K252619 · Qstag, Inc. · JFY · Feb 20, 2026 · Clinical Chemistry
Device Facts
Record ID
K252619
Device Name
QSCHECK UISACR
Applicant
Qstag, Inc.
Product Code
JFY · Clinical Chemistry
Decision Date
Feb 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252619 · Feb 20, 2026
QSCHECK UISACR
Qstag, Inc.
Clinical urine samples from patients with microalbuminuria
The sponsor conducted a method comparison study using 300 unique clinical urine samples to evaluate the performance of the QSCHECK UISACR device against an FDA-cleared comparator device.
Method Comparison Study; Method comparison study using clinical samples
Patients with microalbuminuria; Sample Size: 300 unique urine samples; Number of Sites: 3
FDA cleared comparator device
Microalbumin, creatinine, and ACR agreement
Indications for Use
QSCHECK UISACR is intended for the in vitro semi-quantitative measurement of the following parameters: microalbumin and creatinine in urine specimens and for Albumin/Creatinine ratio (ACR) calculated using those values. The results may be used in conjunction with clinical evaluation as an aid in the assessment for kidney function. The system is intended for prescription use only, in clinical laboratory settings.
Device Story
QSCHECK UISACR is a semi-quantitative urinalysis kit; includes reagent test strips and QSCHECK-app smartphone application. User dips test strip in urine sample; removes residual liquid; waits 60 seconds; captures image of strip using dedicated smartphone (iPhone 14 or Galaxy S23). App analyzes color change on reagent pads (affinity binding for albumin; dye-metal complex for creatinine); calculates Albumin/Creatinine ratio (ACR). Output provided to clinician for kidney function assessment. System performs automatic calibration; requires specific lighting and scan angles; prompts user for rescans if environmental conditions are suboptimal. Benefits include rapid, point-of-care semi-quantitative screening for microalbuminuria in clinical settings.
Clinical Evidence
No clinical trials were performed. Substantial equivalence is supported by analytical performance data, including precision, linearity, interference, and a method comparison study. Method comparison involved 300 patient urine samples compared against an FDA-cleared comparator, showing 89.5-89.7% exact agreement for analytes and 90.6-90.8% for ACR across both smartphone platforms.
Technological Characteristics
Test strips use dye-binding (sulfonephthalein) for microalbumin and dye-metal complex for creatinine. Smartphone-based colorimetric analysis using RGB to HSV conversion. Compatible with iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16). Manufactured per EN ISO 13485:2016. Stability confirmed for 20 months at 2-30°C.
Indications for Use
Indicated for patients requiring assessment of kidney function via semi-quantitative measurement of microalbumin and creatinine in urine. Intended for prescription use in clinical laboratory settings.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K252619
B Applicant
QSTag Inc.
C Proprietary and Established Names
QSCHECK UISACR
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JFY | Class II | 21 CFR 862.1225 - Creatinine Test System | CH - Clinical Chemistry |
| JIR | Class I | 21 CFR 862.1645 - Urinary protein or albumin (nonquantitative) test system | CH - Clinical Chemistry |
| KQO | Class I | 21 CFR 862.2900 - Automated urinalysis system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Measurand:
Microalbumin and creatinine in urine
C Type of Test:
Semi-quantitative urinalysis
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
QSCHECK UISACR is intended for the in vitro semi-quantitative measurement of the following parameters: microalbumin and creatinine in urine specimens and for Albumin/Creatinine ratio (ACR) calculated using those values. The results may be used in conjunction with clinical evaluation as an aid in the assessment for kidney function. The system is intended for prescription use only, in clinical laboratory settings.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
QSCHECK UISACR is limited to clinical laboratory settings.
D Special Instrument Requirements:
Apple iPhone 14 (iOS v26) or Samsung Galaxy S23 (Android v16)
IV Device/System Characteristics:
A Device Description:
QSCHECK UISACR is provided as a kit that consists of:
- Single use test strips. Each test strip contains a minimum of 0.1 mg phenolsulfonphthalein and 30 mg citric acid for microalbumin and a minimum of 0.3 mg picric acid and 20 mg boric acid for creatinine.
- Instructions for use.
- QSCHECK-app smartphone application that runs on a dedicated smartphone (i.e., Apple iPhone 14 (iOS v26), Samsung Galaxy S23 (Android v16)). The test strip must be read only by the QSCHECK-app. The device does not include any tools for visual interpretation.
- Desiccant.
K252619 - Page 2 of 16
{2}
K252619 - Page 3 of 16
## B Principle of Operation:
QSCHECK UISACR urine test strips contain a reagent pad with enzymes and chemicals that change color based on the concentration of each biomarker in the urine. The microalbumin test is based on affinity binding of albumin to a sulfonaphthalein dye at a constant pH to develop a color ranging from pale green to aqua blue. The creatinine test is based on the reaction of creatinine with a dye-metal complex under an alkaline condition to form a purplish brown color complex. After completion of the reaction on the test strip, the QSCHECK-app smartphone application photographs and analyzes the degree of the color change, which are matched to microalbumin and creatinine values that are provided by the app.
## C Instrument Description Information:
1. Instrument Name:
Apple iPhone 14 (iOS v26) or Samsung Galaxy S23 (Android v16)
2. Specimen Identification:
The QSCHECK-app will instruct the user to enter their patient identification information in the application.
3. Specimen Sampling and Handling:
The test is performed using a urine sample collected in a urine cup.
4. Calibration:
The QSCHECK-app performs automatic calibration during the analysis process for each measurement. User calibration is not required.
5. Quality Control:
No external quality controls are provided. The use of commercially available quality control material is recommended.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
ACR | LAB Urine Analysis Test System
### B Predicate 510(k) Number(s):
K182384
{3}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K252619 | K182384 |
| --- | --- | --- |
| Device Trade Name | QSCHECK UISACR | ACR | LAB Urine Analysis Test System |
| General Device Characteristic Similarities | | |
| Intended Use | Intended for the measurement of microalbumin and creatinine in urine | Same |
| Sample Type | Human urine | Same |
| General Device Characteristic Differences | | |
| Reader | Dedicated Apple iPhone 14 (iOS v26) or Samsung Galaxy S23 (Android v16) | Dedicated Apple iPhone 7 (iOS 12.0) |
VI Standards/Guidance Documents Referenced:
- Clinical and Laboratory Standards Institute (CLSI) EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, Third Edition.
- CLSI EP07, 3rd Edition: Interference Testing in Clinical Chemistry.
- CLSI EP09c, 3rd Edition: Measurement Procedure Comparison and Bias Estimation Using Patient Samples.
- CLSI EP12-A2: User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline – Second Edition.
- CLSI EP15-A3: User Verification of Precision and Estimation of Bias; Approved Guideline – Third Edition.
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
- CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.
- CLSI GP16-A3: Urinalysis; Approved Guideline – Third Edition.
- ISO 17511, 2nd Edition: In vitro diagnostic medical devices – Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples.
- ISO 14971, 3rd Edition: Medical Devices – Application of Risk Management to Medical Devices.
- IEC 62304, Edition 1.1: Medical Device Software – Software Life Cycle Processes.
- FDA guidance, “General Principles of Software Validation” issued on January 11, 2002.
- FDA Guidance, “Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023.
- FDA Guidance, “Postmarket Management of Cybersecurity in Medical Devices” issued on December 28, 2016.
K252619 - Page 4 of 16
{4}
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Single-site Precision
Precision testing was conducted at a single site using one Apple iPhone 14 (iOS v26) smartphone and one Samsung Galaxy S23 (Android v16) smartphone by one operator with a single test strip lot. Testing was conducted using human urine spiked with albumin and creatinine at the following concentrations: Sample 1 (~1.4 mg/dL albumin, ~213 mg/dL creatinine), Sample 2 (~8.0 mg/dL albumin, ~115 mg/dL creatinine), and Sample 3 (~15.9 mg/dL albumin, ~13.7 mg/dL creatinine). For within-run precision, each sample was tested in replicates of twenty in a single run over a single day. For between-run precision, each sample was tested in replicates of ten per run with two runs per day, over a single day. For between-day precision, each sample was tested in replicates of four per run with one run per day, over five days. The results are summarized in the tables below.
Apple iPhone 14 (iOS v26)
| Microalbumin | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | | Between-Run | | Between-Day | |
| Sample level (mg/dL) | Measuring Block (mg/dL) | N | Exact Match Rate (%) | N | Exact Match Rate (%) | N | Exact Match Rate (%) |
| 1.4 | 1 | 20 | 100 | 20 | 100 | 20 | 100 |
| 8.0 | 8 | 20 | 100 | 20 | 100 | 20 | 100 |
| 15.9 | 15 | 20 | 100 | 20 | 100 | 20 | 100 |
| Creatinine | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | | Between-Run | | Between-Day | |
| Sample level (mg/dL) | Measuring Block (mg/dL) | N | Exact Match Rate (%) | N | Exact Match Rate (%) | N | Exact Match Rate (%) |
| 13.7 | 10 | 20 | 100 | 20 | 100 | 20 | 100 |
| 115 | 100 | 20 | 100 | 20 | 100 | 20 | 100 |
| 213 | 200 | 20 | 100 | 20 | 100 | 20 | 100 |
K252619 - Page 5 of 16
{5}
Samsung Galaxy S23 (Android v16)
| Microalbumin | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | | Between-Run | | Between-Day | |
| Sample level (mg/dL) | Measuring Block (mg/dL) | N | Exact Match Rate (%) | N | Exact Match Rate (%) | N | Exact Match Rate (%) |
| 1.4 | 1 | 20 | 100 | 20 | 100 | 20 | 100 |
| 8.0 | 8 | 20 | 100 | 20 | 100 | 20 | 100 |
| 15.9 | 15 | 20 | 100 | 20 | 100 | 20 | 100 |
| Creatinine | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | | Between-Run | | Between-Day | |
| Sample level (mg/dL) | Measuring Block (mg/dL) | N | Exact Match Rate (%) | N | Exact Match Rate (%) | N | Exact Match Rate (%) |
| 13.7 | 10 | 20 | 100 | 20 | 100 | 20 | 100 |
| 115 | 100 | 20 | 100 | 20 | 100 | 20 | 100 |
| 213 | 200 | 20 | 100 | 20 | 100 | 20 | 100 |
## Reproducibility
Reproducibility testing was conducted at three sites using three Apple iPhone 14 (iOS v26) smartphone and three Samsung Galaxy S23 (Android v16) smartphones. Each site had one operator performing testing of three spiked urine samples (Sample 1: 1.6 mg/dL albumin, 219 mg/dL creatinine; Sample 2: 9.1 mg/dL albumin, 116 mg/dL creatinine; Sample 3: 17.4 mg/dL albumin, 16 mg/dL creatinine) with one Apple iPhone 14 and one Samsung Galaxy S23. Each sample was tested in replicates of five per run, one run per day over five days using three test strip lots for a total of 225 results per level across sites and lots per smartphone type. The results are summarized in the table below.
Apple iPhone 14 (iOS v26)
| Sample | N | Microalbumin | | Creatinine | |
| --- | --- | --- | --- | --- | --- |
| | | Measuring Block (mg/dL) | Exact Match Rate (%) | Measuring Block (mg/dL) | Exact Match Rate (%) |
| 1 | 225 | 1 | 100 | 200 | 100 |
| 2 | 225 | 8 | 100 | 100 | 100 |
| 3 | 225 | 15 | 100 | 10 | 100 |
K252619 - Page 6 of 16
{6}
Samsung Galaxy S23 (Android v16)
| Sample | N | Microalbumin | | Creatinine | |
| --- | --- | --- | --- | --- | --- |
| | | Measuring Block (mg/dL) | Exact Match Rate (%) | Measuring Block (mg/dL) | Exact Match Rate (%) |
| 1 | 225 | 1 | 100 | 200 | 100 |
| 2 | 225 | 8 | 100 | 100 | 100 |
| 3 | 225 | 15 | 100 | 10 | 100 |
# 2. Linearity:
The reportable range for each analyte was evaluated by measuring five levels of creatinine and albumin in human urine at the following expected concentrations: Level 1 (1.3 mg/dL albumin, 311 mg/dL creatinine), Level 2 (4.8 mg/dL albumin, 217 mg/dL creatinine), Level 3 (9.5 mg/dL albumin, 109 mg/dL creatinine), Level 4 (15.9 mg/dL albumin, 61 mg/dL creatinine), Level 5 (16.3 mg/dL albumin, 17 mg/dL creatinine). The study was conducted by three operators. Each operator tested one Apple iPhone 14 (iOS v26) and one Samsung Galaxy S23 (Android v16) using three test strip lots in replicates of ten per sample per lot for a total of 90 replicates per sample per smartphone type. The results of the study are shown below.
Apple iPhone 14 (iOS v26)
| Sample | N | Microalbumin | | Creatinine | |
| --- | --- | --- | --- | --- | --- |
| | | Measuring Block (mg/dL) | Exact Match Rate (%) | Measuring Block (mg/dL) | Exact Match Rate (%) |
| 1 | 90 | 1 | 100 | 300 | 100 |
| 2 | 90 | 3 | 100 | 200 | 100 |
| 3 | 90 | 8 | 100 | 100 | 100 |
| 4 | 90 | 15 | 100 | 50 | 100 |
| 5 | 90 | 15 | 100 | 10 | 100 |
Samsung Galaxy S23 (Android v16)
| Sample | N | Microalbumin | | Creatinine | |
| --- | --- | --- | --- | --- | --- |
| | | Measuring Block (mg/dL) | Exact Match Rate (%) | Measuring Block (mg/dL) | Exact Match Rate (%) |
| 1 | 90 | 1 | 100 | 300 | 100 |
| 2 | 90 | 3 | 100 | 200 | 100 |
| 3 | 90 | 8 | 100 | 100 | 100 |
| 4 | 90 | 15 | 100 | 50 | 100 |
| 5 | 90 | 15 | 100 | 10 | 100 |
# 3. Analytical Specificity/Interference:
The analytical specificity of the QSCHECK UISACR was assessed using spiked human urine samples at three levels: Level 1 (1.3 mg/dL microalbumin, 211 mg/dL creatinine), Level 2
K252619 - Page 7 of 16
{7}
(8.6 mg/dL microalbumin, 114 mg/dL creatinine) and Level 3 (20.9 mg/dL microalbumin, 14 mg/dL creatinine). Each level sample was further divided into two aliquots: control (with no added interferent) and test (with added interferent). Each sample was measured in replicates of ten using one lot of test strip on Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16).
The results of the test sample with potentially interfering substance were compared to the results of the control sample. No interference was defined as 100% agreement between test and control samples. Interference was defined as a change in output of one or more color blocks between test and control samples. When interference was observed, dose response testing was performed to identify the lowest concentration at which interference was observed.
The substances at the concentrations listed in the table below were found not to interfere with microalbumin, creatinine and ACR.
| Substance | Highest concentration of substance at which no interference was observed (mg/dL unless otherwise indicated) |
| --- | --- |
| Acetaminophen | 300 |
| Albumin (only creatinine was tested as the measurand) | 880 |
| Ascorbic acid | 220 |
| Bilirubin | 6 |
| Caffeine | 37 |
| Blood | 0.05% |
| Calcium Chloride | 210 |
| Captopril | 12 |
| Citric Acid | 100 |
| Creatinine (only microalbumin was tested as the measurand) | 750 |
| Cyanocobalamin (Vitamin B12) | 0.16 |
| Dapagliflozin | 0.03 |
| Fructose | 100 |
| Glucose | 5000 |
| Glycine | 450 |
| Hemoglobin | 6.8 |
| Hydrochlorothiazide | 5 |
| Hydroxychloroquine sulfate | 8 |
| Iron Sulfate | 24 |
| Acetoacetic Acid | 300 |
| Leukocytes | 2500 leukocytes/μL |
| Losartan | 5 |
| Metformin Hydrochloride | 285 |
| Phenolphthalein | 1060 |
K252619 - Page 8 of 16
{8}
The following tables show the substances that interfered with the proposed device. Results are expressed as the lowest concentration of the substance that exhibits interference and the resulting change in color block output.
Albumin
| Interferent | Lowest concentration interference was observed (mg/dL) | Analyte level (mg/dL) | Effect on results | |
| --- | --- | --- | --- | --- |
| | | | iPhone 14 | Galaxy S23 |
| Sodium Bicarbonate | 1300 | 1 | Falsely elevated by one color block | Falsely elevated by one color block |
| | | 8 | Falsely elevated by one color block | Falsely elevated by one color block |
| | | 15 | No change | No change |
| Sodium-2-mercaptoethanesulfonate (Mesna) | 750 | 1 | Falsely elevated by one color block | Falsely elevated by one color block |
| | | 8 | Falsely elevated by one color block | Falsely elevated by one color block |
| | | 15 | No change | No change |
Creatinine
| Interferent | Lowest concentration interference was observed (mg/dL) | Analyte level (mg/dL) | Effect on results | |
| --- | --- | --- | --- | --- |
| | | | iPhone 14 | Galaxy S23 |
| Ascorbic acid | 330 | 200 | No change | Falsely decreased by one color block |
| | | 100 | Falsely decreased by one color block | Falsely decreased by one color block |
| | | 10 | No change | No change |
| Sodium Bicarbonate | 1300 | 200 | Falsely decreased by one color block | Falsely decreased by one color block |
| | | 100 | Falsely decreased by one color block | Falsely decreased by one color block |
| | | 10 | No change | No change |
K252619 - Page 9 of 16
{9}
| Sodium-2-mercaptoethanesulfonate (Mesna) | 750 | 200 | Falsely decreased by one color block | Falsely decreased by one color block |
| --- | --- | --- | --- | --- |
| | | 100 | Falsely decreased by one color block | Falsely decreased by one color block |
| | | 10 | No change | No change |
## Specific gravity
Human urine samples spiked with albumin and creatinine (Level 1: 1.6 mg/dL albumin, 221 mg/dL creatinine; Level 2: 7.9 mg/dL albumin, 127 mg/dL creatinine; Level 3: 16.9 mg/dL albumin, 19 mg/dL creatinine) were adjusted to specific gravities of 1.000, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030, 1.035, and 1.040 and tested. No interference from specific gravity was detected for albumin or creatinine measurement in the examined range on both Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16).
## pH
Human urine samples spiked with albumin and creatinine (Level 1: 1.6 mg/dL albumin, 221 mg/dL creatinine; Level 2: 7.9 mg/dL albumin, 127 mg/dL creatinine; Level 3: 16.9 mg/dL albumin, 19 mg/dL creatinine) were adjusted to pHs of 4, 5, 6, 7, 8, and 9 and tested. No interference from pH was detected for albumin or creatinine measurement in the examined range on both Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16).
## 4. Assay Reportable Range:
The reportable values for microalbumin are 1, 3, 8, and 15 mg/dL and for creatinine are 10, 50, 100, 200 and 300 mg/dL. The reportable values for ACR are Normal (ACR < 30 mg albumin/g creatinine), Abnormal (30-300 mg albumin/g creatinine) and Very Abnormal (> 300 mg albumin/g creatinine).
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Traceability
Traceability is established using an internal control material.
### Sample stability
Sample stability testing was conducted with twenty native human urine samples that were stored under various storage conditions. QSCHECK UISACR urine test strips with Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16) smartphones were used to test samples at baseline and after storage. The sample stability results support the labeling claim that if the test is not performed immediately, samples may be stored in a freezer at -20°C for up to 3 weeks.
## 6. Detection Limit:
A limit of detection study was conducted to determine the cutoff points for every color block for both microalbumin and creatinine. Human urine samples spiked with microalbumin (0.7 to 22.3 mg/dL) or creatinine (3 to 334 mg/dL) were tested using three test strip lots and ten replicates per lot using one Apple iPhone 14 (iOS v26) smartphone and one Samsung Galaxy S23 (Android v16) smartphone. The cutoff for each color block is defined as the lowest
K252619 - Page 10 of 16
{10}
concentration of analyte at which over 50% of the test results were positive for each measuring block. The results are shown below.
Apple iPhone 14 (iOS v26)
| Microalbumin | | | Creatinine | | |
| --- | --- | --- | --- | --- | --- |
| Color Block (mg/dL) | Cut off Value (mg/dL) | Match Rate (%) | Color Block (mg/dL) | Cut off Value (mg/dL) | Match Rate (%) |
| 1 | 0.7 | 100% | 10 | 3 | 100% |
| 3 | 2.6 | 63.3% | 50 | 47 | 83.3% |
| 8 | 6.6 | 83.3% | 100 | 82 | 86.7% |
| 15 | 12.3 | 63.3% | 200 | 188 | 63.3% |
| | | | 300 | 281 | 60.0% |
Samsung Galaxy S23 (Android v16)
| Microalbumin | | | Creatinine | | |
| --- | --- | --- | --- | --- | --- |
| Color Block (mg/dL) | Cut off Value (mg/dL) | Match Rate (%) | Color Block (mg/dL) | Cut off Value (mg/dL) | Match Rate (%) |
| 1 | 0.7 | 100% | 10 | 3 | 100% |
| 3 | 2.6 | 56.7% | 50 | 47 | 80.0% |
| 8 | 6.6 | 80.0% | 100 | 82 | 83.3% |
| 15 | 12.3 | 66.7% | 200 | 188 | 70.0% |
| | | | 300 | 281 | 56.7% |
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
See Section VII.B.1. Method Comparison.
9. Carry-Over:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
A method comparison study was performed at three clinical sites with 300 unique urine samples (100 samples per site) from microalbuminuria patients by one operator. The results from the QSCHECK UISACR were compared to the results from an FDA cleared comparator device. No altered samples were used in this study. The microalbumin, creatinine, and ACR results for Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android OS v16) smartphones, for combined sites, is shown below.
K252619 - Page 11 of 16
{11}
Apple iPhone 14 (iOS v26)
| Albumin | Reported blocks (mg/dL) | Comparator device | | | | Total n |
| --- | --- | --- | --- | --- | --- | --- |
| | | 1 | 3 | 8 | 15 | |
| QSCHECK UISACR | 1 | 69 (90.8%) | 3 (4.4%) | 0 | 0 | 72 |
| | 3 | 7 (9.2%) | 59 (86.8%) | 4 (5.6%) | 0 | 70 |
| | 8 | 0 | 6 (8.8%) | 63 (88.7%) | 7 (8.2%) | 76 |
| | 15 | 0 | 0 | 4 (5.6%) | 78 (91.8%) | 82 |
| Total n | | 76 | 68 | 71 | 85 | 300 |
| % Exact agreement | | 90.8% | 86.8% | 88.7% | 91.8% | 89.5% |
| % Within one block | | 100% | 100% | 100% | 100% | 100% |
| Creatinine | Reported blocks (mg/dL) | Comparator device | | | | | Total n |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | 10 | 50 | 100 | 200 | 300 | |
| QSCHECK UISACR | 10 | 66 (91.7%) | 6 (10.9%) | 0 | 0 | 0 | 72 |
| | 50 | 6 (8.3%) | 48 (87.3%) | 2 (3.9%) | 0 | 0 | 56 |
| | 100 | 0 | 1 (1.8%) | 44 (86.3%) | 2 (3.7%) | 0 | 47 |
| | 200 | 0 | 0 | 5 (9.8%) | 50 (92.6%) | 7 (10.3%) | 62 |
| | 300 | 0 | 0 | 0 | 2 (3.7%) | 61 (89.7%) | 63 |
| Total n | | 72 | 55 | 51 | 54 | 68 | 300 |
| % Exact agreement | | 91.7% | 87.3% | 86.3% | 92.6% | 89.7% | 89.5% |
| % Within one block | | 100% | 100% | 100% | 100% | 100% | 100% |
| ACR | Reported result | Comparator device | | | |
| --- | --- | --- | --- | --- | --- |
| | | Normal | Abnormal | Very Abnormal | Total n |
| QSCHECK UISACR | Normal | 101 (92.7%) | 8 (5.6%) | 0 | 109 |
| | Abnormal | 8 (7.3%) | 128 (89.5%) | 5 (10.4%) | 141 |
| | Very Abnormal | 0 | 7 (4.9%) | 43 (89.6%) | 50 |
| Total n | | 109 | 143 | 48 | 300 |
| % Exact agreement | | 92.7% | 89.5% | 89.6% | 90.6% |
| % Within one block | | 100% | 100% | 100% | 100% |
K252619 - Page 12 of 16
{12}
Samsung Galaxy S23 (Android v16)
| Albumin | Reported blocks (mg/dL) | Comparator device | | | | Total n |
| --- | --- | --- | --- | --- | --- | --- |
| | | 1 | 3 | 8 | 15 | |
| QSCHECK UISACR | 1 | 68 (89.5%) | 3 (4.4%) | 0 | 0 | 71 |
| | 3 | 8 (10.5%) | 61 (89.7%) | 4 (5.6%) | 0 | 73 |
| | 8 | 0 | 4 (5.9%) | 63 (88.7%) | 8 (9.4%) | 75 |
| | 15 | 0 | 0 | 4 (5.6%) | 77 (90.6%) | 81 |
| Total n | | 76 | 68 | 71 | 85 | 300 |
| % Exact agreement | | 89.5% | 89.7% | 88.7% | 90.6% | 89.6% |
| % Within one block | | 100% | 100% | 100% | 100% | 100% |
| Creatinine | Reported blocks (mg/dL) | Comparator device | | | | | Total n |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | 10 | 50 | 100 | 200 | 300 | |
| QSCHECK UISACR | 10 | 65 (90.3%) | 5 (9.1%) | 0 | 0 | 0 | 70 |
| | 50 | 7 (9.7%) | 49 (89.1%) | 3 (5.9%) | 0 | 0 | 59 |
| | 100 | 0 | 1 (1.8%) | 44 (86.3%) | 2 (3.7%) | 0 | 47 |
| | 200 | 0 | 0 | 4 (7.8%) | 50 (92.6%) | 7 (10.3%) | 61 |
| | 300 | 0 | 0 | 0 | 2 (3.7%) | 61 (89.7%) | 63 |
| Total n | | 72 | 55 | 51 | 54 | 68 | 300 |
| % Exact agreement | | 90.3% | 89.1% | 86.3% | 92.6% | 89.7% | 89.6% |
| % Within one block | | 100% | 100% | 100% | 100% | 100% | 100% |
| ACR | Reported result | Comparator device | | | |
| --- | --- | --- | --- | --- | --- |
| | | Normal | Abnormal | Very Abnormal | Total n |
| QSCHECK UISACR | Normal | 101 (92.7%) | 9 (6.3%) | 0 | 110 |
| | Abnormal | 8 (7.3%) | 129 (90.2%) | 5 (10.4%) | 142 |
| | Very Abnormal | 0 | 5 (3.5%) | 43 (89.6%) | 48 |
| Total n | | 109 | 143 | 48 | 300 |
| % Exact agreement | | 92.7% | 90.2% | 89.6% | 90.8% |
| % Within one block | | 100% | 100% | 100% | 100% |
K252619 - Page 13 of 16
{13}
2. **Matrix Comparison:**
Not applicable. This test is only for use with urine samples.
## C Clinical Studies:
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not applicable.
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
The following statements regarding expected values are provided in the device labeling:
"The expected values for the QSCHECK UISACR are as follows:
Albumin: Albumin is normally present in urine at concentrations of less than 2 mg/dL.¹ Moderately increased albuminuria is defined as an albumin excretion rate of 30~299 mg/24 hours.²,³
Creatinine: Creatinine concentrations of 10-300 mg/dL are normally present in urine.
Albumin-to-Creatinine Ratio: Albumin is normally present in urine at concentrations of < 30 mg/g, where mg/g represents mg albumin per g creatinine. Moderately increased albuminuria is indicated at a ratio result of 30 - 300 mg/g and severely increased albuminuria at a ratio of > 300 mg/g.⁴"
## References:
1. Burtis C.A.; and Ashwood ER.: Tietz Textbook of Clinical Chemistry 3rd ed. Philadelphia: Saunders; 1999; pp.483-484.
2. Mangili, R. et al.: Prevalence of Hypertension and Microalbuminuria in Adult Type 1 (Insulin-Dependent) Diabetic patients Without Renal Failure in Italy-Validation of Screening Techniques to Detect Microalbuminuria. Acta Diabetol. 29: 156-166; 1992.
3. American Diabetes Association, Clinical Practice Recommendations, Diabetes Care, Vol. 31, Suppl. 1, January 2008.
4. Position Statement: Diabetic Nephropathy. Diabetes Care 20: S24-S27; 1997.
K252619 - Page 14 of 16
{14}
K252619 - Page 15 of 16
# F Other Supportive Instrument Performance Characteristics Data:
The sponsor conducted flex studies with human urine samples spiked with albumin and creatinine to three levels of albumin, creatinine and ACR. The expected results for the three levels of urine samples were 1 mg/dL, 8 mg/dL and 15 mg/dL for microalbumin; 10 mg/dL, 100 mg/dL and 200 mg/dL for creatinine; Normal, Abnormal, Very Abnormal for ACR. Each of the studies described below was conducted with one Apple iPhone 14 (iOSv26) and one Samsung Galaxy S23 (Android v16) using one lot of test strip. Each sample was tested on each phone in replicates of ten.
1. Dipping Time: This study was conducted to assess how test strip sampling times impact the accuracy of the QSCHECK-UISACR. Test strips were dipped in the urine samples for four different durations (1, 5, 10, or 30 seconds). The residual urine was immediately removed from the test strip and the assay time was held constant at 60 seconds. For both creatinine and albumin, the results support that a dipping time from 1 to 10 seconds does not impact test performance. The instructions for use and QSCHECK-app directs the professional user to dip the strip for 1 second to apply the sample. Users are warned that incorrect results may occur if the 1-second dipping time is not adhered to.
2. Reaction Time: This study was conducted to assess how the reaction time (between dipping the test strip in a urine sample and imaging the test strip) impacts the accuracy of QSCHECK-UISACR. Test strips were dipped in the urine samples for one second and the residual urine was immediately removed from the strip. The operator then waited a predetermined amount of time (30, 45, 60, 90 or 120 seconds) before imaging the test strip. For both creatinine and albumin, the results support that a reaction time from 45 to 90 seconds does not impact test performance. The instructions for use and QSCHECK-app directs the professional user to wait for 60 seconds to allow for the urine sample to react with the test strip prior to scanning the strip for analysis. Users are warned that incorrect results may occur if the 60-second reaction time is not adhered to.
3. Residual sample: This study was conducted to assess how the presence of residual sample on the test strip impacts the accuracy of QSCHECK-UISACR. For the condition without residual sample, test strips were dipped in samples for one second, the residual urine was immediately removed from the strip, and the strip was placed on a flat surface for 60 seconds to allow the reaction to occur. For the condition with residual sample, test strips were dipped in samples for 1 second and, without removing any residual urine, the test strip was then placed on a flat surface for 60 seconds to allow the reaction to occur. For both creatinine and albumin, the presence of residual sample impacted the accuracy of the test results on both iPhone 14 (iOS v26) and Galaxy S23 (Android v16). The instructions for use and QSCHECK-app directs the professional user to lightly tap the test strip on an absorbent surface to remove the residual urine. Users are warned that incorrect results may occur if residual specimen is not immediately removed after dipping the test strip in the urine.
4. Illumination intensity: This study was conducted to evaluate how environmental lighting conditions impact the accuracy of QSCHECK-UISACR. Test strips were dipped in each sample for one second, residual urine was immediately removed, and a 60-second reaction time was employed before imaging the test strip under different simulated light conditions (illumination intensities of 60-100 lx, 100-200 lx, 200-300 lx, 300-400 lx or 400-500 lx). For both creatinine and albumin, test performance was not impacted for any of the illumination
{15}
intensities tested. The instructions for use and QSCHECK-app describe that the test strip may need to be rescanned if the recommended lighting conditions (60x (a dimly lit laboratory) to 500 lx (typical laboratory lighting)) are not met. If lighting conditions fall outside this range, the QSCHECK-app will prompt the user to rescan the test strip (if the maximum 90-second reaction time frame has not been exceeded) or test using a new test strip (if the maximum 90-second reaction time frame has been exceeded).
5. Shadowing: This study was conducted to evaluate how shadows on the test strip impact the accuracy of QSCHECK-UISACR. Test strips were dipped in the urine test samples for one second, residual urine was immediately removed, and a 60-second reaction time was employed before imaging the test strip. Test strips were scanned under three different shadowing conditions: no shadow on the test strip, 25% area under shadow, and 50% area under shadow. For both creatinine and albumin, the results support that 0% to 25% shadowing does not impact the test performance. The instructions for use and QSCHECK-app describe that the test strip may need to be rescanned if more than 25% of the test strip's measurement area is covered by shadow. If more than 25% of the area of the test strip is covered by shadow, the QSCHECK-app will request a rescan of the test strip (if the maximum 90-second reaction time frame has not been exceeded) or test using a new test strip (if the maximum 90-second reaction time frame has been exceeded).
6. Scan Angles: This study was conducted to evaluate how the scan angle on the test strip impacts the accuracy of QSCHECK-UISACR. Test strips were dipped in each sample for one second, residual urine was immediately removed, and a 60-second reaction time was employed before imaging the test strip. Test strips were scanned at different angles (0°, 30°, or 60°). For both creatinine and albumin, the results support that a scan angle from 0° to 30° does not impact the test performance. The instructions for use and QSCHECK-app describe that the test strip may need to be rescanned if the scan angle is more than 30°. If the scan angle is more than 30°, the QSCHECK-app will request a rescan (if the maximum 90-second reaction time frame has not been exceeded) or test using a new test strip (if the maximum 90-second reaction time frame has been exceeded).
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K252619 - Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.