K252675 · Intuitive Surgical, Inc. · NAY · Dec 15, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K252675
Device Name
da Vinci SP Surgical System (SP1098)
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Dec 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical da Vinci SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Device Story
Computer-controlled surgical system; enables minimally invasive surgery via single-port approach. Surgeon console provides 3D visualization and hand/foot controls for instrument manipulation. Vision cart manages video/image processing and networking. Patient cart holds instrument arm with four drives for endoscope and instruments (dissectors, scissors, forceps, needle holders, retractors). Bedside assistant manages instrument/endoscope installation. System translates surgeon inputs into precise instrument movement. Used in operating rooms by trained physicians. Facilitates tissue manipulation (grasping, cutting, dissection, suturing, electrocautery). Benefits include minimally invasive access for complex procedures, including NSM.
Clinical Evidence
Prospective, multi-center RCT comparing SP NSM (n=26 subjects, 46 breasts) to open NSM (n=27 subjects, 48 breasts). Primary endpoints included safety and effectiveness. Results: 0% conversion to open in SP cohort; 0% 'ink on tumor' in SP cohort vs 3.7% in open; SAE rate 3.8% (SP) vs 22.2% (open); NAC preservation 100% (SP) vs 93.5% (open). Bench testing in cadaver model confirmed technical feasibility of NSM and axillary lymph node dissection.
Technological Characteristics
Computer-controlled surgical system; 3D endoscopic visualization; electrosurgical capability. No technical changes from predicate. System components: Surgeon Console, Vision Cart, Patient Cart. Connectivity: Networked hardware for video/image processing. Sterilization: Compatible with standard surgical instrument sterilization protocols.
Indications for Use
Indicated for adult patients undergoing urologic, colorectal, general thoracoscopic, or transoral otolaryngology (oropharynx, benign/T1-T2 malignant tumors) procedures via single-port approach, and nipple sparing mastectomy (NSM).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
da Vinci SP Surgical System, Model SP1098, da Vinci SP Instruments, and Accessories (K243714)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 15, 2025
Intuitive Surgical, Inc.
Mike Yramategui
Fellow Regulatory Engineer
1020 Kifer Road
Sunnyvale, California 94086
Re: K252675
Trade/Device Name: da Vinci SP Surgical System (SP1098)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: NAY
Dated: November 13, 2025
Received: November 13, 2025
Dear Mike Yramategui:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252675 - Mike Yramategui
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252675 - Mike Yramategui
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark
Trumbore -S
Digitally signed by
Mark Trumbore -S
Date: 2025.12.15
08:46:28 -05'00'
Mark Trumbore Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252675 | ? |
| Please provide the device trade name(s). | | ? |
| da Vinci SP Surgical System (SP1098) | | |
| Please provide your Indications for Use below. | | ? |
| da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical da Vinci SP Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
da Vinci SP Instruments:
Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach; and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
Page 1 of 6
# 510(k) Summary [21 CFR § 807.92(c)]
## I. SUBMITTER INFORMATION
Submitter: Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086
Contact: Mike Yramategui
Fellow Regulatory Engineer
Phone Number: 408-523-2145
Fax Number: 408-523-8907
Email: Mike.Yramategui@intusurg.com
Date Summary Prepared: August 21, 2025
## II. SUBMITTER INFORMATION
Trade Name: da Vinci SP® Surgical System, Model SP1098,
da Vinci SP® Instruments, and Accessories
Common Name: System, Surgical, Computer Controlled Instrument
Classification Name: Endoscope and Accessories (21 CFR §876.1500)
Regulatory Class: Class II
Product Code: NAY (System, Surgical, Computer Controlled Instrument)
Submission Type: Traditional 510(k)
## III. PREDICATE DEVICE INFORMATION
Predicate Device: da Vinci SP Surgical System, Model SP1098,
da Vinci SP Instruments, and Accessories (K243714)
{5}
510(k) Summary
Page 2 of 6
# IV. DEVICE DESCRIPTION
The da Vinci SP Surgical System is designed to enable complex surgical procedures using a minimally invasive approach. The system consists of a Surgeon Console, a Vision Cart, and a Patient Cart and is used with an endoscope, instruments, and accessories.
The surgeon seated at the Surgeon Console controls all movement of the instruments and endoscope by using two hand controls and a set of foot pedals. The surgeon views the endoscopic image on a three-dimensional (3D) viewer, which provides a view of patient anatomy and instrumentation, along with icons and other user interface features.
The Vision Cart includes supporting electronic equipment, such as the light source, video and image processing, and the networking hardware. The Vision Cart also has a touchscreen to view the endoscopic image and adjust system settings.
The Patient Cart is the operative component of the da Vinci SP Surgical System. Its primary function is to support the positioning of the surgical port and to manipulate the surgical instruments and endoscope. The Patient Cart is positioned beside the operating room table and contains an instrument arm that is positioned with respect to the target patient anatomy. The instrument arm contains four instrument drives that hold up to three surgical instruments and the endoscope. The bedside surgical assistant installs and removes the endoscope and instruments intra-operatively as needed.
This 510(k) is for a labeling modification only, to add "nipple sparing mastectomy (NSM) procedures" to the indications, and to add "nipple sparing mastectomy" as new representative, specific procedures in the Professional Instructions for Use. There have been no technical changes to the device.
{6}
510(k) Summary
Page 3 of 6
# INDICATIONS FOR USE
da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP® Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical da Vinci SPTM Instruments during urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach, and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
da Vinci SP Instruments:
Intuitive Surgical da Vinci SP Instruments are controlled by the da Vinci SP® Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic, colorectal, and general thoracoscopic surgical procedures that are appropriate for a single port approach, and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is also indicated for use in nipple sparing mastectomy (NSM) procedures. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
# V. COMPARISON OF INTENDED USE, INDICATIONS FOR USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The da Vinci SP Surgical System, Model SP1098 and da Vinci SP Instruments and Accessories are unchanged from the predicate device in terms of intended use, design, performance, and technological characteristics. The labeling has been changed to add "nipple sparing mastectomy (NSM) procedures" to the indication statement and as new representative, specific procedure in the Professional Instructions for Use.
{7}
510(k) Summary
Page 4 of 6
# VI. PERFORMANCE DATA
The addition of nipple sparing mastectomy (NSM) surgical procedures to the SP1098 Indications does not change any of the safety or performance requirements that were previously verified and / or validated for the SP1098 regarding cleaning, sterilization, packaging, shelf life, biocompatibility, software, cybersecurity, electrosurgical performance, electromagnetic compatibility, electrical safety, mechanical and electrical performance, reliability, or human factors for use in urologic (K173906), transoral surgery (K182371), general thoracoscopic surgical procedures (K240502), colorectal surgical procedures (K242318), or transanal local excision (K243714).
# Performance Testing - Bench
Design validation was performed to determine clinical suitability of the SP1098 for use in nipple sparing mastectomy and axillary lymph node dissection (ALND). The SP1098 system was tested in a simulated clinical laboratory setting and a female human cadaver was used for evaluation. A bilateral NSM and bilateral ALND were performed. The cadaver was positioned with their arm above the head for one breast and their arm along the side of the OR table for the other breast. The surgical approach is symmetrical and this patient positioning is the same for either breast. The NSM and ALND were able to be performed through a single incision on each of the breasts. The SP1098 system met all acceptance criteria for all test cases.
# Performance Testing - Clinical
A prospective, multi-center randomized controlled trial (RCT) was conducted to evaluate performance of the da Vinci SP in nipple sparing mastectomy (NSM) procedures, and to provide a direct comparison between SP NSM to open NSM as the current standard of care. The RCT included NSM performed for early-stage cancer (therapeutic breasts) as well as prophylactic breasts if elected by the patient on the contralateral breast. There were 26 subjects (total of 46 breasts) in the SP arm and 27 subjects (total of 48 breasts) in the Open arm. All the NSM procedures were followed by an immediate one- or two-stage implant-based reconstruction procedure via an open approach. Overall safety was demonstrated by showing no differences in short term safety outcomes, and effectiveness was demonstrated by no conversions to an open approach and "no ink on tumor" in the SP cohort. The table below provides a summary of the key outcomes between SP NSM as the Subject Device and open NSM.
{8}
510(k) Summary
Page 5 of 6
Comparison of key outcomes between SP NSM vs. Open NSM
| | SP RCT | Open RCT |
| --- | --- | --- |
| Number of Subjects | 26 | 27 |
| Therapeutic Breasts | 27 | 27 |
| Prophylactic Breasts | 19 | 21 |
| Total Breasts | 46 | 48 |
| Patient Characteristics | | |
| Age (years) Mean (SD) | 49.3 (9.19) | 45.6 (11.45) |
| BMI Mean (SD) | 21.7 (2.48) | 22.9 (2.95) |
| Breast Ptosis n (%) | | |
| Grade-0 | 8 (30.8%) | 7 (25.9%) |
| Grade-1 | 13 (50.0%) | 12 (44.4%) |
| Grade-2 | 5 (19.2%) | 8 (29.6%) |
| Breast Cup Size n (%) | | |
| A | 5 (19.2%) | 4 (14.8%) |
| B | 17 (65.4%) | 15 (55.6%) |
| C | 4 (15.4%) | 8 (29.6%) |
| Clinical T stage, n (%) | | |
| DCIS/Tis | 3 (11.1%) | 7 (25.9%) |
| T0 | 0 (0.0%) | 1 (3.7%) |
| T1 | 15 (55.6%) | 13 (48.1%) |
| T2 | 9 (33.3%) | 6 (22.2%) |
| Tumor Size (cm) Mean (SD) | 1.8 (1.06) | 1.6 (1.22) |
| History of Chemotherapy or other Neo-adjuvant Therapy n(%) | 4 (15.4%) | 7 (25.9%) |
| Clinical Outcomes | | |
| Total Number of Breasts | 46 | 48 |
| NSM Procedure Time/per breast(minutes) Mean (SD) | 128.3 (61.62) | 85.4 (46.62) |
| Reconstruction Procedure Time/per breast (minutes) Mean (SD) | 80.4 (41.12) | 64.2 (31.89) |
| Conversion to Open, n (%) | 0 (0%) | NA |
| SAE Rate^{a} n (%) | 1 (3.8%) | 6 (22.2%) |
| Number of Subjects with SAE classified as Clavien-Dindo Grade III^{b,c}, n (%) | 1 (3.8%) | 6 (22.2%) |
| Number of Subjects with an UADE, n (%) | 0 (0.0%) | 0 (0.0%) |
| Device Related SAEs^{d} | 0 | 2 |
| NAC related SAE | 0 | 0 |
| Skin-Flap Related SAE | 0 | 0 |
| NAC Preservation (42 days) | 100% | 93.5 % |
| Unplanned reoperations n (%)^{e} | 1 (3.8%) | 6 (22.2%) |
| Unplanned readmission n (%)^{e} | 0 | 1 (3.7%) |
| Mortality | 0 | 0 |
| Number of Breasts with Positive Surgical Margin | 4 (14.8%) | 5 (18.5%) |
| Number of breasts with “ink on tumor” | 0 (0%) | 1 (3.7%) |
| Number of breasts with PSM due to DCIS ≤2mm | 4 (14.8%) | 4 (14.8%) |
{9}
510(k) Summary
Page 6 of 6
## Notes
a All SAEs reported were resolved at data cutoff, except one SAE in the Open RCT cohort.
b There were no Grade IV or Grade V adverse events.
c All SAEs across both cohorts were classified as Clavien-Dindo Grade III, requiring unplanned reoperations.
d There were two (2) device-related SAEs in two (2) subjects in the Open RCT Cohort. Both SAE’s were deemed to be related to NSM, the reconstruction procedure, and the devices used during open NSM.
e One (1) subject in the Open RCT Cohort had an unplanned readmission due to infection of the left breast implant received during one-stage reconstruction that required removal, washout, and IV antibiotics.
## Acronyms:
DCIS: Ductal Carcinoma in Situ, NAC: Nipple-Areolar Complex, PSM: Positive Surgical Margin, SAE: Serious Adverse Event, UADE: Unanticipated Adverse Device Effect.
## I. CONCLUSION
The da Vinci SP Surgical System has the same technological characteristics as it is unmodified from the predicate device, and there are no changes to the intended use for these labeling changes.
Bench testing demonstrates the ability of the da Vinci SP Surgical System to perform nipple sparing mastectomy (NSM) procedures. Clinical data from a prospective, randomized clinical trial demonstrates that da Vinci SP is comparable to open NSM, and do not raise different questions of safety or effectiveness.
Thus, these labeling changes to the da Vinci SP Surgical System are substantially equivalent to the cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.