Iso-Gard Filter Straight: The device is intended to remove microbiological and particulate matter from the gases in the breathing circuit. Humid-Vent Filter Compact Angled: The device is a combined Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care.
Device Story
Iso-Gard Filter Straight and Humid-Vent Filter Compact Angled are single-use, non-sterile breathing circuit filters. Iso-Gard removes microbiological and particulate matter from breathing gases. Humid-Vent combines a bacterial/viral filter with a Heat and Moisture Exchanger (HME) to provide passive humidification and filtration. Both devices are placed between the patient and the breathing system. Operation relies on electrostatic charge in the filter media to capture particles; HME function uses microwell paper impregnated with calcium chloride. Used in anesthesia and ventilator care settings by clinicians. Output is filtered, humidified gas (for Humid-Vent) delivered to the patient. Benefits include reduced risk of cross-contamination between patient and equipment and maintenance of airway moisture.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ISO 9360-1 (moisture loss, compliance, resistance to flow), ISO 23328-1/2 (filtration efficiency, gas leakage, pressure drop), ASTM F2101 (BFE >99.9999+%, VFE >99.9967%), and ISO 5356-1/80369-7/20 (mechanical/luer connection integrity). Biocompatibility confirmed per ISO 10993 and ISO 18562.
Technological Characteristics
Housing: Polymethyl Methacrylate (upper), Polypropylene (lower). Luer Cap: ABS. Filter Media: Electrostatic Polypropylene. HME Media: Microwell paper with calcium chloride. Connections: 22mm/15mm per ISO 5356-1. Single-use, non-sterile. Filtration via electrostatic charge. Passive humidification via HME.
Indications for Use
Indicated for adult and pediatric patients (excluding neonates) requiring filtration of breathing circuits to reduce cross-contamination risk during anesthesia and ventilator care. Humid-Vent Filter Compact Angled is additionally indicated for patients receiving respiratory support with dry inspired gases requiring humidification. Contraindicated for patients with excessive airway secretions, pulmonary hemorrhage, or hemoptysis; not for use with conventional heated humidifiers or nebulizers.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 15, 2026
Teleflex Medical Sdn. Bhd.
Suguneswary Subramaniam
Regulatory Affairs Manager - Product Management
Lot Pt. 2577
Jalan Perusahaan 4
Kamunting, Perak 34600
Malaysia
Re: K252643
Trade/Device Name: Iso-Gard Filter Straight; Humid-Vent Filter Compact Angled
Regulation Number: 21 CFR 868.5260
Regulation Name: Breathing Circuit Bacterial Filter
Regulatory Class: Class II
Product Code: CAH
Dated: April 15, 2026
Received: April 15, 2026
Dear Suguneswary Subramaniam:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252643 - Sugunesway Subramaniam
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252643 - Suguneswary Subramaniam
Page 3
Sincerely,
Dolly Singh Digitally signed by Dolly Singh
For Katharine Segars, Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252643 | |
| Device Name Iso-Gard Filter Straight Humid-Vent Filter Compact Angled | |
| Indications for Use (Describe) | |
| Iso-Gard Filter Straight: The device is intended to remove microbiological and particulate matter from the gases in the breathing circuit. Humid-Vent Filter Compact Angled: The device is a combined Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
# K252643 510(k) Summary
## A. Name and Address of Applicant
Teleflex Medical Sdn. Bhd.
Lot PT 2577
Jalan Perusahaan 4
34600 Kamunting, Perak
Malaysia
## B. Contact Person and Phone Number
Suguneswary Subramaniam
Regulatory Affairs Manager – Product Management
Phone: (+60) 10-221 2330
Alternate Contact:
Shah, Jigar
Senior Regulatory Affairs Manager – Product Management
Phone: (919) 503-0322
## C. Date Prepared
May 15, 2026
## D. Device Name
### Subject Device 1
Trade Name: Iso-Gard Filter Straight
Common Name: Filter, Bacterial, Breathing Circuit
Product Code: CAH
Regulation Number: 21 CFR 868.5260
Classification Name: Filter, Bacterial, Breathing-Circuit
Classification: II
Classification Panel: Anesthesia
### Subject Device 2
Trade Name: Humid-Vent Filter Compact Angled
Common Name: Filter, Bacterial, Breathing-Circuit
Product Code: CAH
Regulation Number: 21 CFR 868.5260
Classification Name: Filter, Bacterial, Breathing-Circuit
Classification: II
Classification Panel: Anesthesia
## E. Predicate Device
K965016 Iso-Gard Filter Angled (18212)/Iso-Gard Filter Straight (19212)
K964382 Humid-Vent Filter Compact Angled (18402)/Humid-Vent Filter Compact Straight (19402)
K252643 - Page 1 of 22
{5}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
## F. Reference Device
K202459 Bact-Trap™ Filter, Bact-HME™ Midi HME/Filter
## G. Device Description
The Iso-Gard Filter Straight (Iso-Gard Filter S) is the device intended to remove microbiological and particulate matter from the gases in the breathing circuit. The device is intended for use in anesthesia and ventilator care for filtration of gases. The Iso-Gard Filter S is used in patients requiring filtration of breathing circuits to reduce the risk of bacterial and viral contamination.
The Humid-Vent Filter Compact Angled (HVF Compact A) is a combination of Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care. The HVF Compact A is for patients receiving respiratory support with dry inspired gases, where both humidification and bacterial/viral filtration are required.
Both devices were placed between the patient and breathing system are intended to replace the filtering function of nasopharynx by protecting the patient from contamination within the breathing system and reduce the risk of patients contaminating the interior of breathing systems and equipment. Additionally, HVF Compact A has the additional feature to replace warming and humidifying functions of nasopharynx.
The proposed Teleflex Medical Iso-Gard Filter Straight and Humid-Vent Filter Compact Angled are supplied as clean (non-sterile) and single use devices.
## H. Indications for Use
### Subject Device 1 (Iso-Gard Filter Straight):
The device is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
### Subject Device 2 (Humid-Vent Filter Compact Angled):
The device is a combined Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care.
## I. Contraindications
### Subject Device 1 (Iso-Gard Filter Straight):
- Patients with Excessive Airway Secretions or Bleeding: The device should not be used in patients with copious frothy secretions, significant pulmonary hemorrhage or hemoptysis, as these conditions may lead to filter blockage and increased airway resistance.
- Make sure if you use drugs with surfactant properties, that the hydrophobic pores are not blocked. This can happen if drugs are nebulized in a circuit with a cascade humidifier. Always consult with your drug manufacturer before nebulization of drugs or solutions for the first time.
K252643 - Page 2 of 22
{6}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
## Subject Device 2 (Humid-Vent Filter Compact Angled):
- Patients with Excessive Airway Secretions or Bleeding: The device should not be used in patients with copious frothy secretions, significant pulmonary hemorrhage or hemoptysis, as these conditions may lead to filter blockage and increased airway resistance.
- Use with External Humidification or Nebulization: The device should not be used with conventional heated humidifiers or nebulizers, as excess moisture or aerosolized medications may obstruct the filter media and compromise device function.
## J. Substantial Equivalence
The subject devices are substantially equivalent to the predicate device with respect to intended use, technology characteristics, and construction. The differences between the predicate and the subject devices are minor and any risks have been mitigated through testing. The Reference Device (K202459) is included in this submission to provide a more recently cleared device with similar Indications for Use Technological Characteristics. The table below summarizes the differences between the subject, predicate and the reference device.
K252643 - Page 3 of 22
{7}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Substantial Equivalence Comparative Table: Iso-Gard Filter Straight (Subject Device 1)
| Features | Teleflex Medical Gibeck Iso-Gard Filter Straight (Subject Device 1, K252643) | Gibeck Iso-Gard Filter Straight (Predicate K965016) | Reference Device: Pharma System AB Bact-Trap Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| Classification Name | Filter, Bacterial, Breathing Circuit | Filter, Bacterial, Breathing Circuit | Filter, Bacterial, Breathing Circuit | Same |
| Product Code | CAH | CAH | CAH | Same |
| Regulation | 868.5260 | 868.5260 | 868.5260 | Same |
| Indications for Use | The device is intended to remove microbiological and particulate matter from the gases in the breathing circuit. | Intended to remove microbiological and particulate matter from the gases in the breathing circuit. | Bact Trap filter is a breathing system filter which is designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anaesthetic or ventilator breathing systems. The Bact Trap filter may either be used on the patient side or on the device side of the ventilator anaesthetic device. | Same |
| Principle of Operation | Filtration via the principle of electrostatic charges in the media. | Filtration via the principle of electrostatic charges in the media. | Filtration via the principle of electrostatic charges in the media. | Same |
| Patient Population | Adult and pediatric. Do not use on neonate patients. | Not stated | Adult | Similar to Reference device. The subject device is having broader target population. |
| Contraindications | • Patients with Excessive Airway Secretions or Bleeding: The device | • Iso-Gard Filters should not be used with patients | Not stated | Similar. Clarifying language added for the intended user. |
K252643 - Page 4 of 22
{8}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Iso-Gard Filter Straight (Subject Device 1, K252643) | Gibeck Iso-Gard Filter Straight (Predicate K965016) | Reference Device: Pharma System AB Bact-Trap Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| | should not be used in patients with copious frothy secretions, significant pulmonary hemorrhage or hemoptysis, as these conditions may lead to filter blockage and increased airway resistance. | producing fulminating frothy secretions within their airways, or patients with hemoptysis. | | |
| | • Make sure if you use drugs with surfactant properties, that the hydrophobic pores are not blocked. This can happen if drugs are nebulized in a circuit with a cascade humidifier. Always consult with your drug manufacturer before nebulization of drugs or solutions for the first time. | • Make sure if you use drugs with surfactant properties, that the hydrophobic pores are not blocked. This can happen if drugs are nebulized in a circuit with a cascade humidifier. Always consult with your drug manufacturer before nebulization of drugs or solutions for the first time. | Not stated | Same |
| | Not applicable | • Do not use in conjunction with | Not stated | Different. Clarifying language added |
K252643 - Page 5 of 22
{9}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Iso-Gard Filter Straight (Subject Device 1, K252643) | Gibeck Iso-Gard Filter Straight (Predicate K965016) | Reference Device: Pharma System AB Bact-Trap Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| | | conventional humidifiers and nebulizers. • Do not add moisture to the Filter. | | for the intended user. |
| Single Use | Yes | Yes | Yes | Same |
| Weight | 23.00g | 21.7g | 32g | Different to predicate, there is a slight difference in the weight as a result of the filter media replacement. Difference to Reference device. The subject device has lighter weight when compared to the reference device. |
| Sterile | No | No | No | Same |
| Lifetime/Duration of Use | Maximum usage duration of 24 hours | Replace every 24 hours | Intended for limited or less than 24 hours of use | Same |
| Standard 22mm/15mm connections in compliance with ISO 5356-1 | Connectors (22M/15F–22F/15M) in compliance with ISO 5356-1 | 15/22 – 22/15mm, ISO 5356-1 | Standard conical 15mm/22mm fittings | Same |
| Configurations | Straight with luer port | Straight with luer port | Straight with female luer lock port | Same |
| Component | Upper Part/Dry Side: | Upper Part/Dry Side: | • SBC (styrene-butadiene | Same |
K252643 - Page 6 of 22
{10}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Iso-Gard Filter Straight (Subject Device 1, K252643) | Gibeck Iso-Gard Filter Straight (Predicate K965016) | Reference Device: Pharma System AB Bact-Trap Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| Materials (Housing) | Polymethyl Methacrylate Lower Part/Wet Side: Polypropylene | Polymethyl Methacrylate Lower Part/Wet Side: Polypropylene | copolymer) • Polypropylene • Acrylic and Polypropylene fibers • Polyethylene | |
| Component Materials (Luer Cap) | ABS | ABS | | |
| Component Materials (Filter Media) | Electrostatic (Polypropylene) | Electrostatic (Polypropylene) | | Different. Different grade and weightage of the filter media material for the subject device. |
| Packaging | Individually packed, 25 units per shelf box, 250 units per shipper carton | Individually packed, 25 units per shelf box, 250 units per shipper carton | Not stated | Same |
| Packaging Material | Sealed poly film/paper package | Sealed poly film/paper package | Not stated | Same |
| General Performance | | | | |
| Tidal Volume (Vt) | 250ml – 1000ml | 150ml – 1000ml | 150-1500 ml | Similar (150ml – 1000ml) |
| Filtration Efficiency per ASTM F2101 | BFE: 99.9999+% VFE: 99.9967% | BFE: 99.99+% VFE: 99.99% | BFE: >99.999% VFE: >99.99% | Same |
| Biocompatibility | Materials have been tested as per ISO 10993-1 and ISO 18562 | Materials have been tested as per ISO 10993-1. | Materials have been tested as per ISO 10993-1 and ISO 18562 | Same |
K252643 - Page 7 of 22
{11}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Substantial Equivalence Comparative Table: Humid-Vent Filter Compact Angled (Subject Device 2)
| Features | Teleflex Medical Gibeck Humid-Vent Filter Compact Angled (Subject Device 2, K252643) | Gibeck Humid-Vent Filter Compact Angled (Predicate K964382) | Reference Device: Pharma System AB Bact-HME Midi HME/Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| Classification Name | Filter, Bacterial, Breathing Circuit | Filter, Bacterial, Breathing Circuit | Filter, Bacterial, Breathing Circuit | Same |
| Product Code | CAH | CAH | CAH | Same |
| Regulation | 868.5260 | 868.5260 | 868.5260 | Same |
| Indications for Use | The device is a combined Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care. | Combined Heat and Moisture Exchanger (HME) and Bacterial/Viral Filter for humidification and bacterial/viral filtration during anesthesia and ventilator care. | The Bact HME/Bact-HME Midi/Pharma Mini is a breathing system filter and a Heat and Moisture Exchanger. The combination of a filter and a Heat and Moisture Exchanger offer the benefit of both product features. Heat and Moisture Exchangers are used as a conditioning system for mechanically ventilated patients whose upper airways are bypassed. In almost all cases of mechanical ventilation they are a fully valid alternative to heated humidifiers. The Bact HME/Bact-HME Midi Pharma Mini should be used with patients who have a Tidal Volume between 50 - 1500 ml. | Same as predicate. |
K252643 - Page 8 of 22
{12}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Humid-Vent Filter Compact Angled (Subject Device 2, K252643) | Gibeck Humid-Vent Filter Compact Angled (Predicate K964382) | Reference Device: Pharma System AB Bact-HME Midi HME/Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| Principle of Operation | Filtration via the principle of electrostatic charges in the media. Passive humidification functioning as heat and moisture exchanger to replace normal and humidify the upper airways of patient. | Filtration via the principle of electrostatic charges in the media. Passive humidification functioning as heat and moisture exchanger to replace normal and humidify the upper airways of patient. | Filtration via the principle of electrostatic charges in the media. Passive humidification functioning as heat and moisture exchangers. | Same |
| Patient Population | Adult and pediatric. Do not use on neonate patients. | Not stated | Adult | Similar to Reference device. The subject device has a broader target population. |
| Contraindications | • Patients with Excessive Airway Secretions or Bleeding: The device should not be used in patients with copious frothy secretions, significant pulmonary hemorrhage or hemoptysis, as these conditions may lead to filter blockage and increased airway resistance. | • Humid-Vent Filter Compacts should not be used with patients producing fulminating frothy secretions within their airways, or patients with hemoptysis. | Not stated | Same |
| | • Use with External Humidification or Nebulization: The device should not be used | • Do not use in conjunction with conventional | | Same |
K252643 - Page 9 of 22
{13}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Humid-Vent Filter Compact Angled (Subject Device 2, K252643) | Gibeck Humid-Vent Filter Compact Angled (Predicate K964382) | Reference Device: Pharma System AB Bact-HME Midi HME/Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| | with conventional heated humidifiers or nebulizers, as excess moisture or aerosolized medications may obstruct the filter media and compromise device function. | humidifiers and nebulizers. | | |
| | Not applicable | • Do not add moisture to the Humid-Vent Filter Compact. | | Different. Clarifying language added for the intended user. |
| Single Use | Yes | Yes | Yes | Same |
| Weight | 33.00g | 32.3g | 29.00g | Different to the predicate. A difference in the weight as a result of the filter media replacement. |
| Sterile | No | No | No | Same |
| Lifetime/ Duration of Use | Maximum usage duration of 24 hours | Replace every 24 hours | Intended for limited or less than 24 hours of use | Same |
| Standard 22mm/15mm connections in compliance with ISO 5356-1 | Connectors (22M/15F–22F/15M) in compliance with ISO 5356-1 | 15/22 – 22/15mm, ISO 5356-1 | Standard conical 15mm/22mm fittings | Same |
| Configurations | Angled with luer port | Angled with luer port | Angled with female luer lock port | Same |
K252643 - Page 10 of 22
{14}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibec Humid-Vent Filter Compact Angled (Subject Device 2, K252643) | Gibec Humid-Vent Filter Compact Angled (Predicate K964382) | Reference Device: Pharma System AB Bact-HME Midi HME/Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| Component Materials (Housing) | Upper Part/Dry Side: Polymethyl Methacrylate Lower Part/Wet Side: Polypropylene | Upper Part/Dry Side: Polymethyl Methacrylate Lower Part/Wet Side: Polypropylene | • SBC (styrene-butadiene copolymer) • Polypropylene • Acrylic and Polypropylene fibers • Polyethylene • Cross linkable acrylic polymer • Silicone antifoam emulsion • Polyethylene • Thermally bound Polyester (Polyethylene Terephthalate) | Same |
| Component Materials (Luer Cap) | ABS | ABS | | Same |
| Component Materials (HME Media) | Microwell paper impregnated with hygroscopic salt calcium chloride | Microwell paper impregnated with hygroscopic salt calcium chloride | | Same |
| Component Materials (Filter Media) | Electrostatic (Polypropylene) | Electrostatic (Polypropylene) | | Different. Different grade and weightage of the filter media material for the subject device. |
| Packaging | Individually packed, 25 units per shelf box, 250 units per shipper carton | Individually packed, 25 units per shelf box, 250 units per shipper carton | Not stated | Same |
| Packaging Material | Sealed poly film/paper package | Sealed poly film/paper package | Not stated | Same |
| General Performance | | | | |
| Tidal Volume (Vt) | 150ml – 1000ml | 150ml – 1000ml | 100ml – 1200ml | Same |
| Filtration Efficiency per | BFE: 99.9999+% VFE: 99.9979% | BFE: 99.99+% VFE: 99.99% | BFE: >99.99% VFE: >99.9% | Same |
K252643 - Page 11 of 22
{15}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Features | Teleflex Medical Gibeck Humid-Vent Filter Compact Angled (Subject Device 2, K252643) | Gibeck Humid-Vent Filter Compact Angled (Predicate K964382) | Reference Device: Pharma System AB Bact-HME Midi HME/Filter (K202459) | Comparison of Subject Device to Predicate and/or Reference Device |
| --- | --- | --- | --- | --- |
| ASTM F2101 | | | | |
| Biocompatibility | Materials have been tested as per ISO 10993-1 and ISO 18562 | Materials have been tested as per ISO 10993-1 | Materials have been tested as per ISO 10993-1 and ISO 18562 | Same |
K252643 - Page 12 of 22
{16}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
## K. Comparison to the Predicate
The table above illustrates the similarities and differences between the subject and predicate devices. The basic technological and operating principles are the same for both devices. Additionally, the indication of both subject and predicate devices are consistent, for the bacterial/viral filtration and humidification of the airway within breathing circuits. Both the subject and predicate devices are disposable, clean (non-sterile), single patient use devices. The subject devices are substantially equivalent to the predicate device.
## L. Comparison to the Reference Device
The Reference Device (K202459) is included in this submission to provide a more recently cleared device for both the Iso-Gard Filter Straight (Subject Device 1) and the Humid-Vent Filter Compact Angled (Subject Device 2) to show that there are other devices with similar Indications and Technological Characteristics to the Subject Devices.
## M. Performance Data
Iso-Gard Filter Straight and Humid-Vent Filter Compact Angled have undergone extensive testing to qualify it according to FDA recognized consensus standards and international standard.
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. The testing are summarize in the table below:
K252643 - Page 13 of 22
{17}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Subject Device 1 – Iso-Gard Filter Straight
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Performance Test | | | |
| ISO 9360-1 ISO 23328-2 | Pressure Drop | Product shall be tested according to ISO 9360-1 and ISO 23328-1 to assess resistance to flow (pressure drop) across the device under specified flowrates. • Vt 250ml (15 L/min) • Vt 500ml (30 L/min) • Vt 750ml (30 L/min) • Vt 1000ml (30 L/min) | Pass |
| ISO 9360-1 ISO 23328-2 | Gas Leakage | Gas leakage characteristics of the breathing system filter was carried out in accordance with Clause 5.3 of ISO 23328 2, whereby the measurement of gas leakage is determined in accordance with clause 6.4 of ISO 9360 1. | Pass |
| ISO 9360-1 ISO 23328-2 | Dead Space | The total dead space is determined by the final weight of the product after filling with water minus the initial weight of the product before filling with water. | Pass |
| ISO 23328-1 | Salt Test | An aerosol of sodium chloride particles is generated and passed through the BSF at the specified flow rate of 15 L/min and 30 L/min. The concentrations of particles upstream and downstream of the BSF are measured, and filtration efficiency is calculated in accordance with Clause 5 of ISO 23328-1. | Pass |
| ASTM F2101 | Bacterial Filtration Efficiency (BFE) | Product shall be tested according to ASTM F2101 to assess bacterial filtration efficiency performance, employing a ratio of the upstream bacterial challenge to downstream residual concentration to determine filtration efficiency of materials. | Pass |
| | Virus Filtration Efficiency (VFE) | Product shall be tested according to ASTM F2101 to assess virus | Pass |
K252643 - Page 14 of 22
{18}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Performance Test | | | |
| | | filtration efficiency performance, employing a ratio of the upstream viral challenge to downstream residual concentration to determine filtration efficiency of materials. | |
| ISO 5356-1 | Connector Gauging Test | Leading edge lie between the minimum and maximum diameter step of the gauge. | Pass |
| ISO 80369-7 ISO 80369-20 | Resistance to Separation from Axial Load Test | Luer connector shall not separate from the reference connector over a hold period between 10s and 15s while being subjected to a disconnection applied axial force between 32N and 35N for Luer lock connectors. | Pass |
| | Resistance to Separation from Unscrewing Test | Luer lock shall not separate from the reference connector for a hold period between 10s and 15s while being subjected to disconnection applied axle force between 32N and 35N for Luer lock connectors. | Pass |
| | Resistance to Overriding Test | Luer lock connector shall not override the threads or lugs of the reference connector while being subjected to an applied torque of between 0,15 N·m to 0,17 N·m over a hold period between 5s and 10s. | Pass |
| | Positive Pressure Liquid Leakage Test | Luer connector shall not show signs of leakage, sufficient to form falling drop of water, over a hold period of 30s to 35s while being subjected to an applied pressure of between 300kPa and 330kPa. | Pass |
| | Sub-atmospheric Pressure Air Leakage Test | Luer connector shall not leak by more than 0,005 Pa·m3/s while being subjected to an applied sub-atmospheric pressure of between 80,0 kPa and 88,0 kPa over a hold period of between 15s and 20s. | Pass |
| | Stress Cracking Test | Luer connector shall meet the requirements of fluid leakage after being subjected to stress as per described in Annex E. | Pass |
K252643 - Page 15 of 22
{19}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Subject Device 2 – Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Performance Test | | | |
| ISO 9360-1 | Moisture Loss | Product shall be tested according to ISO 9360-1 to assess moisture loss, whereby the mass of water lost from the test apparatus is recorded. | Pass |
| ISO 9360-1 ISO 23328-2 | Pressure Drop | Product shall be tested according to ISO 9360-1 and ISO 23328-1 to assess resistance to flow (pressure drop) across the device under specified flowrates. • Vt 150ml (15 L/min, 30 L/min, 60 L/min, 90 L/min) • Vt 250ml (15 L/min, 30 L/min, 60 L/min, 90 L/min) • Vt 500ml (15 L/min, 30 L/min, 60 L/min, 90 L/min) • Vt 750ml (15 L/min, 30 L/min, 60 L/min, 90 L/min) • Vt 1000ml (15 L/min, 30 L/min, 60 L/min, 90 L/min) | Pass |
| ISO 9360-1 ISO 23328-2 | Gas Leakage | Gas leakage characteristics of the breathing system filter was carried out in accordance with Clause 5.3 of ISO 23328 2, whereby the measurement of gas leakage is determined in accordance with clause 6.4 of ISO 9360 1. | Pass |
| ISO 9360-1 | Compliance | Product shall be tested according to ISO 9360-1 to assess compliance. | Pass |
| ISO 9360-1 ISO 23328-2 | Dead Space | The total dead space is determined by the final weight of the product after filling with water minus the initial weight of the product before filling with water. | Pass |
| ISO 23328-1 | Salt Test | An aerosol of sodium chloride particles is generated and passed through the BSF at the specified flow rate of 15L/min and 30 L/min. The concentrations of particles upstream and downstream of the BSF are determined in accordance with clause 6.4 of ISO 9360 1. | Pass |
K252643 - Page 16 of 22
{20}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Performance Test | | | |
| | | measured, and filtration efficiency is calculated in accordance with Clause 5 of ISO 23328-1. | |
| ASTM F2101 | Bacterial Filtration Efficiency (BFE) | Product shall be tested according to ASTM F2101 to assess bacterial filtration efficiency performance, employing a ratio of the upstream bacterial challenge to downstream residual concentration to determine filtration efficiency of materials. | Pass |
| | Virus Filtration Efficiency (VFE) | Product shall be tested according to ASTM F2101 to assess virus filtration efficiency performance, employing a ratio of the upstream viral challenge to downstream residual concentration to determine filtration efficiency of materials. | Pass |
| ISO 5356-1 | Connector Gauging Test | Leading edge lie between the minimum and maximum diameter step of the gauge. | Pass |
| ISO 80369-7 ISO 80369-20 | Resistance to Separation from Axial Load Test | Luer connector shall not separate from the reference connector over a hold period between 10s and 15s while being subjected to a disconnection applied axial force between 32N and 35N for Luer lock connectors. | Pass |
| | Resistance to Separation from Unscrewing Test | Luer lock shall not separate from the reference connector for a hold period between 10s and 15s while being subjected to disconnection applied axle force between 32N and 35N for Luer lock connectors. | Pass |
| | Resistance to Overriding Test | Luer lock connector shall not override the threads or lugs of the reference connector while being subjected to an applied torque of between 0,15 N·m to 0,17 N·m over a hold period between 5s and 10s. | Pass |
| | Positive Pressure Liquid Leakage Test | Luer connector shall not show signs of leakage, sufficient to form falling drop of water, over | Pass |
K252643 - Page 17 of 22
{21}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Performance Test | | | |
| | | a hold period of 30s to 35s while being subjected to an applied pressure of between 300kPa and 330kPa. | |
| | Sub-atmospheric Pressure Air Leakage Test | Luer connector shall not leak by more than 0,005 Pa·m3/s while being subjected to an applied sub-atmospheric pressure of between 80,0 kPa and 88,0 kPa over a hold period of between 15s and 20s. | Pass |
| | Stress Cracking Test | Luer connector shall meet the requirements of fluid leakage after being subjected to stress as per described in Annex E. | Pass |
Biological testing was performed according to the requirements of ISO 10993-1 Fifth Edition 2018-08 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process, and ISO 18562-1 Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare Applications – Part 1: Evaluation and Testing within a Risk Management Process. The testing are summarize in the table below:
K252643 - Page 18 of 22
{22}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Subject Device 1 – Iso-Gard Filter Straight
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Biocompatibility Test | | | |
| ISO 10993-5 | Cytotoxicity | Assess whether the final and completed device emits any harmful substances. This evaluation conducted using an in vitro assay on mouse fibroblast (L929) cells. Reactivity scores greater than 2 indicate cytotoxicity for the device. | Pass |
| ISO 10993-10 | Sensitization | Determine if the extracts of the test article trigger an immune response that results in an allergic reaction. Consider a device sensitizing if it scores ≥1 according to Magnusson and Kligman grading scale. | Pass |
| ISO 10993-23 | Intracutaneous Irritation | Determine if the device produces a local irritation response in the dermal tissues. This in vivo assay is carried out on three rabbits. Device is classified as an irritant if the comparative result. Average score of the test animals minus that of the control animals, is greater than 1. | Pass |
| ISO 10993-11 ISO 10993-12 | Acute Systemic Toxicity | The endpoints were evaluated through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 10993-3 | Genotoxicity | Evaluation through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 10993-11 | Subacute Toxicity | Evaluation through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 18562-2 ISO 18562-3 | Particulate Matter Volatile Organic Substance Toxicology Risk Assessment | Particulate and VOC testing followed by toxicological risk assessment is to determine whether VOC and particulate exposure from the device produces toxicological risk per ISO 10993-17. Testing and evaluation is considered acceptable if the toxicological risk assessment concludes that “All the chemical exposures below the conservative TSL ≤30 d of 120μg, TTC of 120μg/day, or derived TIs”. | Pass |
| ISO 18562-4 | Leachable in Condensate | Chemical characterization followed by toxicological risk assessment is to determine whether extractable and | Pass |
K252643 - Page 19 of 22
{23}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| | Toxicological Risk Assessment | leachables exposure from the device produce toxicological risk per ISO 10993-17. Testing and evaluation is considered acceptable if the toxicological risk assessment concludes that “All the chemical exposures below the conservative TSL ≤30 d of 120μg, TTC of 120μg/day, or derived TIs”. | |
K252643 - Page 20 of 22
{24}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
Subject Device 2 – Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Biocompatibility Test | | | |
| ISO 10993-5 | Cytotoxicity | Assess whether the final and completed device emits any harmful substances. This evaluation conducted using an in vitro assay on mouse fibroblast (L929) cells. Reactivity scores greater than 2 indicate cytotoxicity for the device. | Pass |
| ISO 10993-10 | Sensitization | Determine if the extracts of the test article trigger an immune response that results in an allergic reaction. Consider a device sensitizing if it scores ≥1 according to Magnusson and Kligman grading scale. | Pass |
| ISO 10993-23 | Intracutaneous Irritation | Determine if the device produces a local irritation response in the dermal tissues. This in vivo assay is carried out on three rabbits. Device is classified as an irritant if the comparative result. Average score of the test animals minus that of the control animals, is greater than 1. | Pass |
| ISO 10993-11 ISO 10993-12 | Acute Systemic Toxicity | The endpoints were evaluated through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 10993-3 | Genotoxicity | Evaluation through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 10993-11 | Subacute Toxicity | Evaluation through the process of chemical characterization and toxicological risk assessment. | Pass |
| ISO 18562-2 | Particulate Matter | Particulate and VOC testing followed by toxicological risk assessment is to determine whether VOC and particulate exposure from the device produces toxicological risk per ISO 10993-17. Testing and evaluation is considered acceptable if the toxicological risk assessment concludes that “All the chemical exposures below the conservative TSL ≤30 d of 120μg, TTC of 120μg/day, or derived TIs”. | Pass |
| ISO 18562-3 | Volatile Organic Substance | | |
| ISO 18562-4 | Toxicology Risk Assessment | Chemical characterization followed by toxicological risk assessment is to determine whether extractable and toxicological risk assessment is considered acceptable if the toxicological risk assessment concludes that “All the chemical exposures below the conservative TSL ≤30 d of 120μg, TTC of 120μg/day, or derived TIs”. | Pass |
K252643 - Page 21 of 22
{25}
510(k) Submission
Iso-Gard Filter Straight
Humid-Vent Filter Compact Angled
| Test Standard | Test Performed | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| | Toxicological Risk Assessment | leachables exposure from the device produce toxicological risk per ISO 10993-17. Testing and evaluation is considered acceptable if the toxicological risk assessment concludes that “All the chemical exposures below the conservative TSL ≤30 d of 120μg, TTC of 120μg/day, or derived TIs”. | |
# N. Conclusion
The subject device has the same indications for use and technological characteristics and construction as the predicate. Test results demonstrate that the subject device meets its intended use. It is for these reasons that the subject device can be found substantially equivalent to the predicate devices.
The reference device, which has similar indications and technological characteristics, provides a reference to a more recently cleared device.
The conclusion drawn from the non-clinical performance test demonstrate that the device (Iso-Gard Filter Straight and Humid-Vent Filter Compact Angled) is substantially equivalent to the legally marketed device.
K252643 - Page 22 of 22
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.