The Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to establish and maintain pneumoperitoneum by filling the abdominal cavity with carbon dioxide (CO2) gas and distending the cavity.
Device Story
Microprocessor-controlled CO2 insufflator; establishes/maintains pneumoperitoneum for laparoscopic surgery. Inputs: external CO2 gas supply; user-selected abdominal modes (pediatric, adult, bariatric). Operation: internal pressure regulator/flow control system adjusts gas delivery; monitors intra-abdominal pressure; automatically inflates to maintain preset pressure; compensates for leakage. Output: CO2 gas delivered to abdominal cavity; visual/audible warnings for overpressure, gas supply, contamination, and heater defects. Used in clinical/surgical settings; operated by surgical staff. Benefits: maintains surgical workspace/visual field; enhances safety via automated pressure regulation and monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm systems (IEC 60601-1-8), and endoscopic equipment safety (IEC 60601-2-18). Comparative non-clinical testing verified pressure/flow accuracy, leakage compensation, heating performance, and alarm thresholds against the predicate.
Technological Characteristics
Microprocessor-controlled CO2 insufflator; includes control circuit, power supply, pressure regulator, and hoses. Features pediatric, adult, and bariatric modes. Pressure range 0-30mmHg; flow range 0-50 L/min. Includes gas heating, liquid contamination sensing, and overpressure warning. Power: 100-240V, 50/60Hz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-18.
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic laparoscopic procedures requiring pneumoperitoneum via CO2 insufflation.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 25, 2026
Weyo Surgical Technology Ltd.
% Jack Fang
R&D Director
APlus Healthcare Technology (Shanghai) Co., Ltd.
1201 Floor, Zhongzhi Building, #9299 Humin Road, Xuhui Dis
Shanghai, 200030
CHINA
Re: K252584
Trade/Device Name: Insufflator (wAlcor C50A, wAlcor C50B, wAlcor C50C,
wAlcor C35A, wAlcor C35B, wAlcor C35C)
Regulation Number: 21 CFR 884.1730
Regulation Name: Laparoscopic Insufflator
Regulatory Class: II
Product Code: HIF
Dated: May 21, 2026
Received: May 22, 2026
Dear Jack Fang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252584 - Jack Fang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252584 - Jack Fang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252584 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Insufflator (wAlcor C50A, wAlcor C50B, wAlcor C50C, wAlcor C35A, wAlcor C35B, wAlcor C35C) | | |
| Please provide your Indications for Use below. | | ? |
| The Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to establish and maintain pneumoperitoneum by filling the abdominal cavity with carbon dioxide (CO2) gas and distending the cavity. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252584
Page 1 of 5
510(k) Summary---
# 510(k) Summary
## I Submitter
Weyo Surgical Technology Ltd.
Room 313, Building 3 56 Lingzhi Road, Xuanwu District, Nanjing City, Jiangsu Province 210018, China
## Primary Contact person
Mr. Jianfeng Wang
Occupation Title: R&D Director
Tel.: +86-18551626339
E-mail: zhuce@weyosurgical.com
## Submission Correspondent
Mr. Jack Fang
APlus Healthcare Technology (Shanghai) Co., Ltd.
1201 Floor, Zhongzhi Building, No.9299 Humin Road, Xuhui District, Shanghai, China
E-mail: jack.fang@ap-healthcare.com
**Preparation date:** June 24, 2026
## II Proposed Device
**Trade Name of Device:** Insufflator
**Models:** wAlcor C50A, wAlcor C50B, wAlcor C50C, wAlcor C35A, wAlcor C35B, wAlcor C35C
**Regulation Description:** Laparoscopic insufflator
**Regulation Number:** 21 CFR 884.1730
**Regulatory Class:** Class II
**Product codes:** HIF
## III Predicate/Reference Device
**Predicate 510(k) Number:** K222812
**Trade Name:** Insufflator
**Regulation Number:** 21 CFR 884.1730
**Regulatory Class:** Class II
**Product codes:** HIF (Laparoscopic Insufflator)
The predicate device has not been subject to a design-related recall.
---1 / 5
{5}
K252584
Page 2 of 5
510(k) Summary
Reference 510(k): K161554
Trade Name: ENDOFLATOR 40 & 50
### IV Device description
The Insufflator is a microprocessor controlled CO₂ Insufflator used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
The Insufflator consists of a control circuit, power supply section, gas pipeline and accessories (Pressure Regulator, Hose). The control circuit obtains external gas through the connected gas supply pipeline. After pressure and flow adjustment by the internal gas regulation system, the gas is delivered from the gas outlet connected to the insufflation tube to the patient's abdominal cavity. At the same time, the gas pressure is monitored to ensure the safety of the surgery.
The Insufflator has the heating function when used with heatable insufflation tube. Based on the size of the intra-abdominal cavity, the system presets three abdominal modes, each with preset values for pressure, flow rate, and over-pressure time.
### Principles of Operation
The Insufflator is a device used for establishing and maintaining pneumoperitoneum during abdominal endoscopic surgery. The Insufflator can be used to inject medical CO₂ into the abdominal cavity. The gas separates the abdominal wall from the abdominal organs, creating a surgical operation and visual field space. When the preset pressure is reached, the Insufflator can automatically stop the air intake and maintain a certain amount of gas to keep the abdominal cavity in a predetermined pressure inflation state all the time. Whenever the intra-abdominal pressure decreases during the surgical operation, the Insufflator can automatically inflate to maintain the necessary operation and observation space required for the surgical operation.
### V Indications for use
The Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to establish and maintain pneumoperitoneum by filling the abdominal cavity with carbon dioxide (CO₂) gas and distending the cavity.
### VI Comparison of technological characteristics with the predicate device
The comparison between the overall specifications of the predicate device and the proposed device is shown in Table 1. Comparison between the Insufflator and the predicate device was conducted with respect to intended use, technological characteristics and principles of operations, providing more detailed information regarding the bases for the determination of substantial equivalence. The differences in technological characteristics do not raise any different questions of safety or effectiveness.
2 / 5
{6}
K252584
Page 3 of 5
510(k) Summary
**Table 1 Comparison of technological characteristics with the predicate device**
| Item | Proposed Device Insufflator | Predicate Device Insufflator (K222812) |
| --- | --- | --- |
| Product Code | HIF | HIF |
| Regulation No. | 21 CFR 884.1730 | 21 CFR 884.1730 |
| Class | II | II |
| Indications for Use | The Insufflator is intended for use during diagnostic and/or therapeutic laparoscopic procedures to establish and maintain pneumoperitoneum by filling the abdominal cavity with carbon dioxide (CO_{2}) gas and distending the cavity. | The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it. |
| Configuration | Insufflator (control circuit, power supply section), gas pipeline and accessories (Pressure Regulator, Hose) | Insufflator (housing, power supply, pressure reducers, venting system, fluid sensor, gas heater, various setting keys and display elements), high pressure tube and pneumoperitoneum tube |
| Mechanism of Action | Microprocessor controlled CO_{2} insufflator | Microprocessor controlled CO_{2} insufflator |
| Operating Modes | Pediatric mode, adult mode and bariatric mode | Pediatric mode, adult mode and bariatric mode |
| Set Pressure Range | Pediatric mode: 0-15mmHg Adult mode/Bariatric mode: 0-30mmHg | Pediatric mode: 1-10mmHg Adult mode and bariatric mode: 1-30mmHg |
| Accuracy of Pressure | ±2mmHg (266.6Pa) | ±2mmHg |
| Set Gas Flow Range | 0-50 L/min | 1-50L/min |
| Power Supply | 100-240V, 50/60Hz | 100-240V, 50/60Hz |
| Weight | About 8.5kg | About 8.2kg |
| Dimension (W x H x D) | 350mm x 390mm x 145mm | 267mm x 395mm x 134mm |
| Overpressure warning | Has the overpressure warning function; Detected | Has the overpressure warning function; |
3 / 5
{7}
K252584
Page 4 of 5
510(k) Summary
| Item | Proposed Device Insufflator | Predicate Device Insufflator (K222812) |
| --- | --- | --- |
| function | pressure exceeds the set value by more than 4mmHg for 1 second, or the detected pressure exceeds the set value by more than 2mmHg and exceeds the over-pressure time setting, the overpressure warning is initiated. | When the actual pressure is > 3 mmHg above the nominal pressure, the overpressure warning is initiated. |
| Gas supply warning function | Yes | Yes |
| Contamination warning function | Yes, liquid detected in insufflator tubing via a sensor | Yes, via fluid backflow sensor |
| Heater defect warning function | When Heating temperature exceeds 41°C, the gas injection will stop, device warning initiated. | When the gas temperature is greater than 41°C, the gas injection will be interrupted, device warning initiated. |
### VII Performance Data
The following performance data were provided in support of the substantial equivalence determination.
- Electrical safety test was conducted per IEC 60601-1:2005+AMD1:2012+AMD2:2020.
- Electrical safety and electromagnetic compatibility (EMC) testing were conducted per IEC 60601-1-2: 2014+AMD1:2020.
- Alarm system was tested per IEC 60601-1-8:2006+AMD1:2012+AMD2:2020.
- Particular requirements for the basic safety and essential performance of endoscopic equipment per IEC 60601-2-18:2009.
### Non-Clinical Testing
The following comparative non-clinical testing was performed on the subject device and predicate or reference device to demonstrate substantial equivalence:
- Adjustable Range of Set Pressure
- Accuracy of Pressure Preset and Display
- Overpressure Alarm Threshold
- Overpressure Release Time
4 / 5
{8}
K252584
Page 5 of 5
510(k) Summary---
- • Underpressure Compensation Time
- • Compensation of Leakages (Transient)
- • Compensation of Leakages (Continuous)
- • Adjustable Range of Set Flow Rate
- • Accuracy of Flow Set and Flow Display
- • Accuracy of Displayed Gas Consumption
- • Heating (Gas Temperature)
- • Manual Smoke Evacuation
- • Liquid Contamination
### Software
- Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, 'Content of Premarket Submissions for Device Software Functions.' The software for this device required enhanced documentation.
- Cybersecurity
The cybersecurity documentation was provided according to FDA guidance 'Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.'
### VIII Clinical Testing
No clinical study is included in this submission.
### IX Conclusion
The subject device has the same indications for use and has similar design features and technological characteristic as the predicate device. The differences between the predicate device and the subject device do not raise different questions of safety or effectiveness. Performance testing data demonstrates that the proposed device is as safe and effective as the predicate device. Therefore, the subject device is substantially equivalent to the predicate device.
---5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.