgical instrument | Powered laser surgical instrument | Same |
| Indication for use | 1927nm thulium laser system is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos | Lavieen is indicated for use in dermatological procedures requiring coagulation of soft tissue, treatment of actinic keratosis and treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots) and ephelides | JOULE 1927nm Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun | Same |
{8}
510(k) Summary
| | (age spots), solar lentigos (sun spots) and ephelides (freckles). | (freckles). | spots) and ephelides (freckles). | |
| --- | --- | --- | --- | --- |
| Prescription use or not | Prescription use | Prescription use | Prescription use | Same |
Table 2 Performance comparison
| Item | Proposed device | Predicate device | Reference Device | |
| --- | --- | --- | --- | --- |
| Laser Type | Thulium laser | Thulium laser | Thulium laser | Same |
| Laser Wavelength | 1927nm | 1927nm | 1927nm | Same |
| Aiming Beam | 658nm, <5mW | 658nm | -- | Same |
| Beam delivery | Fiber and Handpiece | Fiber and Handpiece | Articulated arm and Handpiece | Same |
| Emission control | Foot Switch | Foot Switch | Foot Switch | Same |
| Laser power | 10W | 10W | 12W | Same |
| Pulse Duration (Pulse Width) | 0.2-20ms | 0.1-20ms | Up to 20ms | Same |
| Pulse Repetition Rate(Frequency) | 45Hz-445Hz | 67-240Hz | 0-3 kHz | Same |
| Spot Size | 300μm | 300μm | 100-620μm | Same |
# SE Analysis:
The spot size of proposed device and predicate device is $300\mu \mathrm{m}$ . The maximum single-pulse energy is $20\mathrm{mJ}$ , which is related to the output power and pulse width., and the maximum energy density is $20 / (0.15^{2}\pi) = 282.9\mathrm{mJ} / \mathrm{mm}2$ .
The radiation energy in the treatment area is the product of the single pulse energy and the number of pulses, which is related to the spot spacing.
These parameters related to clinical use are exactly the same as the predicate device.
{9}
510(k) Summary
Compared with predicate device, the range of pulse repetition frequency of the Proposed device is slightly larger.
$$
\text{Frequency} = \frac{1}{(\text{Pulse width} + \text{Delay})}
$$
Setting a smaller delay, pulse frequency will increase, but at the same time can reduce the single dot matrix output time, will reduce the total operation time in clinical use. In conclusion, the pulse repetition rate does not affect the safety and effectiveness of device use.
In order to further confirm the rationality of the declared device pulse repetition frequency, a reference device was selected with a pulse repetition frequency range of 1-3kHz. The pulse repetition rate of proposed device is within its scope.
Table 3 Safety Comparison
| Item | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- |
| Patient Contact Materials and Biocompatibility | | | |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No evidence of sensitization | No evidence of sensitization | Same |
| Irritation | No evidence of irritation | No evidence of irritation | Same |
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | Same |
## 11. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device K241406.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.