K252677 · Poly Medicure Limited · FOZ · Nov 5, 2025 · General Hospital
Device Facts
Record ID
K252677
Device Name
Polyshield Safety IV Catheters
Applicant
Poly Medicure Limited
Product Code
FOZ · General Hospital
Decision Date
Nov 5, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Polyshield Safety IV Catheters are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. The 14 - 24G catheters may be used intravascularly with power injectors at a maximum pressure of 300 psi.
Device Story
Over-the-needle peripheral safety IV catheter; active spring-activated needle shield prevents needle-stick injuries. Models include 'BC' (blood control feature preventing flow until distal connection) and 'Adva' (quick flash back technology for vein access confirmation). Available in various gauges/lengths; optional hub wings. Used by clinicians in clinical settings for vascular access. Device provides blood flow/pressure monitoring or fluid delivery. Output is physical access to vascular system; clinician observes flash back for placement confirmation. Benefits include reduced needle-stick risk and improved venous access success.
Clinical Evidence
Bench testing only. Verification testing performed on new designs (fresh and end-of-shelf-life samples) including back leakage, flow rate, aspiration, leakage during pressurization, power injection, and tensile force. Testing conducted per ISO 10555-1:2023. Biocompatibility, sterilization, and human factors remain unchanged from predicate.
Technological Characteristics
Polyurethane catheter tube; stainless steel needle with back-cut ground bevel. Active spring-activated needle shield. Polyacetal (PA) slip ring. Female 6% Luer proximal end. Sterile (EO gas, SAL 10⁻⁶). MR Conditional. Complies with ISO 10555-1, ISO 10555-5, and USP <788>.
Indications for Use
Indicated for short-term ( <30 days) vascular access for blood sampling, blood pressure monitoring, or fluid administration in any patient population, considering patient size, solution appropriateness, and therapy duration. 14-24G catheters compatible with power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Polyshield and Polyguard Safety IV Catheters (K230616)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 5, 2025
Poly Medicare Limited
% Roger Gray
VP QA/RA
Donawa Lifescience
Piazza Albania 10
Rome, 00153
Italy
Re: K252677
Trade/Device Name: Polyshield Safety IV Catheters
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular Catheter
Regulatory Class: Class II
Product Code: FOZ
Dated: October 7, 2025
Received: October 7, 2025
Dear Roger Gray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252677 - Roger Gray
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252677 - Roger Gray
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Colleen J. Lawrimore -S
Colleen Lawrimore, Ph.D.
*For David Wolloscheck, Ph.D.*
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252677 | |
| Device Name Polyshield Safety IV Catheters | |
| Indications for Use (Describe) The Polyshield Safety IV Catheters are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. The 14 - 24G catheters may be used intravascularly with power injectors at a maximum pressure of 300 psi. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
POLYMED
510(k) Summary: K252677
Trade Name: Polyshield Safety IV Catheters
Type of 510(k) submission: Special
Date of Preparation: 5 November 2025
510(k) Owner and Submitter: Poly Medicare Ltd
Plot No. 115-117, Sector- 59
HSIIDC Industrial Area, Ballabgarh
Faridabad-121004, Haryana
INDIA
510(k) Contact: Mr Roger Gray
VP Quality and Regulatory
Donawa Lifescience Consulting
Piazza Albania 10
00153 Rome
Italy
Phone: +39 06 578 2665
Email: rgray@donawa.com
Common Name: Catheter, Intravascular, Therapeutic, Short-Term Less than 30 Days
Product Code: FOZ
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular catheter
Classification Panel: General Hospital
Device Class: Class II
Indications for Use:
The Polyshield Safety IV Catheters are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. The 14 - 24G catheters may be used intravascularly with power injectors at a maximum pressure of 300 psi.
Device Description:
The Polyshield Safety IV Catheters are over-the-needle, peripheral safety IV catheters that incorporate an active sharps prevention needle shield. The 'BC' versions also incorporate a blood control feature, and the 'Adva' versions include 'quick flash back' needle technology. The devices are available in multiple gauge sizes and lengths, in both with and without wings on the catheter hub. In more detail:
Polyshield: An over-the-needle, peripheral active safety IV catheter that incorporates a spring-activated needle shield to help prevent needle-stick injuries
Polyshield Adva: An over-the-needle, peripheral active safety IV catheter that incorporates a spring-activated needle shield to help prevent needle-stick injuries. The device is also provided with quick flash back (Adva needle) technology. Instant confirmation of blood flow along catheter body increases clinician's ability to successfully access the vein.
510(k) for Polyshield Safety IV Catheters
{5}
POLYMED
Polyshield BC:
An over-the-needle, peripheral active safety IV catheter with wings that incorporates a spring-activated needle shield to help prevent needle-stick injuries, together with a blood control feature that prevents blood flow until a distal connection is made.
Polyshield BC Adva:
An over-the-needle, peripheral active safety IV catheter with wings that incorporates a spring-activated needle shield to help prevent needle-stick injuries, together with a blood control feature that prevents blood flow until a distal connection is made. The device is also provided with quick flash back (Adva needle). Instant confirmation of blood flow along catheter body increases clinician's ability to successfully access the vein.
Color-coding of the catheter hub (and wings, where fitted) is used to help identify the catheter gauge size, in accordance with ISO 10555-5.
The major features of the available models are indicated in Table 1.
| Table 1: Polyshield major design features | | | | |
| --- | --- | --- | --- | --- |
| Feature | Polyshield | Polyshield Adva | Polyshield BC | Polyshield BC Adva |
| Safety mechanism type | Active | Active | Active | Active |
| Spring activated by button | Yes | Yes | Yes | Yes |
| Push-back mechanism | No | No | No | No |
| Blood control mechanism | No | No | Yes | Yes |
| Quick flash back | No | Yes | No | Yes |
| Hub wings | Optional | Optional | Optional | Optional |
Polyshield Safety IV Catheters are labeled for prescription use only (Rx only) and supplied sterile for single use with a sterility assurance level (SAL) of 10⁻⁶. Sterilization is achieved by ethylene oxide gas exposure.
The purpose of this submission is to introduce minor design changes to the Polyshield BC models and to add new catheter lengths to all four Polyshield models.
## Substantial equivalence
The predicate device selected for comparison with the revised design of Polyshield Safety IV Catheters is the original design of the devices, viz:
Predicate Device...Polyshield and Polyguard Safety IV Catheters
Sponsor...Poly Medicare Ltd
510(k) Number ...K230616
Clearance Date ...5 May 2023
FDA Product Code ...FOZ
Classification Name ...Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Regulation No:...21 CFR 880.5200
Class...II
## Predicate device comparison table:
Table 2 provides evidence of substantial equivalence of the subject device with the selected predicate device:
510(k) for Polyshield Safety IV Catheters
{6}
POLYMED
| Table 2: Predicate device comparison table | | | |
| --- | --- | --- | --- |
| Feature | Subject device (Polyshield Safety IV Catheters) | Predicate device (Polyshield and Polyguard Safety IV Catheters, K230616) | Similarity |
| Device description: | Polyshield: Over-the-needle, peripheral safety IV catheter that incorporates a spring-activated needle shield
'BC' models: Incorporate a blood control feature
'Adva' models: Incorporate a 'quick flash back' feature | Polyshield: Over-the-needle, peripheral safety IV catheter that incorporates a spring-activated needle shield
Polyguard: Over-the-needle, peripheral safety IV catheter that incorporates a push-back needle shield
'BC' models: Incorporate a blood control feature
'Adva' models: Incorporate a 'quick flash back' feature | Same as predicate, but without the Polyguard model |
| Indications for use | The Polyshield Safety IV Catheters are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. The 14 - 24G catheters may be used intravascularly with power injectors at a maximum pressure of 300 psi. | The Polyshield and Polyguard Safety IV Catheters are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. The 14 - 24G catheters may be used intravascularly with power injectors at a maximum pressure of 300 psi. | Same as predicate, but without the Polyguard model |
| Sharps injury protection feature? | Yes – active | Yes – active | Same |
| Sharps injury prevention mechanism | Polyshield and Polyshield BC: User-activated button initiates needle retraction into the needle-shielding barrel. Once the button has been pushed and the needle retracted, the user cannot override the shielding mechanism to re-expose the needle tip. | Polyshield and Polyshield BC: User-activated button initiates needle retraction into the needle-shielding barrel. Once the button has been pushed and the needle retracted, the user cannot override the shielding mechanism to re-expose the needle tip.
Polyguard and Polyguard BC: User-activated push-back mechanism retracts the needle into the needle-shielding barrel. Once activated, the user cannot override the shielding mechanism to re-expose the needle tip. | Same as predicate, but without the Polyguard model |
| Catheter tube material | Polyurethane | Polyurethane | Same |
| X-ray visible | Yes | Yes | Same |
| Needle material | Stainless steel | Stainless steel | Same |
| Needle distal end configuration | Back cut ground beveled needle | Back cut ground beveled needle | Same |
| Flashback visualization | Adva models | Adva models | Same |
510(k) for Polyshield Safety IV Catheters
Page 3 of 5
{7}
POLYMED
| Table 2: Predicate device comparison table | | | |
| --- | --- | --- | --- |
| Feature | Subject device (Polyshield Safety IV Catheters) | Predicate device (Polyshield and Polyguard Safety IV Catheters, K230616) | Similarity |
| Blood control | Yes, in BC models | Yes, in BC models | Same |
| Gauge sizes | 14 - 26G | 14 - 26G | Same |
| Catheter lengths | 14 mm – 48 mm | 19 mm – 45 mm | Different – see Discussion #1 |
| Catheter hub design | Catheter with wings hub design simplified to integrate hub and wings and improve operation | Original design with separate catheter hub and wings | Different – see Discussion #2 |
| Color-coding | Yes, according to ISO 10555-5 | Yes, according to ISO 10555-5 | Same |
| Proximal end configuration | Female 6 % Luer | Female 6 % Luer | Same |
| Single use | Yes | Yes | Same |
| Sterile | Yes, SAL 10^{-6} | Yes, SAL 10^{-6} | Same |
| Sterilization method | Ethylene oxide | Ethylene oxide | Same |
| Shelf life | 3 years | 3 years | Same |
| Physical properties | According to ISO 10555-1, ISO 10555-5 | According to ISO 10555-1, ISO 10555-5 | Same |
| Power injection usage? | Yes, 14 - 24G catheters up to 300 psi. | Yes, 14 - 24G catheters up to 300 psi. | Same |
| Biocompatibility | Biocompatible in accordance with ISO 10993 series and FDA guidance | Biocompatible in accordance with ISO 10993 series and FDA guidance | Same |
| Particulate contamination | Complies with USP <788> | Complies with USP <788> | Same |
| Environment of use | Rx only | Rx only | Same |
| Hub wings? | Yes, as an option | Yes, as an option | Same |
| Metal slip ring material | Polyacetal (PA) | Stainless steel 305 | Different – See Discussion #3 |
| MR compatibility | MR Conditional | MR Conditional | Same |
## Substantial Equivalence Discussion:
The subject device and the predicate device have many identical or similar properties or features, however none of the identified differences introduce new or different aspects of safety or effectiveness. The differences are discussed in the following paragraphs:
**Discussion #1 Extended length range:** The new lengths introduced with this range extension are 14 mm, 30 mm and 48 mm, depending on the gauge size. Device samples of each of the new length/gauge combinations were tested against the same standards as the original size combinations for both freshly manufactured devices and devices that has reached the end of their labeled shelf lives. All tests were successfully completed, therefore it is concluded that no new or different issues of safety or effectiveness have been introduced.
**Discussion #2 Catheter Hub design:** The design of the catheter hub component has been changed to allow the elimination of a previously separate component in the design of the catheter with wings and also to improve catheter operation. Device samples of each of the winged and non-winged designs were tested against the same standards as the original designs for both freshly manufactured devices and devices that has reached the end of their labeled shelf lives. All tests were successfully completed, therefore it is concluded that no new or different issues of safety or effectiveness have been introduced.
510(k) for Polyshield Safety IV Catheters
Page 4 of 5
{8}
POLYMED
Discussion #3 Slip Ring material: The slip ring has changed from a metallic material to a polymeric material. While this component has indirect contract with the patient, the identical material has been used previously in FDA-cleared Poly Medicare IV catheter devices. It is therefore concluded that no new or different issues of safety or effectiveness have been introduced.
## Performance data:
The impact of the design changes has been risk-assessed and certain bench tests have been carried out in response, in accordance with the same protocols as used for the original design, as verification of the acceptability of the design changes. Specifically, the following tests have been repeated on the new design on both freshly manufactured and end of shelf life samples:
- Back leakage of blood control feature
- Flow rate
- Aspiration
- Freedom from leakage during pressurization
- Power injection
- Peak tensile force for catheter
- New catheter lengths verification tests
This testing has been carried out in accordance with the following standard:
- ISO 10555-1:2023, FDA recognition # 6-499
The design changes have not impacted the results of performance testing of the original design for biocompatibility, sterilization, packaging validation or human factors.
## Conclusion
The subject and predicate devices have the same indications for use and fundamental technological characteristics. The design changes planned for introduction of the subject device have been risk assessed and all risks resulting from the design changes have been mitigated and verified in accordance with the same test protocols used for the predicate device. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
510(k) for Polyshield Safety IV Catheters
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.