K252591 · Woorhi Mechatronics Co., Ltd. · QAI · May 13, 2026 · General, Plastic Surgery
Device Facts
Record ID
K252591
Device Name
Automatic Micro Needle System (CODE-X)
Applicant
Woorhi Mechatronics Co., Ltd.
Product Code
QAI · General, Plastic Surgery
Decision Date
May 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Automatic Micro Needle System (CODE-X) is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older.
Device Story
Motorized, AC-powered microneedling device; creates microscopic channels in epidermal/upper dermal skin layers. System comprises reusable handpiece with embedded microprocessor control and sterile, single-use 12-pin or 14-pin needle cartridges. Used in professional healthcare settings by clinicians. Handpiece drives reciprocating needle cartridge to specific depths (0-2.0 mm). Output is mechanical skin penetration; intended to improve facial acne scar appearance. Benefits include controlled, consistent micro-injury to stimulate skin. Clinician adjusts penetration depth via device controls; output informs clinical treatment progression.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including penetration depth accuracy, structural integrity, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), sterilization validation (ISO 11135), and shelf-life/packaging validation.
Technological Characteristics
AC-powered motorized handpiece; 12-pin/14-pin EO-sterilized single-use cartridges. 33G needles, straight cylindrical body with conical tapered point. Adjustable depth 0-2.0 mm (0.20 mm increments). Embedded microprocessor control. Biocompatibility per ISO 10993-1. Shelf life 3 years.
Indications for Use
Indicated for adults aged 22 years or older for the treatment to improve the appearance of facial acne scars.
Regulatory Classification
Identification
A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.
Special Controls
The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 13, 2026
Woorhi Mechatronics Co., Ltd.
Jungsu Lee
Official Correspondent
58, Sanbon-ro, 48beon-gil
Gunpo-si, Gyeonggi-do 15846
Republic Of Korea
Re: K252591
Trade/Device Name: Automatic Micro Needle System (CODE-X)
Regulation Number: 21 CFR 878.4430
Regulation Name: Microneedling device for aesthetic use
Regulatory Class: Class II
Product Code: QAI
Dated: April 12, 2026
Received: April 13, 2026
Dear Jungsu Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252591 - Jungsu Lee
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252591 - Jungsu Lee
Page 3
Sincerely,

For Jodie Giordano, Ph.D.
Assistant Director
DHT4B: Division of Infection Control and Plastic Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252591 | |
| Device Name Automatic Micro Needle System (CODE-X) | |
| Indications for Use (Describe) | |
| The Automatic Micro Needle System (CODE-X) is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(K) Summary
K252591
Submitter
Woorhi Mechatronics Co., Ltd.
Woo Yong Woon
58, Sanbon-ro 48beon-gil, Gunpo-si, Gyeonggi-do, Republic of Korea
Email: cns@cnsbest.co.kr
Tel.: +82-70-4736-1118
Device Information
- Trade Name: Automatic Micro Needle System (CODE-X)
- Common Name: Microneedling Device
- Classification Name: Microneedling Device For Aesthetic Use
- Product Code: QAI
- Panel: General and Plastic Surgery
- Regulation Number: 21 CFR 878.4430
- Device Class: Class II (Special Controls)
- Date Prepared: 04/09/2026
Predicate Devices:
Primary Predicate
- K230420, Dr.pen Microneedling System by GC AMERICA INC. (Guangzhou Ekai Electronic Technology Co., Ltd.)
Indication for Use:
The Automatic Micro Needle System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older.
Device Description:
The CODE-X Automatic Micro Needle System is a motorized, AC-powered microneedling device designed to mechanically create microscopic channels in the epidermal and upper dermal layers of the skin for aesthetic use. The device delivers controlled microneedle penetration through a reciprocating needle cartridge driven by the motorized handpiece with embedded microprocessor control.
The system consists of the following components:
{5}
- Motorized handpiece (reusable) with adjustable needle depth control
- Sterile, single-use disposable microneedle cartridges (12-pin and 14-pin, EO sterilized)
- AC power adapter
The technical specifications of the device are as follows:
| Parameter | Specification |
| --- | --- |
| Power Source | AC Powered |
| Control Mechanism | Microprocessor (Embedded software) |
| Needle Configuration | 12-pin and 14-pin cartridges |
| Needle Gauge/Diameter | 33G (-0.2 mm) |
| Needle Shape/Geometry | Straight, cylindrical body with a conical tapered, sharp point |
| Needle Spacing | 2 mm spacing / 3.96³mm per needle |
| Needle Arrangement | Needles radially arranged |
| Adjustable Penetration Depth | 0 mm to 2.0 mm in 0.20 mm increments |
| Recommended Penetration Depth | 1.5 mm |
| Maximum Penetration Depth | 2.0 mm |
| Sterilization (Cartridge) | Ethylene Oxide (EO), single-use |
| Shelf Life | 3 years |
| Biocompatibility | According to ISO 10993-1 |
| Intended Use Environment | Professional healthcare setting (Rx) |
{6}
# Substantial Equivalence Comparison
| Characteristic | Proposed Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Device Name | Automatic Micro Needle System (CODE-X) | Dr.pen Microneedling System | - |
| Manufacturer | Woorhi Mechatronics Co., Ltd. | GC AMERICA INC. | - |
| 510(k) Number | K252591 | K230420 | - |
| Regulatory Classification | Class II | Class II | - |
| Classification Name | Microneedling Device For Aesthetic Use | Microneedling Device For Aesthetic Use | - |
| Product Code | QAI | QAI | - |
| Indications for Use | A microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older. | A microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older. | Equivalence |
| Intended Location of Use | Skin (Face) | Skin (Face) | Equivalence |
| Power Source | AC Powered | AC Powered | Equivalence |
| Control Mechanism | Microprocessor (Embedded software) | Microprocessor (Embedded software) | Equivalence |
| Microneedling Cartridge | Sterile, Single use | Sterile, Single use | Equivalence |
| No. of Needles | 12-pin, 14-pin | 14-pin | Difference |
| Needle Shape/Geometry | Straight, cylindrical body with a conical tapered, sharp point | Straight, cylindrical body with a conical tapered, sharp point | Equivalence |
| Needle Spacing | 2 mm spacing / 3.96²mm per needle | 2 mm spacing / 3.96²mm per needle | Equivalence |
| Penetration Depth (Recommended) | 1.5 mm | 1.5 mm | Equivalence |
| Max Needle Setting | 2.0 mm | 2.0 mm | Equivalence |
| Penetration Depth Selection | 0 mm to 2.0 mm in 0.20 mm increments | 0 mm to 2.0 mm in 0.25 mm increments | Difference |
| Needle Arrangement | Needles radially arranged | Needles radially arranged | Equivalence |
| Sterility | Ethylene Oxide | Ethylene Oxide | Equivalence |
| Shelf Life | 3 years | 2 years | Difference |
| Clinical Performance | Not applicable | Not applicable | Equivalence |
| Biocompatibility | According to ISO 10993-1 | According to ISO 10993-1 | Equivalence |
# Substantial Equivalence: The Automatic Micro Needle System (CODE-X) is substantially equivalent to the Dr.pen Microneedling System (K230420) predicate device. The devices are under the same product code (QAI), have the same intended use/indication for use, same needle shape and arrangement, same material, recommended penetration depth, and sterilization method. The only technological differences are the number of needle options (additional 12-pin), penetration depth selection increment (0.20 mm vs. 0.25 mm), and shelf life (3 years vs. 2 years). These differences are either verified with performance data or are judged to have no significant impact on achieving the intended use.
{7}
# Non-Clinical Test Data
Below tests were performed on the subject device to demonstrate compliance with 21 CFR 878.4430 special controls:
- Penetration Depth Accuracy Testing (0.5, 1.0, 2.0 mm settings; n=10 per setting; all within ±5% tolerance)
- Maximum Safe Penetration Depth Verification (3 devices × 10 measurements = 30 total; all ≤ 2.1 mm; 95% CI [1.992, 2.008])
- Needle Structural Integrity Testing (1,000 penetration cycles; no fracture or deformation)
- Cartridge Attachment Security Testing (30 attach/detach cycles across 3 devices; no loosening)
- Electrical Safety Testing per IEC 60601-1
- EMC Testing per IEC 60601-1-2
- Biocompatibility Testing per ISO 10993-1
- EO Sterilization Validation per ISO 11135
- Packaging and Shelf Life Validation per ISO 11607 and ASTM D4169
- Reprocessing Validation for reusable handpiece
- Software Verification and Validation
- Human Factors Validation Testing
The results of the above tests demonstrate that the device performs as intended under anticipated conditions of use and meets the special controls requirements of 21 CFR 878.4430. Non-clinical performance data confirm that the device is substantially equivalent as the predicate device.
Clinical testing was not necessary to establish substantial equivalency of the device.
# Conclusion
The Automatic Micro Needle System (CODE-X) constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This device has the same intended use and fundamental scientific technology as its predicate device (Dr.pen Microneedling System, K230420). Therefore, the CODE-X Automatic Micro Needle System and its predicate are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.