LUX-Dx II Insertable Cardiac Monitor (M302); LUX-Dx II+ Insertable Cardiac Monitor (M312)

K252593 · Boston Scientific Corporation · MXD · Sep 12, 2025 · Cardiovascular

Device Facts

Record IDK252593
Device NameLUX-Dx II Insertable Cardiac Monitor (M302); LUX-Dx II+ Insertable Cardiac Monitor (M312)
ApplicantBoston Scientific Corporation
Product CodeMXD · Cardiovascular
Decision DateSep 12, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2

Indications for Use

The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia, such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx ICM is indicated for atrial fibrillation monitoring in patients that have been previously diagnosed or treated for atrial fibrillation. The LUX-Dx has not been tested specifically for pediatric use.

Device Story

Subcutaneously implanted cardiac monitor (ICM) placed in fourth intercostal space of left chest wall; monitors and records subcutaneous ECG (S-ECG). Device evaluates S-ECG waveform data for arrhythmia indications; marks signals for clinical review when algorithm criteria met. System includes ICM, mobile applications (myLUX Patient app, LUX-Dx Clinic Assistant app) on OTS mobile devices, and LATITUDE Clarity server. ICM communicates with mobile app via Bluetooth Low Energy (BLE); app transmits data to server via cellular/Wi-Fi. Clinicians access server to perform remote monitoring, manage patient data, and review marked events. Enables long-term cardiac rhythm monitoring; assists in diagnosis of arrhythmias; facilitates clinical decision-making for patients with suspected or known cardiac conditions.

Clinical Evidence

No clinical data; no performance testing necessary as the device is identical to the predicate.

Technological Characteristics

Implantable device; titanium body, titanium nitride electrodes, implantation-grade polymer/epoxy header, Parylene coating. Battery-powered (LiMnO2); 3-year longevity. Dimensions: 7.2 x 44.8 x 4.0 mm; volume 1.2 cc; mass 3g. Wireless telemetry via BLE, cellular, and Wi-Fi. MR Conditional (1.5T and 3T). Sterilized via EtO. Software-based arrhythmia detection.

Indications for Use

Indicated for patients with known heart conditions at risk of abnormal heart rhythms or symptoms suggesting cardiac arrhythmia (dizziness, palpitations, syncope, chest pain, shortness of breath); also indicated for atrial fibrillation monitoring in patients previously diagnosed or treated for AF. Not tested for pediatric use.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 12, 2025 Boston Scientific Corporation Melissa Klamerus Principal Regulatory Affairs Specialist 4100 Hamline Avenue North Saint Paul, Minnesota 55112 Re: K252593 Trade/Device Name: LUX-Dx II Insertable Cardiac Monitor (M302); LUX-Dx II+ Insertable Cardiac Monitor (M312) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: August 15, 2025 Received: August 15, 2025 Dear Melissa Klamerus: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K252593 - Melissa Klamerus Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252593 - Melissa Klamerus assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JESSICA L. BATISTA-S Digitally signed by JESSICA L. BATISTA-S Date: 2025.09.12 13:25:16 -04'00" Jessica Batista Bertolini Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252593 | | | Device Name LUX-Dx II Insertable Cardiac Monitor (M302); LUX-Dx II+ Insertable Cardiac Monitor (M312) | | | Indications for Use (Describe) The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia, such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx ICM is indicated for atrial fibrillation monitoring in patients that have been previously diagnosed or treated for atrial fibrillation. The LUX-Dx has not been tested specifically for pediatric use. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Boston Scientific 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798 Tel: (651) 582-4000 Fax: (651) 582-5134 # 510(k) Summary ## 1. SUBMITTER Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798 Contact: Emily Bailey Senior Regulatory Specialist Phone: (952) 500-1763 Email: emily.bailey@bsci.com Date Prepared: 11-September-2025 ## 2. DEVICE Trade Names: LUX-Dx II™ Insertable Cardiac Monitor (M302) and LUX-Dx II+™ Insertable Cardiac Monitor (M312) Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025 ## 3. PREDICATE DEVICE Trade Name: LUX-Dx II™ Insertable Cardiac Monitor and LUX-Dx II+™ Insertable Cardiac Monitor Manufacturer: Boston Scientific Corp Clearance Number: K231328, 19-Aug-2023 Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025 510(k) Summary K252593 {5} Boston Scientific 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798 Tel: (651) 582-4000 Fax: (651) 582-5134 # 4. INDICATION FOR USE The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia, such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx ICM is indicated for atrial fibrillation monitoring in patients that have been previously diagnosed or treated for atrial fibrillation. The LUX-Dx has not been tested specifically for pediatric use. # 5. RATIONALE FOR DIFFERENCE FROM PREDICATE PER 21 CFR 807.92(A)(5) The Indication for Use statement includes additional clarifying language that differs from the predicate device. The additional language clarifies the patient population but does not alter it in any way. The intended use also remains the same and there are no new or increased risks associated with the additional language. # 6. DEVICE DESCRIPTION The LUX-Dx II and LUX-Dx II+ ICM devices evaluate S-ECG waveform data for indications of cardiac arrhythmias and "marks" the S-ECG signal for clinical presentation and evaluation when the algorithm criteria are met. The ICM device is inserted into the subcutaneous layer of the fourth intercostal space of the left chest wall. The ICM device is powered by an integrated battery. The LUX-Dx system includes the following main components: - ICM device - a subcutaneously-implanted cardiac monitor device for cardiac arrhythmia event data collection and transmission. In addition, symptom events are collected and transmitted from the device. - Mobile Monitor (MM) - mobile applications (myLUX™ Patient app and LUX-Dx™ Clinic Assistant app) running on an OTS mobile device that communicates with the ICM device (using Bluetooth Low Energy (BLE)) and the LATITUDE Clarity™ server (using cellular/Wi-Fi) for collection and transmission of event, patient, and device data. - LATITUDE Clarity™ server - a server that communicates with the Mobile Monitor for bidirectional data transmission and provides web access for clinicians to perform remote monitoring activities and manage general patient and system parameters and workflow activities. - System Accessories- for insertion of the ICM device, an insertion tool and incision tool are provided. In addition, a magnet is provided to initiate ICM/MM app communication. # 7. SUBSTANTIAL EQUIVALENCE | Category | Predicate (K231328) | Proposed | Comments | | --- | --- | --- | --- | | Device Name (Model) | LUX-Dx II Insertable Cardiac Monitor (M302) LUX-Dx II+ Insertable Cardiac Monitor (M312) | LUX-Dx II Insertable Cardiac Monitor (M302) LUX-Dx II+ Insertable Cardiac Monitor (M312) | Identical | | Pro Code / Regulation | MXD, 21 CFR 870.1025 Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection) | MXD, 21 CFR 870.1025 Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection) | Identical | 510(k) Summary K252593 Page 2 of 4 {6} Boston Scientific 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798 Tel: (651) 582-4000 Fax: (651) 582-5134 | Category | Predicate (K231328) | Proposed | Comments | | --- | --- | --- | --- | | | Special controls: Class II Special Controls Guidance - Arrhythmia Detector and Alarm, Oct 2003 | Special controls: Class II Special Controls Guidance - Arrhythmia Detector and Alarm, Oct 2003 | | | Indication for Use | The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use. | The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx ICM is indicated for atrial fibrillation monitoring in patients that have been previously diagnosed or treated for atrial fibrillation. The LUX-Dx has not been tested specifically for pediatric use. | Substantially Equivalent. | | Use | Single Use | Single Use | Identical | | Sterility | Supplied Sterile (EtO) | Supplied Sterile (EtO) | Identical | | MR Status | MR Conditional (1.5T and 3T) | MR Conditional (1.5T and 3T) | Identical | | Energy Source | Battery powered: Lithium-manganese dioxide (LiMnO2) | Battery powered: Lithium-manganese dioxide (LiMnO2) | Identical | | Longevity | 3 years | 3 years | Identical | | Materials of Construction | Body (Can) Material: Titanium Electrodes: Titanium Nitride Header: Implantation-grade Polymer, Epoxy MC31L Coating: Parylene | Body (Can) Material: Titanium Electrodes: Titanium Nitride Header: Implantation-grade Polymer, Epoxy MC31L Coating: Parylene | Identical | | Dimensions | Volume: 1.2 cc Mass: 3 grams Dimensions 7.2 x 44.8 x 4.0 (mm) | Volume: 1.2 cc Mass: 3 grams Dimensions 7.2 x 44.8 x 4.0 (mm) | Identical | | System Communication | Wireless telemetry using standard communication protocols: - Bluetooth Low Energy (BLE) - Cellular - Wi-Fi | Wireless telemetry using standard communication protocols: - Bluetooth Low Energy (BLE) - Cellular - Wi-Fi | Identical | | Accessories Packaged with the Device | Implant Tools (single use & supplied sterile) - Incision Tool: polycarbonate, stainless steel - Insertion Tool: polycarbonate | Implant Tools (single use & supplied sterile) - Incision Tool: polycarbonate, stainless steel - Insertion Tool: polycarbonate | Identical | 510(k) Summary K252593 Page 3 of 4 {7} Boston Scientific 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798 Tel: (651) 582-4000 Fax: (651) 582-5134 | Category | Predicate (K231328) | Proposed | Comments | | --- | --- | --- | --- | | Other Accessories | Mobile Device Software Apps - myLUX Patient Apps (2925, 2929, 2939) - Clinic Assistant App (2935) LATITUDE Clarity Server Software (7260) Magnet in Shielded Box (6386) myLUX Patient Kit (6385; magnet in shielded box + patient literature) | Mobile Device Software Apps - myLUX Patient Apps (2925, 2929, 2939) - Clinic Assistant App (2935) LATITUDE Clarity Server Software (7260) Magnet in Shielded Box (6386) myLUX Patient Kit (6385; magnet in shielded box + patient literature) | Identical | | Detection Algorithms | Atrial Fibrillation (AF) Tachycardia (Tachy) Bradycardia (Brady) Pause Atrial Tachycardia (AT) Premature Ventricular Contraction (PVC) | Atrial Fibrillation (AF) Tachycardia (Tachy) Bradycardia (Brady) Pause Atrial Tachycardia (AT) Premature Ventricular Contraction (PVC) | Identical | | Monitoring | Remote | Remote | Identical | | Mobile Device Options | Bring Your Own Device (BYOD) - Android - Apple iOS Boston Scientific Provided Mobile Device - Android | Bring Your Own Device (BYOD) - Android - Apple iOS Boston Scientific Provided Mobile Device - Android | Identical | # 8. SUMMARY OF PERFORMANCE TESTING No performance testing necessary. # 9. CONCLUSION Based on the intended use, fundamental technological characteristics, and performance testing, the proposed LUX-Dx II and LUX-Dx II+ have been shown to be appropriate for the intended use and are considered to be substantially equivalent to the predicate device. 510(k) Summary K252593 Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...