Niti-S SPAXUS Stent

K252648 · Taewoong Medical Co., Ltd. · PCU · Mar 13, 2026 · Gastroenterology, Urology

Device Facts

Record IDK252648
Device NameNiti-S SPAXUS Stent
ApplicantTaewoong Medical Co., Ltd.
Product CodePCU · Gastroenterology, Urology
Decision DateMar 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5015
Device ClassClass 2
AttributesTherapeutic

Indications for Use

8mmx20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution. 10mmx20mm and 16mmx20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the Niti-S SPAXUS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.

Device Story

Fully covered, self-expanding nitinol metallic stent; bi-flange design for lumen apposition; silicone coating prevents leakage; radiopaque markers (Pt/Ir, stainless steel) for fluoroscopic guidance. Used in endoscopic ultrasound-guided transluminal drainage; placed via non-electrocautery delivery system requiring auxiliary tools (e.g., cystotome) for tissue penetration. Functions as access port for endoscopes to perform debridement, irrigation, and cystoscopy. Used in clinical settings by physicians; implanted up to 60 days. Benefits include effective drainage of fluid collections and gallbladder; reduces surgical risk for cholecystitis patients.

Clinical Evidence

Prospective, multi-center, single-arm study (n=35) for pancreatic pseudocyst drainage. Primary endpoints: safety (SAEs) and effectiveness (≥50% size reduction at 30/60 days). Results: 97.1% clinical success (33/34), 97.1% technical success, 100% stent patency, 0% device-related SAEs. Supported by three meta-analyses comparing SPAXUS to AXIOS, showing comparable or superior success rates and lower adverse event profiles (e.g., bleeding 1.5% vs 7.5%).

Technological Characteristics

Self-expanding nitinol stent with silicone coating; bi-flange design. Radiopaque markers (Pt/Ir and 316L stainless steel). MR Conditional (up to 3T). Sterilization: Ethylene Oxide. Delivery system: non-electrocautery, disposable. No electrical components or software.

Indications for Use

Indicated for patients with symptomatic pancreatic pseudocysts (≥6 cm) or walled-off necrosis (≥6 cm) adherent to gastric/bowel walls, and patients with acute cholecystitis at high surgical risk requiring gallbladder drainage.

Regulatory Classification

Identification

A pancreatic drainage stent is a prescription device that consists of a self-expanding, covered, metallic stent, intended for placement to facilitate transmural endoscopic drainage of pancreatic pseudocysts. This stent is intended to be removed upon confirmation of pseudocyst resolution. This device may also include a delivery system.

Special Controls

In combination with the general controls of the FD&C Act, the AXIOS Stent and Delivery System is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The device and elements of the delivery device that may contact the patient must be demonstrated to be biocompatible. (2) Performance data must demonstrate the sterility of patient-contacting components of the device. (3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the requested shelf life. (4) Non-clinical testing data must demonstrate that the stent and delivery system perform as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Deployment testing of the stent and delivery system must be conducted under simulated use conditions. (ii) Removal force testing must be conducted. The removal force testing must demonstrate that the stent can be safely removed, and that the stent will remain in place when subjected to forces encountered during use. (iii) Expansion force testing must be conducted. The expansion force must demonstrate that the forces exerted by the stent will not damage the tissue surrounding the stent. (iv) Compression force testing must be conducted. The compression force must demonstrate that the stent will withstand the forces encountered during use. (v) Dimensional verification testing must be conducted. (vi) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (vii) Fatigue testing must be conducted. Material strength must demonstrate that the stent will withstand forces encountered during use. (viii) Corrosion testing must be conducted. Corrosion resistance must demonstrate that the stent will withstand conditions encountered during use. (5) Non-clinical testing must evaluate the compatibility of the stent in a magnetic resonance (MR) environment. (6) Well-documented clinical experience must demonstrate safe and effective use, and capture any adverse events observed during clinical use. (7) Labeling must include the following: (i) Appropriate instructions, warnings, cautions, limitations, and information related to the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device. (ii) A warning that the safety and patency of the stent has not been established beyond the duration of the documented clinical experience. (iii) Specific instructions and the qualifications and clinical training needed for the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device. (iv) Information on the patient population for which the device has been demonstrated to be effective. (v) A detailed summary of the clinical experience pertinent to use of the device. (vi) A detailed summary of the device technical parameters. (vii) A detailed summary of the device- and procedure-related complications pertinent to use of the device. (viii) An expiration date/shelf life.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 13, 2026 Taewoong Medical Co., Ltd. % Matthew Krueger Principal Consultant Biologics Consulting Group 100 Daingerfield Road, Suite 101 Alexandria, Virginia 22314 Re: K252648 Trade/Device Name: Niti-S SPAXUS Stent Regulation Number: 21 CFR 876.5015 Regulation Name: Pancreatic Drainage Stent And Delivery System Regulatory Class: Class II Product Code: PCU, QXH Dated: February 13, 2026 Received: February 13, 2026 Dear Matthew Krueger: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252648 - Matthew Krueger Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252648 - Matthew Krueger Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ANTHONY LEE -S Anthony Lee, Ph.D., MBA Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K252648 Device Name Niti-S SPAXUS Stent Indications for Use (Describe) 8mmx20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution. 10mmx20mm and 16mmx20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the Niti-S SPAXUS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} 510(k) Summary Taewoong Medical K252648 # 1. SUBMITTER INFORMATION Applicant: Taewoong Medical Contact: Yongjin Jeff Kim Phone: +82 70 4649 1543 Email: jinjeff@stent.net Address: 14, Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do, 10022, Republic of Korea # 2. CORRESPONDENT INFORMATION Contact: Matthew Krueger Title: Principal Consultant Firm: Biologics Consulting Group # 3. DATE PREPARED: FEBRUARY 13, 2026 # 4. DEVICE INFORMATION Device Name: Niti-S SPAXUS™ Stent Common Name: Pancreatic drainage stent and delivery system Gallbladder drainage stent and delivery system Regulation Number: 21 CFR 876.5015 21 CFR 876.5016 Regulation Name: Pancreatic drainage stent and delivery system Product Code: PCU, QXH Regulatory Class: Class II # 5. PREDICATE DEVICE INFORMATION Primary Predicate Device Name: AXIOS Stent and Electrocautery-Enhanced Delivery System 510(k) Number: K233318 Manufacturer: Boston Scientific Reference Device Name: Esophageal TTS Stent 510(k) Number: K240522 Manufacturer: Taewoong Medical Page 1 of 7 {5} 510(k) Summary Taewoong Medical K252648 Neither the predicate nor the reference devices have been subject to a design related recall. ## 6. DEVICE DESCRIPTION The Niti-S SPAXUS™ Stent is a fully covered, self-expanding metallic stent. The stent is constructed from nickel-titanium alloy (nitinol), which provides flexibility, durability, and self-expansion properties. Its bi-flange design provides stable lumen apposition and helps mitigate migration risk. The complete silicone coating helps prevent leakage during drainage and facilitates the safe removal of the device. Radiopaque markers made of platinum/iridium and stainless steel enhance visibility under fluoroscopic guidance, aiding precise placement and monitoring. The Stent Delivery System (SDS) is a disposable, non-electrocautery system for the delivery and deployment of the stent at the target position. It is similar in design to the SDS used for the Esophageal TTS Stent (K240522). ## 7. INDICATIONS FOR USE 8mm×20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution. 10mm×20mm and 16mm×20mm stents: The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥6 cm in size and walled-off necrosis ≥6 cm in size that are adherent to the gastric or bowel wall. Once placed, the Niti-S SPAXUS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery. ## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The Indications for Use for the subject Niti-S SPAXUS™ Stent are identical to a subset of those cleared for the predicate devices. The following tables compare some of the key technological characteristics between the subject and predicate devices. {6} 510(k) Summary Taewoong Medical K252648 Table 1: Pancreas Comparison – Niti-S SPAXUS Stent vs. AXIOS Stent 510(k) Predicates | Feature | Niti-S SPAXUS Stent | AXIOS Stent and Electrocautery-Enhanced Delivery System | | --- | --- | --- | | 510(k) Number | K252648 (subject device) | K203132, K220112, K233318 | | Regulation | 21 CFR 876.5015 (Pancreatic drainage stent and delivery system) | 21 CFR 876.5015 (Pancreatic Drainage Stent and Delivery System) | | Product Code(s) | PCU | PCU, KNS | | Regulatory Class | Class II | Class II | | Stent Sizes (Diameter × Length) | 8mm × 20mm, 10mm × 20mm, 16mm × 20mm (The listed length refers to the total stent length.) | 6mm × 8mm 8mm × 8mm 10mm × 10mm 15mm × 10mm 15mm × 15mm 20mm × 10mm (The listed length refers to the saddle length) | | Primary Stent Materials | Nitinol with a silicone coating | Nitinol with a silicone coating | | Delivery System Type | Standard delivery system (requires auxiliary tools, e.g., cystotome, for tissue penetration). | Available with both a standard delivery system and an electrocautery-enhanced delivery system. | | Stent MRI Compatibility | MR Conditional | MR Conditional | | Visualization | Radiopaque markers (Pt/Ir and 316L stainless steel) for fluoroscopic visibility. | Radiopaque markers (Pt/Ir) for fluoroscopic visibility. | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | | Packaging | Provided sterile for single-patient use. | Provided sterile for single-patient use. | {7} 510(k) Summary Taewoong Medical K252648 Table 2: Gallbladder Comparison – Niti-S SPAXUS Stent vs. AXIOS De Novo and 510(k) Predicates | Feature | Niti-S SPAXUS Stent | AXIOS Stent and Electrocautery-Enhanced Delivery System | | --- | --- | --- | | 510(k) Number | K252648 (subject device) | DEN230019, K233318 | | Regulation | 21 CFR 876.5016 - Gallbladder drainage stent and delivery system | 21 CFR 876.5016 - Gallbladder drainage stent and delivery system | | Product Code | QXH | QXH | | Regulatory Class | Class II | Class II | | Stent Sizes (Diameter × Length) | 10mm × 20mm and 16mm × 20mm (The listed length refers to the total stent length.) | 10mm × 10mm 15mm × 10mm (The listed length refers to the saddle length) | | Primary Stent Materials | Nitinol with a silicone coating | Nitinol with a silicone coating | | Delivery System Type | Standard delivery system (requires auxiliary tools, e.g., cystotome, for tissue penetration). | Available with both a standard delivery system and an electrocautery-enhanced delivery system. | | Stent MRI Compatibility | MR Conditional | MR Conditional | | Visualization | Radiopaque markers (Pt/Ir and 316L stainless steel) for fluoroscopic visibility. | Radiopaque markers (Pt/Ir) for fluoroscopic visibility. | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | | Packaging | Provided sterile for single-patient use. | Provided sterile for single-patient use. | # 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING # Biocompatibility Testing The stent materials of the subject devices are identical to those of the Niti-S Biliary Stent cleared in K240522 including its formulation, processing, and sterilization process, with no additional chemicals (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents) introduced. Similarly, the SDS material of the subject device is identical to the Esophageal TTS Stent cleared in K240522 by Taewoong Medical, including its formulation, processing, and sterilization, with no additional chemicals added. Therefore, additional biocompatibility testing for the Stent Delivery System is not required. # Electrical Safety Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type. {8} 510(k) Summary Taewoong Medical K252648 ## Electromagnetic Compatibility (EMC) Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type. ## Software Not applicable. The subject device contains no software. ## Performance Testing Additional performance testing was conducted to demonstrate that the device meets its design requirements and performs as intended. The performance tests include: - Galvanic corrosion - Corrosion - Deployment Force - Expansion force - Compression force - Dimensional - Stent pull-out force - Implant anchor function-retention - Sterility verification - Shelf life - Package integrity ## 10. SUMMARY OF CLINICAL TESTING A prospective, multi-center, single arm, non-inferiority, open label was conducted on 35 patients in South Korea using the Niti-S SPAXUS™ Stent. The purpose of the study was to evaluate safety and effectiveness of the Niti-S SPAXUS™ Stent in endoscopic ultrasound-guided transluminal drainage for the treatment of pancreatic pseudocysts ≥ 6 cm in size. Twenty-seven (27) males and 8 females comprised the 35 subjects enrolled in the study. The mean age was 53 ± 13 years (range, 22-77 years). The primary safety endpoint was the incidence of serious adverse events (SAEs) related to the procedure or the Niti-S SPAXUS™ stent during the period of the study (i.e., from stent placement to 20 days ("Day 20") after the stent removal). The primary effectiveness endpoint was clinical success, defined as reduction in size of the pancreatic pseudocyst by at least 50% at 30 days (± 10 days; "Day 30") or 60 days (± 10 days; "Day 60") after the stent placement. One subject was excluded from the final effectiveness analysis. Secondary endpoints evaluated technical success, stent lumen patency, stent removal success, and procedure time. There were no procedural or device-related SAEs reported during the study follow-up period. Clinical success was achieved in 33 of 34 subjects (97.1%) and non-inferiority was achieved. {9} 510(k) Summary Taewoong Medical K252648 Technical success was reported in 33 of 34 subjects (97.1%). Thirty-four (34) subjects (100%) maintained stent lumen patency at stent removal (Day 30 or 60 (± 10 days)). All stents were successfully removed (100%). The mean procedure time was 17.9 minutes ± 10.3 minutes. In addition to this prospective study, three meta-analyses were referenced to help support the safe and effective use of the Niti-S SPAXUS™ Stent for the complete indications for use, including gallbladder drainage. These meta-analyses are described in: - Suresh Kumar VC, Singh S, Moond V, Mohan BP, Aswath G, Khan HMA, Sapkota B, Adler DG. Safety and efficacy of lumen-apposing metal stents for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a systematic review and meta-analysis. Endoscopy. 2025 Mar;57(3):282-290. - Mangiavillano B, Lakhtakia S, Samanta J, Auriemma F, Vargas-Madrigal J, Arcidiacono PG, Barbera C, Ashhab H, Song TJ, Pham KD, Teoh AYB, Moon JH, Crinò SF, Kongkam P, Aragona G, De Lusong MA, Dhar J, Ofosu A, Ventra A, Paduano D, Franchellucci G, Repici A, Larghi A, Facciorusso A; PFC LAMS study group. Lumen-apposing metal stents for the treatment of pancreatic and peripancreatic fluid collections and bleeding risk: a propensity matched study. Endoscopy. 2024 Apr;56(4):249-257. - Singh S, Suresh Kumar VC, Aswath G, Akbar Khan HM, Sapkota B, Vinayek R, Dutta S, Dahiya DS, Inamdar S, Mohan BP, Sharma N, Adler DG. Indirect comparison of various lumen-apposing metal stents for EUS-guided biliary and gallbladder drainage: a systematic review and meta-analysis. Gastrointest Endosc. 2024 Nov;100(5):829-839.e3. ## 11. BENEFIT-RISK ASSESSMENT The benefit-risk assessment for the Niti-S SPAXUS™ Stent is based on nonclinical and clinical tests demonstrating substantial equivalence to the predicate AXIOS Stent (K233318). Nonclinical tests, including galvanic corrosion, potentiodynamic polarization, deployment force, expansion/compression force, dimensional verification, stent pull-out force, implant anchor retention, torque strength, trackability, tensile strength, fatigue, and MR compatibility, confirmed that all acceptance criteria were met. These results show the device maintains structural integrity, resists corrosion, deploys reliably, and is MR Conditional up to 3T/2000 gauss/cm, with risks such as material degradation or migration mitigated through design and testing comparable to the predicate. Clinical evidence from a prospective study (35 patients; 97.1% technical/clinical success, no device-related SAEs) and three meta-analyses supports the safety and effectiveness of the SPAXUS™ Stent. Singh et al. (2024) reported SPAXUS™ gallbladder drainage success rates of 95.9% technical/94.2% clinical, with 9.5% AEs vs. AXIOS 23.6%. Kumar et al. (2024) showed SPAXUS™ PFC drainage success of 98.2% technical/93.5% clinical, with 7.6% AEs vs. AXIOS 20.4%. Mangiavillano et al. (2024) propensity-matched study (132 patients/group) indicated SPAXUS™ bleeding (1.5%) and overall AEs (3.0%) lower than AXIOS (7.5% and 9.8%). The OUS marketing history (no recalls, low complaints) further affirms long-term safety. Benefits include effective drainage of pancreatic pseudocysts/WON and gallbladder (high success rates) outweighing risks like bleeding (1.5-1.8%), migration (0.9-2.4%), or occlusion {10} 510(k) Summary Taewoong Medical K252648 (14.3%), which are lower or comparable to the predicate and are mitigated via labeling/instructions. The device is as safe and effective as the predicate. ## 12. CONCLUSION The results of the performance testing described above demonstrate that the Niti-S SPAXUS™ Stent performs equivalently as compared to the predicate and reference devices and supports a determination of substantial equivalence. Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...